Search Filters

Search Results

Found 52 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    Wrist Blood Pressure Monitor (AOJ-35A); Wrist Blood Pressure Monitor (AOJ-35B); Wrist Blood Pressure
    Monitor (WRS-35B); Wrist Blood Pressure Monitor (AOJ-35D); Wrist Blood Pressure Monitor (AOJ-35E); Wrist
    Blood Pressure Monitor (AOJ-35F); Wrist Blood Pressure Monitor (AOJ-35G); Wrist Blood Pressure Monitor
    (WRS-35G); Wrist Blood Pressure Monitor (WRS-35H); Wrist Blood Pressure Monitor (WRS-35K); Wrist Blood
    Pressure Monitor (WRS-35N); Wrist Blood Pressure Monitor (WRS-35P); Wrist Bloo

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wrist Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique by an inflatable cuff wrapped around the wrist at medical facilities or at home.

    Device Description

    The Wrist Blood Pressure Monitor is designed as a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at wrist within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa.

    All the models included in this submission follow the the same intended use, same measurement principle, same blood pressure core algorithm and similar product design. All the models can be used with one cuff size 13.5~19.5 cm (5.3-7.7inches).

    The main differences are appearance, Dimensions and some specifications which will not affect the safety and effectiveness of the device.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and associated summary pertain to a Wrist Blood Pressure Monitor, which is a medical device for measuring blood pressure and pulse rate. It is not an AI/Software as a Medical Device (SaMD). Therefore, many of the typical acceptance criteria and study designs associated with AI/SaMD (such as multi-reader multi-case studies, ground truth establishment by experts, training set details, or effect sizes of AI assistance) are not applicable to this device.

    The acceptance criteria and study details provided are tailored to the performance of a non-invasive blood pressure measurement system (hardware device), focusing on accuracy, safety, and effectiveness.

    Here's a breakdown of the requested information based on the provided document, addressing the device's specific characteristics as a hardware blood pressure monitor:


    Acceptance Criteria and Device Performance (Wrist Blood Pressure Monitor)

    1. Table of Acceptance Criteria and Reported Device Performance

    As per the 510(k) summary, the device's accuracy is a key performance metric. The acceptance criteria are based on the international standard ISO 81060-2 Third edition 2018-11 [Including AMD1:2020].

    Performance MetricAcceptance Criteria (from ISO 81060-2)Reported Device Performance
    Blood Pressure AccuracyMean error and standard deviation of differences for systolic and diastolic pressure not over the limits specified in ISO 81060-2.All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2.
    Heart Rate Accuracy± 5% of reading± 5% of reading (Same as Predicate, implying met for proposed)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set (Clinical Accuracy Study):
      • Three groups of clinical accuracy research were conducted. Each group included 100 subjects, for a total of 300 subjects across the 13 models.
      • Group 1: 100 subjects (47 Male, 53 Female)
      • Group 2: 100 subjects (54 Male, 46 Female)
      • Group 3: 100 subjects (44 Male, 56 Female)
      • Minimum subjects for each group was 85, as per ISO 81060-2.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical data. However, the manufacturer is "Shenzhen AOJ Medical Technology Co., Ltd." in Shenzhen, Guangdong, China. It is highly probable the data was collected in China.
    • Retrospective or Prospective: The clinical accuracy study, designed to meet ISO 81060-2, is typically conducted prospectively to collect new data for device validation. The wording "clinical accuracy research" and "clinical accuracy test report and data analysis followed the requirements" implies a prospective study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • This question is not directly applicable in the context of this device because "ground truth" for a blood pressure monitor's accuracy is established against a reference standard, not through expert consensus on interpretations of images or signals (as would be the case for AI/SaMD).
    • For blood pressure monitors, the "ground truth" or reference measurement is typically taken by trained medical professionals using a standardized reference sphygmomanometer (e.g., mercurial or auscultatory method), as per the ISO 81060-2 standard. The document states "The Same Arm Sequential Method was chosen for all studies," which is a standard procedure comparison method against a reference device. The qualifications of the individuals performing these reference measurements would be trained clinicians (e.g., physicians, nurses).

    4. Adjudication Method for the Test Set

    • This question is not applicable for a blood pressure monitor's accuracy testing. Adjudication methods (like 2+1 or 3+1) are used to resolve discrepancies in human expert interpretations, especially in image-based diagnostics.
    • For blood pressure accuracy, deviations are quantified statistically between the device reading and the reference measurement, not through an adjudication process among multiple "readers."

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, this was not done. This type of study is relevant for AI/SaMD devices where AI assists human interpretation and is a key component for assessing the AI's clinical utility. The Wrist Blood Pressure Monitor is a standalone hardware device that provides a measurement; it does not involve human "readers" interpreting data or AI assistance.

    6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    • Yes, in essence. The entire clinical accuracy study described in Section 8 ("Clinical data") evaluates the device's performance (which incorporates its internal oscillometric algorithm) in a standalone manner against a reference standard. The "algorithm" here refers to the embedded software that processes the oscillometric signals to derive blood pressure and pulse rate. The study directly assesses how accurately the device (with its integrated algorithm) measures blood pressure readings compared to the reference.
    • Performance Metrics: The evaluation was based on the "mean error and standard deviation of differences for systolic, diastolic pressure" as per ISO 81060-2.

    7. The Type of Ground Truth Used

    • The ground truth for the clinical accuracy testing was established through concurrent measurements using a standardized reference method (e.g., auscultatory method with a mercurial sphygmomanometer or another validated reference device) on the same arm, sequentially with the test device. This is the standard for blood pressure monitor validation as per ISO 81060-2.
    • It is not "expert consensus" in the sense of subjective medical interpretation, but rather an objective, standardized measurement performed by trained personnel using a calibrated reference instrument.

