Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K042967
    Date Cleared
    2004-12-03

    (36 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The A&D LifeSource UB-511 & UB-512 digital blood pressure monitors are intended for use by adults for measuring the systolic and diastolic blood pressure and pulse rate.
    Measure blood pressure (systolic and diastolic) and pulse rate.

    Device Description

    The A&D LifeSource UB-511 & UB-512 digital blood pressure monitors are intended for use by adults for measuring the systolic and diastolic blood pressure and pulse rate. UB-511 & UB-512 use an inflated cuff which is wrapped around the wrist. The cuff is inflated automatically by the air pump. The systolic and diastolic blood pressures are determined by oscillometric method while the cuff is inflated. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for one minute. The blood pressure results are compared with WHO (World Health Organization) BP classifications, which are Sever Hypertension, Moderate Hypertension, Mild Hypertension, High Normal, and Optimal. The corresponding LCD segment will be turned on along with the systolic, diastolic, and pulse rate information. UB-511 & UB-512 measure blood pressure and pulse rate even when an irregular heartbeat occurs. After one minute without operation, UB-511 & UB-512 turns off automatically.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the A&D LifeSource UB-511 & UB-512 Digital Blood Pressure Monitors, based on the provided document:

    Acceptance Criteria and Device Performance

    Acceptance Criteria (Standard / Section)Reported Device Performance
    NIST/AAMI SP-10 Section 4.1.1 (General)Conformed
    NIST/AAMI SP-10 Section 4.1.2.1 (Device labeling)Conformed
    NIST/AAMI SP-10 Section 4.1.2.2 (Outer container)Conformed
    NIST/AAMI SP-10 Section 4.1.3 (Information manual)Conformed
    NIST/AAMI SP-10 Section 4.1.4.1 (Component replacement)Conformed
    NIST/AAMI SP-10 Section 4.1.4.2 (Power system labeling)Conformed
    NIST/AAMI SP-10 Section 4.1.4.3 (Labeling for battery-powered devices)Conformed
    NIST/AAMI SP-10 Section 4.2.1 (Storage conditions)Conformed
    NIST/AAMI SP-10 Section 4.2.2 (Operating conditions)Conformed
    NIST/AAMI SP-10 Section 4.2.3 (Vibration and shock)Conformed
    NIST/AAMI SP-10 Section 4.2.4.1 (Voltage range)Conformed
    NIST/AAMI SP-10 Section 4.2.4.2 (Life)Conformed
    NIST/AAMI SP-10 Section 4.3.1.1 (Maximum cuff pressure)Conformed
    NIST/AAMI SP-10 Section 4.3.1.2 (Cuff deflation)Conformed
    NIST/AAMI SP-10 Section 4.3.2 (Electrical safety)Conformed
    NIST/AAMI SP-10 Section 4.3.3 (Conductive components)Conformed
    NIST/AAMI SP-10 Section 4.4.1 (Pressure indicator accuracy)Conformed
    NIST/AAMI SP-10 Section 4.4.2 (Overall system efficacy)Conformed
    NIST/AAMI SP-10 Section 4.4.2.1 (Auscultatory method as the reference standard)Conformed
    NIST/AAMI SP-10 Section 4.4.2.2 (Intra-arterial method as the reference standard)Not applicable
    NIST/AAMI SP-10 Section 4.4.3 (Battery-powered devices)Conformed
    NIST/AAMI SP-10 Section 4.5 (Requirements for devices with manual inflation systems)Conformed
    FDA guidance "Non-invasive Blood Pressure (NIBP) Monitor Guidance"Conformed

    Study Details

    1. Sample size used for the test set and the data provenance:
      The document states, "UB-511 & UB-512 are not clinically tested." This implies that there was no separate test set used for clinical validation of blood pressure accuracy from human subjects. The device's performance against the AAMI SP-10 and FDA guidance was reported as "Conformed" for various technical and safety aspects, which would involve testing the device itself, not necessarily on a human subject test set.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      Not applicable, as no clinical test set on human subjects was conducted for this specific device. The device's equivalence was based on its identical software codes and pressure detection hardware to predicate devices.

    3. Adjudication method for the test set:
      Not applicable, as no clinical test set on human subjects was conducted for this specific device.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
      No, an MRMC study was not done. The device was "not clinically tested" and its substantial equivalence was based on technological characteristics and conformity to standards rather than comparative clinical effectiveness with human readers or devices.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
      The document indicates that the UB-511 & UB-512 "use the identical software codes and pressure detection related hardware as the predicate devices to determine systolic, diastolic, and pulse rate." This implies that the algorithm's performance (specifically the oscillometric method) was relied upon based on the predicate devices' established performance. While not explicitly stated as a standalone study for this specific device, the technical conformity relies on the standalone performance of the underlying technology shared with the predicates.

    6. The type of ground truth used:
      For the technical tests (e.g., pressure indicator accuracy, overall system efficacy), the ground truth would have been established by calibrated instruments and reference methods as dictated by the AAMI SP-10 standard (e.g., auscultatory method as a reference standard for overall system efficacy). For the general performance of the blood pressure measurement technology, the established performance of the predicate devices implicitly serves as the ground truth.

    7. The sample size for the training set:
      Not applicable. As the device was not clinically tested, and its equivalence relies on predicate devices, there is no mention of a training set for an algorithm developed specifically for this device. The shared software codes and hardware indicate that any "training" (if applicable to the algorithm's development) would have occurred for the predicate devices.

    8. How the ground truth for the training set was established:
      Not applicable, as no training set for a new algorithm was utilized for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1