Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K060776
    Date Cleared
    2006-05-25

    (64 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Surgical Face Mask is Device that is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operation room personnel from transfer of microorganisms, body fluids and particulate material.

    Device Description

    Hong Ya Non-woven Products Co., Ltd. Surgical Face Mask (Yellow, White, Green, Pink) are flat pleated 3-ply (at least) masks with an inner and outer layer (spunbonded I mk) are nat produce o psy (a a meltblown polypropylene filter material, also with elastic loops and/or strip. The nosepiece for all Hong Ya Non-woven Products Co., Ltd. Surgical Face Mask (Yellow, White, Green, Pink) is malleable aluminum wire. All the materials used in the construction of the Hong Ya Non-woven Products Co., Ltd. Surgical Face Mask (Yellow, White, Green, Pink) are being used in currently marked devices.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and the study used to demonstrate the device meets those criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance (Hong Ya Surgical Face Mask)Predicate Device Performance (K022256)
    Fluid ResistanceNo Visual Penetration (implied by predicate)Fluid Resistance (No Visual Penetration implied by comparison)No Visual Penetration
    Particulate Filtration Efficiency Performance (%)Not explicitly stated, but predicate's 2.0 microns is the benchmark.Average 96.8% at 0.1 micron2.0 microns (likely indicating filtration down to 2.0 microns, not a percentage at a specific size)
    Bacterial Filtration Efficiency (BFE) (%)Not explicitly stated, but predicate's 97.9% is the benchmark.Higher than 99.9%97.9%
    Differential Pressure (Delta-P)Not explicitly stated, but predicate's 1.8 is the benchmark.Average 2.341.8
    Flammability ClassNot explicitly stated, but predicate's Class 2 is the benchmark.Class I (No Flame Spread)Class 2

    Note: The document sometimes lists the predicate's performance in the "Acceptance Criteria" column implicitly by comparing the new device's performance to it. For particulate filtration, it seems the predicate's "2.0 microns" is the filtration size it achieves, and the new device achieves a much higher percentage at a smaller particle size (0.1 micron), indicating superior performance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for the test set for any of the performance criteria (Fluid Resistance, Particulate Filtration, BFE, Delta-P, Flammability).

    The data provenance is not explicitly stated. However, given that Hong Ya Non-Woven Products Co., Ltd. is located in P.R. China and the submission is to the FDA, it's highly probable the testing was conducted retrospectively on samples of the manufactured device. The location where the tests were performed is not mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is Not Applicable in this context. The study is a series of non-clinical, laboratory-based performance tests for a surgical mask. Ground truth is established by the standardized methods of the tests themselves, not by expert consensus on clinical data. For example, BFE is measured by a specific laboratory procedure, not by a panel of experts reviewing cases.

    4. Adjudication Method for the Test Set

    This information is Not Applicable. As with the ground truth, the tests are objective laboratory measurements, not subjective evaluations requiring adjudication among multiple observers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic or screening devices where human interpretation is involved. The listed tests are non-clinical performance evaluations of a physical product.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, a standalone performance evaluation was done. All the tests (Fluid Resistance, Particulate Filtration Efficiency, BFE, Delta-P, Flammability) are standalone assessments of the device's physical properties and performance characteristics, without human intervention in the loop for interpreting results.

    7. The Type of Ground Truth Used

    The ground truth for the device's performance is established by standardized laboratory test methods. The specific standards mentioned include:

    • Fluid Resistance: ASTM F1862-05 (Synthetic Blood Penetration Resistance Test)
    • Bacterial Filtration Efficiency (BFE): ASTM F2101-01
    • Particulate Filtration Efficiency: ASTM F1215 (Latex Particle Challenge)
    • Differential Pressure (Delta-P): MIL M 36954 C
    • Flammability: 16CFR 1610
    • Biocompatibility: ISO 10993

    8. The Sample Size for the Training Set

    This information is Not Applicable. Surgical masks are not typically "trained" in the sense of machine learning algorithms. The device's performance is inherent to its design and materials.

    9. How the Ground Truth for the Training Set Was Established

    This information is Not Applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1