    8. The Sample Size for the Training Set

    • This concept is not applicable to this type of medical device clearance. The Wrist Blood Pressure Monitor is a hardware device with an embedded algorithm (oscillometric technique) that is based on established physiological principles. It doesn't use machine learning or deep learning in a way that requires a separate "training set" of patient data for an AI model to learn from, as would be the case for AI/SaMD devices. The device's "training" (development and calibration) would involve engineering principles and laboratory testing, rather than a data-driven machine learning process.

    9. How the Ground Truth for the Training Set Was Established

    • As the concept of a "training set" in the context of machine learning is not applicable here (see point 8), the establishment of ground truth for such a set is also not applicable. The device's underlying measurement principle is well-established oscillometric technology. Development and calibration rely on physical models and engineering validation.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241634
    Date Cleared
    2025-02-11

    (250 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Wrist Blood Pressure Monitor (XY-W01A, XY-W01B)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wrist Blood Pressure Monitor is intended to measure the systolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home. The intended wrist circumference is 13.5-21.5 cm.

    Device Description

    The Wrist Blood Pressure Monitor, including XY-W01B, can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person with wrist circumference of 135-215mm through oscillometric method. The initial inflation pressure of the cuff is zero pressure. When start the cuff will be inflated and deflated. The Wrist Blood Pressure Monitor is composed of monitor unit and wrist cuff. Of which the monitor unit contains main control circuit board, air pump, deflation valve, LCD and shell. The two models XY-W01A and XY-W01B have the same intended use, working principle, measuring range, accuracy, cuff, and conformance standard, except for product appearance. The devices are powered by 2 AAA batteries. The Wrist Blood Pressure Monitor has a memory function that automatically stores 90 sets data of the latest measurements.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the Wrist Blood Pressure Monitor (models XY-W01A, XY-W01B).

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard: ISO 81060-2:2018)Reported Device Performance (Clinical Test)
    Mean error for systolic pressure ≤ ±5 mmHgMean error within ±5 mmHg
    Standard deviation for systolic pressure ≤ 8 mmHgStandard deviation within 8 mmHg
    Mean error for diastolic pressure ≤ ±5 mmHgMean error within ±5 mmHg
    Standard deviation for diastolic pressure ≤ 8 mmHgStandard deviation within 8 mmHg
    All data not over the limits.All data is not over the limits.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 120 subjects (53 female and 67 male)
    • Data Provenance: The document does not explicitly state the country of origin.
    • Retrospective or Prospective: Not explicitly stated, but clinical tests for device approval are typically prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided in the document. For blood pressure monitors, the "ground truth" (reference measurements) is typically established by trained technicians using a standardized auscultatory method, not necessarily by "experts" in the sense of physicians or radiologists establishing a diagnostic truth.

    4. Adjudication Method

    This information is not provided in the document, as it relies on direct comparative measurements rather than expert adjudication of imaging or diagnostic results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is more relevant for diagnostic imaging AI systems where human readers interpret images. For a blood pressure monitor, the performance is compared against a reference measurement.

    6. Standalone Performance

    Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The clinical test evaluated the device's accuracy in measuring blood pressure and pulse rate against a reference standard.

    7. Type of Ground Truth Used

    The ground truth implicitly used is reference blood pressure measurements, likely obtained through a rigorously standardized auscultatory method as per the ISO 81060-2:2018 standard.

    8. Sample Size for the Training Set

    This information is not provided in the document. Given that this is a hardware device with an embedded algorithm (oscillometric technique), there isn't typically a separate "training set" in the way an AI/ML model would have for image recognition. The core algorithm's parameters are likely developed through extensive engineering and testing rather than a distinct training phase on a large dataset.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided. As mentioned above, the concept of a "training set ground truth" as understood in AI/ML is not directly applicable here. The device's underlying oscillometric algorithm and its parameters would have been developed and refined through engineering principles and testing against various blood pressure profiles and body types.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242568
    Date Cleared
    2025-01-30

    (154 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Wrist Blood Pressure Monitor (OHMS11, OHMS12)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use in measuring blood pressure in adult patient population with wrist circumference ranging from 13.0 cm to 21.0 cm.

    Device Description

    The Wrist Blood Pressure Monitor is a reusable, active, non-sterile, non-invasive and non-implantable device for use in measuring blood pressure in adult patient population with wrist circumference ranging from 13.0 cm to 21.0 cm. It is a home-use device prohibited from use in MRI (magnetic resonance imaging) environment. This device has the systolic and diastolic display features. The device is of the watch type and powered by a rechargeable lithium-polymer battery. An AC adapter is used for charging the device, but the device cannot be operated while charging. The device wrist cuff inflates using an integral pump, and deflates via an electric valve. During inflation, the wrist cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines systolic and diastolic blood pressure. The systolic and diastolic blood pressures are measured using the oscillometric method. The cuff can measure pressure range from 0 to 300mmHg. Product includes device (preinstalled with a strap and large-sized cuff: 16.6-21.0cm), strap with a medium-sized cuff: 13.0-16.5cm, wrist measuring tape, charging base (with the power cable). The product functional configurations of OHMS11 and OHMS12 are completely identical, and the difference between the two is reflected in the color (OHMS11 is the dark gray color, OHMS12 is the champagne color).

    AI/ML Overview

    The provided text describes the acceptance criteria and a study for a Wrist Blood Pressure Monitor (OHMS11, OHMS12).

