Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K161492
    Device Name
    Juno VPAP ST-A
    Manufacturer
    Date Cleared
    2017-01-19

    (232 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Juno VPAP ST-A

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Juno VPAP ST-A is indicated to provide noninvasive ventilation for patients weighing more than 30lbs (13 kg) with respiratory insufficiency or obstructive sleep apnoea (OSA).

    The iVAPS mode is indicated for patients weighing more than 66lbs (30 kg).

    The Juno VPAP ST-A is intended for home and hospital use.

    The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment.

    Device Description

    The subject device Juno VPAP ST-A retains all the same hardware, technologies and manufacturing characteristics as previously cleared in K153061. The device is used in a wider breathing circuit which typically comprises patient tubing and a mask (patient interface) to deliver a prescribed positive airway pressure treatment to patients.

    Some of the kev features of the device include an in-line power supply: fully integrated humidifier: alarms module; heater controller; colour LCD and simple controls for ease of use. The device also allows data transfer/connectivity via an integrated wireless module (When used in a hospital setting, remote changes may not be appropriate for certain patients, as these setting changes may not be communicated to all hospital personnel treating the patient. Hospital staff should liaise with the patient's regular care provider such that the desired therapy outcome is achieved).

    Juno VPAP ST-A utilizes a Micro-processor controlled blower system that generates positive airway pressure (CPAP) between 4-20 cmHJO as required to maintain an "air splint" for effective treatment of OSA and (Bilevel) pressures between 3-30 cmH2O for the treatment respiratory insufficiency.

    The therapy modes available in the Juno VPAP ST-A include CPAP, Spontaneous, Spontaneous/Timed, Timed, PAC and iVAPS. In this subject device now features an optional "AutoEPAP" function on iVAPS mode. AutoEPAP automatically adjusts EPAP pressure (within set values) in response to flow limitations of the upper airway.

    Juno VPAP ST-A is intended to be used under the conditions and purposes indicated in the labelling provided with the product.

    It is a prescription device, supplied non-sterile.

    AI/ML Overview

    The document describes the Resmed Juno VPAP ST-A device (K161492), which is an update to a previously cleared device (K153061). The key change in the subject device is the addition of an optional "AutoEPAP" function to the iVAPS therapy mode.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets in the provided text. Instead, suitability for market is based on demonstrating substantial equivalence to the predicate device (Juno VPAP ST-A, K153061) and demonstrating that the new AutoEPAP algorithm provides equivalent therapy to the fixed EPAP setting of the predicate.

    The reported device performance is that the AutoEPAP iVAPS therapy mode was as efficacious as iVAPS with fixed EPAP in terms of several clinical metrics.

    Acceptance Criterion (Inferred from Substantial Equivalence Claim)Reported Device Performance (from Clinical Data)
    Equivalence in maintaining upper airway patencyAutoEPAP iVAPS was as efficacious as iVAPS with fixed EPAP.
    Equivalence in Apnea-Hypopnea Index (AHI)Equivalent
    Equivalence in Oxygen Desaturation Index (ODI)Equivalent
    Equivalence in Blood Oxygen Saturation (SPO2)Equivalent
    Equivalence in Transcutaneous Carbon Dioxide (PtcCO2)Equivalent
    Equivalence in Sleep QualityEquivalent
    No serious adverse events or complicationsNo serious adverse events or complications related to the study device were recorded.
    Compliance with relevant standards and specificationsThe subject device meets all requirements of the Juno VPAP ST-A System Specification and complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11, FDA Draft Reviewer Guidance for Ventilators, and FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample size for the test set: Not explicitly stated. The document mentions "patients with respiratory insufficiency" but does not provide the number of participants.
    • Data provenance: Not explicitly stated, but the study was a "clinical trial" which implies prospective data collection. The country of origin is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. As this is a study evaluating the clinical efficacy of a ventilator mode, the "ground truth" would be established by direct physiological measurements and clinical assessments rather than expert review of images or data.

