510(k) Data Aggregation

    K Number
    K122519
    Manufacturer
    Date Cleared
    2013-05-21

    (277 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DIO UF HSA Internal Sub-Merged Implant System is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment to restore a patient's chewing function. The smaller (03.8 ~ (05.5) implants can be placed with a conventional two stage surgival process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability is achieved with appropriate occlusal loading. The larger (Ø6.0 ~ Ø7.0) implants can be placed with a conventional two stage surgical process with an option for transmucosal healing and are indicated for the molar region with delayed loading.

    Device Description

    The DIO UF HSA Internal Sub-Merged Implant System is comprised of dental implants, superstructures, instruments for prosthetics and surgical instruments. The DIO UF HSA Internal Sub-Merged Implant System is specially designed for use in dental implant surgery. A successfully osseointegrated implant will achieve a firm implant when surgically implanted under controlled conditions, per well known clinical studies. There are intended for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations. The DIO UF HSA Internal Sub-Merged Implant System, fixtures are made of pure titanium, grade 4(ASTM F67) which have a a S.L.A commercial (Sand-blasted Large grit Acid-etched) treated surface. These fixtures can be used one-stage surgery method or two-stage surgery method. And that are surgically inserted into the upper and/or lower jaw bone. The fixtures replace tooth root as providing a stable foundation for restorations. The fixtures have the diameter(3.8 ~ 7.0mm)and length.(7 ~ 16mm). Geometrically, the implant is screw-type. An abutment is connected to the implant through a tapered-joint. The device functions by being surgically implanted in the bone of the upper or lower iaw arches in order to provide support for a prosthetic device, such as an artificial tooth, in order to restore a patient's chewing function.

    AI/ML Overview

    Acceptance Criteria and Study Details for DIO UF HSA Internal Sub-Merged Implant System

    This submission pertains to the DIO UF HSA Internal Sub-Merged Implant System, which is a dental implant device. The provided document details a 510(k) submission, indicating a focus on demonstrating substantial equivalence to a legally marketed predicate device rather than outright safety and effectiveness from de novo studies.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criterion for this device is substantial equivalence to its predicate device, Implantium II (K060501). This is primarily assessed through non-clinical performance testing (fatigue testing) and comparison of material, design, intended use, and other characteristics.

    Acceptance CriterionReported Device Performance
    Mechanical Fatigue LimitMust be over 250N with no fracture, cracks, or severe distortion of any parts after 5x10⁶ cycles at 14Hz frequency and 35° tilting angle (matching predicate device). A "worst case scenario" of the DIO HSA STEADY Internal Sub-Merged Fixture and an angled abutment was tested.
    Material CompositionCP Ti Gr4 (ASTM F67) for fixtures, CP Ti Gr3, Gr4 (ASTM F67) & Ti Alloy (ASTM F136:Ti-6Al-4V ELI) for abutment.
    Design CharacteristicsInternal Type and Morse Tapered screw-type implant.
    Implant Diameters3.8 ~ 7.0 mm
    Implant Lengths7 ~ 16 mm
    Surface TreatmentS.L.A (Sand-blasted Large grit Acid-etched) for implant fixture. Machined type for abutment surface.
    Sterilization MethodGamma for implant fixture. Non-sterilization for abutment.
    Intended UseSurgical placement in upper/lower jaw arches to provide root form for single/multiple prosthetic units to restore chewing function. Smaller implants (Ø3.8 ~ Ø5.5) for conventional two-stage or single-stage immediate loading. Larger implants (Ø6.0 ~ Ø7.0) for conventional two-stage in molar region with delayed loading.

    Reported Performance: The document explicitly states, "The fatigue limit is over 250N (Fracture or cracks or severe distortion of any parts were not detected.). It is the same fatigue limit of predicate device." It also claims that "The DIO UF HSA Internal Sub-Merged Implant System has same material and characteristics as design and technological the indication for used, similar predicate device." and "There are no known technological differences between the DIO UF HSA Internal Sub-Merged Implant System and Implantium II." This indicates that the device met the stated acceptance criteria for substantial equivalence to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The primary test set mentioned is for fatigue testing.

    • Sample Size: Not explicitly stated as a number of devices/implants tested, but implies that "the worst case scenario of the DIO HSA STEADY Internal Sub-Merged Fixture and an angled abutment" was tested. This suggests a representative sample of configurations were subjected to the specified fatigue conditions.
    • Data Provenance: The testing was conducted by the manufacturer, DIO Corporation, in Korea, as indicated by the submitter information. The data is thus prospective, generated specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This submission does not involve clinical studies with human subjects or image-based diagnostics requiring expert interpretation for ground truth. The "ground truth" for the non-clinical fatigue testing is the physical failure analysis of the tested devices against predefined engineering criteria (e.g., absence of fracture, cracks, or severe distortion, and exceeding the 250N fatigue limit). This evaluation would typically be performed by qualified engineers or technicians in a laboratory setting, rather than clinical experts. The number and qualifications of such individuals are not specified in this document.

    4. Adjudication Method for the Test Set

    Not applicable. This is a non-clinical, objective assessment of mechanical properties (fatigue testing), not a subjective clinical or diagnostic evaluation requiring an adjudication process for varying expert opinions.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This device is a dental implant, and the submission is a 510(k) for substantial equivalence, primarily relying on non-clinical testing and comparison to a predicate device, not on assessing human reader performance or AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No, a standalone study (in the context of an algorithm's performance) was not done. This device is a physical medical device (dental implant), not an AI algorithm or software. Therefore, the concept of algorithm-only performance is not applicable.

    7. Type of Ground Truth Used

    The ground truth for the non-clinical fatigue testing was engineering data/physical failure analysis. This includes objective measurements of load endured, number of cycles, and visual/microscopic inspection for material integrity (absence of fracture, cracks, severe distortion).

    8. Sample Size for the Training Set

    Not applicable. This submission is for a physical medical device (dental implant), not an AI algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.