Search Results
Found 2 results
510(k) Data Aggregation
(135 days)
ZYNO MEDICAL ADMINISTRATION SET
Zyno Medical Administration Set is a device used to administer fluids from a container to a patient's vascular system through a needle or a catheter inserted into a vein.
Zyno Medical Administration Set is a device used to administer fluids from a container to a patient's vascular system through a needle or a catheter inserted into a vein. The major modifications are: 1) extended selection of interchangeable components manufactured by additional suppliers; 2) additional contract manufacturer for the device: 3) Gamma Irradiation as additional sterilization method.
The Zyno Medical Administration Set is an intravenous administration set used to deliver fluids from a container to a patient's vascular system.
Here's an analysis of the acceptance criteria and study information provided:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Category) | Reported Device Performance |
---|---|
Safety & Effectiveness | |
Biocompatibility | Successfully met acceptance criteria. |
Pyrogenicity | Successfully met acceptance criteria. |
Chemical Testing | Successfully met acceptance criteria. |
Microbial Ingress | Successfully met acceptance criteria. |
DEHP Testing | Successfully met acceptance criteria. |
Sterilization Testing | Successfully met acceptance criteria. |
Functional Performance | |
Bench Testing (physical & mechanical) | Successfully met acceptance criteria. |
Integration Testing (with Z-800 series infusion pumps) | Successfully met acceptance criteria. |
Aging Testing | Successfully met acceptance criteria. |
Material & Design Standards | |
Tubing Material | Approved Standard PVC (Same as predicate) |
Single Use | Yes (Same as predicate) |
Sterile | Yes (Same as predicate) |
ISO 8536-4:2010 Compliant | Yes (Same as predicate) |
ISO 10993 Compliant | Yes (Same as predicate) |
Operation Mode | For use with Z-800 series of infusion pumps or gravity feed (Same as predicate) |
Conclusion of Nonclinical Testing: "All nonclinical testing results successfully demonstrated that the Zyno Medical Administration Sets performed as intended. Conclusions drawn from the nonclinical testing is that the testing data demonstrated that the device is as safe, as effective, and performs as well as the legally marketed device cleared in K120685."
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes for each of the nonclinical tests (Biocompatibility, Pyrogenicity, Chemical, Microbial Ingress, DEHP, Sterilization, Bench, Integration, and Aging Testing). It only states that "extensive verification and validation testing" was performed and that "All results successfully met acceptance criteria."
The data provenance is not specified regarding country of origin, nor is it explicitly stated whether the data was retrospective or prospective. Given these are bench and lab-based tests for a medical device administration set, they are inherently prospective controlled tests conducted by the manufacturer.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided. The nonclinical tests appear to rely on established testing methods and standards (e.g., ISO standards, FDA guidance documents) rather than subjective expert consensus for establishing ground truth. Compliance with these standards is the metric, not expert opinion.
4. Adjudication Method for the Test Set
Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving interpretation of data by multiple experts. Since no human clinical studies were performed, and the testing was nonclinical (bench, lab-based), an adjudication method is not applicable and therefore not mentioned.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Human clinical studies were deemed unnecessary to evaluate the safety or effectiveness of the Zyno Medical Administration Sets." The submission focuses on demonstrating substantial equivalence through nonclinical testing.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This question is not applicable to the Zyno Medical Administration Sets. This device is a physical medical administration set, not an AI algorithm or software device. Therefore, the concept of "standalone performance" in the context of an algorithm does not apply. The device's performance stands alone in its physical and functional testing.
7. The Type of Ground Truth Used
The ground truth for the nonclinical tests was based on established industry standards and regulatory guidance. For example:
- Biocompatibility: Likely ISO 10993 series.
- Sterilization: Compliance with relevant sterility assurance levels.
- Physical/Mechanical Testing: Adherence to performance specifications to ensure proper function and safety.
- Integration Testing: Successful operation with the Z-800 series infusion pumps, demonstrating compatibility and correct fluid delivery.
- Pyrogenicity, Chemical, Microbial Ingress, DEHP, Aging: Compliance with specific test methods and limits defined in relevant standards and guidance documents (e.g., "Guidance for Industry and FDA Staff - Intravascular Administration Sets Premarket Notification Submissions [510(k)]" and "Total Product Life Cycle: Infusion Pump - Premarket Notification [510(k)] Submissions").
8. The Sample Size for the Training Set
This question is not applicable. The Zyno Medical Administration Set is a physical medical device, not a machine learning or AI model that requires a "training set."
9. How the Ground Truth for the Training Set was Established
This question is not applicable, as there is no training set for a physical medical device.
Ask a specific question about this device
(185 days)
ZYNO MEDICAL ADMINISTRATION SET
The Zyno Medical Administration Set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein.
