Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K022764
    Manufacturer
    Date Cleared
    2002-08-28

    (7 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medtronic® Launcher Guiding Catheter is designed to provide a pathway through which therapeutic devices are introduced. The Launcher catheter is intended to be used in the coronary or peripheral vascular system.

    Device Description

    The Medtronic Launcher Guide Catheter is constructed with an inner liner, stainless steel braid, outer shaft jacket, sleeve, marker band and a soft distal tip. The inner lumen of the Launcher Guide Catheter has a thin lubricious coating.

    AI/ML Overview

    The provided document, K022764, describes the Medtronic 7F Launcher Guide Catheter, its intended use, and its substantial equivalence to a predicate device (Medtronic® 6F Z3 Guide Catheter - K021256). The primary method for demonstrating safety and effectiveness relies on a comparison to the predicate device and the results of in-vitro functional and biocompatibility testing. The document does not describe a study involving human subjects or AI performance analysis.

    Here's an analysis of the provided information based on your requested criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Biocompatibility:
    Acute Intracutaneous ReactivityPassed (tests successfully performed)
    Acute Systemic ToxicityPassed (tests successfully performed)
    CytotoxicityPassed (tests successfully performed)
    HemolysisPassed (tests successfully performed)
    SensitizationTest in process at the time of submission (implies expectation of passing)
    Functional Testing:
    Hub to shaft tensileMeets applicable specifications
    Shaft tensileMeets applicable specifications
    Distal tip to shaft tensileMeets applicable specifications
    Torque responseMeets applicable specifications
    Lumen lubricityMeets applicable specifications
    RadiopacityMeets applicable specifications
    Pressure shaft burstMeets applicable specifications
    Hub leakMeets applicable specifications
    Tip compression (softness)Meets applicable specifications

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not detail specific sample sizes for each functional or biocompatibility test beyond stating "tests successfully performed." The data provenance is derived from in-vitro laboratory testing of the Medtronic 7F Launcher Guide Catheter's materials and functional components. There is no indication of country of origin for the data or whether it was retrospective or prospective in the context of human data. The testing described is prospective in nature as it evaluates the new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This submission relies on in-vitro testing against predetermined specifications, not on expert-established ground truth from human data. The "experts" would be the scientists and engineers conducting the tests and interpreting results against established ISO 10993 standards for biocompatibility and internal engineering specifications for functional testing. Their qualifications are not specified but are implied to be appropriate for conducting such tests.

    4. Adjudication Method for the Test Set

    Not applicable. There was no Adjudication method as the study was not based on human readings or interpretations that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No MRMC comparative effectiveness study was conducted. This submission is for a medical device (catheter), not an AI diagnostic or assistive tool. Therefore, there is no mention of human readers, AI assistance, or effect size improvements related to AI.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No standalone algorithm performance study was done. This document concerns a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on:

    • Standards Compliance: Biocompatibility testing against ISO 10993 standards.
    • Engineering Specifications: Functional testing against internal design and performance specifications for parameters like tensile strength, lubricity, radiopacity, and burst pressure.
    • Predicate Device Equivalence: The overall safety and effectiveness are established through substantial equivalence to the legally marketed Medtronic® 6F Z3 Guide Catheter (K021256), implying that if the new device meets the same functional and material requirements, it will perform similarly to a device already deemed safe and effective.

    8. The Sample Size for the Training Set

    No training set was used. This is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1