    Here's an analysis of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Clinical Accuracy (ISO 81060-2.2019): Specific numerical criteria for mean difference and standard deviation between device and reference measurement for both systolic and diastolic blood pressure (as outlined in ISO 81060-2)."The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2." (Specific numerical values for mean difference and standard deviation are not provided in the document, only a statement of compliance).
    Electrical Safety (IEC 60601-1:2020 and IEC 80601-2-30)"shown to comply with IEC 60601-1:2020 and IEC 80601-2-30 for electrical safety."
    Electromagnetic Compatibility (EMC) (IEC 60601-1-2.2020)"shown to comply with IEC 60601-1-2.2.2020 for electromagnetic compatibility."
    Performance Testing (Product Design Specifications)"Data generated from the test met the predetermined acceptance criteria." (Specific acceptance criteria and performance data are not provided).
    Software Verification and Validation (FDA Guidance)"in compliance with FDA Guidance-Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
    Cybersecurity (FDA Guidance)"in compliance 'Content of Premarket Submissions for Management of Cybersecurity in Medical Devices'."
    Biocompatibility (ISO10993-1, including Cytotoxicity, Sensitization, Irritation)Biocompatibility evaluation was conducted for Cytotoxicity, Sensitization, and Irritation. (Implied compliance with ISO10993-1 as it refers to the guidance directly related to this standard).

    2. Sample size used for the test set and the data provenance

    • Sample Size: 85 patients (49 males and 36 females)
    • Data Provenance: Clinical study, conducted prospectively as patients "volunteered to take part in the subjects completed the clinical study." The country of origin is not explicitly stated in the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number of experts used or their qualifications for establishing the ground truth. It mentions that "manual Mercury Sphygmomanometer was used as a reference device" and that the study followed ISO 81060-2. This standard generally requires trained observers for the reference measurements, but specific expert details are absent.

    4. Adjudication method for the test set

    The document states, "Same arm sequential method was adopted during the clinical testing." This indicates a specific measurement protocol where both the test device and the reference device take measurements sequentially on the same arm. There is no mention of an adjudication method involving experts for discrepancies in ground truth, as the ground truth is established by the reference device measurement.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool requiring human reader evaluation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance study was done. The clinical testing evaluates the device's accuracy (algorithm only) against a reference standard (manual mercury sphygmomanometer) without direct human intervention in interpreting the device's output. The device itself performs the measurement and provides the blood pressure readings.

    7. The type of ground truth used

    The ground truth used was comparison to a reference device, specifically a "manual Mercury Sphygmomanometer," following the methodology of ISO 81060-2. This is a clinical standard for validating automated blood pressure monitors.

    8. The sample size for the training set

    The document does not provide information regarding a training set sample size. This type of submission (510(k) for a hardware device with embedded algorithms) typically focuses on validation data rather than detailed training information for AI/ML models.

    9. How the ground truth for the training set was established

    Since no training set information is provided, there is no information on how its ground truth would have been established. It is assumed the device relies on established oscillometric principles rather than a deep learning model requiring a large labeled training set.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Wrist Blood Pressure Monitor (222); Wrist Blood Pressure Monitor (222BT); Wrist Blood Pressure Monitor
    (W1681BL); Wrist Blood Pressure Monitor (W1681BT); Wrist Blood Pressure Monitor (W1681R)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 3/8 inches (13.5cm to 19.5cm). The device detects the appearance of irregular hearbeats during measurement and gives a warning signal with reading.

    Device Description

    The Wrist Blood Pressure Monitor,Models: 222,222BT,W1681BL, W1681BT, W1681R are powered by a rechargable Li-ion battery, automatic, non-invasive blood pressure system intended for home use. The devices are intended for use in adult patient population with wrist circumference ranging from 5 3/8 inches to 7 2/3 inches (13.5 cm). The devices' wrist cuff inflates using an integral pump and deflates via an electric valve. During deflation, the wrist cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic blood pressure. The systolic and diastolic blood pressures are measured using the oscillometric method. The cuff can measure pressure range from 0 to 295 mmHg, and the pulse rate range from 40 to 199 beats/min.

    When the device detected irregular rhythms, " W " will display on screen. An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while your monitor is measuring blood pressure. The WHO blood pressure indicator bar can classify by WHO and ISH recommendation. The devices display the latest blood pressure reading, while up to 2 x 99 readings can be stored in memory.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study proving the device meets them, based on the provided text:

    Device Name(s): Wrist Blood Pressure Monitor (Models: 222, 222BT, W1681BL, W1681BT, W1681R)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard / Requirement)Reported Device Performance (Subject Device)
    ISO 81060-2:2018 Criterion 1:
    Mean difference (device vs. reference)Systolic BP: 1.0 mmHg
    Diastolic BP: 0.9 mmHg
    ISO 81060-2:2018 Criterion 1:
    Standard deviation of differences (device vs. reference)Systolic BP: 3.1 mmHg
    Diastolic BP: 3.4 mmHg
    ISO 81060-2:2018 Criterion 2:
    Standard deviation (device vs. reference)Systolic BP: 2.4 (from Table 1, assumed to be an overall standard deviation)
    Diastolic BP: 2.6 (from Table 1, assumed to be an overall standard deviation)
    Cuff Pressure Accuracy±3 mmHg
    Pulse Accuracy±5% of reading
    ISO 81060-2:2018 standard acceptanceMet
    IEC 60601-1:2005+A1:2012+A2:2020 (Basic Safety & Essential Performance)Complies
    IEC 60601-1-11:2015 (Home Healthcare Environment)Complies
    IEC 80601-2-30:2018 (Automated Non-Invasive Sphygmomanometers)Complies
    IEC 60601-1-2:2014 (Electromagnetic Compatibility)Complies
    ISO 10993-10:2010 (Biocompatibility - Irritation/Sensitization)Complies
    ISO 10993-5:2009 (Biocompatibility - Cytotoxicity)Complies
    FDA Software Validation Guidance (General Principles of Software Validation)Complies (for moderate level of concern)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 86 subjects.
    • Data Provenance: Clinical investigation conducted by Shenzhen Cihai Hospital (China), prospective study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number of experts or their specific qualifications (e.g., years of experience for physicians/nurses performing the reference measurements). It states that the "auscultatory method (mercury sphygmomanometer)" was used as the reference standard, implying trained human operators.

