Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wrist Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique by an inflatable cuff wrapped around the wrist at medical facilities or at home.

    Device Description

    The Wrist Blood Pressure Monitor is designed as a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at wrist within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa.

    All the models included in this submission follow the the same intended use, same measurement principle, same blood pressure core algorithm and similar product design. All the models can be used with one cuff size 13.5~19.5 cm (5.3-7.7inches).

    The main differences are appearance, Dimensions and some specifications which will not affect the safety and effectiveness of the device.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and associated summary pertain to a Wrist Blood Pressure Monitor, which is a medical device for measuring blood pressure and pulse rate. It is not an AI/Software as a Medical Device (SaMD). Therefore, many of the typical acceptance criteria and study designs associated with AI/SaMD (such as multi-reader multi-case studies, ground truth establishment by experts, training set details, or effect sizes of AI assistance) are not applicable to this device.

    The acceptance criteria and study details provided are tailored to the performance of a non-invasive blood pressure measurement system (hardware device), focusing on accuracy, safety, and effectiveness.

    Here's a breakdown of the requested information based on the provided document, addressing the device's specific characteristics as a hardware blood pressure monitor:


    Acceptance Criteria and Device Performance (Wrist Blood Pressure Monitor)

    1. Table of Acceptance Criteria and Reported Device Performance

    As per the 510(k) summary, the device's accuracy is a key performance metric. The acceptance criteria are based on the international standard ISO 81060-2 Third edition 2018-11 [Including AMD1:2020].

    Performance MetricAcceptance Criteria (from ISO 81060-2)Reported Device Performance
    Blood Pressure AccuracyMean error and standard deviation of differences for systolic and diastolic pressure not over the limits specified in ISO 81060-2.All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2.
    Heart Rate Accuracy± 5% of reading± 5% of reading (Same as Predicate, implying met for proposed)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set (Clinical Accuracy Study):
      • Three groups of clinical accuracy research were conducted. Each group included 100 subjects, for a total of 300 subjects across the 13 models.
      • Group 1: 100 subjects (47 Male, 53 Female)
      • Group 2: 100 subjects (54 Male, 46 Female)
      • Group 3: 100 subjects (44 Male, 56 Female)
      • Minimum subjects for each group was 85, as per ISO 81060-2.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical data. However, the manufacturer is "Shenzhen AOJ Medical Technology Co., Ltd." in Shenzhen, Guangdong, China. It is highly probable the data was collected in China.
    • Retrospective or Prospective: The clinical accuracy study, designed to meet ISO 81060-2, is typically conducted prospectively to collect new data for device validation. The wording "clinical accuracy research" and "clinical accuracy test report and data analysis followed the requirements" implies a prospective study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • This question is not directly applicable in the context of this device because "ground truth" for a blood pressure monitor's accuracy is established against a reference standard, not through expert consensus on interpretations of images or signals (as would be the case for AI/SaMD).
    • For blood pressure monitors, the "ground truth" or reference measurement is typically taken by trained medical professionals using a standardized reference sphygmomanometer (e.g., mercurial or auscultatory method), as per the ISO 81060-2 standard. The document states "The Same Arm Sequential Method was chosen for all studies," which is a standard procedure comparison method against a reference device. The qualifications of the individuals performing these reference measurements would be trained clinicians (e.g., physicians, nurses).

    4. Adjudication Method for the Test Set

    • This question is not applicable for a blood pressure monitor's accuracy testing. Adjudication methods (like 2+1 or 3+1) are used to resolve discrepancies in human expert interpretations, especially in image-based diagnostics.
    • For blood pressure accuracy, deviations are quantified statistically between the device reading and the reference measurement, not through an adjudication process among multiple "readers."

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, this was not done. This type of study is relevant for AI/SaMD devices where AI assists human interpretation and is a key component for assessing the AI's clinical utility. The Wrist Blood Pressure Monitor is a standalone hardware device that provides a measurement; it does not involve human "readers" interpreting data or AI assistance.

    6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    • Yes, in essence. The entire clinical accuracy study described in Section 8 ("Clinical data") evaluates the device's performance (which incorporates its internal oscillometric algorithm) in a standalone manner against a reference standard. The "algorithm" here refers to the embedded software that processes the oscillometric signals to derive blood pressure and pulse rate. The study directly assesses how accurately the device (with its integrated algorithm) measures blood pressure readings compared to the reference.
    • Performance Metrics: The evaluation was based on the "mean error and standard deviation of differences for systolic, diastolic pressure" as per ISO 81060-2.

