510(k) Data Aggregation

    K Number
    K063254
    Date Cleared
    2006-12-07

    (41 days)

    Product Code
    Regulation Number
    882.5950
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TRUFILL® DCS Syringe II is indicated for use with the TRUFILL® family of Detachable Coils.

    Device Description

    The TRUFILL® DCS Syringe II consists of a 14-cc barrel with a pressure gauge, a threaded plunger assembly with a locking mechanism, and a flexible high-pressure extension tube with a male luer connector. The gauge faceplate is calibrated for use with the TRUFILL DCS ORBIT™ Detachable Coil and the TRUFILL® DCS Detachable Coil; i.e., the TRUFILL® family of Detachable Coils. The TRUFILL® DCS Syringe II is used to generate controllable pressure for preparation and coil detachment of the TRUFILL™ family of Detachable Coils.

    AI/ML Overview

    The provided document describes the TRUFILL® DCS Syringe II, an artificial embolization device, and its substantial equivalence to a predicate device. The information details the performance and safety testing, as well as design validation, conducted to demonstrate its safety and effectiveness.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the "Performance and Safety Testing" and "Design Validation Testing" conducted. The reported device performance is that these tests were conducted and the device performed as intended and met specifications.

    Acceptance Criteria (Implied from Test Titles)Reported Device Performance
    Gauge Accuracy and Pressure Cycling AccuracyTesting conducted; device performs as intended
    Luer Lock Connector Dimensional Verification adherenceTesting conducted; device performs as intended
    Chemical CompatibilityTesting conducted; device performs as intended
    Joint Pull StrengthTesting conducted; device performs as intended
    Device IntegrityTesting conducted; device performs as intended
    Torque FunctionalityTesting conducted; device performs as intended
    Device Flush Particulate LevelsTesting conducted; device performs as intended
    BiocompatibilityTesting conducted; device performs as intended
    Syringe Packaging IntegrityTesting conducted; device specifications conform to user needs
    Purge and Detachment of TRUFILL® DCS Detachable Coils FunctionalityTesting conducted; device specifications conform to user needs
    Purge and Detachment of TRUFILL DCS ORBIT™ Detachable Coils FunctionalityTesting conducted; device specifications conform to user needs
    Syringe IntegrityTesting conducted; device specifications conform to user needs
    Ability to Pressurize SyringeTesting conducted; device specifications conform to user needs

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for each specific test or the data provenance (e.g., country of origin of the data, retrospective or prospective). It generally states that "in-vitro laboratory performance testing" and "in-vitro testing" were conducted.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. The studies described are in-vitro laboratory tests, which typically rely on measurement instruments and adherence to technical specifications rather than human expert interpretation for ground truth.

    4. Adjudication Method for the Test Set

    This information is not applicable and is not provided. Given the nature of in-vitro engineering and performance testing, there's no mention of human adjudication methods like 2+1 or 3+1.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The studies described are in-vitro performance and design validation tests of the device itself, not studies comparing human reader performance with or without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    The device is a physical medical instrument (a syringe) used to manipulate medical coils. It does not involve an "algorithm" or "AI" in the context of image analysis or diagnostic support. Therefore, the concept of "standalone (algorithm only)" performance is not applicable. The tests performed are standalone in the sense that they assess the device's technical specifications and functionality independent of a human operator, but it's not an AI or algorithm.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The ground truth for the in-vitro tests would have been established by engineering specifications, validated measurement techniques, and established industry standards for device performance, material compatibility, and sterile processing. For example, gauge accuracy would be compared against a known standard, and dimensional verification against engineering drawings.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical medical instrument, not an AI or machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as above; there is no training set for a physical medical device like the TRUFILL® DCS Syringe II.