Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K052065
    Date Cleared
    2005-08-22

    (24 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Reprocessed external fixation devices are indicated for use in patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects.

    Device Description

    External fixation devices are specially designed frames, clamps, rods, rod-to-rod couplings, pins, posts, fasteners, wire fixations, fixation bolts, washers, nuts, hinges, sockets, connecting bars and screws used for the management of bone fractures and reconstructive, as well as corrective, orthopedic surgery. Materials used include metal alloys, plastic and composites. These materials are chosen to address a wide range of fractures and applications as well as to allow for the appropriate amount of rigidity and stability.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a reprocessed medical device, specifically "Reprocessed External Fixation Devices."

    For such devices, the "acceptance criteria" and "study" typically focus on demonstrating that the reprocessed device performs equivalently to the original, new device and the predicate device, especially after the reprocessing procedures (cleaning, disinfection/sterilization, functional testing).

    Here's an analysis based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred)Reported Device Performance and Evidence
    Functional EquivalenceReprocessed devices must perform functionally as originally intended, comparable to new devices and predicate devices."Performance testing demonstrates that Reprocessed External Fixation Devices perform as originally intended."
    "The mechanism of action of the Reprocessed External Fixation Device is identical to the predicate devices in that the same standard mechanical design, materials, shapes and sizes are utilized."
    "Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed External Fixation Devices."
    Material/Design IdentityMaterials, design, shapes, and sizes of the reprocessed devices must remain identical to the original and predicate devices."The design, materials, and intended use of the Reprocessed External Fixation Devices are identical to the predicate devices."
    Intended Use IdentityThe intended use and indications for use of the reprocessed devices must be identical to the original and predicate devices."Intended use: External Fixation Devices are intended to be used for the fixation of supracondylar, or condylar fractures of the femur; for fusion of a joint; for surgical procedures that involve cutting the bone, for fixation of bone fractures; bone reconstruction; as a quide pin for insertion of other implants; or may be implanted through the skin so that a pulling force or traction may be applied to the skeletal system; and others may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system."
    "The only change is the modified external fixation devices will be provided non-sterile. Sterilization of the devices will occur in the hospital prior to use." (Note: this is a change in sterilization delivery, not intended use).
    Safety and EffectivenessThe reprocessed devices must be safe and effective for their intended use."Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed External Fixation Devices."
    "Alliance Medical Corporation concludes that the modified device (the Reprocessed External Fixation Device) is safe, effective and substantially equivalent to the predicate devices, as described herein."
    No change in claimsThere should be no changes to claims, clinical applications, patient population, or performance specifications."There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation."
    Reprocessing ValidationThe reprocessing procedures themselves must be validated to ensure the device performs as intended after reprocessing."Validation of reprocessing Function Testing" (This implicitly means the reprocessing process was validated such that the device maintains its function).

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed External Fixation Devices." and "Performance testing demonstrates that Reprocessed External Fixation Devices perform as originally intended."

    • Sample Size: The specific sample size for the "bench and laboratory testing" is not explicitly mentioned in the provided text.
    • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. Given it's pre-market testing for a reprocessed device, it would be prospective testing conducted by Alliance Medical Corporation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This information is not applicable to this type of submission. For medical devices like reprocessed External Fixation Devices, the "ground truth" is typically defined by engineering specifications, material properties, and mechanical performance standards, not by expert clinical consensus on diagnostic accuracy. The testing involves physical and mechanical properties, not interpretation of clinical data by experts.

    4. Adjudication Method for the Test Set

    • This information is not applicable to this type of submission. Adjudication methods (like 2+1, 3+1) are used when establishing ground truth from multiple expert readings of diagnostic images or clinical assessments. For mechanical performance testing of a device, the results are typically quantitative measurements against predefined specifications.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret output (e.g., medical images) and their performance with and without AI assistance is compared. This submission is for a reprocessed surgical implant (external fixation device), not a diagnostic tool requiring human interpretation of data where AI assistance would be relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This device is a physical medical device, not an algorithm or AI system.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • For this device, the "ground truth" for performance testing is generally based on mechanical and material specifications derived from the original device design, relevant ASTM/ISO standards, and engineering performance requirements for external fixation devices. It involves quantifiable measurements of properties like strength, fatigue resistance, torsional stability, and material integrity, ensuring they meet the same standards as new, original devices. It's not based on expert consensus, pathology, or outcomes data in the traditional sense, but rather established engineering benchmarks.

    8. The Sample Size for the Training Set

    • Not applicable. As this is a physical reprocessed medical device, there is no "training set" in the context of machine learning. The "training" for the reprocessing process would involve establishing and validating specific cleaning, disinfection, and functional testing protocols based on engineering principles and regulatory guidelines.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As mentioned above, there's no "training set" in the AI/ML sense. The "ground truth" for the reprocessing process (which ensures the device meets performance criteria post-reprocessing) is established through rigorous validation protocols that demonstrate the ability to consistently clean, sterilize, and restore functional performance to predefined engineering specifications for external fixation devices. This involves chemistry, microbiology, and mechanical engineering testing rather than expert-derived ground truth.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1