    4. Adjudication Method for the Test Set

    The document states that the mean value and standard deviation of differences between the device and the mercury sphygmomanometer were calculated according to ISO 81060-2:2018. This suggests a direct comparison rather than a multi-reader adjudication process in the traditional sense for medical imaging or diagnostic algorithms. For blood pressure measurement, the reference standard (auscultatory method) intrinsically relies on trained observers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs. Without AI Assistance

    No MRMC comparative effectiveness study was done. This type of study is not applicable to a standalone wrist blood pressure monitor, which directly measures physiological parameters rather than assisting human readers in interpreting complex data.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance study was done for the device's accuracy in measuring blood pressure and pulse rate. The device operates independently to provide these measurements.

    7. The Type of Ground Truth Used

    The ground truth for the clinical accuracy study was established using the auscultatory method with a mercury sphygmomanometer, as described in ISO 81060-2:2018. This is considered a gold standard for non-invasive blood pressure measurement.

    8. The Sample Size for the Training Set

    The document does not specify a separate training set or its sample size. The clinical study described appears to be for validation/testing, not for training a model. Blood pressure monitors typically implement algorithms based on established physiological principles (oscillometry) and are calibrated, rather than "trained" in the machine learning sense with a distinct training dataset.

    9. How the Ground Truth for the Training Set Was Established

    As no specific "training set" for an AI algorithm is mentioned in the context of the device's accuracy claims, this question is not directly applicable. The device's underlying measurement principle (oscillometric method) and its calibration would rely on established scientific and engineering practices, not on ground truth established from a training dataset in the AI sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232815
    Date Cleared
    2023-12-11

    (89 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Wrist Blood Pressure Monitor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Wrist Blood Pressure Monitor is used to measure adult systolic blood pressure and pulse rate. The values are for diagnostic reference only. Suitable for medical institutions (such as hospitals, clinics, health centers, etc.) and home use.

    Device Description

    Wrist Blood Pressure Monitor mainly consist of pressure sensor, air pump, valve, cuff, main board and plastic case. Wrist Blood Pressure Monitor that uses the oscillometric principle to measure your blood pressure and pulse rate. The radial artery in the arm changes from blocked to open as the pressure in the cuff tied around the arm changes from high to low, causing the pressure in the cuff to be superimposed on a series of small pressure pulses. The sphygmomanometer senses these signals and, after certain calculations, finds the systolic and diastolic pressures of the radial artery in the body. The Wrist Blood Pressure Monitor include two models: YJ110, YJ111

    AI/ML Overview

    The provided text describes the non-clinical testing and a clinical validation study for a Wrist Blood Pressure Monitor (models YJ110, YJ111) to demonstrate its substantial equivalence to a predicate device (K223291).

    It is important to note that the document primarily focuses on non-clinical testing (e.g., appearance, safety indices, performance indices, electrical safety, EMC, software verification) and a clinical validation study that demonstrates the accuracy of the device against a reference method, rather than a comparative effectiveness study involving human readers with and without AI assistance. The device is a blood pressure monitor, not an AI-powered diagnostic tool requiring expert interpretation of images. Therefore, some of the requested information regarding AI-specific studies (e.g., MRMC studies, expert adjudication, AI standalone performance) would not be applicable in this context.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a detailed table (Table VII-1) of "Inspecting item" with "Inspection content and acceptance standard" and the "Conclusion". Below is a summary, focusing on performance-related criteria:

    Inspecting Item / Acceptance StandardReported Device Performance (Conclusion)
    Appearance and Structure Test:Pass
    1.1 Correct shape, clean surface, no obvious scratches/damage/deformationPass
    1.2 Text/logo accurate, clear, firmPass
    1.3 Function clearly marked/indicatedPass
    1.4 Reliable connections, free key activityPass
    1.5 Display number clear, no missing/breakingPass
    Identification Requirement:Pass
    2.1 Device Identification (traceability, warnings, instructions, parameters)Pass
    2.2 External Packing (wrist circumference, battery info)Pass
    2.3 Specification (manual content, instructions, common arrhythmias, recalibration, measurement method, affected by posture, storage conditions, warranty, neonates, airbag risks, failure method, disinfection, correlation)Pass
    2.4 Low voltage prompt function (battery label, wristband ID)Pass
    Safety Index:Pass
    3.1 Maximum band pressure (= 0.267kPa/s)Pass
    4.5.2.3 Blow-by (time to drop 34.67kPa to 2kPa
    Ask a Question

    Ask a specific question about this device

    K Number
    K213503
    Date Cleared
    2022-02-18

    (109 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Wrist Blood Pressure Monitor, models AOJ-35A, AOJ-35B, AOJ-35C, AOJ-35D, AOJ-35E

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wrist Blood Pressure Monitor is intended to measure and diastolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique by an inflatable cuff wrapped around the wrist at medical facilities or at home.

