Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150756
    Manufacturer
    Date Cleared
    2015-07-20

    (119 days)

    Product Code
    Regulation Number
    884.6160
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in-vitro fertilisation, handling and embryo culture. It may be used for the culturing of embryos using drop culture

    Device Description

    The Micro well group culture dishes are injection molded and disposable. The dishes are round and supplied with a lid. The dishes holds 9 small wells respectively 16 small wells. The polymers used for the dishes are virgin crystal grade polystyrene, which has successfully passed 1- cell embryotoxicity test. The dishes are packed 1 unit in a sleeve, 10 sleeves in a case. The dishes are terminally sterilized by irradiation to achieve a SAL of 10-6. The dishes are non-pyrogenic as tested by LAL, and non-embryotoxic as tested by one cell Mouse Embryo Assay (MEA). The Dishes are single use devices.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a new medical device, not a study evaluating a device's performance against detailed acceptance criteria for an AI algorithm. Therefore, many of the requested fields are not applicable to the provided text.

    The document discusses two new devices: a "Micro well group culture dish, 9-well" and a "Micro well group culture dish, 16-well," for use in in-vitro fertilization (IVF). The submission focuses on demonstrating substantial equivalence to a legally marketed predicate device (Vitrolife Micro Droplet Culture Dish, K123641) rather than presenting a performance study of a novel AI system.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria here are based on the characteristics of the predicate device and established standards for IVF labware. The performance is reported as meeting these standards.

    Acceptance Criteria (from Predicate Device/Standards)Reported Device Performance (Micro well group culture dish, 9-well & 16-well)
    Sterility: SAL 10⁻⁶Sterile (SAL 10⁻⁶)
    Non-pyrogenic: LAL of
    Ask a Question