    7. The Type of Ground Truth Used

    • The ground truth for the clinical accuracy testing was established through concurrent measurements using a standardized reference method (e.g., auscultatory method with a mercurial sphygmomanometer or another validated reference device) on the same arm, sequentially with the test device. This is the standard for blood pressure monitor validation as per ISO 81060-2.
    • It is not "expert consensus" in the sense of subjective medical interpretation, but rather an objective, standardized measurement performed by trained personnel using a calibrated reference instrument.

    8. The Sample Size for the Training Set

    • This concept is not applicable to this type of medical device clearance. The Wrist Blood Pressure Monitor is a hardware device with an embedded algorithm (oscillometric technique) that is based on established physiological principles. It doesn't use machine learning or deep learning in a way that requires a separate "training set" of patient data for an AI model to learn from, as would be the case for AI/SaMD devices. The device's "training" (development and calibration) would involve engineering principles and laboratory testing, rather than a data-driven machine learning process.

    9. How the Ground Truth for the Training Set Was Established

    • As the concept of a "training set" in the context of machine learning is not applicable here (see point 8), the establishment of ground truth for such a set is also not applicable. The device's underlying measurement principle is well-established oscillometric technology. Development and calibration rely on physical models and engineering validation.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241634
    Date Cleared
    2025-02-11

    (250 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wrist Blood Pressure Monitor is intended to measure the systolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home. The intended wrist circumference is 13.5-21.5 cm.

    Device Description

    The Wrist Blood Pressure Monitor, including XY-W01B, can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person with wrist circumference of 135-215mm through oscillometric method. The initial inflation pressure of the cuff is zero pressure. When start the cuff will be inflated and deflated. The Wrist Blood Pressure Monitor is composed of monitor unit and wrist cuff. Of which the monitor unit contains main control circuit board, air pump, deflation valve, LCD and shell. The two models XY-W01A and XY-W01B have the same intended use, working principle, measuring range, accuracy, cuff, and conformance standard, except for product appearance. The devices are powered by 2 AAA batteries. The Wrist Blood Pressure Monitor has a memory function that automatically stores 90 sets data of the latest measurements.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the Wrist Blood Pressure Monitor (models XY-W01A, XY-W01B).

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard: ISO 81060-2:2018)Reported Device Performance (Clinical Test)
    Mean error for systolic pressure ≤ ±5 mmHgMean error within ±5 mmHg
    Standard deviation for systolic pressure ≤ 8 mmHgStandard deviation within 8 mmHg
    Mean error for diastolic pressure ≤ ±5 mmHgMean error within ±5 mmHg
    Standard deviation for diastolic pressure ≤ 8 mmHgStandard deviation within 8 mmHg
    All data not over the limits.All data is not over the limits.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 120 subjects (53 female and 67 male)
    • Data Provenance: The document does not explicitly state the country of origin.
    • Retrospective or Prospective: Not explicitly stated, but clinical tests for device approval are typically prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided in the document. For blood pressure monitors, the "ground truth" (reference measurements) is typically established by trained technicians using a standardized auscultatory method, not necessarily by "experts" in the sense of physicians or radiologists establishing a diagnostic truth.

    4. Adjudication Method

    This information is not provided in the document, as it relies on direct comparative measurements rather than expert adjudication of imaging or diagnostic results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is more relevant for diagnostic imaging AI systems where human readers interpret images. For a blood pressure monitor, the performance is compared against a reference measurement.

    6. Standalone Performance

    Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The clinical test evaluated the device's accuracy in measuring blood pressure and pulse rate against a reference standard.

    7. Type of Ground Truth Used

    The ground truth implicitly used is reference blood pressure measurements, likely obtained through a rigorously standardized auscultatory method as per the ISO 81060-2:2018 standard.

    8. Sample Size for the Training Set

    This information is not provided in the document. Given that this is a hardware device with an embedded algorithm (oscillometric technique), there isn't typically a separate "training set" in the way an AI/ML model would have for image recognition. The core algorithm's parameters are likely developed through extensive engineering and testing rather than a distinct training phase on a large dataset.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided. As mentioned above, the concept of a "training set ground truth" as understood in AI/ML is not directly applicable here. The device's underlying oscillometric algorithm and its parameters would have been developed and refined through engineering principles and testing against various blood pressure profiles and body types.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242568
    Date Cleared
    2025-01-30

    (154 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use in measuring blood pressure in adult patient population with wrist circumference ranging from 13.0 cm to 21.0 cm.

