510(k) Data Aggregation

    K Number
    K141141
    Date Cleared
    2014-08-14

    (104 days)

    Product Code
    Regulation Number
    888.3027
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kiva VCF Treatment System is indicated for use in the reduction and treatment of spinal fractures in the thoracic and/or lumbar spine from T6-L5. It is intended to be used in combination with the Benvenue Vertebral Augmentation Cement Kit.

    Device Description

    The Kiva® VCF Treatment System is provided as a sterile, single use, implantable device which may be used in percutaneous procedures for the reduction and treatment of spinal fractures. The device consists of an implant, a deployment system for the implant, and a set of accessory access instruments. The deployment system component is a single-use, non- implantable device that is used to properly position and deliver the implant. It consists of a nitinol Kiva Coil, which is guided through a deployment cannula into the bone via a hand- operated mechanism. The Kiva Implant, made from PEEK Optima with barium sulfate for radiopacity, is guided over the nitinol Kiva Coil. As the Kiva Implant is pushed over the Kiva Coil, it reduces the fracture via height distraction of the vertebral body. Up to 5 loops of implant may be deployed, corresponding with a maximum height of 15mm. Once the Implant is placed into the vertebral body, PMMA bone cement is deployed into the Kiva Implant. The Kiva Implant contains the PMMA bone cement and helps extravasation. A collection of manual surgical orthopedic instrumentation (Class I needles, stylets, cannulas) is used to gain access to the vertebral body at the start of the procedure, then again later for bone cement deployment. Other instruments are optional for the physician and not provided as part of the Kiva® VCF Treatment System including a Bone Drill, a Bone Biopsy Needle and Kiva Pilot.

    AI/ML Overview

    The provided text describes the Kiva® VCF Treatment System, a medical device for treating spinal fractures. However, it does not include information about a study proving the device meets acceptance criteria related to AI/algorithm performance, nor does it mention any AI components, human readers, or ground truth establishment in the context of an AI-powered diagnostic or treatment device.

    The document is a 510(k) summary for a physical medical device that augments vertebral bodies with cement. The "acceptance criteria" and "study" described in the provided text relate to the mechanical and biomechanical performance of an implantable device, specifically comparing a 5-loop version to a 4-loop version.

    Therefore, most of your requested information (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC comparative effectiveness, standalone performance, training set sample size, ground truth for training set) is not applicable or present in this document, as it pertains to AI/algorithm studies, not the mechanical testing of a physical implant.

    I can, however, extract the acceptance criteria and performance data relevant to the physical device's mechanical integrity and deployment performance as described in the document.


    Acceptance Criteria and Device Performance (for the physical Kiva® VCF Treatment System)

    Acceptance Criteria CategoryReported Device Performance
    Biomechanical Performance (Static & Dynamic)The 5-loop system met all pre-defined criteria per ASTM F2077-11 and ASTM F1877-05 (evaluated without PMMA cement present, as a worst-case scenario). This was equivalent to the original 4-loop system.
    Deployment PerformanceThe 5-loop system met all pre-defined criteria in a simulated clinical-use protocol. This was equivalent to the original 4-loop system.

    Study Information (related to the physical device's mechanical and deployment performance):

    1. Sample size used for the test set and the data provenance:

      • Biomechanics: Not explicitly stated as a "sample size" of devices evaluated, but indicates testing per ASTM F2077-11 and ASTM F1877-05 for both the 4-loop and 5-loop systems. These standards specify how many samples are typically tested.
      • Deployment Performance: "a total of 30 units" for the 5-loop system.
      • Data Provenance: The cadaveric spines and urethane foam analogs would be considered laboratory/bench testing data. The origin isn't specified (e.g., country), but it's a controlled test environment. This is prospective testing for the device modification.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The "ground truth" here is meeting engineering specifications (e.g., structural integrity, successful deployment), not expert clinical interpretation.

    3. Adjudication method for the test set: Not applicable. This refers to mechanical testing against engineering standards.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document does not describe a study involving human readers or AI.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This document is about a physical medical device, not an algorithm.

    6. The type of ground truth used:

      • Biomechanical Performance: Adherence to established ASTM standards (material and mechanical engineering specifications).
      • Deployment Performance: Successful completion of a predefined simulated clinical-use protocol (i.e., successful deployment of the implant within specified parameters).
    7. The sample size for the training set: Not applicable. There is no algorithm or "training set" mentioned.

    8. How the ground truth for the training set was established: Not applicable. There is no algorithm or "training set" mentioned.


    Summary of the document's relevance:

    This document is a regulatory submission demonstrating the substantial equivalence of a modified physical medical device (Kiva® VCF Treatment System with 5 loops) to a previously cleared version (4 loops). The "study" described is a series of non-clinical, benchtop, and simulated-use tests to ensure the modified device maintains the same safety and performance characteristics as its predicate. It is not related to software, AI, or diagnostic image analysis.