    Device Description

    AOJ-35 series wrist blood pressure monitor is designed as a battery driven automatic on-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at wrist within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg. The device also has low voltage indication, which will be triggered when the battery is low. All the models included in this submission follow the same software, same measurement principle and same specifications. The main differences are color of the face shell and keys, which will not affect the safety and effectiveness of the device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (from ISO 80601-2-30 & ISO 81060-2:2013)Reported Device Performance (as per Clinical Data)
    Blood Pressure Accuracy: Mean value of difference between device and reference standard should be less than 5.0 mmHg.Mean value is less than 5.0 mmHg
    Blood Pressure Accuracy: Standard deviation of difference between device and reference standard should be less than 8.0 mmHg.Standard deviation is less than 8.0 mmHg
    Cuff Size Distribution Requirement (ISO 81060-2:2013 clause 5.1.4):
    - Upper half of the cuff: At least 40%54% (Complied)
    - Lower half of the cuff: At least 40%46% (Complied)
    - Upper quarter of the cuff: At least 20%29% (Complied)
    - Lower quarter of the cuff: At least 20%24% (Complied)
    - Upper octal of the cuff: At least 10%11% (Complied)
    - Lower octal of the cuff: At least 10%10% (Complied)
    Blood Pressure Distribution Requirement (ISO 81060-2:2013 clause 5.1.5):
    - Systolic Blood Pressure ≤100 mmHg: At least 5%12% (Complied)
    - Systolic Blood Pressure ≥160 mmHg: At least 5%13% (Complied)
    - Systolic Blood Pressure ≥140 mmHg: At least 20%33% (Complied)
    - Diastolic Blood Pressure ≥100 mmHg: At least 5%16% (Complied)
    - Diastolic Blood Pressure ≤60 mmHg: At least 5%10% (Complied)
    - Diastolic Blood Pressure ≥85 mmHg: At least 20%38% (Complied)
    Non-Invasive Sphygmomanometer Performance EffectivenessComplies with ISO 80601-2-30
    Electrical SafetyComplies with IEC 60601-1, IEC 60601-1-11, and IEC 60601-1-2
    BiocompatibilityComplies with ISO 10993-1, ISO 10993-5, ISO 10993-10 (Cytotoxicity, Skin Sensitization, Skin Irritation)
    Software Verification and ValidationPassed according to FDA Guidance for "major" level of concern

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 100 subjects (49 female, 51 male).
    • Data Provenance: The document does not explicitly state the country of origin of the data. It is a prospective clinical study, as it was conducted "per ISO 81060-2: 2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    The document mentions that the clinical testing was conducted "per ISO 81060-2: 2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type." This standard typically outlines the requirements for validation of automated non-invasive sphygmomanometers against a reference standard. While the document doesn't explicitly state the number of experts, ISO 81060-2 typically requires measurements by trained observers using a mercury-column or auscultatory device as the reference ground truth. The qualifications of these observers would be that they are trained to accurately perform manual blood pressure measurements according to the standard. No specific details about the number or precise qualifications (e.g., "radiologist with 10 years of experience") are provided in this summary, but the reference to the ISO standard implies adherence to its requirements for ground truth establishment.

    4. Adjudication Method for the Test Set:

    The document does not explicitly mention an adjudication method like 2+1 or 3+1. For blood pressure validation, ISO 81060-2 typically involves simultaneous or closely timed measurements by multiple trained observers (often two) and the device under test. The differences between the device readings and the average or adjudicated readings of the observers (or a specific protocol for handling discrepancies) form the basis of the accuracy assessment.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted for this device. The study described focuses on the standalone accuracy of the blood pressure monitor against a reference standard, not its impact on human reader performance or diagnostic accuracy.

    6. Standalone Performance Study:

    Yes, a standalone study (algorithm only without human-in-the-loop performance) was done. The clinical data section explicitly details the performance of the "monitor" (the device) in measuring blood pressure and pulse rate against a reference standard. The "Bench Testing" section also describes tests conducted on the device's technical performance.

    7. Type of Ground Truth Used:

    The ground truth used for the clinical study was established according to ISO 81060-2: 2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. This standard mandates a clinical validation process where the automated device's measurements are compared to measurements taken by trained observers using a validated reference method (e.g., auscultation with a mercury sphygmomanometer). This constitutes an expert-based reference standard rather than pathology or outcomes data.

    8. Sample Size for the Training Set:

    The document does not provide information on the sample size for the training set. This is a premarket notification for a medical device, and the focus of the provided summary is on the clinical validation (testing) and non-clinical data, not on the development or training of an AI algorithm (though the device itself likely contains algorithms, their training data is not detailed here).

    9. How the Ground Truth for the Training Set was Established:

    As no training set information is provided, how its ground truth was established is not detailed in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212312
    Date Cleared
    2022-01-27

    (188 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    RA-T59 Wrist blood pressure monitor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a digital monitor for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 inches to 7 1/4 inches (12.8 cm to 18.5 cm).

    Device Description

    RA-T59 Wrist blood pressure monitor is designed to measure the blood pressure and pulse rate of adults within the range of the designated wrist cuff (12.8 to 18.5 cm). RA-T59 Wrist blood pressure monitor is based on an oscillometric method to measure blood pressure and pulse rate from wrist. The device uses automatic inflation mode. It starts to inflate from 0 mmHg and automatically stops the motor after 25 seconds and quickly deflates, with a steady inflation speed detecting the measurement during inflation. The device utilizes the oscillometric method whereby the electronic pressure sensor converts variation in cuff pressure to electrical signals. The signal from the electronic pressure sensor is conditioned with a circuit before data conversion by an analog-to-digital converter (ADC). The systolic pressure, diastolic pressure, and pulse rate are then calculated in the digital domain. The resulting systolic, diastolic, and pulse-rate measurements are displayed on a LCD and stored in the device's memory.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the RA-T59 Wrist blood pressure monitor, based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The device's performance for blood pressure measurement is evaluated against the requirements of ISO 81060-2:2018. The document does not explicitly present a table of "acceptance criteria" and "reported device performance" in a quantitative manner as one might find in a detailed study report. However, it states that "The results of this clinical investigation show that the proposed device fulfills the requirement of ISO 81060-2:2018."