    4. Adjudication Method for the Test Set

    This information is not provided in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, an MRMC study was not done. This device is a ventilator, not an AI-assisted diagnostic tool that would typically involve human readers interpreting data or images. The study focuses on the direct efficacy of the ventilator's automated function.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the clinical trial described appears to be a standalone performance evaluation of the AutoEPAP iVAPS algorithm. It directly compares the efficacy of the new automated algorithm with the fixed EPAP mode, which also functions automatically (without real-time human intervention during therapy delivery).

    7. The Type of Ground Truth Used

    The ground truth was established using physiological measurements and clinical outcomes data such as:

    • Apnea-Hypopnea Index (AHI)
    • Oxygen Desaturation Index (ODI)
    • Blood Oxygen Saturation (SPO2)
    • Transcutaneous Carbon Dioxide (PtcCO2)
    • Sleep Quality (presumably based on polysomnography or other clinical assessments)

    8. The Sample Size for the Training Set

    The document describes a clinical trial (double-blinded, randomized, crossover study) which evaluates the performance of the implemented algorithm. It does not mention a separate "training set" in the context of an AI/machine learning model development lifecycle. The AutoEPAP algorithm would have been developed and internally validated by the manufacturer, but the specific details of its training data (if any for a machine learning component) are not provided. The clinical trial serves as the validation or test set for the device's efficacy.

    9. How the Ground Truth for the Training Set Was Established

    As explained above, a "training set" in the machine learning sense is not explicitly discussed. For the clinical trial (validation data), the ground truth was established through direct physiological measurements and clinical assessments as listed in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153061
    Device Name
    Juno VPAP ST-A
    Manufacturer
    Date Cleared
    2016-04-13

    (175 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Juno VPAP ST-A

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Juno VPAP ST-A is indicated to provide noninvasive ventilation for patients weighing more than 30lbs (13 kg) with respiratory insufficiency or obstructive sleep apnoea (OSA).

    The iVAPS mode is indicated for patients weighing more than 66lbs (30 kg).

    The Juno VPAP ST-A is intended for home and hospital use.

    The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment.

    Device Description

    The Juno VPAP ST-A System is a positive pressure device which design is based on the previously cleared ResMed S9 WANDA VPAP ST (K140159) platform. The device is used in a wider breathing circuit which typically comprises patient tubing and a mask (patient interface) to deliver the prescribed positive pressure treatment to patients.

    Some of the key features of the platform include: in-line power supply; fully integrated humidifier; heater controller; colour LCD and simple controls for ease of use. An integrated wireless module for data transfer/connectivity is also included (note: Remote changes in a hospital setting may not be appropriate, as these setting changes may not be communicated to all hospital personnel treating the patient. Hospital staff should liaise with the patient's reqular care provider such that the desired therapy outcome is achieved).

    Juno VPAP ST-A contains a Micro-processor controlled blower system that generates positive airway pressure (CPAP) between 4-20 cmHyO as required to maintain an "air splint" for effective treatment of OSA and (Bilevel) pressures between 3-30 cmH2O for the treatment respiratory insufficiency.

    The device also incorporates an Alarm function as a fully integrated module.

    The therapy modes available in the Juno VPAP ST-A come from the predicate ResMed S9 VPAP ST-A (K113288). These include CPAP, Spontaneous, Spontaneous/Timed, Timed, PAC and iVAPS.

    Juno VPAP ST-A is intended to be used under the conditions and purposes indicated in the labelling provided with the product.

    It is a prescription device, supplied non-sterile.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or a study proving that the device meets these criteria in the manner you've requested (e.g., performance metrics, sample sizes, expert ground truth).

    The document is a 510(k) summary for a medical device (Juno VPAP ST-A), which focuses on demonstrating substantial equivalence to a predicate device (ResMed S9 VPAP ST-A K113288), rather than presenting a detailed clinical study with performance metrics against pre-defined acceptance criteria.

    The "Non Clinical data submitted" section (page 8) mentions "Bench test data is presented to demonstrate that the new device meets its published specifications and back-to-back testing shows substantial equivalence to the predicate device S9 VPAP ST-A (K113288)." It further states, "All bench tests confirmed that the product met the predetermined acceptance criteria. this included Pressure, Flow, Pressure Support, Trigger and Cycling, Hypopnea and Apnea tests against the predicate devices using common protocols for both devices."

