Search Results
Found 4 results
510(k) Data Aggregation
(62 days)
Vinyl Co-Polymer Powder-free Examination Gloves, Black
Vinyl Co-Polymer Powder-free Examination Gloves, Black is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Vinyl Co-Polymer Powder-free Examination Gloves, Black are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl and oil-based liquid nitrile rubber materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-19 Standard Specification for Vinyl Examination Gloves for Medical Application.
The provided text is related to the FDA 510(k) premarket notification for "Vinyl Co-Polymer Powder-free Examination Gloves, Black". This document describes the device's technical specifications and how it compares to a predicate device. It also includes non-clinical performance data to demonstrate that the device meets established standards.
However, the request asks for specific information regarding acceptance criteria and a study proving a medical device meets those criteria, particularly in the context of AI/ML-assisted diagnostic tools (implied by questions about MRMC studies, human readers, ground truth establishment for training/test sets, etc.). The provided document is for examination gloves, which are Class I medical devices and do not involve AI, image analysis, or diagnostic capabilities in the way the questions are framed. Therefore, most of the requested information is not applicable to this submission.
I will fill in the relevant information that is present in the document and explicitly state when the requested information is not applicable due to the nature of the device.
Here's the breakdown based on the provided document:
1. A table of acceptance criteria and the reported device performance
Test Method | Purpose | Acceptance Criteria | Reported Device Performance (Results) |
---|---|---|---|
ASTM D5250-19 | Length | Minimum 230mm | All size ≥230mm |
ASTM D5250-19 | Palm Width (XS) | 75 ± 5 mm | 76-78mm |
ASTM D5250-19 | Palm Width (S) | 85 ± 5 mm | 86-88mm |
ASTM D5250-19 | Palm Width (M) | 95 ± 5 mm | 96-98mm |
ASTM D5250-19 | Palm Width (L) | 105 ± 5 mm | 106-108mm |
ASTM D5250-19 | Palm Width (XL) | 115 ± 5 mm | 116-118mm |
ASTM D5250-19 | Thickness (Finger) | 0.08mm (min) | 0.08mm |
ASTM D5250-19 | Thickness (Palm) | 0.08mm (min) | 0.08-0.10mm |
ASTM D5250-19, ASTM D412-16 | Tensile Strength, Before Aging | 11MPa, min | 13.5-17.8 MPa |
ASTM D5250-19, ASTM D412-16 | Tensile Strength, After Accelerated Aging | 11MPa, min | 13.1-17.6 MPa |
ASTM D5250-19, ASTM D412-16 | Ultimate Elongation, Before Aging | 300%, min | 320-430% |
ASTM D5250-19, ASTM D412-16 | Ultimate Elongation, After Accelerated Aging | 300%, min | 310-420% |
ASTM D5151-19, ASTM D65250-19 | Freedom from holes | G-I, AQL 2.5 | Meets AQL 2.5 requirements |
ASTM D6124-06(2017), ASTM D5250-19 | Powder-Content | ≤2 mg per glove | 0.35-0.68 mg |
10993-10:2010 | Skin Irritation Study | Under the conditions of the study, not an irritant | Under the conditions of the study, not an irritant |
10993-10:2010 | Maximization Sensitization Study | Under the conditions of the study, not a sensitizer | Under the conditions of the study, not a sensitizer |
10993-5:2009 | In Vitro Cytotoxicity Test | No potential cytotoxicity to L929 cells | No potential cytotoxicity to L929 cells |
ISO 10993 Part 11 | Acute Systemic Toxicity Test | No systemic toxicity | No systemic toxicity |
2. Sample sizes used for the test set and the data provenance
The document refers to "non-clinical tests" and "performance test data" but does not specify the exact sample sizes (e.g., number of gloves tested for each individual parameter). The tests conducted are standard physical and biocompatibility tests for examination gloves as per ASTM and ISO standards.
- Sample Size: Not explicitly stated for each test, but implied to be sufficient to meet the specifications of the referenced ASTM and ISO standards.
- Data Provenance: The manufacturing entity is Grand Work Plastic Products Co., Ltd in Donggao Industrial Zone, Zanhuang, Hebei, China. The testing was presumably conducted by or for this manufacturer. The document doesn't specify if the data is retrospective or prospective, but for device development and regulatory submission, it would reflect tests performed on batches of the device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This device is an examination glove, not an AI/ML-based diagnostic system that requires expert interpretation for ground truth establishment. The "ground truth" for these tests is based on objective physical, chemical, and biological measurements against established international standards (ASTM, ISO).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not an AI/ML diagnostic system requiring expert adjudication of results. Testing is based on laboratory measurements and adherence to specified tolerances.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an examination glove and does not involve human readers, AI assistance, or diagnostic interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is an examination glove and does not involve algorithms or standalone performance testing in the context of AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device is based on:
- Physical and Mechanical Standards: Measurements of length, palm width, thickness, tensile strength, and elongation against specifications in ASTM D5250-19 and ASTM D412-16.
