Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K030526
    Manufacturer
    Date Cleared
    2004-12-29

    (679 days)

    Product Code
    Regulation Number
    882.1870
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PediGuard is indicated for use during pedicle screw pilot hole drilling to provide feedback to the surgeon via visual and audible alerts that indicate a change in impedance at the tip of the probe and may indicate contact of the tip with soft tissues and possible vertebral cortex perforation. The PediGuard also is specifically indicated for use in intraoperative electromyographic ("EMG") surveillance to assist in the location and evaluation of spinal nerves during surgery of the spine, by administration of low voltage electrical energy to tissues and nerves at the operative site, and EMG monitoring of muscle groups associated with those nerves.

    Device Description

    The PediGuard System is a hand held, battery operated bi-polar probe that is comprised of the following components: (1) a stainless steel shaft; (2) a hollow plastic handle; and (3) an electronic cartridge. Its components are manufactured from stainless steel and plastics according to ASTM and USP requirements.

    AI/ML Overview

    The provided 510(k) summary for the PediGuard™ Nerve Detector System does not contain the detailed information necessary to complete most of the requested sections regarding acceptance criteria and study design.

    Here's a breakdown of what can be inferred and what is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Criteria CategoryAcceptance Criteria (As reported in 510(k))Reported Device Performance (As reported in 510(k))
    Functional EquivalenceNot explicitly stated as quantifiable metrics. Implied: Device performs nerve monitoring function similar to predicate devices."Testing was performed to demonstrate the equivalence of the construct design to currently marketed spinal systems." (No specific results provided)
    SafetyNot explicitly stated as quantifiable metrics. Implied: Device safe for intended use."Its components are manufactured from stainless steel and plastics according to ASTM and USP requirements." (Material safety, not functional safety)
    Impedance FeedbackNot explicitly stated. Implied: Provides visual and audible alerts for impedance changes."provide feedback to the surgeon via visual and audible alerts that indicate a change in impedance at the tip of the probe..." (Function description, not performance metric)
    EMG SurveillanceNot explicitly stated. Implied: Assists in location and evaluation of spinal nerves via EMG monitoring."...assist in the location and evaluation of spinal nerves... and EMG monitoring of muscle groups associated with those nerves." (Function description, not performance metric)

    Missing Information:

    • Specific numerical or statistical acceptance criteria (e.g., sensitivity, specificity, accuracy, signal-to-noise ratio, response time, threshold values for impedance change).
    • Quantifiable results of the device's performance against any established criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    Missing Information:

    • Sample Size: The document does not mention any specific test set, number of cases, or even the type of study (e.g., animal, cadaver, human).
    • Data Provenance: Not mentioned. It's unclear if any data used for "testing" was retrospective or prospective, or the country of origin.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Missing Information:

    • No mention of a "test set" in the context of expert review or ground truth establishment. Therefore, no information on the number or qualifications of experts.

    4. Adjudication Method for the Test Set

    Missing Information:

    • Given the absence of a described test set with expert review, there is no information on any adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Missing Information:

    • The document does not indicate that an MRMC comparative effectiveness study was performed. It focuses on substantial equivalence to predicate devices, but not on human reader improvement with or without AI assistance. This device is a nerve stimulator, not an AI-powered diagnostic tool, so such a study would not typically be applicable.

    6. Standalone Performance Study (Algorithm Only)

    Missing Information:

    • This device is a hardware probe with electronics for impedance sensing and electrical stimulation. It is not an "algorithm-only" device that would typically have a standalone performance study in the AI/software context. The "testing" mentioned is likely related to its functional performance and material equivalence, not an algorithm's diagnostic accuracy. No specific standalone performance study results are provided.

    7. Type of Ground Truth Used

    Missing Information:

    • No specific ground truth establishment method (e.g., pathology, outcomes data, expert consensus) is described, as the detailed testing methodology is absent. The "testing for equivalence" likely relied on comparing electrical outputs or functional behaviors to predicate devices, rather than a diagnostic ground truth.

    8. Sample Size for the Training Set

    Missing Information:

    • This device is not described as an AI/machine learning device that would typically have a "training set." Therefore, no information on training set sample size is provided.

    9. How the Ground Truth for the Training Set Was Established

    Missing Information:

    • As no training set is mentioned (see point 8), there is no information on how its ground truth was established.

    Summary of what can be gleaned from the document:

    The 510(k) summary focuses on establishing substantial equivalence to predicate devices based on:

    • Similar design and materials: "components of the PediGuard Nerve Monitoring System are similar in design, material, and intended use to other nerve monitoring systems..."
    • Intended use: The indications for use are similar to predicate devices.
    • Functional testing for equivalence: "Testing was performed to demonstrate the equivalence of the construct design to currently marketed spinal systems." (No details about this testing are provided).
    • Compliance with material standards: "Its components are manufactured from stainless steel and plastics according to ASTM and USP requirements."

    The provided text is a high-level summary for regulatory filing, not a detailed scientific study report. It demonstrates that the device's fundamental design, materials, and intended functions are comparable to existing cleared devices, which is the cornerstone for 510(k) clearance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1