    Why did this record match?
    Device Name :

    Cepheid Xpert MRSA NxG Control Panel

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cepheid Xpert® MRSA NxG Control Panel is intended for use as an external assayed positive and negative quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Methicillin-Resistant Staphylococus aureus performed with the Cepheid Xpert® MRSA NxG Assay on the GeneXpert® Instrument System. The controls comprise cultured and inactivated Methicillin-Resistant Staphylococcus aureus as the positive control and Staphylococcus epidermidis as the negative control.

    The Cepheid Xpert® MRSA NxG Control Panel is not intended to replace manufacturer controls provided with the Cepheid Xpert® MRSA NxG Assay.

    Device Description

    The Cepheid Xpert® MRSA NxG Control Panel is used to monitor the DNA extraction, amplification and detection processes of the Cepheid Xpert® MRSA NxG Assay. The Cepheid Xpert® MRSA NxG Control Panel contains cultured microorganisms inactivated by heat treatments. Each Cepheid Xpert® MRSA NxG Control Panel consists of 6 individually packaged positive control swabs and 6 individually wrapped negative control swabs. Each positive control swab contains cultured and inactivated Methicillin-Resistant Staphylococcus aureus (MRSA) at arget level that is designed to provide reproducible performance above the limit of detection for each of the genes targeted by the Cepheid Xpert® MRSA NxG Assay: methicillin resistance gene mecA (mec) and the Staphylococcal cassette chromosome (SCC). Each negative contains Staphylococcus epidermidis (MSSE) that is not detected by the Cepheid Xpert® MRSA NxG Assay. Each swab is individually wrapped with a desiccant in a heat-sealed foil pouch.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Cepheid Xpert® MRSA NxG Control Panel, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Positive Control100% agreement for Methicillin-Resistant Staphylococcus aureus (mec, SCC) detection across all sites100% (93/93) overall agreement
    Negative Control100% agreement for Staphylococcus epidermidis (no detection) across all sites100% (92/92) overall agreement

    (Note: The provided text doesn't explicitly state the acceptance criteria as separate from the performance, but the 100% agreement achieved suggests this was the expected outcome for a control panel's precision and reproducibility study.)

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Sample Size:
        • For Positive Controls: 93 tests (32 at Site 1, 31 at Site 2, 30 at Site 3).
        • For Negative Controls: 92 tests (31 at Site 1, 31 at Site 2, 30 at Site 3).
      • Data Provenance: The study was conducted at three different testing locations (referred to as Site 1, Site 2, and Site 3). The document does not specify the country of origin, but it implies a prospective study since it describes "a precision and reproducibility study was conducted."
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The concept of "ground truth" as established by experts (e.g., radiologists) is not applicable here. This is a study for a quality control material for a diagnostic assay, not an AI diagnostic device evaluating patient data. The "ground truth" for the control material is inherently defined by its composition (e.g., inactivated Methicillin-Resistant Staphylococcus aureus for the positive control, Staphylococcus epidermidis for the negative control) and the known expected results from the Cepheid Xpert® MRSA NxG Assay.
    3. Adjudication method for the test set:

      • Not applicable. As this is a study for a quality control material, the results are objective (positive/negative detection) based on the performance of the assay with the control. There's no subjective interpretation requiring adjudication among experts. The study reports raw agreement percentages. Any "ERROR" responses were retested with a new control, indicating a procedural re-evaluation rather than expert adjudication of a result.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This type of study (MRMC for AI) is not applicable to a quality control panel. This product is a control material for an in-vitro diagnostic assay, not an AI-powered diagnostic device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, in essence. The "device performance" refers to the performance of the control panel itself, which is then used by human operators with the GeneXpert® Instrument System and Cepheid Xpert® MRSA NxG Assay. While human operators are involved in running the test, the evaluation of the control material's performance against the assay's expected results is a standalone assessment of the control material's ability to consistently produce those results. This isn't an AI algorithm; rather, it's a quality control for an existing diagnostic system.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" for the control panel is based on its known composition and the expected analytical performance when run on the target assay. For the positive control, the ground truth is the presence of Methicillin-Resistant Staphylococcus aureus (mec, SCC) at a level detectable by the assay. For the negative control, the ground truth is the presence of Staphylococcus epidermidis which should not be detected by the target assay.
    7. The sample size for the training set:

