Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K243677
    Device Name
    iovera° System
    Date Cleared
    2024-12-26

    (29 days)

    Product Code
    Regulation Number
    882.4250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    iovera° System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iovera® System is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The iovera System is not indicated for treatment of central nervous system tissue.

    When stimulation compatible components are used, the iovera System can also facilitate target nerve location by conducting electrical nerve stimulation from a compatible 3rd party nerve stimulator.

    Device Description

    The iovera System is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through application of extreme cold to the selected site. The device is based on introduction of a Smart Tip internally cooled by the cryogenic fluid (nitrous oxide, NoO) to a selected area. The Smart Tip is cooled by the Joule-Thomson Effect and/or latent heat of vaporization. The iovera8 System may be used in conjunction with standard of-the-shelf nerve stimulator device in applications where precise nerve location is desired.

    The device is comprised of four main components:

    • A reusable Handpiece .
    • A Charging Dock ●
    • An assortment of single-patient-use Smart Tips ●
    • A Cartridge containing nitrous oxide ●
    AI/ML Overview

    The provided text describes the iovera® System, a cryogenic surgical device cleared by the FDA (K243677). The document focuses on demonstrating substantial equivalence to predicate devices (K220656 and K211334) for a new Smart Tip configuration (STT21180STIM) that also facilitates nerve location.

    Here's a breakdown of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily relies on adherence to recognized standards and satisfactory completion of design verification testing. Specific quantitative acceptance criteria for each test (e.g., specific thresholds for biocompatibility, precise tolerances for mechanical tests) are not explicitly detailed in this summary, but the general outcome of "meets all ISO10993-1 and USP acceptance criteria" or "complied with standards" is stated.

    Acceptance Criterion CategoryReported Device Performance (Summary)
    BiocompatibilityConformed to ISO 10993-1:2022. No signs of acute systemic toxicity and pyrogenic response, sensitization, or skin irritation. All extractables below Analytical Evaluation Threshold (AET), no unexpected contaminants or material degradation. Negligible toxicological risk with a significant margin of safety.
    Electrical Safety & EMCComplied with IEC 60601-1:2005+AMD1:2012+AMD2:2020 Edition 3.2, IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 Edition 3.2, and IEC 60601-1-8:2006+AMD1:2012+AMD2:2020 Edition 2.2 for safety (e.g., electric shock), usability, and alarms. Met all EMC requirements per IEC TS 60601-4-2:2024 (specific clauses listed).
    Software V&VDocumentation provided per FDA guidance. Software classified as "Class B" concern. Followed IEC 62304:2006/AMD1:2015 and relevant FDA guidance documents for software development, risk management, software life cycle processes, and cybersecurity. Specific performance metrics are not provided but the implication is that all V&V activities were completed satisfactorily.
    Functional & Product Perf.Testing performed on Handpiece, stim cable, cartridges, and Smart Tip STT21180STIM. Indicated that the device meets product specifications and is safe and effective. Specific performance metrics (e.g., temperature ranges, cooling times) are listed in the comparison table for different Smart Tips, and implied to be met for the new STT21180STIM tip. For example, cooling time for STT21180STIM is 70 seconds. Post-heat time is 15 seconds. Skin Warmer Set Point is 15°C.
    Usability Engineering/UIConformed to IEC 62366-1 Ed. 1.1. Assessed usability by identifying hazards and hazardous situations associated with the user interface, ensuring safety features are intuitive and support proper usage. Simulated use testing (Human Factors Study) with the Smart Tip and Stim cable assembly attached was performed.
    Mechanical & ResistanceResistance Testing (difficulty in securing connection between needle and PCBA), Flex Introducer and Leak Test (to ensure needle does not break), Heater Block-to-Needle Bond Strength Test. All tests were conducted successfully, indicating mechanical strength and reliability. Specific quantitative acceptance criteria are not mentioned beyond "ensuring mechanical strength and reliability."
    Packaging EvaluationConformed to ASTM D4332-22 and ASTM D4169. Package Integrity Peel Strength Testing, Inspection, and bubble test performed. Validated packaging integrity and compliance with ISO 11607 requirements for sterile barrier systems. ASTM F1886/F1886M-16 and ASTM F88/F88M-23 are referenced.
    Sterility AssuranceConformed to ASTM F1980-21 and ISO 11135:2014. Sterilization validation performed (EO). Smart Tip individually packaged and sterile with a SAL of 10-6. Ensured compliance. Also referenced EN ISO 10993-7:2008+A1:2022 (Ethylene oxide sterilization residuals) and BS EN ISO 11737-1:2018+A1:2021 (Microbiological methods-Determination of a population of microorganisms on products).