The Zyno Medical's I.V. Administration Set is a single use, latex-free, non-DEHP device sterilized with Ethylene Oxide gas. It is utilized to administer fluids from a container to a patient's vascular system through a catheter or needle inserted into a vein. Depending on configuration, the device may include the following components: tubing, protective cap(s), drip chamber (s) with spike, tubing, back check valve(s), needleless Y-site(s), slide clamp(s), roller clamp(s), pinch clamp(s), 0.22 micron filter or 1.2 micron filter, male luer lock (with cap), and pressure activated valve(s). Seven (7) representative administration sets and one (1) representative extension set are described in the application package to illustrate the device confiqurations.
Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Zyno Medical Administration Set (K120685).
Important Note: The provided text is a 510(k) summary for a medical device that is substantially equivalent to existing predicate devices. This type of submission relies heavily on demonstrating that the new device is as safe and effective as a legally marketed device, rather than proving novel performance against specific, quantifiable acceptance criteria in a clinical study. Therefore, many of the typical "AI/software device" study questions (like MRMC, training sets, expert ground truth) are not applicable to this type of submission.
Acceptance Criteria and Reported Device Performance
Given the nature of this 510(k) summary, the "acceptance criteria" are primarily related to demonstrating substantial equivalence to predicate devices and meeting relevant safety standards. There aren't specific numerical performance metrics (like sensitivity/specificity for a diagnostic AI) listed in the summary.
Acceptance Criteria Category | Reported Device Performance | Comments |
---|---|---|
Intended Use | "administer fluids from a container to a patient's vascular system through a needle or a catheter inserted into a vein." | This is met by the device's design and intended function, as described. The study demonstrated functional equivalence to predicate devices. |
Material Composition | Constructed of high-grade extruded DEHP-free Polyvinyl Chloride (PVC). | Verified through material specifications. Substantially equivalent or identical to predicate devices. |
Sterilization | Sterilized with Ethylene Oxide gas. | Verified through sterilization testing, ensuring sterility for patient safety. |
Components | May include: tubing, protective cap(s), drip chamber (s) with spike, tubing, back check valve(s), needleless Y-site(s), slide clamp(s), roller clamp(s), pinch clamp(s), 0.22 micron filter or 1.2 micron filter, male luer lock (with cap), and pressure activated valve(s). | Device configurations are described and are identical/similar to predicate devices. Bench testing confirms components function as expected. |
Safety and Effectiveness | Complies with safety and effectiveness criteria. | Demonstrated through bench testing, sterilization testing, and biocompatibility testing, showing substantial equivalence to predicate devices. |
Performance Standards | "met its intended use requirements" and "complies with the safety and effectiveness criteria." | Non-clinical testing included evaluation studies and testing to relevant safety standards. Predicate devices were also tested for direct comparison where appropriate. |
Substantial Equivalence | Demonstrated substantial equivalence to predicate devices: Baxter Healthcare Corporation (K981792), Amsino International, Inc (K973107), Health Line International Corporation (K060352). | This is the overarching "acceptance criterion" for a 510(k) submission like this. The device was found to be substantially equivalent based on intended use, basic design, function, materials, and non-clinical test results. |
Study Details (Applicable to this type of device/submission)
This 510(k) is for a basic medical device (intravascular administration set), not an AI/software device. Therefore, the study conducted is primarily non-clinical bench testing for substantial equivalence. Clinical studies in the AI/software sense were not necessary or performed.
-
Sample size used for the test set and the data provenance:
- Sample Size: The document states, "When appropriate, predicate devices were tested using the exact same method and sample size, for direct comparison of results." However, the specific number of units tested for the Zyno Medical Administration Set itself is not explicitly stated in the provided summary. This information would typically be found in the full 510(k) submission, not the summary.
- Data Provenance: This refers to bench testing. The testing was conducted by or on behalf of Zyno Medical, LLC in Natick, MA, USA. It is prospective bench testing based on design specifications to confirm performance.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. For a physical device like an I.V. administration set, "ground truth" isn't established by human experts in the same way it would be for an AI diagnostic. Ground truth is derived from engineering specifications, industry standards, and measurable physical properties (e.g., flow rates, leak integrity, material strength).
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. As there are no human interpretive "readings" involved, there's no adjudication method in play for the test set. Performance is objectively measured against specifications.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This is a physical medical device, not an AI or software-assisted diagnostic tool. No MRMC study was conducted.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is not an algorithm or AI device. The device's performance is standalone in the sense that its physical properties and function are tested independently of a human "loop" for its intended use (fluid administration).
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for this device's testing is based on engineering specifications, recognized industry standards, and performance characteristics of legally marketed predicate devices. For example, ground truth for sterility is a sterile state, ground truth for material properties is adherence to chemical/physical specifications, and ground truth for functional performance (e.g., integrity, flow) is meeting pre-defined acceptable ranges.
-
The sample size for the training set:
- Not Applicable. This is not an AI/machine learning device; therefore, there is no "training set."
-
How the ground truth for the training set was established:
- Not Applicable. See point 7.
Ask a specific question about this device
Page 1 of 1