Search Results
Found 3 results
510(k) Data Aggregation
(184 days)
X-CUBE 70, X-CUBE 90
The X-CUBE 70 and X-CUBE 90 diagnostic ultrasound systems are intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic: Adult Cephalic: Trans-raginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Trans-esoph(Cardiac), Peripheral Vessel(PV); and Urology(including prostate).
And, the imaging modes are 2D(B) mode; Harmonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode: High PRF Doppler mode: Tissue Doppler Imaging(TDI) Mode: 3D/4D mode.
The X-CUBE 70 and X-CUBE 90 are intended to be used in a hospital or medical clinic.
The X-CUBE 70 and X-CUBE 90 products are general purpose ultrasound imaging system for medical diagnosis assistance.
These products are used as an aid tool to diagnosis, such as a commonly used ultrasound diagnostic device.
Also X-CUBE 70 and X-CUBE 90 have the same operating principles, intended use, risk grade and design/manufacturing characteristics as reference/predicate devices.
This systems platform provide patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, image quality.
The provided text describes the regulatory clearance of the ALPINION Medical Systems X-CUBE 70 and X-CUBE 90 diagnostic ultrasound systems. It does not contain information about acceptance criteria and a study proving a device meets these criteria in the context of an AI/ML-driven medical device, as the device itself is a diagnostic ultrasound system and not an AI/ML product.
However, the document lists several standards and regulations that the device conforms to to establish its safety and effectiveness relative to predicate devices. These can be interpreted as the high-level criteria for substantial equivalence.
Here's a breakdown of the information that is available, reframed to address your request where possible, and noting where information is explicitly not present for an AI/ML context:
1. A table of acceptance criteria and the reported device performance:
The document does not specify performance acceptance criteria in terms of metrics like sensitivity, specificity, or accuracy that would be typical for an AI/ML device study. Instead, it states the device's conformance to various safety and performance standards. The "New Technologies" are features of the ultrasound system itself, not standalone AI algorithms with performance metrics.
Acceptance Criteria (Standards Conformed To) | Reported Device Performance (Conformance) |
---|---|
IEC60601-1:2005 (General Requirements for Safety) | Conforms |
IEC60601-1-2:2014 (Electromagnetic Compatibility) | Conforms |
IEC60601-2-37:2007 (Basic Safety and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment) | Conforms |
AAMI/ANSI/ISO10993-1:2009 (Biological Evaluation of Medical Devices - Biocompatibility) | Conforms |
AAMI/ANSI/ISO14971:2007 (Application of Risk Management to Medical Devices) | Conforms |
AIUM MUS, Third edition (Medical Ultrasound Safety) | Conforms |
NEMA UD 2-2004 (Acoustic Output Measurement Standard) | Conforms |
NEMA UD 3-2004 (Real Time Display of Thermal and Mechanical Acoustic Output Indices) | Conforms |
21 CFR 820 (Quality System Regulation) | Conforms |
ISO 13485 (Medical devices — Quality management systems) | Conforms |
2. Sample size used for the test set and the data provenance:
- Not applicable / Not provided. The document states: "The subject of this premarket submission, X-CUBE 70 and X-CUBE 90, did not require clinical studies to support substantial equivalence." This means there was no specific clinical test set used to evaluate performance in the way an AI/ML device would be tested. The evaluation was based on conformance to engineering and safety standards and comparison to existing predicate devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable / Not provided. As no clinical study or test set for an AI/ML algorithm was conducted, there was no need for experts to establish ground truth in this context.
4. Adjudication method for the test set:
- Not applicable / Not provided. No clinical test set.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable / Not provided. This is not an AI-assisted device, so no such study was conducted.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable / Not provided. This is an ultrasound imaging system, not a standalone AI algorithm.
7. The type of ground truth used:
- Not applicable / Not provided. As no clinical evaluation study requiring ground truth was performed, this information is not available. The "ground truth" for this submission was based on the technical specifications and validated performance against established standards for ultrasound devices, as well as comparison to predicate devices.