    2. Sample Size Used for the Test Set and Data Provenance:

    The document summarizes design verification testing rather than a clinical study with a "test set" of patient data. Therefore, information about sample size for a test set and data provenance (country of origin, retrospective/prospective) is not applicable or provided in this regulatory summary. The testing focuses on engineering verification (e.g., number of units tested for mechanical properties, number of samples for biocompatibility, etc.), not patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    Again, this pertains to a clinical study with patient data and does not apply to the design verification testing described. The "ground truth" for the engineering tests would be the compliance with recognized standards and specified performance parameters.

    4. Adjudication Method for the Test Set:

    Not applicable, as this is design verification testing rather than a clinical study requiring consensus from experts on case interpretation.

    5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study was not done. The submission focuses on demonstrating substantial equivalence through non-clinical performance testing and comparison to predicate devices, not on evaluating human reader performance with or without AI assistance. The device in question is a physical cryogenic surgical device, not an AI diagnostic tool that assists human readers with interpretations.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done:

    Not applicable. The iovera® System is a physical medical device, not an algorithm. Performance is assessed through engineering tests of the device itself. While it facilitates nerve location, its primary function is direct tissue destruction via freezing, with a human operator always in the loop.

    7. The Type of Ground Truth Used:

    For the design verification testing, the "ground truth" used is:

    • Compliance with recognized international and national standards (e.g., ISO 10993-1, IEC 60601-1, IEC 62366-1, ISO 11135, ASTM standards).
    • Pre-defined product specifications and design input requirements for physical, electrical, mechanical, software, and biological properties.
    • Laboratory test results (e.g., chemical analysis for extractables, temperature measurements for cooling performance, force measurements for bond strength).

    8. The Sample Size for the Training Set:

    Not applicable. This is not a machine learning or AI device that requires a training set of data.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set for this type of medical device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220656
    Device Name
    iovera System
    Date Cleared
    2022-05-20

    (74 days)

    Product Code
    Regulation Number
    882.4250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    iovera System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iovera System is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The iovera" System is not indicated for treatment of central nervous system tissue.

    When stimulation compatible components are used , the iovera System can also facilitate target nerve location by conducting electrical nerve stimulation from a compatible 3rd party nerve stimulator.

    Device Description

    The iovera? System is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through application of extreme cold to the selected site. The device is based on introduction of a Smart Tip internally cooled by the cryogenic fluid (nitrous oxide, N2O) to a selected area. The Smart Tip is cooled by the Joule-Thomson Effect and/or latent heat of vaporization. The iovera" System may be used in conjunction with standard of-the-shelf nerve stimulator device in applications where precise nerve location is desired.

    The device comprises four main components:

    • A reusable Handpiece
    • A Charging Dock
    • An assortment of single-patient-use Smart Tips
    • A Cartridge containing nitrous oxide

    The iovera® System Handpiece is battery powered and provides feedback to the user during device preparation and use. The Handpiece connects to both the Cartridge and to the Smart Tip. The user activates a treatment cycle through a control on the Handpiece, which starts and stops the treatment. The Handpiece contains an LCD display for providing feedback to the user when the device is ready to use. The Charging Dock stores the Handpiece between uses and provides power for charging the battery.

    An assortment of Smart Tips is available for the ioveraº system. All Smart Tip needles are made of stainless steel and have a closed-end design that fully contains the cryogen so that it does not enter the target tissue. The Smart Tip is the only patient contacting component of the iovera® System. The user removes the Smart Tip from its sterile packaging and attaches it to the Handpiece immediately before use. Certain Smart Tip designs provide a connection to an external nerve stimulator to facilitate nerve location prior to treatment.

    The iovera System uses a commercially available nitrous oxide cylinder. The Cartridge is filled with pure N20.