    Device Description

    The Wrist Blood Pressure Monitor is a reusable, active, non-sterile, non-invasive and non-implantable device for use in measuring blood pressure in adult patient population with wrist circumference ranging from 13.0 cm to 21.0 cm. It is a home-use device prohibited from use in MRI (magnetic resonance imaging) environment. This device has the systolic and diastolic display features. The device is of the watch type and powered by a rechargeable lithium-polymer battery. An AC adapter is used for charging the device, but the device cannot be operated while charging. The device wrist cuff inflates using an integral pump, and deflates via an electric valve. During inflation, the wrist cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines systolic and diastolic blood pressure. The systolic and diastolic blood pressures are measured using the oscillometric method. The cuff can measure pressure range from 0 to 300mmHg. Product includes device (preinstalled with a strap and large-sized cuff: 16.6-21.0cm), strap with a medium-sized cuff: 13.0-16.5cm, wrist measuring tape, charging base (with the power cable). The product functional configurations of OHMS11 and OHMS12 are completely identical, and the difference between the two is reflected in the color (OHMS11 is the dark gray color, OHMS12 is the champagne color).

    AI/ML Overview

    The provided text describes the acceptance criteria and a study for a Wrist Blood Pressure Monitor (OHMS11, OHMS12).

    Here's an analysis of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Clinical Accuracy (ISO 81060-2.2019): Specific numerical criteria for mean difference and standard deviation between device and reference measurement for both systolic and diastolic blood pressure (as outlined in ISO 81060-2)."The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2." (Specific numerical values for mean difference and standard deviation are not provided in the document, only a statement of compliance).
    Electrical Safety (IEC 60601-1:2020 and IEC 80601-2-30)"shown to comply with IEC 60601-1:2020 and IEC 80601-2-30 for electrical safety."
    Electromagnetic Compatibility (EMC) (IEC 60601-1-2.2020)"shown to comply with IEC 60601-1-2.2.2020 for electromagnetic compatibility."
    Performance Testing (Product Design Specifications)"Data generated from the test met the predetermined acceptance criteria." (Specific acceptance criteria and performance data are not provided).
    Software Verification and Validation (FDA Guidance)"in compliance with FDA Guidance-Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
    Cybersecurity (FDA Guidance)"in compliance 'Content of Premarket Submissions for Management of Cybersecurity in Medical Devices'."
    Biocompatibility (ISO10993-1, including Cytotoxicity, Sensitization, Irritation)Biocompatibility evaluation was conducted for Cytotoxicity, Sensitization, and Irritation. (Implied compliance with ISO10993-1 as it refers to the guidance directly related to this standard).

    2. Sample size used for the test set and the data provenance

    • Sample Size: 85 patients (49 males and 36 females)
    • Data Provenance: Clinical study, conducted prospectively as patients "volunteered to take part in the subjects completed the clinical study." The country of origin is not explicitly stated in the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number of experts used or their qualifications for establishing the ground truth. It mentions that "manual Mercury Sphygmomanometer was used as a reference device" and that the study followed ISO 81060-2. This standard generally requires trained observers for the reference measurements, but specific expert details are absent.

    4. Adjudication method for the test set

    The document states, "Same arm sequential method was adopted during the clinical testing." This indicates a specific measurement protocol where both the test device and the reference device take measurements sequentially on the same arm. There is no mention of an adjudication method involving experts for discrepancies in ground truth, as the ground truth is established by the reference device measurement.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool requiring human reader evaluation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance study was done. The clinical testing evaluates the device's accuracy (algorithm only) against a reference standard (manual mercury sphygmomanometer) without direct human intervention in interpreting the device's output. The device itself performs the measurement and provides the blood pressure readings.

    7. The type of ground truth used

    The ground truth used was comparison to a reference device, specifically a "manual Mercury Sphygmomanometer," following the methodology of ISO 81060-2. This is a clinical standard for validating automated blood pressure monitors.

    8. The sample size for the training set

    The document does not provide information regarding a training set sample size. This type of submission (510(k) for a hardware device with embedded algorithms) typically focuses on validation data rather than detailed training information for AI/ML models.

    9. How the ground truth for the training set was established

    Since no training set information is provided, there is no information on how its ground truth would have been established. It is assumed the device relies on established oscillometric principles rather than a deep learning model requiring a large labeled training set.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 3/8 inches (13.5cm to 19.5cm). The device detects the appearance of irregular hearbeats during measurement and gives a warning signal with reading.

    Device Description

    The Wrist Blood Pressure Monitor,Models: 222,222BT,W1681BL, W1681BT, W1681R are powered by a rechargable Li-ion battery, automatic, non-invasive blood pressure system intended for home use. The devices are intended for use in adult patient population with wrist circumference ranging from 5 3/8 inches to 7 2/3 inches (13.5 cm). The devices' wrist cuff inflates using an integral pump and deflates via an electric valve. During deflation, the wrist cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic blood pressure. The systolic and diastolic blood pressures are measured using the oscillometric method. The cuff can measure pressure range from 0 to 295 mmHg, and the pulse rate range from 40 to 199 beats/min.