      • Not applicable. This study is for a quality control material, not a machine learning or AI algorithm that requires a training set. The "samples" used are the control materials themselves tested on the assay system.
    8. How the ground truth for the training set was established:

      • Not applicable. As there is no training set for an AI algorithm, this question is not relevant to the described device and study.
    Ask a Question

    Ask a specific question about this device

    K Number
    K162444
    Device Name
    Xpert MRSA NxG
    Manufacturer
    Date Cleared
    2016-11-23

    (84 days)

    Product Code
    Regulation Number
    866.1640
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Xpert MRSA NxG

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Xpert® MRSA NxG Assay, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro diagnostic test intended for the detection of methicillin-resistant Staphylococcus aureus (MRSA) DNA directly from nasal swabs in patients at risk for nasal colonization. The test utilizes automated real-time polymerase chain reaction (PCR) for the amplification of MRSA-specific DNA targets and fluorogenic target-specific hybridization probes for the real-time detection of the amplified DNA. The Xpert MRSA NxG Assay is intended to aid in the prevention and control of MRSA infections in healthcare settings. The Xpert MRSA NxG Assay is not intended to diagnose, guide, or monitor treatment for MRSA infections, or provide results of susceptibility to methicillin. A negative result does not preclude MRSA nasal colonization. Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing.

    Device Description

    The Xpert MRSA NxG Assay is an automated real-time polymerase chain reaction (PCR) in vitro diagnostic test for qualitative detection of methicillin-resistant Staphylococcus aureus (MRSA) directly from nasal swab specimens of patients at risk for colonization with MRSA in a healthcare setting. The Xpert MRSA NxG Assay is performed on the Cepheid GeneXpert® Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48, GeneXpert Infinity-48s, and GeneXpert Infinity-80 systems). The GeneXpert Instrument System platform automates sample preparation, amplification and real-time detection. The GeneXpert Instrument Systems require the use of single-use, disposable cartridges (the Xpert MRSA NxG cartridges) that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained and specimens never come into contact with working parts of the instrument modules, cross-contamination between samples is minimized. The Xpert MRSA NxG Assay includes reagents for the detection of target MRSA DNA. The primers and probes in the Xpert MRSA NxG Assay detect proprietary sequences for the gene for methicillin/oxacillin resistance (mecA and mecC), and staphylococcal cassette chromosome mec (SCCmec), which is inserted into the SA chromosomal attB site. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are internal controls utilized by the GeneXpert Instrument System platform. The SPC is present to control for adequate processing of the target bacteria and to monitor for the presence of inhibitor(s) in the PCR assay to avoid false-negative results. The Probe Check Control verifies reagent rehydration, real-time PCR tube filling in the cartridge, probe integrity, and dye stability. The single-use, multi-chambered fluidic cartridges are designed to complete sample preparation and real-time PCR for the detection of genomic DNA from methicillinresistant Staphylococcus aureus (MRSA) in 70 minutes or less. The GeneXpert Instrument Systems, comprised of the GeneXpert Dx Systems and the GeneXpert Infinity Systems, have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR and RT-PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and RT-PCR and detection. Nasal swab specimens are collected using the Cepheid Sample Collection Device and are transported to the laboratory or designated GeneXpert testing area. The nasal swab specimen is placed in a tube containing 2 mL of Elution Reagent. Following a brief vortexing of the sample, the entire contents of the Elution Reagent tube are transferred to the sample chamber of the Xpert MRSA NxG cartridge using a transfer pipette. For nasal swab specimens that are collected using Copan Liquid Amies Elution Swab (ESwab) collection and transport system, 300 uL of liquid sample is transferred to a tube containing 2 mL of Elution Reagent. Following a brief vortexing of the sample, the entire contents of the Elution Reagent tube are transferred to the sample chamber of the Xpert MRSA NxG cartridge using a transfer pipette. The user initiates a test from the system user interface and places the cartridge into the GeneXpert instrument platform, which performs hands-off real-time, multiplex PCR for detection of DNA. The results are automatically generated at the end of the process in a report that can be viewed and printed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Xpert MRSA NxG device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" for the clinical performance in a numerical format that would be directly comparable to the predicate device. Instead, it aims to demonstrate "substantial equivalence" to the predicate device (BD MAX™ MRSA XT assay). The primary performance metrics for this type of in vitro diagnostic test are analytical sensitivity (Limit of Detection), clinical sensitivity, and clinical specificity. I've extracted the relevant performance metrics from the clinical studies to serve as the reported device performance.

