Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133255
    Manufacturer
    Date Cleared
    2014-08-12

    (293 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ZERAMEX T AND ZERAMEX PLUS DENTAL IMPLANT SYSTEMS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zeramex® T Dental Implant System is intended to be surgically placed in the bone of the upper and lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore aesthetics and chewing function.

    The Zeramex® T Dental Implant System can be used for single or multiple unit restorations.

    Zeramex® T implants are intended for delayed loading.

    The Zeramex® T dental implants are specially indicated for patients with metal allergies and chronic illnesses due to metal allergies.

    The Zeramex® T (Ø 3.5 mm) implant may only be used in the anterior teeth in the lower jaw and lateral teeth in the upper jaw.

    Device Description

    The Zeramex® T System is an endosseous dental implant/abutment system. The Zeramex® T implants are placed using the Zeramex® T surgical tools. The implants are provided in three diameters (Ø 3.5 mm, 4.2 mm, and 5.5 mm) and three lengths (8 mm, 10 mm, and 12 mm). The Ø 4.2 mm implant is also offered in a 14 mm length. Dentalpoint AG offers several compatible abutments and other accessories including straight and angular (15°) abutments in small, regular/wide, and wide sizes, healing caps, and gingiva formers. The abutments are bonded to the implants using dental cements which have been tested and validated for use with the Zeramex system: Panavia (K032455) and 3M ESPE RelyX (K111185).

    Class I, 510(k) exempt accessories are provided in the Zeramex® T tray and include drill bits, screwdrivers, countertorque devices, placement and removal tools, laboratory pieces used for fabrication or dental prosthetics, and trial abutments. The surgical tools are intended to be reprocessed (cleaned, disinfected, and sterilized) after each procedure.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the Zeramex® T Dental Implant System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria for the clinical study endpoints. However, it does report the device performance for key safety and effectiveness measures. The "acceptance criteria" listed below are inferred as the performance observed in the study.

    Acceptance Criteria (Inferred)Reported Device Performance
    Minimum 2-year cumulative survival rate for implants87%
    Maximum soft-tissue complication rate0%
    Maximum technical complication rate4%
    Maximum complication rate for bone loss greater than 2 mm0%
    Maximum aesthetic complication rate0%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Study): 49 implants
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the manufacturer is Dentalpoint AG, located in Zurich, Switzerland. The clinical study was a "single-center" study.
      • Retrospective or Prospective: Prospective

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide details on the number of experts or their qualifications for establishing ground truth in the clinical study. The "ground truth" for the clinical study would likely be the direct clinical observation and measurements by the treating clinicians and study staff.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method used for the test set (clinical study). Outcomes were presumably assessed by the clinical site directly.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The clinical study described is a single-arm study evaluating the device's performance, not comparing it against human readers or other devices.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The Zeramex® T Dental Implant System is a physical medical device (dental implant), not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

    7. The Type of Ground Truth Used

    For the clinical study, the ground truth was based on:

    • Outcomes data: Specifically, implant survival rates, and reported complication rates (soft-tissue, technical, bone loss, aesthetic). These are direct measures from the patients and their clinical follow-ups.

    8. The Sample Size for the Training Set

    This question is not applicable. The Zeramex® T Dental Implant System is a physical medical device, not an AI/ML algorithm that requires a training set. The "training" for such a device involves engineering design, material science, and manufacturing processes, supported by pre-clinical testing, rather than data-driven machine learning.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as there is no "training set" in the context of an AI/ML algorithm. The development of the device involves pre-clinical testing:

    • Fatigue testing in accordance with ISO 14801
    • Surface chemical analysis
    • Validation of cleaning, disinfection, and sterilization processes
    • Packaging and transport validation

    The "ground truth" for these pre-clinical tests would be the established scientific standards, protocols, and measurable outcomes (e.g., meeting specific force thresholds in fatigue testing, chemical composition analysis results, sterility assurance levels).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1