Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K212797
    Manufacturer
    Date Cleared
    2022-01-06

    (126 days)

    Product Code
    Regulation Number
    880.5725
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    iPrime Viscodelivery System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iPRIMETM Viscodelivery System is a manually operated device for delivery of small amounts of viscoelastic fluid, for example HEALON® PRO from Johnson & Johnson Vision, Amvisc@ from Bausch & Lomb, or PROVISC® from Alcon, during ophthalmic surgery.

    Device Description

    The iPRIMET™ Viscodelivery System is a sterile, single-use ophthalmic surgical instrument for dispensing cohesive viscoelastic fluid (supplied separately, at point of use) during ophthalmic surgery. The iPRIMET™ Viscodelivery System is a delivery device for delivering viscoelastic fluid. The procedure is performed by a trained ophthalmic professional in a sterile surgical setting. The iPRIME device is filled by the user, at the point of use, with FDA approved commercially available cohesive viscoelastic fluid (e.g. HEALON® PRO, Amvisc® or PROVISC®; sold and supplied separately). The iPRIME device consists of a handpiece which includes a reservoir, dispense trigger, cannula, slide button, rotatable hub, and microcatheter manufactured from medical grade materials. The slide button allows the user to adjust the microcatheter length. The dispense trigger dispenses the viscoelastic fluid. The rotatable hub allows the user to adjust the angle of the cannula in order to dispense viscoelastic fluid into other areas within the anterior chamber. The iPRIMET™ Viscodelivery System serves as dispensary means to deliver cohesive viscoelastic fluid. The OVD device containing the viscoelastic fluid is connected to the iPRIME luer fitting and viscoelastic is pumped into the iPRIME device. After the device is fully primed and the microcatheter has been extended to the desired location, the user advances forward the dispense trigger delivering a small amount of viscoelastic fluid into the desired location within the anterior chamber of the eye.

    AI/ML Overview

    The iPRIME™ Viscodelivery System was subjected to various non-clinical tests to demonstrate its performance and substantial equivalence to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategorySpecific TestAcceptance CriteriaReported Device Performance
    Functional & PerformanceJoint Strength TestingAble to perform according to intended use and predicate device characteristicsMet all acceptance criteria and performs as intended
    Microcatheter and Cannula Extension/RetractionAble to perform according to intended use and predicate device characteristicsMet all acceptance criteria and performs as intended
    Priming/Dispense Volume TestingAble to perform according to intended use and predicate device characteristicsMet all acceptance criteria and performs as intended
    Corrosion TestingAble to perform according to intended use and predicate device characteristicsMet all acceptance criteria and performs as intended
    Human Factors Engineering EvaluationAble to perform according to intended use and predicate device characteristicsMet all acceptance criteria and performs as intended
    BiocompatibilityCytotoxicity (ISO 10993-1: 2018)Components contacting directly or indirectly with patient are biocompatibleDemonstrated biocompatibility
    Sensitization (ISO 10993-1: 2018)Components contacting directly or indirectly with patient are biocompatibleDemonstrated biocompatibility
    Intracutaneous Irritation (ISO 10993-1: 2018)Components contacting directly or indirectly with patient are biocompatableDemonstrated biocompatibility
    Systemic Toxicity (ISO 10993-1: 2018)Components contacting directly or indirectly with patient are biocompatibleDemonstrated biocompatibility
    Package IntegrityPackaging Visual InspectionSterile barrier maintained throughout the three-month shelf lifeMaintained sterile barrier
    Packaging Peel TestSterile barrier maintained throughout the three-month shelf lifeMaintained sterile barrier
    Packaging Bubble TestSterile barrier maintained throughout the three-month shelf lifeMaintained sterile barrier
    Dispensed VolumeViscoelastic Fluid (per activation)2.7 µL of cohesive viscoelastic fluid per activation of dispense trigger2.7 µL per activation (No total volume limit specified for the proposed device)

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify the exact sample sizes for each non-clinical test. It only states that the device was "subjected to the following functional and performance tests." The data provenance is from non-clinical testing performed on the iPRIME™ Viscodelivery System by the manufacturer, Glaukos Corporation. This is not patient data, so concepts like country of origin or retrospective/prospective do not apply in the same way.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for these non-clinical tests is based on objective measurements and established engineering and biological standards, not expert clinical interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. The non-clinical tests involve objective measurements against predefined acceptance criteria, rather than subjective interpretation requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document explicitly states: "Clinical data are not included in this submission and are not required. Substantial equivalence is based on technological comparison." Therefore, no MRMC comparative effectiveness study was performed.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. The iPRIME™ Viscodelivery System is a medical device, not an algorithm or AI software. Its performance is evaluated through non-clinical functional, performance, and biocompatibility testing.

    7. Type of Ground Truth Used

    The ground truth for the non-clinical tests used objective measurements and adherence to established standards and criteria for device functionality, material biocompatibility (ISO 10993-1: 2018), and package integrity. For the dispensed volume, the ground truth is the measured volume of viscoelastic fluid dispensed per activation.

    8. Sample Size for the Training Set

    Not applicable. This is a physical medical device, not a software algorithm, so there is no "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1