Found 3 results
510(k) Data Aggregation
(78 days)
WATERLASE MILLENNIUM
Class I, II, III, IV and V cavity preparation.
Caries removal.
Hard tissue surface roughening or etching.
Enameloplasty, excavation of pits and fissures for placement of sealants.
Cutting, shaving, contouring and resection of oral osseous tissues (bone).
The Waterlase™ Millennium® hydrokinetic tissue cutting system is a diverse instrument for performing several dental applications. Waterlase™ Millennium® utilizes advanced laser and water atomization technologies to incise, excise and ablate intraoral soft and hard tissues safely and effectively. An erbium, chromium, yttrium, scandium, gallium garnet (Er, Cr.YSGG) solid state laser provides optical energy to a user controlled distribution of atomized water droplets. As the water droplets absorb the optical energy hydrokinetic cutting effects result. The hydrokinetic process refers to the removal of tissues with laser energized water particles. Strong absorption of laser energy by atomized water droplets results in an intense yet controlled water particle micro-expansion and acceleration. The resulting hydrokinetic forces induce mechanical separation of surface material, yielding quick and clean mechanical tissue removal. A flexible fiberoptic handpiece delivers the Waterlase™ Millennium's® unique hydrokinetic tissue cutting technology. A visible light emitted from the handpiece distal end pinpoints the area of treatment. The optical power output and atomized water spray distribution may be adjusted to specific user requirements.
The provided FDA 510(k) summary for the BioLase Waterlase Millennium focuses on demonstrating substantial equivalence to predicate devices, primarily traditional dental handpieces. As such, it does not contain the typical information one would expect for a study proving a device meets specific acceptance criteria using quantitative metrics like sensitivity, specificity, or accuracy.
The summary references "clinical results reported in this Premarket Notification," but it does not detail these clinical studies in a granular way that would allow for the extraction of the requested information regarding acceptance criteria, sample sizes, ground truth establishment, or expert adjudication.
Therefore, for almost all of the requested categories, the answer is "Not provided in the document."
Here is the breakdown based on the provided text:
Acceptance Criteria and Device Performance Study (Not Provided in Detail)
The document asserts "Waterlase Millennium is substantially equivalent to the Dental handpiece in terms of safety and efficacy," based on "clinical results reported in this Premarket Notification and Feature Comparison Table." However, specific quantitative acceptance criteria (e.g., minimum percentage of successful cavity preparations, a defined tissue removal rate, or reduction in patient discomfort scores) are not explicitly stated. Likewise, the "reported device performance" against such criteria is also not detailed.
Acceptance Criteria | Reported Device Performance |
---|---|
Not explicitly stated and quantified in this document. The overarching criterion is "substantially equivalent in terms of safety and efficacy" to predicate dental handpieces. | Not explicitly reported in quantifiable metrics against specific acceptance criteria in this document. |
Additional Study Information:
-
Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
- Not provided in the document. The document mentions "clinical results reported in this Premarket Notification" but does not detail the size or nature of the test set, nor its provenance.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Not provided in the document. The means by which any clinical study data was evaluated or ground truth established (if applicable to the study type) are not described.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not provided in the document.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a surgical instrument (hydrokinetic tissue cutting system) and not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study related to readers improving with AI assistance would not be relevant.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a medical device for direct clinical use, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not provided in the document. The clinical results mentioned are not detailed enough to ascertain the type of ground truth (e.g., direct observation of tissue removal, patient outcomes, success of cavity preparations, etc.).
-
The sample size for the training set:
- Not applicable / Not provided in the document. As this is not an AI/machine learning device, there wouldn't typically be a "training set" in the sense of data used to train an algorithm. If "training set" refers to data used for initial development or pilot studies, it is not described.
-
How the ground truth for the training set was established:
- Not applicable / Not provided in the document. As per point 7.
Ask a specific question about this device
(165 days)
WATERLASE MILLENNIUM
Tooth preparation to obtain access to the root canal Pulpotomy Pulp extirpation Pulpotomy as an adjunct to root canal therapy Root canal debridement and cleaning Root canal preparation including enlargement
The Waterlase Millennium dental laser system may be used to perform several dental applications. For hard tissue procedures the Waterlase Millennium utilizes the Erbium, Chromium: Yttrium, Scandium, Gallium, Garnet (Er,Cr:YSGG) laser in combination with advanced water atomization technology to cut, remove, roughen and etch tissues. Soft tissue procedures are performed using a different mode of operation where direct Er,Cr.YSGG laser energy is applied to incise, excise or ablate the tissues. In soft tissue procedures the water spray is applied for hydration, cooling or to keep tissues clean. A flexible fiberoptic handpiece delivers the Waterlase Millennium laser energy. A visible light emitted from the handpiece distal end pinpoints the area of treatment. In both hard and soft tissue applications the power output, pulse energy, repetition rate and air and water flow rates are adjustable to specific user requirements.