    AI/ML Overview

    This document is a 510(k) premarket notification for the iovera® System, a cryogenic surgical device. The purpose of this notification is to demonstrate that the modified iovera® System is substantially equivalent to previously cleared iovera® Systems (K211334 and K173763).

    Based on the provided text, the device is evaluated against its product specifications. The document describes a series of engineering and performance tests rather than a clinical study evaluating diagnostic or prognostic performance with human readers. Therefore, several of the specific questions related to clinical studies (e.g., sample size for training/test sets, expert ground truth, MRMC studies) are not directly applicable or are not detailed in this submission.

    Here's a breakdown of the acceptance criteria and the study (design verification testing) that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states that "The subject iovera® System device met the design verification and validation inputs, passing all predetermined acceptance criteria." However, it does not provide a specific table with numerical acceptance criteria and corresponding reported performance values for each parameter. Instead, it lists categories of testing performed to demonstrate compliance.

    Category of TestingAcceptance Criteria (Implicit)Reported Device Performance
    Functional & Product PerformanceDevice operates according to design specifications for its intended use (e.g., proper cooling, cryogen flow).Met product specifications for intended clinical use. Passed all predetermined acceptance criteria.
    Smart Tip Functional TestingSmart Tips perform as designed (e.g., proper cooling, nerve stimulation capability for compatible tips).Met product specifications. Passed all predetermined acceptance criteria.
    User Interface Testing (IEC 62366-1)User interface is safe and effective, and meets usability standards.Conforms to IEC 62366-1:2015. Passed all predetermined acceptance criteria.
    Software Testing (IEC 62304)Software functions correctly and safely.Conforms to IEC 62304:2006/A. Passed all predetermined acceptance criteria.
    Electrical & EMC Safety (IEC 60601-1, -1-2)Device meets electrical safety and electromagnetic compatibility standards.Conforms to IEC-60601-1: A1 2012 and 60601-1-2:2014. Passed all predetermined acceptance criteria.
    Mechanical & Thermal Safety (IEC 60601-1)Device meets mechanical and thermal safety standards.Conforms to IEC 60601-1: A1:2012. Passed all predetermined acceptance criteria.
    Sterility Testing (ASTM F1980-16, ISO 11135:2014/A)Smart Tips achieve and maintain required sterility assurance level (SAL).Conforms to ASTM F1980-16 and ISO 11135:2014/A. Smart Tip individually packaged and sterile with a SAL of 10-6. Passed all predetermined acceptance criteria.
    Packaging Testing (ASTM D4332-14)Packaging protects the device and maintains sterility.Conforms to ASTM D4332-14. Passed all predetermined acceptance criteria.
    Biocompatibility Testing (ISO 10993-1)Patient-contacting materials are biocompatible.Conforms to ISO 10993-1. All patient contacting materials tested to ISO 10993-1; found biocompatible. Passed all predetermined acceptance criteria.
    Clinical Effect (Cryoneurolysis)Device achieves cellular death through cryoneurolysis and Wallerian degeneration, leading to second-degree nerve injury (axonotmesis).Stated as the Clinical Effect of the device. (This is a principle of operation rather than an acceptance criterion met by a specific test in this submission).
    Treatment TemperatureAchieves a temperature between -55°C to -75°C (-67°F to -103°F) during treatment.Stated as the achieved treatment temperature range. (This is a performance characteristic).
    Timing Parameters (Pre-heat, Cooling, Post-heat)Meets specified pre-heat, cooling, and post-heat durations for different Smart Tips.12 seconds (8.5mm), 1 second (90mm) pre-heat; 33 seconds (8.5mm), 60 seconds (90mm) max cooling; 20 seconds (8.5mm), 45 seconds (90mm) post-heat.
    Skin Warmer Set PointMaintains Skin Warmer Set Point at 30°C.30°C.

    2. Sample size used for the test set and the data provenance:

    The document describes "Design Verification testing" and "Verification and validation testing." These are engineering and performance validation tests, not clinical efficacy or diagnostic performance studies involving patient data. Therefore, the concept of a "test set" in the context of a dataset with patient outcomes or images is not applicable here. The "samples" would be units of the device or its components tested in a lab setting. The document does not specify the number of device units or components tested for each category.