- Performance Standards: Freedom from holes (ASTM D5151-19, ASTM D65250-19) and powder content (ASTM D6124-06(2017), ASTM D5250-19).
- Biocompatibility Standards: Results from tests for skin irritation, sensitization (ISO 10993-10:2010), in vitro cytotoxicity (ISO 10993-5:2009), and acute systemic toxicity (ISO 10993 Part 11).
8. The sample size for the training set
Not applicable. This device is an examination glove and does not involve AI/ML models that require training sets.
9. How the ground truth for the training set was established
Not applicable. As there is no training set for an AI/ML model, no ground truth establishment method for a training set is relevant.
Ask a specific question about this device
(100 days)
Vinyl Co-Polymer Powder Free Examination Gloves
A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Vinyl Co-Polymer Powder Free Examination Gloves, Blue Color are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl and oil-based liquid nitrile rubber materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-06 (Reapproved 2011) Standard Specification For Poly (Vinyl Chloride) Gloves For Medical Application.
The provided document describes the Grand Work Plastic Products Co., Ltd.'s Vinyl Co-Polymer Powder Free Examination Gloves, Blue Color (K142409) and seeks to prove its substantial equivalence to a predicate device (K061562). The study conducted is a non-clinical performance study comparing the subject device against the predicate device based on established standards and FDA requirements.
Here's the breakdown of the acceptance criteria and study details:
1. A table of acceptance criteria and the reported device performance:
Characteristic | Acceptance Criteria (Stated Standard ASTM D5250-06(2011) or other) | Reported Device Performance (Subject Device K142409) | Result of Comparison to Predicate & Acceptance |
---|---|---|---|
Material | Vinyl and oil-based liquid Nitrile compound (plasticizer type different) | Vinyl and oil-based liquid Nitrile compound (plasticizer type different) | Similar |
Color | Blue | Blue | Similar |
Tensile Strength (Mpa) | |||
- Before aging | 11 Mpa min | 15.2 Mpa min | Pass (Exceeds standard and similar to predicate) |
- After aging | 11 Mpa min | 14.6 Mpa min | Pass (Exceeds standard and similar to predicate) |
Ultimate Elongation (%) | |||
- Before aging | 300% min | 390% min | Pass (Exceeds standard and similar to predicate) |
- After aging | 300% min | 360% min | Pass (Exceeds standard and similar to predicate) |
Freedom from Pinholes | AQL 2.5 (G-I) | AQL 1.5 (G-I) | Pass (Exceeds standard and similar to predicate, lower AQL is better) |
Overall Length (mm) | 230 for all sizes min (S-2, AQL4.0) | 239 mm (for all sizes) | Pass (Exceeds standard and similar to predicate) |
Width (mm) | S: 85±5, M: 95±5, L: 105±5, XL: 115±5 (S-2, AQL4.0) | S: 87, M: 97, L: 107, XL: 118 | Pass (Within range and similar to predicate) |
Palm Thickness (mm) | 0.08 mm min (S-2, AQL4.0) | 0.08 mm | Pass (Meets standard and similar to predicate) |
Finger Thickness (mm) | 0.05 mm min (S-2, AQL4.0) | 0.10 mm | Pass (Exceeds standard and similar to predicate) |
Residual Powder | Not more than 2mg per glove (N=5) | Not more than 2mg per glove | Pass (Meets standard and similar to predicate) |
Primary Skin Irritation Test | Under conditions of the study, not an irritant (ISO 10993-10) | Under conditions of the study, not an irritant | Similar (Pass) |
Dermal Sensitization Assay | Under conditions of the study, not an irritant (ISO 10993-10) | Under conditions of the study, not an irritant | Similar (Pass) |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size for Test Set: The sample sizes for each test are indicated by the "Inspection Level and AQL" or "N" values.
- For physical dimensions (Overall Length, Width, Palm Thickness, Finger Thickness), Tensile Strength, and Ultimate Elongation: S-2, AQL4.0.
- For Pinhole: G-I, AQL1.5.
- For Residual Powder: N=5 (5 gloves tested).