8. The sample size for the training set:
- Not applicable / Not provided. This is not an AI/ML device.
9. How the ground truth for the training set was established:
- Not applicable / Not provided. This is not an AI/ML device.
In summary, the provided document is a 510(k) clearance letter for a diagnostic ultrasound system. It outlines the regulatory process for demonstrating "substantial equivalence" to existing devices, primarily through adherence to safety and performance standards and a comparison of technological characteristics. It is not a submission for an AI/ML product and therefore does not contain the specific types of acceptance criteria and study details you would expect for such a device.
Ask a specific question about this device
(103 days)
X-CUBE 90
The X-CUBE 90 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Trans-vaginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiatric); Peripheral Vessel(PV); and Urology(including prostate).
And, in the imaging modes of 2D(B) mode; Harmonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode; 3D/4D mode. The X-CUBE 90 is intended to be used in a hospital or medical clinic.
X-CUBE 90 product is an ultrasound imaging system for medical diagnosis. This innovative system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.
The provided text describes the Alpinion Medical Systems Co., Ltd. X-CUBE 90 ultrasound system and its equivalence to predicate devices, but it does not include information about acceptance criteria or a study proving device performance in the context of AI/ML.
The document is a 510(k) summary for a traditional medical device (ultrasound system), not an AI/ML-powered device. Therefore, it focuses on demonstrating substantial equivalence to previously cleared ultrasound systems based on technical characteristics, indications for use, and compliance with general medical device safety standards.
Here's what can be extracted from the document, with the understanding that the request for AI/ML-specific details cannot be fully met from the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not specify quantitative clinical performance acceptance criteria or report device performance in terms of metrics like sensitivity, specificity, or AUC, which are common for AI/ML devices. Instead, it demonstrates performance by comparing its features and characteristics to legally marketed predicate devices, asserting that it is as safe and effective.
The acceptance criteria are implicitly based on compliance with recognized medical device standards and having equivalent specifications and functionalities to the predicate devices.
Acceptance Criterion (Implicit) | Reported Device Performance (Implicit) |
---|---|
Compliance with IEC 60601-1 (General Safety) | X-CUBE 90 conforms to IEC60601-1:2005(Third Edition)+CORR.1:2006+CORR.2:2007+A1:2012 |
Compliance with IEC 60601-1-2 (EMC) | X-CUBE 90 conforms to IEC60601-1-2:2014 |
Compliance with IEC 60601-2-37 (Ultrasonic Specific Safety) | X-CUBE 90 conforms to IEC60601-2-37:2007/AMD1:2015 |
Biocompatibility of patient contact materials | Transducer materials and other patient contact materials are biocompatible (conforming to AAMI/ANSI/ISO10993-1:2009(R)2013) |
Compliance with risk management standards | Conforms to AAMI/ANSI/ISO14971:2007/(R)2010 |
Compliance with Acoustic Output Measurement Standard | Conforms to NEMA UD 2-2004(R2009) |
Compliance with Real Time Display of Thermal and Mechanical AI | Conforms to NEMA UD 3-2004(R2009) |
Indications for Use (equivalent to predicate) | The X-CUBE 90 has the same clinical applications and imaging modes as its primary predicate (K200449 X-CUBE 90). |
Technical Specifications (dimensions, electrical power, etc.) | Specifications are comparable or slightly improved to predicate devices, not affecting safety or effectiveness. |
Regarding the sections specific to AI/ML devices (points 2-9 in your request), the provided text does not contain this information as it describes a conventional ultrasound system, not an AI/ML powered one.
Therefore, the following cannot be answered from the provided document:
- 2. Sample size used for the test set and the data provenance: Not applicable, as no AI/ML test set is mentioned.
- 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
- 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
- 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
- 8. The sample size for the training set: Not applicable, as no AI/ML training set is mentioned.