    • Data Provenance: Not applicable in the context of patient data. The tests are likely conducted in-house or by accredited testing laboratories.
    • Retrospective or Prospective: Not applicable as it's not a clinical study involving patient data. These are design verification tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. The ground truth for these engineering performance tests is established by product specifications, recognized standards (e.g., ISO, IEC, ASTM), and internal design requirements, not by expert consensus on clinical data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. This is for performance testing against established engineering specifications, not for interpretation of clinical data requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This submission is for a medical device (cryogenic surgical system), not an AI/CADe or CADx device that assists human readers in interpreting medical images.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. The iovera® system is a physical device used by a human practitioner. It's not a standalone algorithm.

    7. The type of ground truth used:

    The ground truth for this device's performance is based on:

    • Engineering Specifications: Pre-defined design requirements and performance parameters for the device and its components (e.g., temperature range, timing, electrical properties).
    • Recognized Standards: Compliance with international and national standards (IEC, ISO, ASTM) for medical devices, software, electrical safety, biocompatibility, and sterility.
    • Predicate Device Equivalence: The claim of substantial equivalence to previously cleared iovera® Systems (K211334 and K173763) implies that their performance characteristics serve as a benchmark.

    8. The sample size for the training set:

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. As above, this is not an AI/machine learning device, so there is no "training set." The performance verification relies on meeting established design specifications and adhering to relevant industry standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211334
    Device Name
    Iovera System
    Date Cleared
    2021-09-10

    (130 days)

    Product Code
    Regulation Number
    882.4250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Iovera System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iovera? system is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The ioveraº system is not indicated for treatment of central nervous system tissue.

    Device Description

    The iovera° system is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through application of extreme cold to the selected site. The device is based on introduction of a Smart Tip internally cooled by the cryogenic fluid (nitrous oxide, N₂O) to a selected area. The Smart Tip is cooled by the Joule-Thomson Effect and/or latent heat of vaporization. The device comprises four main components: A reusable Handpiece, A Charging Dock, An assortment of single-patient-use Smart Tips, A Cartridge containing nitrous oxide.

    AI/ML Overview

    The provided document K211334 is a 510(k) Premarket Notification for the Iovera System. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, not for proving significant clinical efficacy or performing novel clinical trials to establish acceptance criteria for a new device's performance.

    Therefore, the document does not contain the detailed information necessary to answer all parts of your request, especially regarding acceptance criteria for a new device's analytical performance, a groundbreaking study proving its independent performance, reader study details (MRMC), or specifics about training and test sets.

    Here's an analysis of what can and cannot be extracted from the provided text:

    Key Takeaway from the Document:
    The Iovera System (subject device) is being cleared as substantially equivalent to a previously cleared Iovera System (predicate device, K161835). This means the focus of the submission is to show that any modifications to the device since its last clearance do not raise new questions of safety or effectiveness and that the core performance remains the same as the already approved predicate. It does not describe a new study to establish performance metrics against a novel set of acceptance criteria for a first-of-its-kind device.

    Information Extracted/Deduced from the Document:

    1. A table of acceptance criteria and the reported device performance:
      The document states:
      "Design Verification testing was performed for the iovera° system to demonstrate that the device meets product specifications."
      "The subject ioveraº system device met the design verification and validation inputs, passing all predetermined acceptance criteria."

      However, the document does not provide a specific table of acceptance criteria with numerical targets or the reported device performance outcomes for these criteria. It lists the categories of testing performed:

      • Functional Testing
      • Product Performance (Handpiece and Cartridge)
      • Smart Tip Testing
      • User Interface Testing
      • Electrical /EMC safety Testing

      Given this is a 510(k) for substantial equivalence based on modifications, the "acceptance criteria" likely refer to internal engineering and design verification specifications for the modified components to ensure they perform as intended and don't negatively impact the device's established safety and effectiveness. These are typically pass/fail for specific tests rather than statistical performance metrics like sensitivity/specificity for an AI or diagnostic algorithm.