    However, this section does not provide specific numerical acceptance criteria values or the reported device performance against those criteria. It also doesn't describe an AI/ML-based device that would typically involve a "test set" with ground truth established by experts, MRMC studies, or standalone algorithm performance.

    Therefore, most of the requested information cannot be extracted from this document, as it pertains to a different type of device assessment (bench testing for substantial equivalence of a continuous ventilator) than what your questions imply (e.g., an AI/ML diagnostic device).

    Here's what can be inferred or stated based on the document's content, and what cannot:

    1. A table of acceptance criteria and the reported device performance:

    • Cannot provide. The document states that bench tests confirmed the product met "predetermined acceptance criteria" for parameters like pressure, flow, pressure support, trigger and cycling, hypopnea, and apnea tests. However, the specific numerical acceptance criteria (e.g., "pressure within +/- X cmH2O") and the actual performance results are not detailed in this summary.

    2. Sample sized used for the test set and the data provenance:

    • Cannot provide. The document refers to "bench tests" and "back-to-back testing" against a predicate device. This is not a "test set" in the context of clinical data for an AI/ML device. There are no details on sample size, data provenance, or whether the data was retrospective or prospective, as this was not a clinical study involving patients for direct performance evaluation against ground truth.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device is a continuous ventilator, and its performance was evaluated through engineering bench tests, not clinical evaluation against expert-adjudicated ground truth as would be done for a diagnostic AI/ML system.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No expert adjudication was described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI-assisted diagnostic device, so an MRMC study comparing human readers with and without AI assistance was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is a hardware medical device (ventilator) with integrated software/firmware, not a standalone diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Implied ground truth for bench testing: Engineering specifications and performance of the predicate device (S9 VPAP ST-A (K113288)). This is a "technical" ground truth, not a "clinical" one. The document states: "this included Pressure, Flow, Pressure Support, Trigger and Cycling, Hypopnea and Apnea tests against the predicate devices using common protocols for both devices."

    8. The sample size for the training set:

    • Not applicable. This document is for a hardware device, not an AI/ML model that undergoes a training phase with a specific dataset.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.

    In summary: The provided 510(k) summary focuses on demonstrating substantial equivalence through bench testing against a predicate device's specifications and performance, rather than clinical study results of an AI/ML device against expert-established ground truth. Therefore, very little of the information typically sought for AI/ML device evaluation is available in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140159
    Device Name
    S9 WANDA VPAP ST
    Manufacturer
    Date Cleared
    2014-06-17

    (146 days)

    Product Code
    Regulation Number
    868.5905
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    S9 WANDA VPAP ST

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The S9 WANDA VPAP ST is indicated for the treatment of Obstructive Sleep Apnea (OSA) in patients weighing more than 66 lb (30 kg). It is intended for use in the hospital and home.

    Device Description

    S9 WANDA VPAP ST System is similar to the predicate device (S9 VPAP ST (K102513)). Key features include: inline power supply; integrated humidifier; tubing; colour LCD; and simple controls. The S9 WANDA VPAP ST contains a Micro-processor controlled blower system that generates Continuous Positive Airway Pressure (CPAP) from 3-25 cmHzO as required to maintain an "air splint" for treatment of OSA. The system comprises the flow generator, patient tubing, mask (patient interface) and humidifier. The S9 WANDA VPAP ST flow generator includes the same therapy modes as the S9 VPAP ST system (K102513) predicate device. These base therapy modes include: CPAP and CPAP with expiratory pressure relief (EPR) modes the device delivers a continuous positive airway pressure throughout the therapy session Bilevel mode the flow generator augments any breath initiated by the patient by detecting the onset of inspiration or expiration and delivering the set Inspiratory Positive Airway Pressure (IPAP) and Expiratory Positive Airway Pressure (EPAP) VAuto mode - the mean air-way pressure (between IPAP and EPAP) will alter based upon breathing events The functional characteristics of the S9 WANDA VPAP ST system includes all the clinician and user friendly features of the predicate device which have been verified during usability studies in accordance with IEC 62366 Medical devices - Application of usability engineering to medical devices.