    When the device detected irregular rhythms, " W " will display on screen. An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while your monitor is measuring blood pressure. The WHO blood pressure indicator bar can classify by WHO and ISH recommendation. The devices display the latest blood pressure reading, while up to 2 x 99 readings can be stored in memory.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study proving the device meets them, based on the provided text:

    Device Name(s): Wrist Blood Pressure Monitor (Models: 222, 222BT, W1681BL, W1681BT, W1681R)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard / Requirement)Reported Device Performance (Subject Device)
    ISO 81060-2:2018 Criterion 1: Mean difference (device vs. reference)Systolic BP: 1.0 mmHg Diastolic BP: 0.9 mmHg
    ISO 81060-2:2018 Criterion 1: Standard deviation of differences (device vs. reference)Systolic BP: 3.1 mmHg Diastolic BP: 3.4 mmHg
    ISO 81060-2:2018 Criterion 2: Standard deviation (device vs. reference)Systolic BP: 2.4 (from Table 1, assumed to be an overall standard deviation) Diastolic BP: 2.6 (from Table 1, assumed to be an overall standard deviation)
    Cuff Pressure Accuracy±3 mmHg
    Pulse Accuracy±5% of reading
    ISO 81060-2:2018 standard acceptanceMet
    IEC 60601-1:2005+A1:2012+A2:2020 (Basic Safety & Essential Performance)Complies
    IEC 60601-1-11:2015 (Home Healthcare Environment)Complies
    IEC 80601-2-30:2018 (Automated Non-Invasive Sphygmomanometers)Complies
    IEC 60601-1-2:2014 (Electromagnetic Compatibility)Complies
    ISO 10993-10:2010 (Biocompatibility - Irritation/Sensitization)Complies
    ISO 10993-5:2009 (Biocompatibility - Cytotoxicity)Complies
    FDA Software Validation Guidance (General Principles of Software Validation)Complies (for moderate level of concern)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 86 subjects.
    • Data Provenance: Clinical investigation conducted by Shenzhen Cihai Hospital (China), prospective study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number of experts or their specific qualifications (e.g., years of experience for physicians/nurses performing the reference measurements). It states that the "auscultatory method (mercury sphygmomanometer)" was used as the reference standard, implying trained human operators.

    4. Adjudication Method for the Test Set

    The document states that the mean value and standard deviation of differences between the device and the mercury sphygmomanometer were calculated according to ISO 81060-2:2018. This suggests a direct comparison rather than a multi-reader adjudication process in the traditional sense for medical imaging or diagnostic algorithms. For blood pressure measurement, the reference standard (auscultatory method) intrinsically relies on trained observers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs. Without AI Assistance

    No MRMC comparative effectiveness study was done. This type of study is not applicable to a standalone wrist blood pressure monitor, which directly measures physiological parameters rather than assisting human readers in interpreting complex data.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance study was done for the device's accuracy in measuring blood pressure and pulse rate. The device operates independently to provide these measurements.

    7. The Type of Ground Truth Used

    The ground truth for the clinical accuracy study was established using the auscultatory method with a mercury sphygmomanometer, as described in ISO 81060-2:2018. This is considered a gold standard for non-invasive blood pressure measurement.

    8. The Sample Size for the Training Set

    The document does not specify a separate training set or its sample size. The clinical study described appears to be for validation/testing, not for training a model. Blood pressure monitors typically implement algorithms based on established physiological principles (oscillometry) and are calibrated, rather than "trained" in the machine learning sense with a distinct training dataset.

    9. How the Ground Truth for the Training Set Was Established

    As no specific "training set" for an AI algorithm is mentioned in the context of the device's accuracy claims, this question is not directly applicable. The device's underlying measurement principle (oscillometric method) and its calibration would rely on established scientific and engineering practices, not on ground truth established from a training dataset in the AI sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182166
    Date Cleared
    2018-11-08

    (90 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5.3 inches (13.5cm to 21.5cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

    Device Description

    The Wrist Blood Pressure Monitor Model BP4350 ("BP4350") is a battery-powered, automatic, noninvasive, wrist-worn blood pressure measuring system intended for over-the-counter (OTC) home use. BP4350 is designed for wrist circumference ranging from 5.3 inches to 8.5 inches (13.5cm to 21.5cm). The systolic and diastolic blood pressures are measured using the oscillometric method, where the cuff is inflated with an integral controllable Piezoelectric pump and deflates via an electric automatic rapid deflation valve. During inflation, the cuff pressure is monitored, and pulse waveform data is extracted. The extracted pulse waveform data is then further analyzed by software which determines pulse rate, as well as systolic and diastolic blood pressure. The cuff can measure pressure range from 0 to 299mmHg, and the pulse rate range from 40 to 180 beats/min.