    2. Sample sizes used for the test set and the data provenance:
      The document does not provide information on sample sizes for any "test sets" in the context of clinical performance or algorithm evaluation. The testing described is design verification and validation, which often involves engineering tests on a defined number of units, but these are not typically referred to as "test sets" in the context of clinical data for performance evaluation.
      Data provenance (country of origin, retrospective/prospective) is also not mentioned as this type of clinical data is not the focus of this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      This information is not applicable and not provided. The device is a cryogenic surgical device, not a diagnostic or AI device that relies on expert interpretation to establish ground truth for performance evaluation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      This information is not applicable and not provided for the same reasons as point 3.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      This information is not applicable and not provided. The Iovera System is a therapeutic device (cryogenic surgical device), not a diagnostic device or AI software that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      This information is not applicable and not provided. The Iovera System is a physical device used by a clinician, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      As this is a physical therapeutic device, the concept of "ground truth" for performance evaluation in the context of diagnostic accuracy does not apply in the way it would for an AI or diagnostic device. The performance is assessed through meeting engineering specifications and demonstrating safe and effective tissue destruction/lesion creation as intended.

    8. The sample size for the training set:
      This information is not applicable and not provided. The device is not an AI model that requires a "training set."

    9. How the ground truth for the training set was established:
      This information is not applicable and not provided for the same reasons as point 8.

    In summary, the provided document relates to a 510(k) submission for a physical medical device (cryogenic surgical device) that is being cleared based on substantial equivalence to a predicate device after modifications. It does not contain the type of performance evaluation study details (e.g., test sets, ground truth methodology, reader studies) that would be relevant for an AI/diagnostic device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173763
    Device Name
    iovera system
    Manufacturer
    Date Cleared
    2018-02-28

    (79 days)

    Product Code
    Regulation Number
    882.4250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    iovera system

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iovera? system is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The iovera" system is not indicated for treatment of central nervous system tissue.

    The ioveraº system's "1x90" Smart Tip configuration (indicating one needle which is 90 mm long) can also facilitate target nerve location by conducting electrical nerve stimulation from a separate nerve stimulator.

    Device Description

    The iovera system is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue creating a nerve block through application of extreme cold to the selected site. The device is based on introduction of a Smart Tip internally cooled by the cryogenic fluid (nitrous oxide, N2O) to a selected area. The Smart Tip is cooled by the Joule-Thomson Effect and/or Latent Heat of Vaporization. The iovera® system may be used in conjunction with a standard off-the-shelf nerve stimulator device in applications where precise nerve location is desired.

    The device is comprised of four main components:

    1. A reusable Handpiece
    2. A Charging Dock
    3. An assortment of single-patient use Smart Tips
    4. A Cartridge (Nitrous Oxide)
    AI/ML Overview

    Based on the provided text, the document is a 510(k) Premarket Notification for the Myoscience iovera® system, primarily focused on an extended Smart Tip configuration. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" refer to the performance data submitted to demonstrate substantial equivalence to predicate devices, rather than a clinical outcome study in the traditional sense of proving clinical efficacy for a new indication.

    Here's an breakdown of the requested information based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this submission are primarily focused on demonstrating the safety and performance of the new Smart Tip configuration (1x90mm) as being substantially equivalent to existing predicate devices, particularly regarding its physical characteristics and operational performance within the iovera® system. The acceptance criteria are essentially implied by the "PASS" results for various tests.

    Acceptance Criteria CategorySpecific Test PerformedReported Device Performance
    BiocompatibilityCytotoxicityPASS
    SensitizationPASS
    Intracutaneous ReactivityPASS
    Acute System ToxicityPASS
    Material-Mediated PyrogenicityPASS
    Software FunctionalitySoftware verification for new Smart Tip descriptorPASS
    Bench/Hardware PerformanceElectrical Safety Testing per IEC 60601-1PASS
    EMC/Immunity Testing per IEC 60601-1-2PASS
    Visual and Dimensional Inspection of Smart Tip needlePASS
    Verification of temperature reproducibilityPASS
    Cryozone TestingPASS
    Needle IntegrityPASS
    Tip Descriptor verification to confirm treatment parametersPASS

