Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K070406
    Device Name
    ENDODRAPE
    Date Cleared
    2007-07-30

    (168 days)

    Product Code
    Regulation Number
    878.4370
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Endodrape™ surgical and diagnostic procedure drape manufactured by Vortek Surgical is for patients receiving and/or medical professionals performing a non-sterile diagnostic procedure for colonoscopies to help protect the patient and staff from bodily secretions and to maintain a cleaner procedural site. The device is a single use disposable drape that is provided non-sterile. The device can, however, be EtO (ethylene oxide) sterilized by the end user prior to use.

    Device Description

    The Endodrape™ surgical and diagnostic procedure drapes manufacture by Vortek Surgical, Inc. are non-sterile, single use, disposable barriers intended to be used by medical professionals as a protective patient covering; to protect clinical staff and equipment from patient secretions; and help maintain a cleaner procedural site. The drapes are blue in color and are a non-woven fabric that will be manufactured under private label to Vortek Surgical. The Endodrape™ fabric itself is a repellant treated, non-woven fabric manufactured under the trade name of Sontara by the contract manufacturer to Vortek Surgical, Inc. There are no reinforced areas of the drape. The drape is 0.13 inch thick and measures 43" x 49" and is without any areas of reinforcement. The fibers used in the drape are mechanically bonded together.

    AI/ML Overview

    The provided document is a 510(k) summary for the Endodrape™ Surgical and Diagnostic Procedure Drape, a physical medical device (non-sterile surgical drape), not a software or AI-powered device. Therefore, many of the requested categories related to AI performance, such as multi-reader multi-case studies, standalone algorithm performance, training set details, and expert ground truth establishment for a test set, are not applicable.

    However, I can extract information related to the acceptance criteria and the studies performed to demonstrate equivalence for this specific device.


    1. Table of Acceptance Criteria and Reported Device Performance

    The device is assessed for substantial equivalence to predicate devices based on various physical, mechanical, liquid barrier penetration, and biocompatibility characteristics. The acceptance criterion for each characteristic is generally "Same or similar" to the predicate devices.

    Product CharacteristicAcceptance Criteria (compared to predicate)Reported Device Performance (Endodrape™)
    Physical/Mechanical
    WeightSame or similar1.9 oz/yd²
    Grab Tensile StrengthSame or similar22 lb/in
    Water ImpactSame or similar1.5 g
    Water HydrostaticSame or similar22 cm
    Surface Wetting SpraySame or similar90 (range: 0-100)
    Alcohol RepellencySame or similar9 (range: 0-10)
    Flame RetardancySame or similarDNI - Class I
    Biocompatibility
    CytotoxicityNon cytotoxicNon cytotoxic
    Skin IrritationNon skin irritantNon skin irritant
    Skin SensitivityNon skin sensitizing agentNon skin sensitizing agent
    Liquid Barrier Classification
    AAMI Liquid Barrier ClassificationSame or similarLevel 1 (≤ 4.5 g)
    Other Characteristics
    Indications for UseSame or similarFor non-sterile diagnostic colonoscopies to protect patient/staff from bodily secretions and maintain a cleaner procedural site. Single-use, disposable, non-sterile (can be EtO sterilized by end user).
    ContraindicationsNA (No contraindications)NA
    Environment Where UsedSame or similarSurgical and diagnostic centers
    Number of UsesSame or similarSingle
    Provided & Used Non-SterileYes/No (EtO for predicate)Yes/Yes
    SizesSame or similar43" x 49" (only)
    ThicknessSame or similar0.13 inch
    Material DescriptionSame or similarNon-woven fabric; various man-made and synthetic fibers (e.g., Sontara), Color: blue.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size for each specific bench test performed on the Endodrape™. However, it indicates that "All physical and mechanical testing was performed in accordance with the applicable standards and regulations" set forth by ASTM, INDA, US CFR, AATCC, and ISO.

    The data provenance is from non-clinical performance (bench testing) conducted by the contract manufacturer to Vortek Surgical, Inc. The location of the testing is not specified beyond being performed by the contract manufacturer. The data is prospective in the sense that these tests were specifically carried out for the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is a physical device being evaluated through bench testing against industry standards and comparison to predicate devices, not requiring expert ground truth for a test set in the conventional sense of clinical or diagnostic accuracy.

    4. Adjudication Method for the Test Set

    Not applicable. The "test set" here refers to the bench testing of the physical characteristics. Adjudication in the context of expert consensus is not relevant. The performance is objectively measured against established standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI device or an imaging device requiring human reader performance evaluation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical performance (bench testing) is based on:

    • Established Industry Standards: Reference to ASTM, INDA, US CFR, AATCC, and ISO standards (e.g., for physical, mechanical, and liquid barrier properties).
    • Predicate Device Characteristics: The comparison of the Endodrape™'s performance against the known characteristics of the predicate devices (HVO, Inc. surgical drapes and Medical Concepts Development's ColoShield Colonoscopy Drape Model E2000).
    • Biocompatibility Testing: Results from specific biocompatibility tests (Cytotoxicity, Sensitization, Irritation) conducted in accordance with ISO 10993-1:2003.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical device. There is no training set in the context of machine learning.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no training set for a physical device.


    Study Proving Device Meets Acceptance Criteria:

    The study that proves the device meets the acceptance criteria is the Non-Clinical Performance (Bench Testing).

    • Study Design: Comprehensive bench testing covering Physical, Mechanical, Liquid Barrier Penetration, and Biocompatibility.
    • Standards Followed: Testing was performed in accordance with applicable industry-recognized test methods and standards from ASTM, INDA, US CFR, AATCC, and ISO, and followed recommendations from the FDA Guidance Document for Surgical Gowns and Surgical Drapes, August 1993. Biocompatibility testing followed ISO 10993-1:2003.
    • Comparison Basis: The results of these tests were compared to the characteristics of legally marketed predicate devices (HVO, Inc. surgical drapes and Medical Concepts Development's ColoShield Colonoscopy Drape Model E2000).
    • Key Findings: The Endodrape™ surgical and diagnostic procedure drapes were found to have physical, mechanical, and liquid barrier properties that are "Same or similar" to the predicate devices. Furthermore, biocompatibility testing (cytotoxicity, skin irritation, and skin sensitivity) confirmed the device was non-cytotoxic, non-irritating, and non-skin sensitizing, even under worst-case EtO sterilized conditions.
    • Conclusion of Study: The results of the non-clinical performance testing demonstrated that the Endodrape™ drapes are acceptable for their intended use and are substantially equivalent to the predicate devices, not raising any new questions regarding safety or effectiveness.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1