    The device also detects the appearance of irregular heartbeats during the blood pressure measurement process, which is defined as the appearance of two (2) or more heartbeat intervals which differ by greater than 25% from the average heartbeat rhythm. Detection of such irregular rhythms would result in an "irregular heartbeat symbol" displayed along with the blood pressure and pulse rate readings. In addition, a "HIGH" indication appears if the blood pressure recorded is greater than 130/80 based on the AHA/ACC High Blood Pressure Clinical Practice Guideline criterion for Stage 1 Hypertension, published in 2017.

    In addition, the device includes an Advanced Positioning Sensor (APS) feature known as the Heart Zone Indicator, which aids the user in determining if the Wrist Cuff is at the correct height in relation to the heart. It makes this determination based on the reading of an accelerometer (integral to the device) to measure the angle of the arm in relation to the table. The APS feature is similar to the predicate device.

    The device displays the latest blood pressure reading, while up to 100 readings can be stored in memory. The operation of the device is intended for home use. Functions and other features that are controlled by the end user include applying the wrist cuff to the wrist, powering on/off the system, starting or stopping the blood pressure and pulse measurement cycle, and installing and changing the batteries as needed. As an optional feature, the user can also pair the BP4350 to a smartphone when employing the "Omron connect" app. This app is an optional feature and is only intended to display trend graphs of measured systolic and diastolic blood pressure, and pulse rate. This app does not provide any diagnostic or measurement functions and does not interpret or analyze the data for medical decision making. Unlimited readings can be stored in the app for archiving and review by the user. Aside from this optional app for smartphones, BP4350 does not connect with other collateral devices.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Omron Healthcare, Inc. Wrist Blood Pressure Monitor Model BP4350. This document primarily focuses on establishing substantial equivalence to a predicate device rather than detailing a study that proves the device meets specific acceptance criteria in the context of an AI/ML-driven device.

    Therefore, much of the requested information regarding acceptance criteria for an AI/ML device, sample sizes for test sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone algorithm performance, and training set details cannot be extracted from this document as it is not an AI/ML device and the study described is a traditional clinical validation for a blood pressure monitor.

    However, I can extract the information relevant to a traditional medical device clinical validation study as presented in the document:


    1. A table of acceptance criteria and the reported device performance

    The document refers to the acceptance criteria implicitly by stating adherence to the ANSI/AAMI/ISO 81060-2:2013 standard for non-invasive sphygmomanometers. This standard dictates the accuracy requirements for blood pressure monitors.

    Acceptance Criteria (from ANSI/AAMI/ISO 81060-2:2013, implied)Reported Device Performance
    Blood Pressure Accuracy: Mean difference (device - reference) ≤ ±5 mmHg. Standard deviation of difference ≤ 8 mmHg. (These are the primary statistical requirements of the standard)"The results demonstrated that BP4350 performed equivalently to the auscultation method and is in conformance with ANSI/AAMI/ISO 81060-2:2013." (Specific numerical results are not provided in this summary, but conformance implies meeting the standard's criteria).
    Pulse Rate Accuracy: Within ±5% of reading (stated as a specification in comparison table)"Accuracy of pulse rate is ±5% in both devices." (Implies the proposed device meets this).
    Irregular Heartbeat Detection: The device detects and gives a warning signal when two or more heartbeat intervals differ by greater than 25% from the average heartbeat rhythm."The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings." (Functionality confirmed, but specific performance metrics for this feature like sensitivity/specificity are not provided in this summary).
    AHA/ACC Hypertension Indicator Threshold: "HIGH" indication if BP > 130/80 mmHg."a 'HIGH' indication appears if the blood pressure recorded is greater than 130/80 based on the AHA/ACC High Blood Pressure Clinical Practice Guideline criterion for Stage 1 Hypertension, published in 2017." (Functionality confirmed).

