510(k) Data Aggregation

    K Number
    K143692
    Manufacturer
    Date Cleared
    2015-08-21

    (240 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GranOS™ devices are indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. GranOS™ is indicated to be gently placed into bony voids or gaps of the skeletal system (ie the extremities, posterolateral spine and pelvis). These defects may be surgically created osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

    Device Description

    GranOS™ is a porous, resorbable osteoconductive scaffold constructed of highly pure beta tricalcium phosphate granules for use in the repair of bony defects.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a medical device called GranOS™ - beta tricalcium phosphate synthetic bone granules. It establishes substantial equivalence to legally marketed predicate devices, rather than a study proving the device meets specific acceptance criteria in the context of diagnostic or prognostic performance. The information provided does not describe a clinical study in the typical sense of measuring diagnostic accuracy or treatment effectiveness with human subjects.

    Therefore, the requested information elements related to diagnostic performance, sample sizes for test/training sets, expert adjudication, MRMC studies, and ground truth establishment for a diagnostic AI device are not applicable to this document. This document focuses on the technical and safety characteristics for regulatory approval as a bone void filler.

    However, I can extract the relevant information from the document regarding the device's technical characteristics and how it meets regulatory standards for substantial equivalence.


    1. Table of Acceptance Criteria and Reported Device Performance

    For medical devices like bone void fillers, "acceptance criteria" are typically defined by recognized standards and comparison to predicate devices, focusing on material properties, biocompatibility, and intended use. The performance is then demonstrated through non-clinical testing to meet these standards and show equivalence to predicates.

    Acceptance Criteria / Characteristic TypeSpecific Criteria / Predicate ValuesReported Device Performance (GranOS™)
    Chemical CompositionBeta tricalcium phosphate Ca3(PO4)2 meeting ASTM F 1088-04a (2010)Beta tricalcium phosphate Ca3(PO4)2 meeting ASTM F 1088-04a (2010), same as predicates
    Patient PopulationIndividuals with bony defects resulting from surgery or traumaIndividuals with bony defects resulting from surgery or trauma, same as predicates
    Anatomical LocationsBony voids/gaps in extremities, posterolateral spine, and pelvisTo be packed into irregular shaped bony voids of the skeletal system (extremities, spine, pelvis), same as predicates
    Physical StructureInterconnected porosity (most predicates) / Trabecular structure (one predicate)Interconnected porosity
    PorosityApproximately 70% (two predicates) / 60-70% (one predicate) / 90% (one predicate)Approximately 70%
    Pore SizeMicropores 0-