Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K041971
    Date Cleared
    2005-05-20

    (302 days)

    Product Code
    Regulation Number
    892.5050
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    WIMRT MODULE FOR ARTP PLANNING SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TOPSLANE product ARTP with WIMRT is a Radiation Therapy Treatment Planning System for radiation dose planning of patients undergoing external beam treatment in the oncology clinic. ARTP with WIMRT is used to plan radiation treatments with linear accelerators and other similar teletherapy devices with x-ray energies from 1 to 25MV, as well as Cobalt-60, and electron energies from 1 to 25 MeV.

    Device Description

    The TOPSLANE product ARTP with WIMRT is a Radiation Therapy Treatment Planning System for radiation dose planning of patients undergoing external beam treatment in the oncology clinic. ARTP with WIMRT is used to plan radiation treatments with linear accelerators and other similar teletherapy devices with x-ray energies from 1 to 25MV, as well as Cobalt-60, and electron energies from 1 to 25 MeV.

    AI/ML Overview

    The provided document, K041971, is a 510(k) premarket notification for the WIMRT Radiation Treatment Planning System. It does not contain information about the acceptance criteria or a study proving the device meets acceptance criteria in the manner typically expected for diagnostic AI devices with quantifiable performance metrics (e.g., sensitivity, specificity, AUC).

    Instead, this submission focuses on establishing substantial equivalence to predicate devices based on design, construction, intended use, and performance characteristics. The "performance" discussed here refers to functional capabilities and algorithms rather than statistical accuracy against a ground truth.

    Therefore, the following information cannot be extracted from the provided text:

    • A table of acceptance criteria and the reported device performance: No specific performance metrics (like sensitivity, specificity, accuracy) are reported for WIMRT against a defined ground truth, nor are explicit acceptance criteria for such metrics stated. The document compares technical specifications of WIMRT with predicate devices.
    • Sample size used for the test set and the data provenance: No test set is described.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: No ground truth establishment process for a test set is described.
    • Adjudication method for the test set: Not applicable as no test set or ground truth adjudication is described.
    • Multi-reader multi-case (MRMC) comparative effectiveness study: Not conducted or reported in this document.
    • If a standalone performance (algorithm only without human-in-the-loop performance) was done: Not reported. The device is a treatment planning system, implying human interaction.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): No ground truth for performance evaluation is mentioned.
    • The sample size for the training set: No training set is mentioned.
    • How the ground truth for the training set was established: No training set or ground truth establishment is mentioned.

    Summary of Device Features (from the provided table, which serves as a comparison of specifications rather than performance metrics against a ground truth):

    SpecificationWIMRT Module
    Application (Use)Inverse Planning
    Imaging Modalities SupportedCT, MR, PET-Spec
    Dose Calculation AlgorithmSuper-position Convolution
    Optimization EngineSimulated Annealing and Descending Gradient
    IMRT Modalities SupportedSliding Window and Step and Shoot
    Operating SystemWindows
    Image RegistrationAutomatic
    Automated PlanningYes
    Inhomogeneity CorrectionsYes
    Forward IMRT via Field within a Field techniqueYes
    Forward IMRT via MLC Based compensatorsYes
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1