The provided text is a 510(k) Summary for the Waterlase Millennium dental laser system. This type of submission focuses on demonstrating "substantial equivalence" to predicate devices already on the market, rather than conducting new clinical trials to establish acceptance criteria for novel device performance. As such, the document does not contain the detailed study information typically associated with acceptance criteria and clinical performance studies for AI/ML-based devices.
Here's an analysis based on the information provided, highlighting why many aspects of your request cannot be fulfilled:
Acceptance Criteria and Reported Device Performance
The document states:
"Evidence of equivalence has been demonstrated through:
- Equivalent performance specifications
- Promotional materials for equivalent systems
- Equivalent intended uses"
This indicates that the "acceptance criteria" are based on matching the performance of predicate devices, rather than defining new, quantifiable performance metrics for this specific device in a standalone study. The document does not provide a table of specific acceptance criteria or reported device performance metrics in the way one would expect for an AI/ML device (e.g., sensitivity, specificity, AUC).
Instead, the "performance" demonstrated is the ability to achieve the same results as existing tools.
Study Information (as requested for AI/ML devices)
Many of the requests for study information are not applicable or cannot be answered from this 510(k) summary, as it is a submission for a physical medical device (laser system) and not an AI/ML diagnostic or prognostic tool.
-
A table of acceptance criteria and the reported device performance:
- Acceptance Criteria: Not explicitly stated as quantifiable metrics. The overarching criterion is "substantial equivalence" to predicate devices in performance specifications and intended uses.
- Reported Device Performance: Not reported in terms of specific clinical metrics (e.g., success rates, complication rates) from a dedicated study for the Waterlase Millennium. The document implies performance is equivalent to predicate devices.
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not applicable / Not provided. The 510(k) summary does not describe a test set or data provenance from a prospective or retrospective study to validate performance against specific criteria. It relies on the established performance of predicate devices.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable / Not provided. No ground truth establishment is described for a test set, as this is not a study aiming to establish the accuracy of a diagnostic or prognostic output.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable / Not provided. No adjudication method is described, as there is no test set for diagnostic accuracy assessment.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a laser system for dental procedures, not an AI-assisted diagnostic tool. Therefore, an MRMC study related to human reader improvement with AI assistance is irrelevant and not performed.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical device that requires direct human operation; it is not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable. No ground truth in this context is described, as the submission relies on substantial equivalence rather than primary clinical data collection and comparison against a ground truth.
-
The sample size for the training set:
- Not applicable / Not provided. This is not an AI/ML device that undergoes training with a dataset.
-
How the ground truth for the training set was established:
- Not applicable / Not provided. This is not an AI/ML device that requires ground truth for training.
Summary of How the Device Meets Acceptance Criteria
The Waterlase Millennium dental laser system meets its "acceptance criteria" by demonstrating "substantial equivalence" to legally marketed predicate devices through:
- Equivalent Performance Specifications: This implies that the technical specifications (e.g., laser wavelength, power output, pulse energy, repetition rate, and air/water flow rates) of the Waterlase Millennium are comparable to those of the predicate devices for the same intended uses.
- Equivalent Intended Uses: The indications for use for the Waterlase Millennium (e.g., tooth preparation, pulpotomy, root canal debridement and cleaning, root canal preparation including enlargement) are the same as those cleared for the predicate devices.
- Promotional materials for equivalent systems: While not a "study" in the traditional sense, this likely refers to referencing existing marketing claims and clinical literature (not provided in this summary) that support the predicate devices' efficacy for the stated indications.
Conclusion: The provided text is a 510(k) summary for a physical medical device (dental laser), not a software or AI/ML device. Therefore, the questions related to AI/ML device study design (e.g., test sets, training sets, ground truth establishment by experts, MRMC studies) are not applicable to this document. The device meets its "acceptance criteria" by demonstrating substantial equivalence to pre-existing, legally marketed predicate devices based on comparable performance specifications and intended uses, rather than through a dedicated clinical study with defined performance metrics and ground truth.