    The key performance criterion for blood pressure monitors, based on the Accuracy or pressure indicator and Accuracy Pulse Rate sections in the comparison table (page 4), is:

    Acceptance CriteriaReported Device Performance
    Accuracy (Pressure): ±3 mmHg or 2% of readingMeets: The device was validated to ISO 81060-2:2018, which includes these accuracy requirements.
    Accuracy (Pulse Rate): ±5%Meets: The device was validated to ISO 81060-2:2018, which includes these accuracy requirements.

    Note: While the document confirms compliance with ISO 81060-2:2018, specific numerical results (e.g., mean difference and standard deviation between device and reference measurements) that demonstrate this compliance are not provided within this summary for either systolic or diastolic blood pressure, or pulse rate.


    Study Information

    2. Sample size used for the test set and the data provenance:

    • Sample Size: 85 qualified healthy adult subjects.
    • Data Provenance: Not explicitly stated regarding country of origin. The study is described as a "clinical investigation," which generally implies a prospective study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. For blood pressure clinical validation studies like ISO 81060-2, ground truth is typically established by trained and certified observers using a validated reference sphygmomanometer, not usually by "experts" in the context of interpretation (like radiologists). The standard specifies requirements for the observers.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not provided in the document. For blood pressure measurements in clinical validation, ground truth is typically established by two (or more) independent observers. The standard ISO 81060-2:2018 outlines the methodology for taking concurrent auscultatory measurements by these observers.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, an MRMC comparative effectiveness study was not done. This device is an automated blood pressure monitor and does not involve human readers interpreting results or AI assistance in that context. The study performed was a clinical validation of the device's measurement accuracy.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Yes, a standalone performance study was done. The clinical investigation directly assessed the RA-T59 Wrist blood pressure monitor's (algorithm and hardware) ability to accurately measure blood pressure and pulse rate against a reference standard in a clinical setting, without human interpretation of the device's output.

    7. The type of ground truth used:

    The type of ground truth used is implied by the adherence to ISO 81060-2:2018. This standard typically establishes ground truth through concurrent auscultatory measurements performed by trained observers using a reference sphygmomanometer. This is a form of expert-derived, objective measurement used as the "truth" against which the automated device's readings are compared.

    8. The sample size for the training set:

    The document does not mention a training set sample size. This is a clinical validation document for a medical device that measures physiological parameters, not a machine learning or AI algorithm in the typical sense that would have an explicit "training set" for model development. The "clinical investigation" mentioned is for validation, not training.

    9. How the ground truth for the training set was established:

    As no training set is mentioned in the context of this device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Wrist Blood Pressure Monitor, Model Name:W203, Wrist Blood Pressure Monitor, Model Name:W202, Wrist Blood
    Pressure Monitor, Model Name:W1681

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device are digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging form 5 3/8 inches to 7 2/3 inches (13.5cm to 19.5cm) .

    The devices detect the appearance of irregular heatbeats during measurement and gives a warning signal with readings.

    Device Description

    The Wrist Blood Pressure Monitor (model: W203,W202,W1681) are battery-powered, automatic, non-invasive blood pressure system intended for home use. The Wrist Blood Pressure Monitor (model: W203,W202,W1681) are intended for use in adult patient population with wrist circumference ranging from 5 3/8 inches to 7 2/3 inches (13.5 cm). These devices are powered by 2 X AAA battery. The device wrist cuff inflates using an integral pump and deflates via an electric valve. During inflation, the wrist cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic blood pressure. The systolic and diastolic blood pressures are measured using the oscillometric method. The cuff can measure pressure range from 0 to 295 mmHg, and the pulse rate range from 40 to 195 beats/min. When the device detected irregular rhythms, " Y " will display on screen. An irreqular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while your monitor is measuring blood pressure. The WHO blood pressure indicator bar can classify by WHO and ISH recommendation. The devices display the latest blood pressure reading, while up to 2) 99 readings can be stored in memory.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study proving that the device meets these criteria. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for accuracy are derived from ISO 81060-2:2013, which is referenced for the clinical study. The specific criteria are:

    Acceptance CriteriaReported Device Performance
    Accuracy of Pressure: ±3mmHg±3mmHg
    Accuracy of Pulse Rate: ±5% of reading±5% of reading

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The document mentions "clinical accuracy should be checked by application of the tests of ISO 81060-2:2013, we conducted the the clinical study in accordance with this standard." ISO 81060-2 specifies subject recruitment criteria but not a fixed sample size for all studies.
    • Data Provenance: Not explicitly stated. However, since Shenzhen Yuezhongxing Technology Co., Ltd. is based in China, it is reasonable to infer the data provenance is likely China, and the study was conducted to support their submission to the FDA, suggesting a prospective design for regulatory purposes.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not available in the provided text, as the device is a blood pressure monitor, not an AI diagnostic tool requiring expert interpretation for ground truth. Blood pressure monitors establish ground truth through direct physical measurements (e.g., using a auscultatory reference method against which the oscillometric device is compared).

    4. Adjudication Method for the Test Set

    This information is not available in the provided text. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in image analysis or subjective interpretations, which is not applicable to a blood pressure monitor's accuracy testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • Was it done? No. MRMC studies are relevant for evaluating AI systems with human interaction. This device is a standalone blood pressure monitor and does not involve human readers in the output interpretation in the manner an AI diagnostic system would.
    • Effect size of human readers improvement with AI vs. without AI assistance: Not applicable, as no MRMC study was conducted.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    Yes, a standalone performance evaluation was done. The device (Wrist Blood Pressure Monitor) operates independently to measure blood pressure and pulse rate. Its accuracy was tested against established standards (ISO 81060-2:2013, which describes clinical validation of automated sphygmomanometers).

    7. The Type of Ground Truth Used

    The ground truth for blood pressure monitor accuracy studies typically involves:

    • Simultaneous auscultatory measurements performed by trained observers (manual sphygmomanometry) as a reference method, as prescribed by standards like ISO 81060-2.
    • Comparison of the device's readings against these reference measurements to determine accuracy.