    Study Details

    1. Sample sizes used for the test set and the data provenance:

      • The document does not specify numerical sample sizes for each of the performance tests (e.g., number of Smart Tips tested for dimensional inspection, or number of cycles for temperature reproducibility).
      • Data Provenance: The tests were conducted by Myoscience, Inc. as part of their premarket notification for the device. The document does not specify a country of origin for the data or whether it was retrospective or prospective, but given it's a premarket submission, it would be prospective testing conducted for the purpose of the submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • For this type of device and submission (a 510(k) for a modification to an existing device, emphasizing engineering and material performance), "ground truth" is established through standardized engineering tests, material testing protocols (like ISO-10993), and software verification, rather than clinical expert consensus on complex diagnostic images. Therefore, the concept of "experts establishing ground truth" in the clinical sense is not directly applicable. The "experts" would be qualified engineers, test technicians, and quality assurance personnel performing the described tests according to recognized standards.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Adjudication methods like 2+1 or 3+1 are typical for clinical studies involving multiple human readers assessing medical images or outcomes. This document details engineering and biocompatibility testing, not a clinical reader study. Therefore, no such adjudication method was used or is relevant to the data presented.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not performed. This device is a cryogenic surgical system, not an AI-powered diagnostic or assistive tool for human readers. The concept of "human readers improve with AI vs without AI assistance" is not relevant to this device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No, this term typically applies to AI algorithms. The "software testing" mentioned is for the device's embedded software (e.g., for controlling treatment parameters), not a standalone AI algorithm for medical image analysis or diagnosis.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the performance data provided, the "ground truth" is based on:
        • Standardized Test Specifications: For electrical safety, EMC, dimensional accuracy, temperature reproducibility, cryozone formation, and needle integrity, the ground truth is defined by pre-established engineering specifications and international standards (e.g., IEC 60601-1, ISO-10993).
        • Chemical/Biological Standards: For biocompatibility, the ground truth is determined by the results of specific standardized biological tests (Cytotoxicity, Sensitization, etc.) against accepted pass/fail criteria from ISO-10993.
        • Software Design Requirements: For software verification, the ground truth is the confirmed meeting of the software design requirements for the new Smart Tip descriptor.
    7. The sample size for the training set:

      • This document describes performance testing for a 510(k) submission, not an AI model development. Therefore, there is no "training set" as understood in machine learning. The device itself (the iovera® system) has undergone extensive development and testing, but the term "training set" is not applicable here.
    8. How the ground truth for the training set was established:

      • As there is no "training set" in the context of this 510(k) submission, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161835
    Device Name
    iovera system
    Manufacturer
    Date Cleared
    2017-03-24

    (262 days)

    Product Code
    Regulation Number
    882.4250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    iovera system

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iovera? system is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The iovera" system is not indicated for treatment of central nervous system tissue.

    Device Description

    The ioveras system is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue creating a nerve block through application of extreme cold to the selected site. The device is based on introduction of a Smart Tip internally cooled by the cryogenic fluid (nitrous oxide, N2O) to a selected area. The Smart Tip is cooled by the Joule-Thomson Effect and/or Latent Heat of Vaporization. The ioveraº system may be used in conjunction with a standard off-the-shelf nerve stimulator device in applications where precise nerve location is desired.

    The device is comprised of four main components:

    1. A reusable Handpiece
    2. A Charging Dock
    3. An assortment of single-patient use Smart Tips
    4. A Cartridge (Nitrous Oxide)

    The iovera Handpiece is battery powered and provides feedback to the user during device preparation and use. The Handpiece connects to both the Cartridge and to the Smart Tip. The user activates a treatment cycle through a control on the Handpiece, which starts and stops the treatment. The Handpiece also contains LEDs for providing feedback to the user when the device is ready to use. The Charging Dock stores the Handpiece between uses and provides power for charging the battery.

    An assortment of Smart Tips is available for the iovera system. All Smart Tip needles are made of stainless steel and have a closed-end that fully contains the cryogen so that it does not enter the target tissue. The Smart Tip is the only patient contacting component of the iovera° system. The user removes the Smart Tip from the sterile packaging and attaches it to the Handpiece.

    The ioveras system uses a commercially available nitrous oxide cylinder. The Cartridge is filled with pure N2O.

    AI/ML Overview

    This document describes the FDA 510(k) clearance for the ioveraº system, a cryogenic surgical device, for an expanded indication related to the relief of pain and symptoms associated with osteoarthritis of the knee.