    2. Sample sized used for the test set and the data provenance

    • Sample Size: The document does not explicitly state the exact sample size (number of subjects) used for the clinical investigation. ANSI/AAMI/ISO 81060-2:2013 typically requires a minimum of 85 subjects for full validation.
    • Data Provenance: Not specified in the provided text (e.g., country of origin). The study was a "clinical investigation" which implies it was prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • The ground truth was established by "trained medical staff" using an "auscultation method" with a "calibrated sphygmomanometer."
    • The exact number of experts and their specific qualifications (e.g., medical doctors, nurses, with how much experience) are not detailed in this summary. The ANSI/AAMI/ISO 81060-2:2013 standard often requires multiple observers to mitigate bias.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • The document does not specify any adjudication method for the clinical investigation results. For blood pressure validation, direct comparison to reference measurements from trained observers is the primary method, and the standard outlines protocols for this.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC study was not done. This study is for a blood pressure monitor, not an AI/ML-driven diagnostic device that relies on human readers interpreting images with or without AI assistance. The device directly measures blood pressure.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is a standalone device that provides direct blood pressure readings. Its performance is measured independently against a reference standard (auscultation). So, in a sense, its "standalone" performance is what was evaluated. There is no "human-in-the-loop" interaction in the measurement process itself that would alter the device's reading.

    7. The type of ground truth used

    • The ground truth was established by expert auscultation (manual blood pressure measurement by trained medical staff) using a calibrated sphygmomanometer. This is the gold standard for clinical validation of automated blood pressure devices.

    8. The sample size for the training set

    • Not applicable. This device is a traditional blood pressure monitor, not an AI/ML device that requires a "training set" in the machine learning sense. Its design and algorithm are based on established oscillometric principles and engineering, not on learning from a large dataset.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182127
    Date Cleared
    2018-11-02

    (88 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5.3 inches to 8.5 inches (13.5 cm to 21.5 cm).

    The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

    Device Description

    The Wrist Blood Pressure Monitor (BPM) Model BP6100 ("BP6100") is an automatic, non-invasive, wrist-worn blood pressure measurement system for over-the-counter (OTC) home use. During measurements, an electric pump within the main unit slowly inflates the wrist cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The systolic and diastolic pressures are determined via the oscillometric cuff method. The cuff pressure range is 0 to 299mmHg and the pulse rate range is 40 to 180 beats/min.

    The BP6100 is intended for use in the adult patient population, in wrist circumferences ranging between 5.3 - 8.5 inches (13.5 - 21.5 cm). During measurement, the device also detects the appearance of irregular heartbeat rhythms. BP6100 further includes administrative functions for automatic storage of the 60 most recent readings taken by the device. The device also provides an "average" function, which can calculate an average reading based on the latest two (2) or three (3) readings, when taken within a 10-minute span.

    The operations of BP6100 are primarily through use of buttons. Functions and other features that are controlled by the end user include applying the wrist cuff to the wrist, powering on/off the system, starting or stopping measurements, and installing and changing batteries as needed. It has no potential connection to external power. There are no additional parts or accessories to the BP6100. The BP6100 does not connect to any other devices.

    AI/ML Overview

    The provided document is a 510(k) summary for the Omron Healthcare, Inc. Wrist Blood Pressure Monitor Model BP6100. It details the device's indications for use, technological characteristics, and performance data to demonstrate substantial equivalence to a predicate device (Omron Healthcare, Inc. HEM-6131).

    Based on the provided information, here's a description of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for the Omron BP6100 are primarily aligned with the standards used for the predicate device, K131742 (HEM-6131), and the ANSI/AAMI/ISO 81060-2:2013 standard.