    Performance MetricAcceptance Criteria (Implied for Substantial Equivalence to Predicate)Reported Device Performance (Xpert MRSA NxG Assay)Comments
    Analytical Studies
    Limit of Detection (LoD)Not explicitly stated (demonstrated reproducibility at 95% confidence)Rayon Swab: 95% positive at 302 CFU/swabThe study determined the lowest concentration of MRSA that can be reproducibly distinguished from negative samples 95% of the time with 95% confidence. This is a measure of analytical sensitivity.
    ESwab: 95% positive at 812 CFU/swab
    Analytical Specificity (Cross-reactivity)100% specificity for tested non-MRSA organisms100% for 152 potentially cross-reactive microorganisms and human cellsNo cross-reactivity with methicillin-susceptible Staphylococcus aureus (MSSA), coagulase-negative staphylococci, or other nasal flora.
    Inclusivity100% detection of diverse MRSA strains100% (all 196 MRSA strains correctly detected)Assessed with a wide range of MRSA strains, including various SCCmec types and known USA types.
    Microbial InterferenceNo interference or cross-reactivity observedNo interference or cross-reactivity observedTested with 9 common commensal bacteria found in nasal cavities.
    Interfering SubstancesNo false negatives; delays in Ct values acceptable if results remain correctIdentified 3 substances causing Ct delays, but no false negatives.Addressed as limitations in the package insert.
    Carry-Over Contamination100% prevention of crossover contamination100% (no carry-over in 42 negative samples)Demonstrated in a study with high positive and negative samples.
    Clinical Studies
    Clinical SensitivitySubstantially equivalent to predicate deviceRayon Swab: 91.0% (95% CI: 84.6-94.9)Based on two prospective, multi-site studies against reference culture and susceptibility testing.
    ESwab: 92.9% (95% CI: 86.0-96.5)
    Combined: 91.8% (95% CI: 87.4-94.8)
    Clinical SpecificitySubstantially equivalent to predicate deviceRayon Swab: 96.9% (95% CI: 95.7-97.8)
    ESwab: 97.6% (95% CI: 96.2-98.5)
    Combined: 97.2% (95% CI: 96.3-97.9)
    Positive Predictive Value (PPV)Not explicitly stated (reported for context)Rayon Swab: 78.7% (95% CI: 71.3-84.7)
    ESwab: 83.5% (95% CI: 75.4-89.3)
    Combined: 80.8% (95% CI: 75.5-85.2)
    Negative Predictive Value (NPV)Not explicitly stated (reported for context)Rayon Swab: 98.9% (95% CI: 98.0-99.4)
    ESwab: 99.1% (95% CI: 98.1-99.5)
    Combined: 98.9% (95% CI: 98.3-99.3)
    ReproducibilityHigh agreement across sites, days, lots, operators98.6-100% agreement for low-positive and negative samplesDemonstrated excellent reproducibility in % agreement and low variability in Ct values.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Clinical Test Sets:
      • Rayon Swab Study: 1103 eligible nasal swab specimens.
      • ESwab Study: 846 eligible nasal swab specimens.
      • Combined: 1949 total specimens.
    • Data Provenance: The clinical studies were prospective, multi-site investigational studies.
      • The Rayon Swab study involved "eight investigational sites within the US and outside of the US."
      • The ESwab study involved "six investigational sites within the US."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The ground truth for the clinical test set was established using a reference culture and susceptibility testing method, not expert consensus in the traditional sense of medical image interpretation (e.g., radiologists). The document does not specify the number or qualifications of laboratory personnel performing these culture and susceptibility tests. However, the FDA's acceptance of this method implies that these are standard, qualified laboratory procedures.