    8. The Sample Size for the Training Set

    This information is not available in the provided text. Blood pressure monitors, especially those using oscillometric methods, are typically designed based on physics and signal processing principles rather than extensive machine learning "training sets" in the modern AI sense. While there might be internal development and calibration data, it's not described as a "training set" like in deep learning models.

    9. How the Ground Truth for the Training Set was Established

    This information is not available in the provided text, and the concept of a "training set" and "ground truth" in that context is not directly applicable to a traditional blood pressure monitor as it would be for an AI-powered diagnostic device. Device calibration and validation would typically involve reference measurements against known standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210768
    Date Cleared
    2021-07-30

    (137 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Wrist Blood Pressure Monitor, Blood Pressure Monitor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

    Device Description

    Fully Automatic Electronic Blood Pressure Monitor (KD-743V, KD-743B, KD-752) is designed and manufactured according to IEC 80601-2-30.

    The operational principle is based on oscillometric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure, and display the result on the LCD. The measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user.

    AI/ML Overview

    This document is a 510(k) summary for a blood pressure monitor, not an AI/ML device. Therefore, the detailed information typically required for describing an AI/ML model's acceptance criteria and validation study (such as training/test set sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for AI models) is not provided in this document.

    The document discusses the substantial equivalence of the "Fully Automatic Electronic Blood Pressure Monitor" (models KD-743V, KD-743B, KD-752) to predicate devices (KD-753, K183535 and KD-721, K163276). The "acceptance criteria" here refer to meeting established performance standards for non-invasive blood pressure monitors, rather than the performance metrics of an AI algorithm.

    Here's a breakdown of what can be extracted and what information is not available from the provided text, in the context of your specific questions:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for blood pressure monitors are generally based on international standards like ISO 81060-2, which specifies requirements for the accuracy of automated non-invasive sphygmomanometers. The document states:

    "Accuracy of the blood pressure monitors for the clinical test report was verified by meeting criteria 1 and criteria 2 of ISO 81060-2."

    While the specific numeric criteria (e.g., mean difference and standard deviation of differences between device and reference measurements) are not explicitly listed in a table within this document, meeting ISO 81060-2 criteria 1 and 2 is the acceptance benchmark.

    Acceptance Criteria (General for NIBP Devices per ISO 81060-2)Reported Device Performance
    Criteria 1: Mean difference between the device measurement and the reference measurement, and the standard deviation of differences, for systolic and diastolic blood pressure. (Specific thresholds are in the ISO standard, typically mean difference within ±5 mmHg and standard deviation within 8 mmHg for at least 85 subjects)."Accuracy of the blood pressure monitors for the clinical test report was verified by meeting criteria 1 and criteria 2 of ISO 81060-2." (Implies the device met these statistical requirements, but the specific statistical values are not given in this summary.)
    Criteria 2: Cumulative percentage of subjects for whom the device measurement difference from the reference measurement is within specific ranges (e.g., within ±5 mmHg, ±10 mmHg, ±15 mmHg). (Specific thresholds are in the ISO standard)."Accuracy of the blood pressure monitors for the clinical test report was verified by meeting criteria 1 and criteria 2 of ISO 81060-2." (Implies the device met these threshold requirements, but the specific percentages are not given in this summary.)
    Pulse Rate Accuracy: Within ±5%Within ±5%
    Pressure Accuracy: Within ±3mmHgWithin ±3mmHg

    Detailed Information as Requested for AI/ML Studies:

    1. A table of acceptance criteria and the reported device performance: See table above. This is based on NIBP standards, not AI performance metrics.

    2. Sample sizes used for the test set and the data provenance:

      • The document refers to "clinical test report of KD-7961" and "clinical test report of KD-972" being used as reference for the subject devices. These reports verified accuracy by meeting ISO 81060-2 criteria.
      • ISO 81060-2 typically recommends a minimum of 85 subjects for validation. While not explicitly stated here for the specific test, it's implied that the reference clinical tests adhered to this.
      • Data Provenance: Not specified (e.g., country of origin). The document states the company is based in Tianjin, China.
      • Retrospective or Prospective: Not specified. Clinical validation studies for medical devices are typically prospective.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable in the context of this device study. Ground truth for blood pressure monitors is established by simultaneously measuring blood pressure using a calibrated reference method (e.g., auscultation by trained observers) and the device under test. It's not a consensus-based reading like for imaging. ISO 81060-2 requires a minimum of three trained observers for reference measurements. Their specific qualifications (e.g., years of experience) are not detailed in this summary, beyond "trained observers."

    4. Adjudication method for the test set: Not applicable in the AI/ML sense. For NIBP validation, the "adjudication" is the comparison of device readings against the mean of the reference observers' readings, as per ISO 81060-2. There's no consensus or 2+1/3+1 adjudication of images.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a blood pressure monitor, not an AI-assisted diagnostic imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable in the AI/ML sense. The device itself is "standalone" in that it performs the measurement. The "algorithm" here refers to the oscillometric method of blood pressure determination, not a separate AI algorithm that provides diagnostic interpretations.

    7. The type of ground truth used: Reference blood pressure measurements obtained by trained observers using a reference method (e.g., auscultation), according to ISO 81060-2.

    8. The sample size for the training set: Not applicable. This is a conventional medical device, not an AI/ML device that requires a distinct "training set" for model development. The "algorithm" for blood pressure measurement is based on established oscillometric principles and does not "learn" from a training set in the way an AI model does.

    9. How the ground truth for the training set was established: Not applicable, as there's no training set for this type of device.

