Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151960
    Date Cleared
    2016-01-07

    (175 days)

    Product Code
    Regulation Number
    870.1650
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.

    Device Description

    The proposed devices are available in packs, which may include different configurations of syringes and accessories. The syringes are plastic, disposable syringes, which are available in various models and configurations. They are intended to be used with an U.S. legally marketed angiography injector, compatibilities are shown in Table 1.

    AI/ML Overview

    The provided document is a 510(k) summary for the Shenzhen Baoan Medical Supplies Co., Ltd. Sterile High-pressure Angiographic Syringes for Single-use. This type of submission is for medical devices, and the "acceptance criteria" and "device performance" primarily refer to compliance with recognized standards and biocompatibility, rather than clinical efficacy or diagnostic accuracy.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists various international and national standards that the device complies with, which serve as the acceptance criteria for its non-clinical performance and safety. The reported device performance is that it "complies with" these standards.

    Acceptance Criteria (Standards Complied With)Reported Device Performance
    ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residualsComplies
    ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materialsComplies
    ASTM F1140/F1140M-13 Standard test methods for internal pressurization failure resistance of unrestrained packagesComplies
    ASTM F1929-12 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye PenetrationComplies
    USP37-NF32 <85> Bacterial Endotoxins LimitComplies
    ISO 11737-2:2009 Sterilization of medical devices - Microbiological methods - Part 2: Test of sterilityComplies
    ISO 7886-1:1993 Sterile hypodermic syringes for single use - Part 1: Syringes for manual useComplies
    ISO 7886-2:1996 Sterile hypodermic syringes for single use - Part 2: Syringes for use with power-driven Syringe PumpsComplies
    ISO594:1986 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1Complies
    ISO594:1998 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2Complies
    ISO 10993-4:2002 AMD 2006 Biological evaluation of medical devices -- Part 4: Selection of tests for interactions with bloodComplies
    ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicityComplies
    ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationComplies
    ISO 10993-11:2006 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicityComplies
    ASTM F 756-08, Standard practice for assessment of hemolytic properties of materialComplies
    USP37-NF32 <151> Pyrogen testComplies
    Biocompatibility requirements of ISO 10993 series StandardsConforms to the requirements (for both proposed and predicate devices)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states "Non clinical tests were conducted to verify that the proposed device met all design specifications..." and lists various standards for these tests. However, it does not specify the sample sizes used for these non-clinical tests (e.g., number of syringes tested for seal strength or internal pressurization). The data provenance is generally from laboratory testing performed by the manufacturer to demonstrate compliance with the listed standards. No country of origin is explicitly stated for the testing itself, but the manufacturer is based in China. The data is prospective in the sense that the manufacturer conducted these tests to support the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This type of information is not applicable to this 510(k) submission. The document focuses on the engineering and material safety aspects of a medical device (angiographic syringes), not on diagnostic accuracy where expert ground truth would be established. The "ground truth" here is compliance with established engineering and biological safety standards, which are verified through objective testing methods rather than expert consensus on images or case outcomes.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set:

    Not applicable. Adjudication methods like "2+1" typically refer to the process of resolving discrepancies in expert interpretations (e.g., in reading medical images). Since no expert interpretations are involved in establishing the "ground truth" for these engineering and biological tests, no such adjudication method was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical study is included in this submission." This type of study is relevant for assessing the clinical utility or diagnostic effectiveness of devices, often involving human readers interpreting medical data. This syringe device is a non-active, sterile, single-use product, and its evaluation focuses on its physical and biological integrity, not its interaction with clinical interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    Not applicable. This device is a physical medical product (angiographic syringe), not an algorithm or software. Therefore, the concept of "standalone performance" for an algorithm is not relevant here.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's evaluation is primarily based on compliance with recognized national and international consensus standards for medical device physical performance, sterility, and biocompatibility. This involves objective measurements against predefined thresholds and specifications outlined in standards like ISO 7886, ISO 10993, ASTM F88, etc.

    8. The Sample Size for the Training Set:

    Not applicable. The concept of a "training set" is relevant for machine learning algorithms. This submission is for a physical medical device, not an AI/ML product. The manufacturer did not develop this device using machine learning.

    9. How the Ground Truth for the Training Set was Established:

    The concept of a training set and its ground truth is not applicable for this device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1