Search Results
Found 1 results
510(k) Data Aggregation
(44 days)
Powder Free Vinyl Patient Examination Gloves, Colored (Yellow), is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free Vinyl Patient Examination Gloves, Colored (Yellow) that meets all of the requirements of ASTM standard D 5250-06.
The provided text describes a 510(k) summary for "Powder Free Vinyl Patient Examination Gloves, Colored (Yellow)". This is a medical device and not an AI/ML powered device, therefore most of the requested information (e.g., sample size for test/training set, number of experts, adjudication method, MRMC study, standalone performance) is not applicable or not found in the document.
Here's an attempt to answer the questions based on the provided text, indicating where information is not applicable or not found:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimension | ASTM standard D 5250-06 | Meets |
| Physical Properties | ASTM standard D 5250-06 | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Residual | ASTM standard D 5250-06 and D6124-01 | Meets (<2mg/glove) |
| Biocompatibility | Primary Skin Irritation in rabbits (ISO10993-10) | Passes (Not a Primary Skin Irritation) |
| Biocompatibility | Dermal sensitization in the guinea pig (ISO10993-10) | Passes (Not a Dermal sensitization) |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not specified. The document refers to meeting standards (ASTM, FDA regulations for pinholes, ISO for biocompatibility tests), which typically involve specific sampling plans outlined in those standards, but the specific sample sizes used for this submission are not detailed.
- Data Provenance: Not explicitly stated. The testing was conducted to meet international and US standards. The submitter is from China, so the testing was likely conducted in or for the company in China, but this is an inference rather than explicitly stated. The studies are non-clinical, essentially laboratory-based testing of material properties, rather than human subject data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a physical product (gloves) and its performance is assessed against physical and chemical standards, not through expert interpretation of data.
4. Adjudication method for the test set
Not applicable. Adjudication methods are relevant for subjective assessments, particularly in medical imaging or diagnostics, where multiple experts might disagree. This device's performance is objectively measured against established standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-powered device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an AI-powered device.
7. The type of ground truth used
The "ground truth" for this device's performance is defined by the objective metrics and thresholds established in the referenced standards:
- ASTM standard D 5250-06 for dimension, physical properties, and powder residual.
- 21 CFR 800.20 for freedom from pinholes (waterleak test on pinhole AQL).
- ASTM D6124-01 for powder residual.
- ISO10993-10 for biocompatibility (Primary Skin Irritation and Dermal sensitization methods).
8. The sample size for the training set
Not applicable. This is not an AI/ML device; there is no "training set."
9. How the ground truth for the training set was established
Not applicable. There is no training set for this device.
Ask a specific question about this device
Page 1 of 1