    FeatureAcceptance Criteria (based on predicate and standards)Reported Device Performance (BP6100)
    Blood Pressure AccuracyWithin ±3 mmHg (as per predicate and general BP monitor accuracy standards like ANSI/AAMI/ISO 81060-2:2013)Clinical Investigation: "The results demonstrated that BP6100 performed equivalently to the auscultation method and is in conformance with ANSVAAMI/ISO 81060-2:2013." Bench Testing: "Comparative blood pressure and pulse rate testing to the predicate device" was conducted and deemed acceptable, and "performance verification testing of BP6100 to confirm acceptable performance of device features and functions" was also performed. The "Accuracy" row in the comparison table explicitly states "Blood Pressure: Within ±3mmHg".
    Pulse Rate AccuracyWithin ±5% of reading (as per predicate)The "Accuracy" row in the comparison table explicitly states "Pulse Rate: Within ±5 % of reading". Bench testing included "Comparative blood pressure and pulse rate testing to the predicate device" and "Performance verification testing of BP6100 to confirm acceptable performance of device features and functions."
    Cuff Pressure Range0 to 299 mmHg0 to 299 mmHg (stated in Device Description and Comparison Table)
    Pulse Rate Range40 to 180 beats/min40 to 180 beats/min (stated in Device Description and Comparison Table)
    Wrist Circumference Range5.3 inches to 8.5 inches (13.5 cm to 21.5 cm)5.3 inches to 8.5 inches (13.5 cm to 21.5 cm) (stated in Indications for Use and Device Description)
    Irregular Heartbeat DetectionDevice detects irregular heartbeats and provides a warning signal."The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings." (Stated in Indications for Use and Device Description)
    BiocompatibilityCompliance with ISO 10993-1 requirements for patient-contacting materials."Biocompatibility of patient-contacting materials per ISO 10993-1 requirements" was conducted. The collective results demonstrate "the materials chosen... meet the established specifications necessary for consistent performance."
    Electrical Safety / EMCMeet applicable electrical safety, electromagnetic compatibility, and electrostatic discharge standards."Electrical safety, electromagnetic compatibility, and electrostatic discharge testing" was conducted. The collective results demonstrate "the materials chosen... meet the established specifications necessary for consistent performance."
    Software V&VSoftware verification and validation to ensure proper functionality."Software verification and validation" was conducted. The collective results demonstrate "the materials chosen... meet the established specifications necessary for consistent performance."
    Cleaning VerificationDevice retains performance when cuff is cleaned with household detergents."Cleaning verification testing to confirm device retains its performance when cuff is cleaned with household detergents as may be required in home use environment" was conducted. The collective results demonstrate "the materials chosen... meet the established specifications necessary for consistent performance."
    Equivalence to PredicateThe device should be substantially equivalent to the predicate device (HEM-6131) in terms of safety and effectiveness."The proposed BP6100 device and the HEM-6131 predicate device have similar technological characteristics... The conclusions drawn from the nonclinical and clinical tests performed in support of BP6100 demonstrate that the device is safe and effective, and performs at least as safely and effectively as the legally marketed HEM-6131 predicate device." Minor differences "do not raise any different questions of safety or effectiveness."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not explicitly state the numerical sample size for the clinical investigation. It mentions that the study was conducted "in accordance with guideline per ANSI/AAMI/ISO 81060-2:2013 Non-invasive sphygmomanometers — Part 2: Clinical investigation of automated measurement type." This standard typically specifies minimum sample sizes (e.g., 85 subjects for validation in the general population). While the exact number isn't present, compliance with this standard implies a sufficient sample size was used.
    • Data Provenance: The document does not specify the country of origin of the data. It also does not explicitly state if the study was retrospective or prospective, but clinical investigations for device validation like this are almost universally prospective studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • The ground truth for blood pressure measurement in the clinical investigation was established using an "auscultation method using a calibrated sphygmomanometer by trained medical staff."
    • The document does not specify the exact number of experts (medical staff) or their specific qualifications (e.g., years of experience, specific medical degree beyond "trained medical staff"). However, for such studies, it implies that the personnel performing the auscultation measurements are qualified and trained according to standard medical practices for blood pressure measurement.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • The document does not describe a specific adjudication method like "2+1" or "3+1." The ground truth was established by "trained medical staff" using the auscultation method. In the context of a clinical investigation for a blood pressure monitor validating against an auscultatory reference, the comparison is typically direct, and an adjudication process beyond the core measurement methodology itself is not usually required or described in a 510(k) summary for this type of device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool that would typically involve human readers interpreting images or data with and without AI assistance. The study was a clinical investigation validating the accuracy of the device against a reference method (auscultation), not evaluating human performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone performance evaluation was indeed done. The clinical investigation directly validated the accuracy of the BP6100 device's oscillometric measurements ("algorithm only") against the auscultation method. This represents the device's performance without "human-in-the-loop" assistance in the measurement itself, although a human operates the device. The non-clinical bench testing also evaluated the device's standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The primary ground truth for the clinical investigation was expert-performed auscultation using a calibrated sphygmomanometer. This is considered the reference standard for non-invasive blood pressure measurement.

    8. The sample size for the training set

    • The document describes a clinical investigation and bench testing for device validation. It does not mention a "training set" or "test set" in the context of machine learning (AI/ML models). This is a traditional medical device, not an AI device. Therefore, there is no AI model training set.

    9. How the ground truth for the training set was established

    • As this is not an AI/ML device, there was no "training set" in the AI sense, and therefore no ground truth established for such a set. The studies described are for validation of the device's inherent accuracy and performance.
    Ask a Question

    Ask a specific question about this device

    K Number
    K031582
    Date Cleared
    2003-08-01

    (72 days)

    Product Code
    Regulation Number
    870.2770
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Combination unit for measuring systolic and diastolic blood pressures and pulse rates in adult patients with arm circumference between 4.9" to 8" inches as well as body composition, i.e. estimation of body fat, a noninvasive bioimpedance analyzer for use in estimating the human's body fat volume (mass) and percentage by weight. Age range 10-80 years normal subjects, 18-80 years for athletic subjects.