- For Biocompatibility (Primary Skin Irritation, Dermal Sensitization): The specific number of samples (e.g., animals, human subjects, in-vitro replicates) is not explicitly stated, but the tests were performed "Under conditions of the study".
- Data Provenance: The study was conducted by Grand Work Plastic Products Co., Ltd., which is located in Zanhuang, Hebei, China. The data originates from this manufacturer. It is a prospective study as the tests were performed on the subject device to demonstrate compliance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This study is a physical and chemical performance test of a medical device (examination gloves), not an AI/clinical diagnostic device. Therefore, it does not involve experts establishing "ground truth" in the clinical sense (e.g., radiologists, pathologists). The "ground truth" is established by the pre-defined technical standards (e.g., ASTM D5250-06(2011), ISO 10993-10) and FDA requirements. The tests are performed by technicians in a laboratory setting according to specified protocols.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. This is not a study involving human readers or clinical adjudication. The results are quantitative measurements or qualitative observations against a set standard.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is a performance study for a physical device (examination gloves), not a comparative effectiveness study involving human readers or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a performance study for a physical device (examination gloves), not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for this study is based on established industry standards and regulatory requirements. Specifically:
- ASTM D5250-06 (Reapproved 2011) Standard Specification For Poly (Vinyl Chloride) Gloves For Medical Application.
- ISO 10993-10 for biocompatibility (Primary Skin Irritation and Dermal Sensitization).
- FDA requirements (e.g., for pinhole AQL).
8. The sample size for the training set:
Not applicable. This is not a machine learning or AI study; there is no training set involved.
9. How the ground truth for the training set was established:
Not applicable. There is no training set.
Ask a specific question about this device
(107 days)
VINYL CO-POLYMER POWDER FREE EXAMINATION GLOVES, BLUE COLOR
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Classified by FDA's General and Plastic Surgery Device panel as Class I. 21 CFR 880.6250, Power-Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-00E4
The provided text describes the acceptance criteria and the study conducted for the "Shijiazhuang Hongxiang Plastic Products Co., Ltd. Vinyl Co-Polymer Powder Free Examination Gloves, Blue Color." This is a Class I medical device (patient examination glove). The evaluation focuses on demonstrating substantial equivalence to a predicate device, as per FDA 510(k) requirements.
Here's an analysis of the provided information in the requested format:
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria (Standard / Test) | Reported Device Performance |
---|---|
Physical and Dimensions Testing | Met requirements based on ASTM D-5250-00E4 (Inspection level S-2, AQL 4.0) |
Integrity (Pinhole Test) | Met FDA 1000 ml. Water Fill Test requirements based on ASTM D-5151-99 (AQL 2.5, inspiration level G-1) |
Biocompatibility (Irritation) | No primary skin irritant reactions in Primary Skin Irritation testing |
Biocompatibility (Sensitization) | No sensitization reactions in Skin Sensitization testing |
Residual Powder | Met "powder-free" claims (contains no more than 2 mg powder per glove) based on ASTM D-6124-06 |
2. Sample Size Used for the Test Set and Data Provenance:
- Physical and Dimensions Testing: Inspection level S-2, AQL 4.0. The specific number of gloves tested is not provided, but AQL (Acceptable Quality Limit) indicates the maximum acceptable percentage of defective items.
- Pinhole Test (Water Fill Test): AQL 2.5, inspection level G-1. The specific number of gloves tested is not provided, but AQL indicates the maximum acceptable percentage of defective items.
- Primary Skin Irritation and Skin Sensitization Testing: The sample size for these biological tests is not specified in the document.
- Residual Powder Test: The sample size is not specified in the document.
- Data Provenance: The origin of the data is from testing performed by Shijiazhuang Hongxiang Plastic Products Co., Ltd. in China. The study appears to be prospective as it involves new testing of the device for submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Not applicable. This device is a Class I examination glove, and the evaluation does not involve diagnostic interpretation that would require expert human readers to establish ground truth from medical images or clinical observations. The ground truth is objective measurements against established ASTM standards and FDA requirements.
4. Adjudication Method for the Test Set:
- Not applicable. As the tests are objective measurements against defined physical, chemical, and biological standards (e.g., AQL, mg of powder, presence/absence of irritation), there is no adjudication process involving multiple human reviewers.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No. An MRMC study is not relevant for this type of device. Such studies are typically conducted for diagnostic imaging devices or algorithms where human interpretation is a key component.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, in essence. The testing performed (physical, dimensional, pinhole, biocompatibility, residual powder) is equivalent to a "standalone" evaluation of the device's inherent properties without human interaction or interpretation as part of its primary function. The device itself is not an algorithm, but its performance is measured directly.