    4. Adjudication Method (for the test set)

    The adjudication method was based on the reference culture and susceptibility testing.

    • A specimen was considered positive for MRSA if the presence of MRSA was confirmed in either direct culture on MRSA selective chromogenic medium or via an enriched culture (Trypticase Soy Broth (TSB) with 6.5% Sodium Chloride followed by subculture on Blood Agar (BA) and MRSA selective chromogenic medium).
    • Identification of presumptive S. aureus colonies was confirmed with Gram stain, catalase, and coagulase testing.
    • MRSA confirmation was done by susceptibility testing with a Cefoxitin disk (30µg).
    • For discrepancies (e.g., false positives/negatives by the device), repeat subculture of the enrichment broth was performed, as noted in the footnotes to Tables 8-7 and 8-8. This acts as a form of "adjudication" for discrepant results.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is an automated, standalone diagnostic assay (real-time PCR) for detecting MRSA DNA. It does not involve human "readers" in the interpretation of results in a way that would be typical for image-based AI tools. Therefore, there's no "human improvement with AI vs without AI assistance" to report.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was conducted. The Xpert MRSA NxG Assay is an automated qualitative in vitro diagnostic test, and its performance (sensitivity, specificity) was directly compared against the reference method without human interpretation of the assay's output itself. The device itself performs sample preparation, amplification, and real-time detection, automatically generating results.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used was laboratory culture and susceptibility testing, specifically:

    • Direct culture on MRSA selective chromogenic medium.
    • Enriched culture (TSB with 6.5% NaCl, followed by subculture on BA and MRSA selective chromogenic medium).
    • Confirmation of S. aureus by Gram stain, catalase, and coagulase testing.
    • Confirmation of MRSA by Cefoxitin disk susceptibility testing.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning model development. This is an in vitro diagnostic device based on PCR technology, which typically does not involve machine learning training in the same way an AI imaging model would. The "training" here would refer to the development and optimization of the PCR primers and probes, and the cartridge and instrument system, which is part of the product development process rather than a distinct "training set" for an algorithm.

    However, if we consider "training" in the sense of analytical development, the following data points could be relevant to the development and optimization of the assay's internal logic and performance parameters:

    • Analytical Sensitivity (LoD) Study: Used 13 individual MRSA strains (representing various SCCmec types).
    • Analytical Specificity (Cross-reactivity) Study: Tested 152 potentially cross-reactive microorganisms and human cells.
    • Analytical Reactivity (Inclusivity) Study: Evaluated 196 methicillin-resistant Staphylococcus aureus strains.

    These studies technically inform and confirm the performance of the developed assay, which is analogous to a validation set in traditional software, but not a training set for an AI/ML algorithm.

    9. How the Ground Truth for the Training Set Was Established

    As noted in point 8, the concept of a "training set" in the context of an AI/ML algorithm, along with its ground truth, is not applicable here. The assay's analytical parameters (e.g., LoD, specificity, inclusivity) were established through rigorous laboratory testing using well-characterized microbial strains and samples, where the "ground truth" for each analytical study (e.g., presence/absence of a specific MRSA strain, concentration, presence of interfering substance) was defined by standard microbiological and chemical methods.