    Device Description

    The Model WT-20 Wrist Blood Pressure Monitor & Fat Meter combines the familiar wrist blood pressure monitor found over the counter in drug stores with a body composition/body fat meter.

    AI/ML Overview

    Here's an analysis of the provided text to extract the requested information about acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes the device, the Nissei WT-20 Wrist Blood Pressure Monitor & Fat Meter, as substantially equivalent to predicate devices. This implies that the 'acceptance criteria' for the Nissei WT-20 are that its performance is comparable to or meets the specifications of the predicate devices. The study provided focuses on this comparison rather than defining explicit numerical acceptance criteria for a new, unique performance claim.

    CharacteristicPredicate Device (OMRON Body Fat Analyzer HBF-306)Nissei WT-20 (Reported Performance)Acceptance Criteria (Implied)
    Operating PrincipleBioelectrical Impedance Analysis (BIA)SAME (Also measures blood pressure)Must be BIA for fat measurement
    Display (Body Fat %)4.0 to 50.0%5.0 to 50.0%Functionally similar range (within acceptable variance)
    Display (BMI)7.0 to 90.05 to 100Functionally similar range (within acceptable variance)
    BMI Classification4 levels4 levelsSame number of classification levels
    Measurement Time7 Seconds5 to 10 secondsSimilar measurement time frame
    Height Range3'4" to 6'6" (101cm. to 198cm.)36 to 79 inches (appears to be a slightly wider range: 3 ft to 6ft 7in)Covers a similar range of heights
    Weight Range23lbs. to 440 1/2 lbs (10kg.to220kg.)22 to 440 pounds (appears slightly different at the lower end)Covers a similar range of weights
    Age Range (Normal)10 to 80 years old10 to 80 (age)Covers the same age range
    Age Range (Athlete)18 to 60 years oldIndicated 18-80 years for athletic subjects in Indications for Use. The table only says "10 to 80 (age)".Covers the same or broader age range
    GenderMale / FemaleMale or female (sex)Supports both genders
    Power Supply2 AAA batteries (R03)SAME (Alkaline recommended)Uses 2 AAA batteries
    Memory9 personal profile memories4 sets of weight, height, age, sex and 10 measurement results of blood pressure and body fatProvides memory for user profiles and historical measurements
    Indications for UseNot explicitly stated for HBF-306, but implied for body fat estimation.Measuring systolic and diastolic blood pressures and pulse rates, and body composition (body fat volume and percentage).Performance for body fat: comparable to predicate. Blood pressure: comparable to WS-500

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "clinical testing data" but does not explicitly state the sample size (number of participants) used for the test set for the body fat measurement or blood pressure features. The provenance of the data (country of origin, retrospective or prospective) is also not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not provide information on the number or qualifications of experts used to establish ground truth for the test set. Given the nature of a 510(k) submission for substantial equivalence, the "ground truth" for the body fat measurement would likely be established through comparison to a well-accepted clinical standard (e.g., DEXA or underwater weighing), and for blood pressure, comparison to a validated manual sphygmomanometer. However, the details of these comparisons are not in the provided text.

    4. Adjudication Method for the Test Set:

    The document does not describe any adjudication method for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study is mentioned. The device is a direct measurement tool, not an interpretive one typically evaluated by MRMC studies (which often involve human readers interpreting images, for example).

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    The provided text details a comparison of the device's technical specifications and intended use against predicate devices. The "Summary of Safety and Effectiveness" states that "The results of bench and user testing indicate that the new device is as safe and effective as the predicate devices." This implies a standalone performance evaluation of the device itself (without human interpretation of results, beyond the user reading the display), but the details of this study are not elaborated. The comparison matrix directly addresses the device's inherent capabilities against the predicate.

    7. Type of Ground Truth Used:

    For the body fat measurement component, the ground truth would typically be established against a recognized gold standard for body composition (e.g., DEXA scan, hydrostatic weighing, or a dual-energy X-ray absorptiometry). For the blood pressure measurement component, the ground truth would typically be established against a medical-grade, validated oscillometric or auscultatory blood pressure measurement device. However, the document does not explicitly state which methods were used as ground truth. It focuses on comparison to a legally marketed predicate device rather than an ultimate gold standard.

    8. Sample Size for the Training Set:

    The document does not mention the use of a "training set" in the context of an algorithm or AI. This device appears to be based on established bioelectrical impedance analysis (BIA) and oscillometric blood pressure measurement principles, likely using pre-defined algorithms rather than machine learning models that require a distinct training phase.

    9. How the Ground Truth for the Training Set Was Established:

    As no training set is indicated, this information is not applicable and is not provided in the document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1