- 9. How the ground truth for the training set was established: Not applicable.
Summary of Non-Clinical and Clinical Tests (from the document):
- Summary of Non-Clinical Tests: The X-CUBE 90 was evaluated for biocompatibility, acoustic output, and thermal, electrical, electromagnetic, and mechanical safety. It was found to conform to various medical device safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, AAMI/ANSI/ISO10993-1, AAMI/ANSI/ISO14971, AIUM MUS, NEMA UD 2, NEMA UD 3). Quality management system measures were also applied, including risk management, requirements reviews, design reviews, component verification, integration review, performance testing, safety testing, and design validation.
- Summary of Clinical Tests: The submission explicitly states: "The subject of this premarket submission, X-CUBE 90, did not require clinical studies to support substantial equivalence." This is typical for 510(k) submissions of conventional devices demonstrating equivalence to existing ones.
Ask a specific question about this device
(137 days)
X-Cube 90
The X-CUBE 90 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Trans-vaginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Peripheral Vessel(PV); and Urology(including prostate).
And, in the imaging modes of 2D(B) mode; Harnonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode; 3D/4D mode. The X-CUBE 90 is intended to be used in a hospital or medical clinic.
X-CUBE 90 product is an ultrasound imaging system for medical diagnosis. This innovative system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.
The provided document is an FDA 510(k) premarket notification for the Alpinion Medical Systems Co., Ltd. X-CUBE 90 ultrasound system. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving efficacy through clinical studies for novel devices.
Therefore, the acceptance criteria and study detailed here are focused on demonstrating that the new device (X-CUBE 90) performs equivalently to previously cleared predicate devices and complies with relevant safety and performance standards. It does not involve a traditional clinical study with patient data in the sense of an Artificial Intelligence (AI) device seeking de novo authorization or PMA.
Based on the provided text, here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a "table of acceptance criteria" related to specific performance metrics for the ultrasound imaging capabilities beyond what is implied by the comparison to predicate devices and adherence to standards. Instead, substantial equivalence is demonstrated by a comparison of features, technical specifications, and compliance with recognized safety standards. The "performance" is implicitly deemed equivalent if these factors are comparable to or better than the predicate devices.
Here's a breakdown of the comparison presented in the "Determination of Substantial Equivalence: Comparison table with Predicate devices" on page 25:
Feature | Proposed X-CUBE 90 Performance | Predicate E-CUBE 12 (K181277) Performance (and other Reference Devices) |
---|---|---|
Indications for Use | Fetal, Abdominal (renal & GYN/pelvic), Pediatric, Small Organ (breast, testes, thyroid), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional), Musculo-skeletal (Superficial), Cardiac (adult & pediatric), Peripheral Vessel (PV), Urology (including prostate). | All listed indications are present ("√") for the predicate E-CUBE 12, suggesting that the X-CUBE 90 offers the same range of clinical applications. |
Dimensions & Weight | Weight: 90kg, Height: 1325/1560 mm, Width: 554 mm, Depth: 815 mm | Weight: 94 kg, Height: 1,420/1,520 mm, Width: 590 mm, Depth: 895 mm (Comparable physical characteristics to predicate devices). |
Electrical Power | Voltage: 100-120V~, 200-240V~, Frequency: 50-60 Hz, Power: Max. 700VA | Voltage: 100-120V~, 200-240V~, Frequency: 50/60 Hz, Power: Max. 600VA (The X-CUBE 90 has slightly higher max power, but this would be evaluated against safety standards). |
Imaging Modes | 2D(B) mode, Harmonic mode (HAR), M mode, Color M mode, Anatomical M mode, Color Flow Doppler (CF) Mode, Power Doppler (PD) Mode, Directional PD mode, Pulsed Wave Doppler (PWD) Mode, Continuous Wave Doppler (CWD) Mode, High PRF Doppler mode, Tissue Doppler Imaging (TDI) Mode, 3D/4D mode. | All listed imaging modes are present ("√") for the predicate E-CUBE 12, indicating functionally equivalent imaging capabilities. |