    In summary: The provided document is a regulatory submission for a conventional medical device (blood pressure monitor), not an AI/ML product. Therefore, many of the detailed questions regarding AI/ML model validation are not addressed, as they are not relevant to this type of device. The "acceptance criteria" revolve around established performance standards for blood pressure measurement accuracy (ISO 81060-2).

    Ask a Question

    Ask a specific question about this device

    K Number
    K202599
    Date Cleared
    2020-12-06

    (89 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Transtek Wrist Blood Pressure Monitor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Transtek Blood Pressure Monitor intended for use in measuring blood pressure and heartbeat rate with a wrist circumference ranging from 13.5cm to 23cm. It is intended for adult, indoor use only.

    Device Description

    Transtek Wrist Blood Pressure Monitor, TMB-2072 is designed to measure the systolic and diastolic blood pressure and heartbeat rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. Measurement method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating heartbeat rate, which is a wellknown technique in the market called the "Oscillometric method". Transtek Wrist Blood Pressure Monitor is single-mounted devices of the main unit and cuff unit. The preformed cuff unit, which is applicable to wrist circumference approximately between 13.5 and 23 cm, includes the inflatable bladder and nylon shell. The device consists of the microprocessor, the pressure sensor, the operation keys, the pump and the electromagnetic deflation control valve. The subject device is powered by 3.7V Built-in rechargeable lithium-ion battery. The device also compares the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%. Transtek Wrist Blood Pressure Monitor TMB-2072 embeds a Bluetooth module that allows it to connect to nearby BT receiving end. The mobile application displays results. And once measurement is over, the device will start transmission data by BT. Thus, users can receive, and display/storage, measurement data from TMB-2072 unit through their end devices that embedded BT module.

    AI/ML Overview

    The provided text describes the 510(k) submission for the Transtek Wrist Blood Pressure Monitor (Model TMB-2072). Here's an analysis of the acceptance criteria and study information:

    1. Acceptance Criteria and Reported Device Performance

    The primary clinical performance acceptance criteria for the blood pressure monitor are based on the ISO 81060-2:2018 standard for clinical validation of automated measurement type. While the specific numerical acceptance criteria (e.g., mean difference and standard deviation thresholds) are not explicitly stated in the document, the conclusion indicates that the device met these criteria.

    Acceptance Criterion (Based on ISO 81060-2:2018)Reported Device Performance
    Mean error and standard deviation of differences for systolic blood pressure not exceeding specified limits."All data's mean error and standard deviation of differences for systolic... pressure is not over the limits of ISO 81060-2: 2018"
    Mean error and standard deviation of differences for diastolic blood pressure not exceeding specified limits."...and diastolic pressure is not over the limits of ISO 81060-2: 2018"
    Accuracy better than the predicate device."and the accuracy of subject device is better than predicate device."
    No adverse effects/complications during the study."No adverse effect and/or complication is found in this study."

    Additional Performance Criteria (Non-Clinical):
    The device also claims compliance with various non-clinical standards for electrical safety, electromagnetic compatibility, usability, and wireless testing. These are not presented in a table with specific acceptance values, but rather as statements of compliance.

    2. Sample Size and Data Provenance

    • Sample Size for the Test Set: 85 qualified healthy adult subjects.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical study data or whether it was retrospective or prospective. However, based on the manufacturer's location (Guangdong, China) and the submission's context, it's highly likely the study was conducted prospectively in China.

    3. Number of Experts and Qualifications for Ground Truth

    The document specifies that the clinical study was conducted according to ISO 81060-2:2018. This standard typically requires "reference blood pressure measurements obtained by two trained observers using a reference sphygmomanometer against which the automated device measurements are compared."

    • Number of Experts: The document does not explicitly state the number of experts used to establish the ground truth. However, based on the adherence to ISO 81060-2, it would involve at least two trained observers (likely medical professionals or technicians specifically trained in auscultatory blood pressure measurement).
    • Qualifications of Experts: Not specified in the document, but ISO 81060-2 mandates "trained observers." These observers would be qualified to perform accurate auscultatory blood pressure measurements.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an "adjudication method" in the context of expert consensus (like 2+1 or 3+1). For blood pressure measurement validation under ISO 81060-2, the ground truth is established by concurrent measurements from reference sphygmomanometers read by trained observers. The standard outlines specific methods for comparing the automated device readings against these observer readings. Discrepancies between the two observers' readings typically have pre-defined limits within the standard, and if these limits are exceeded, the measurements might be discarded or re-taken. This is more of a concordance assessment rather than adjudication to resolve differing opinions on a single diagnosis.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a standalone performance validation of the device against a reference standard (observers), not an assessment of human readers' improvement with or without AI assistance. The device in question is an automatic measuring device, not an AI-assisted diagnostic tool for interpretation by human readers.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The entire clinical study described, based on ISO 81060-2:2018, is a standalone (algorithm only without human-in-the-loop performance) assessment of the device's accuracy in measuring blood pressure. The device's measurements were compared directly to the reference measurements obtained by trained observers.

    7. Type of Ground Truth Used

    The type of ground truth used was expert consensus (from trained observers) using reference sphygmomanometers, as mandated by the ISO 81060-2:2018 standard. This is considered a gold standard for blood pressure measurement validation.

    8. Sample Size for the Training Set

    The document does not provide information on the sample size for a training set. This submission is for a medical device that measures blood pressure using the "Oscillometric method," which is a well-established technique. While the internal algorithms of such devices are developed and likely refined using data, the document focuses on the clinical validation of the final device, not the development data. Therefore, details about a "training set" in the context of, for instance, a deep learning AI model, are not applicable or provided here.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, information on a "training set" in the context of an AI model is not provided. For the clinical validation (test set) ground truth, it was established by trained human observers using reference sphygmomanometers following the protocols of ISO 81060-2:2018.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 6