    AI/ML Overview

    Acceptance Criteria and Study for S9 WANDA VPAP ST

    The submission for the S9 WANDA VPAP ST primarily relied on bench testing to demonstrate substantial equivalence to its predicate device, the S9 VPAP ST (K102513). Therefore, the "acceptance criteria" discussed below primarily refer to the performance standards met during these bench tests, which were deemed sufficient to prove a similar level of safety and effectiveness to the predicate. No clinical studies are mentioned.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a formal "acceptance criteria" table with specific quantitative thresholds for each test. Instead, it states that "All tests confirmed the product met the predetermined acceptance criteria" and that these criteria were "traced to the S9 WANDA VPAP ST System Specification and to the predicate device's performance."

    However, the "Characteristics (differences) between predicate and new device" and "Flow Generator comparison" sections provide some comparative performance data, which can be interpreted as demonstrating the device's fulfillment of implied acceptance criteria based on equivalence to the predicate.

    CharacteristicPredicate S9 VPAP ST (K102513) PerformanceS9 WANDA VPAP ST PerformanceAcceptance Criteria (Implied)Met/Not Met (as per document)
    Intended UseTreatment of Obstructive Sleep Apnea (OSA) in patients weighing > 66 lb (30 kg), for hospital and home use.Treatment of Obstructive Sleep Apnea (OSA) in patients weighing > 66 lb (30 kg), for hospital and home use.Equivalent to PredicateMet (Explicitly stated "Equivalent")
    Operating System/ControllerMicrium uC/OS-II Software/digitalMicrochip: STM32F405ZG micro-controller with ARM 32-bit Cortex™-M4 CPU Software/digitalUpdated micro in line with latest technologies appropriate for processing power needed, maintaining functional equivalence.Met (Explicitly stated "Equivalent")
    Flow Generator Weight1.7 lb2.5 lbBiologically Equivalent (accounts for integrated humidifier)Met (Explicitly stated "Equivalent")
    Dimensions H x W x D (inches)Flow generator unit: 3.4 x 5.5 x 6.0Flow generator unit: 4.5 x 9.6 x 6.0Biologically Equivalent (accounts for integrated humidifier, larger width)Met (Explicitly stated "Equivalent")
    Data Transfer MediumSD CardSD Card, WirelessEquivalent to Predicate (Wireless data transfer equivalent to EasyCare Online (K132371))Met (Explicitly stated "Equivalent")
    Humidifier Output (for > 8 hours)12.7 mg/L @ 20cm H2O (50 L/min)12.0 mg/L @ 20cm H2O (50 L/min)Not clinically significant difference compared to predicate.Met (Explicitly stated "The difference is not clinically significant and acceptance criteria was met (refer to bench testing (section 18)).")
    Pressure StabilityNot explicitly quantified but implied performance of predicate.Demonstrated in bench testing.Equivalent to PredicateMet (Stated "This performance testing demonstrate successful implementation of therapy modes into the S9 WANDA VPAP ST")
    Response to ApneasNot explicitly quantified but implied performance of predicate.Demonstrated in bench testing.Equivalent to PredicateMet (Stated "This performance testing demonstrate successful implementation of therapy modes into the S9 WANDA VPAP ST")
    Response to Flow Limitations & SnoreNot explicitly quantified but implied performance of predicate.Demonstrated in bench testing.Equivalent to PredicateMet (Stated "This performance testing demonstrate successful implementation of therapy modes into the S9 WANDA VPAP ST")
    Reporting of Closed Airway Detection (CAD)Not explicitly quantified but implied performance of predicate.Demonstrated in bench testing.Equivalent to PredicateMet (Stated "This performance testing demonstrate successful implementation of therapy modes into the S9 WANDA VPAP ST")

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable as no human-subject clinical test set was used. The study relied on bench testing using a "breathing machine" and the device itself.
    • Data Provenance: The bench testing was performed by ResMed Ltd., an Australian company. The testing is therefore attributed to the manufacturer's internal testing procedures. The data would be considered prospective in relation to the submission, as it was specifically generated for the purpose of demonstrating substantial equivalence.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Number of Experts: Not applicable. The "ground truth" for the bench testing was established by comparing the S9 WANDA VPAP ST's performance to the "S9 WANDA VPAP ST System Specification" and the "predicate device's performance." These are engineering specifications and previously established performance metrics, not expert human interpretations of data.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. There was no human "test set" requiring adjudication. The assessment was based on objective measurements against engineering specifications and comparison to the predicate device's measured performance during bench testing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. The submission explicitly states "Bench testing alone is sufficient to demonstrate Substantial Equivalence."
    • Effect Size of Human Readers with/without AI Assistance: Not applicable, as no human-in-the-loop study with AI assistance was conducted.