Imaging Functions | Xpeed™, Full SRI™, Spatial Compounding Image (SCI), Frequency Compounding image (FCI), Panoramic, Stress Echo, Cube Strain™, Live HQ ™, Needle Vision™ / Needle Vision™ Plus, Elastography, Cube view™, Contrast Enhanced Ultrasound (CEUS), Cube Note, B-STIC, Point Shear Wave Elastography (PSWE), Microvascular Imaging (MVI), Volume Advance™ (Free Angle MSV, AnySlice™, Volume Analysis). | Many of these advanced imaging functions (Xpeed , Full SRI , SCI , FCI , Panic , Stress Echo , Cube Strain , Live HQ , Needle Vision , Elastography , Cube view , CEUS , Cube Note , B-STIC , Volume Advance ) are also present in the predicate devices, though some (like PSWE, MVI) are new to this device and would be evaluated for safety and effectiveness on their own, or by comparison to similar functions. |
Safety Standards | IEC60601-1, IEC60601-1-2, IEC60601-2-37, AAMI/ANSI/ISO10993-1, AAMI/ANSI/ISO14971, AIUM MUS, NEMA UD 2-2004(R2009), NEMA UD 3-2004(R2009). | The document states the X-CUBE 90 "has been found to conform to applicable medical device safety standards" and "comply with voluntary standards as detailed in this premarket submission." The predicate device also conforms to these same standards, indicating equivalent safety profiles. |
Acceptance Criteria (Implicit from Substantial Equivalence):
The acceptance criteria for this 510(k) submission are implicitly defined by demonstrating that the X-CUBE 90 is "substantially equivalent" to legally marketed predicate devices (primarily K181277 E-CUBE 12). This means:
- Same intended use: The X-CUBE 90 has the same indications for use as the predicate device.
- Same technological characteristics: The device uses the same fundamental technology (ultrasound, various imaging modes and functions).
- No new questions of safety or effectiveness: Any differences in technological characteristics do not raise new questions of safety or effectiveness. This is primarily addressed through compliance with established safety and performance standards.
2. Sample size used for the test set and the data provenance
The document explicitly states: "The subject of this premarket submission, X-CUBE 90, did not require clinical studies to support substantial equivalence."
This means there was no dedicated clinical test set of patient data in the typical sense for demonstrating diagnostic accuracy or efficacy. The "study" for this submission primarily consists of non-clinical performance testing, compliance with international safety and performance standards, and comparisons of technical specifications and features to predicate devices.
Therefore, there is no sample size for an an imaging test set, and no data provenance relevant to patient-derived images.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
As there was no clinical study involving a test set of patient data to establish ground truth diagnoses, this information is not applicable.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as there was no clinical test set for which ground truth adjudication was required.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was conducted. This device is an ultrasound system, not an AI-assisted diagnostic tool that would typically be evaluated for reader improvement.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as no clinical ground truth was established for a test set. The "ground truth" for this type of submission is adherence to recognized performance standards and demonstrated technological equivalence to already cleared devices.
8. The sample size for the training set
Not applicable. This device is an ultrasound imaging system, not an AI/machine learning model whose performance is enhanced through training on a dataset.
9. How the ground truth for the training set was established
Not applicable, as there was no training set for an AI/machine learning model.
In summary, this 510(k) submission for the X-CUBE 90 ultrasound system primarily relies on non-clinical testing and comparison to predicate devices, rather than clinical studies with patient data and ground truth establishment, to demonstrate substantial equivalence. The "acceptance criteria" are met by demonstrating that the device's technical specifications, imaging modes, and overall performance fall within the established safety and effectiveness profile of already cleared ultrasound systems, and by proving compliance with relevant international and national medical device standards.
Ask a specific question about this device
Page 1 of 1