    1. Table of Acceptance Criteria (Primary and Secondary Endpoints) and Reported Device Performance

    Acceptance Criteria (Primary/Secondary Endpoints)Reported Device Performance (ioveraº treatment group vs. Sham control group)
    Primary Endpoint
    Superiority of ioveraº treatment over sham treatment for reducing pain and symptoms due to osteoarthritis in the knee as assessed by the Total WOMAC scale from baseline to Day 30 (p
    Ask a Question

    Ask a specific question about this device

    K Number
    K142866
    Manufacturer
    Date Cleared
    2015-01-21

    (112 days)

    Product Code
    Regulation Number
    882.4250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Myoscience iovera system

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The myoscience ioveraº system is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the blocking of pain. The ioveraº system is not indicated for treatment of central nervous system tissue.

    Device Description

    The iovera° system is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through application of extreme cold to the selected site. The device is based on introduction of a Smart Tip internally cooled by the cryogenic fluid (nitrous oxide, N₂O) to a selected area. The Smart Tip is cooled by the Joule-Thomson Effect and/or Latent Heat of Vaporization. iovera° may be used in conjunction with a standard off-the-shelf nerve stimulator device in applications where precise nerve location is desired.

    The device is comprised of four main components:

    1. A reusable Handpiece
    2. A Charging Dock
    3. An assortment of single-patient use Smart Tips
    4. A Cartridge (Nitrous Oxide)
    AI/ML Overview

    The provided text is a 510(k) Summary for the Myoscience iovera® system, a cryogenic surgical device. It focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and study data for a novel device or AI algorithm.

    Therefore, much of the requested information regarding acceptance criteria for an AI device, sample sizes for test/training sets, expert ground truth adjudication, MRMC studies, and standalone performance is not present in this document.

    However, I can extract the information that is available:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes several bench tests and software tests with a pass/fail outcome. It doesn't explicitly state quantitative acceptance criteria beyond "PASS," but it implies meeting design requirements and pertinent standards.

    Test PerformedAcceptance Criteria (Implied)Reported Device Performance
    Tip Descriptor verification to confirm treatment parameters (Software testing)Treatment parameters confirmed correctlyPASS
    Visual and dimensional inspection of Smart Tip needle (Bench testing)Met design specificationsPASS
    Verification of temperature reproducibility (Bench testing)Consistent and reproducible temperature (within specified range)PASS
    Validation of cryozone size (Bench testing)Cryozone formed within specified parametersPASS
    Validation of needle integrity in simulated use conditions (Bench testing)Maintained integrityPASS
    After flexing, needle shall return to straight condition (Bench testing)Needle returns to straight conditionPASS
    Needle shall not leak after kink failure (Bench testing)No leakage after kink failurePASS
    Sterility Testing (Bench testing)Compliant with sterility standards (e.g., ISO 11135-1)PASS
    Transit/Shelf Life Testing (Bench testing)Maintained performance over transit/shelf lifePASS

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not specified in the document. The tests performed are described but not the number of units or instances tested for each.
    • Data Provenance: Not applicable in the context of this device's testing. The tests are primarily bench and software tests of the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is not an AI/diagnostic device requiring expert ground truth for image interpretation or diagnosis. The ground truth for device performance is based on direct measurement and adherence to engineering specifications and standards.

    4. Adjudication method for the test set:

    • Not applicable. The tests are objective measurements or functional verifications, not subjective expert reviews.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI algorithm. The device performance tests mentioned are standalone tests of the physical device and its software.

    7. The type of ground truth used:

    • The ground truth for the device's performance tests appears to be based on:
      • Engineering specifications and design requirements: For dimensional integrity, temperature reproducibility, cryozone size, and needle integrity.
      • International standards: Such as ISO 11135-1 for sterility and ISO-10993 for biocompatibility.
      • FDA guidance documents: For software verification testing.

    8. The sample size for the training set:

    • Not applicable. This is a physical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set for an AI model.

    Summary based on the Provided Text:

    The provided document is a 510(k) summary for a medical device (Myoscience iovera® system) seeking substantial equivalence to a predicate device. It focuses on the safety and effectiveness of the device itself through bench testing, software verification, and biocompatibility evaluations, especially for the modified "Smart Tip" component. It explicitly states that "No preclinical testing was deemed necessary for this modification" and "No clinical testing was deemed necessary for this modification," indicating that the decision for substantial equivalence was based on non-clinical data. The document does not describe the evaluation of an AI algorithm or diagnostic performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1