7. The Type of Ground Truth Used:
- Objective Standards and Measurement: The ground truth is established by:
- ASTM Standards: D-5250-00E4 (physical and dimensions), D-5151-99 (pinhole/water fill), D-6124-06 (residual powder).
- FDA Requirements: Specific pinhole requirements and biocompatibility expectations.
- Defined Limits: E.g., no more than 2 mg powder per glove for "powder-free" claim.
- Direct Biological Observation: For skin irritation and sensitization tests.
8. The Sample Size for the Training Set:
- Not applicable. This device is not an AI/ML algorithm or a diagnostic tool that requires a "training set." The testing is for a physical medical product.
9. How the Ground Truth for the Training Set was Established:
- Not applicable. There is no training set for this type of device.
Ask a specific question about this device
(36 days)
VINYL CO-POLYMER POWDER FREE EXAMINATION GLOVES, BLUE COLOR
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Power-Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-00E4
The device described is the Grand Work Plastic Products Co., Ltd.'s Vinyl Co-Polymer Powder Free Examination Gloves, Blue Color. The submission is a 510(k) for substantial equivalence to a predicate device. This type of submission generally relies on non-clinical performance data rather than extensive clinical studies.
Here's an analysis of the provided information concerning acceptance criteria and supporting studies:
1. Table of Acceptance Criteria and Reported Device Performance
Test/Characteristic | Acceptance Criteria (from ASTM D-5250-00E4 and FDA) | Reported Device Performance |
---|---|---|
Physical and Dimensions Testing | Requirements of ASTM D-5250-00E4 | Meets all requirements (implicitly, as stated "All testing meets requirements") |
Pinhole Detection (Water Fill Test) | AQL 2.5, Inspection Level G-1 (based on ASTM D-5151-99) | Meets these requirements |
Primary Skin Irritation | No primary skin irritant reactions | No primary skin irritant reactions |
Skin Sensitization (Allergic Contact Dermatitis) | No sensitization reactions | No sensitization reactions |
Residual Powder | No more than 2 mg powder per glove (ASTM D-6124-06 for Starch) | Meets "powder-free" claims (contains no more than 2 mg powder per glove) |
2. Sample Size Used for the Test Set and Data Provenance
- Physical and Dimensions Testing: Inspection level S-2, AQL 4.0. The specific number of items sampled is not explicitly stated but is determined by the AQL and inspection level. Data provenance is not specified but would typically be from the manufacturer's internal testing.
- Pinhole Detection (Water Fill Test): AQL 2.5, Inspection Level G-1. The specific number of items sampled is not explicitly stated. Data provenance is not specified.
- Primary Skin Irritation and Skin Sensitization: The sample size for these biological tests is not specified in the provided document. These tests are typically conducted in vitro or in vivo on animal models or human volunteers according to standard protocols. Data provenance is not specified.
- Residual Powder Test: The sample size is not specified. Data provenance is not specified.
All testing would be considered retrospective in the context of this submission, as they were performed prior to the submission for FDA clearance. The country of origin of the data is implied to be from the manufacturer's testing facilities, likely in China, given the submitter's location.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
Not applicable for this type of device and submission. The "ground truth" for the performance tests (e.g., tensile strength, pinholes) is established by the objective physical and chemical testing methods and their predefined acceptance criteria, not by expert consensus on interpretations of complex data like medical images.
4. Adjudication Method for the Test Set
Not applicable. The tests are objective measurements against predetermined standards, not subjective assessments requiring adjudication among experts.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret medical data (e.g., images), and the AI's impact on reader performance is evaluated. Examination gloves do not involve such human interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
No, a standalone algorithm performance study was not done. The device is a physical product (gloves), not an algorithm or AI system.
7. The Type of Ground Truth Used
The "ground truth" for this device is based on objective measurements and predefined standards.
- For physical properties and dimensions, the ground truth is derived from the ASTM D-5250-00E4 standard specifications.
- For pinhole detection, the ground truth is determined by the FDA 1000 ml. Water Fill Test based on ASTM D-5151-99.
- For biocompatibility (skin irritation and sensitization), the ground truth is established by the results of standardized biological assays that demonstrate no adverse reactions.
- For residual powder, the ground truth is based on the ASTM D-6124-06 standard for starch content.
8. The Sample Size for the Training Set
Not applicable. This device is not an AI/ML algorithm that requires a training set. The performance is based on the manufacturing process and material characteristics, assessed through quality control and testing.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
Page 1 of 1