    6. Standalone (Algorithm Only) Performance Study

    • Standalone Study: Yes, in a sense. The "bench testing" represents a standalone performance evaluation of the device's algorithms and hardware without a human operator directly influencing its core therapeutic function during the test. The "breathing machine" simulates patient input, and the device's response is objectively measured. This is not "algorithm only" in the modern sense of a software-as-a-medical-device (SaMD) study, but rather the performance of the integrated device's therapeutic functions.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for the bench testing was based on:
      • Engineering Specifications: The S9 WANDA VPAP ST System Specification defined the expected performance.
      • Predicate Device Performance: Direct comparison to the measured performance of the legally marketed S9 VPAP ST (K102513) further served as a "ground truth" for demonstrating equivalence. This is an objective, measured ground truth.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This device is a hardware-based medical device with integrated software that controls its mechanical functions (like pressure generation and response to breathing patterns). It does not employ machine learning or AI models that require a separate "training set" of data in the way modern AI/ML devices do. The software development likely involved traditional engineering verification and validation.

    9. How Ground Truth for the Training Set Was Established

    • How Ground Truth for Training Set Was Established: Not applicable, as there was no training set in the context of machine learning. The device's operational parameters and control logic were developed through engineering design process, adhering to established medical device standards (e.g., IEC 60601-1, IEC 62304) and based on the predicate device's proven functionality.
    Ask a Question

    Ask a specific question about this device

    K Number
    K113288
    Device Name
    VPAP ST-A
    Manufacturer
    Date Cleared
    2012-03-29

    (143 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    VPAP ST-A

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VPAP ST-A is indicated to provide non-invasive ventilation for patients weighing more than 30 lbs (13kg) or more than 66 lbs (30kg) in iVAPS mode with respiratory insufficiency or obstructive sleep apnea (OSA). The VPAP ST-A is intended for use in the hospital or home.

    Device Description

    VPAP ST-A System (VPAP ST-A with H5) is similar to the predicate device(s), using a blower based positive pressure system with an integrated heated humidifier and heater controller. The device platform is similar to the S9 VPAP ST (K102513) and contains a Micro-processor controlled blower system that generates controlled positive airway pressure (CPAP) between 4-20 cmHzO as required to maintain an "air splint" for effective treatment of OSA and (Bilevel) pressures between 3-30 cmHzO for the treatment respiratory insufficiency. The system comprises the flow generator, patient tubing, mask (patient interface), alarm functions and integrated humidifier. Therapy modes contained in the VPAP ST-A are CPAP, Spontaneous, Spontaneous/Timed, Timed, PAC and iVAPS. Therapy modes come from the S9 VPAP ST (K102513) and Stellar 150 (K103167). The functional characteristics of the VPAP ST-A system includes all the clinician and user friendly features of the predicate devices.

    AI/ML Overview

    The provided document is a 510(k) summary for the ResMed VPAP ST-A. It discusses the device's substantial equivalence to predicate devices and outlines non-clinical testing. However, it does not contain information about a study that used a test set with established ground truth, MRMC studies, or training sets.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance from a study with a test set, nor can I answer questions related to sample sizes for test/training sets, expert involvement for ground truth, adjudication methods, or MRMC studies, as this information is not present in the document.

    Based on the provided text, here's what can be inferred and what cannot:

    1. A table of acceptance criteria and the reported device performance

    • Cannot be provided directly. The document states: "All bench tests confirmed the product met the predetermined acceptance criteria, this included Pressure. Flow, Pressure Support, Trigger and Cycling, Hypopnea and Apnea tests against the predicate devices using common protocols for both devices." However, the specific acceptance criteria values and the quantitative reported device performance for these criteria are not detailed in the summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Cannot be provided. The document describes "bench tests" and "validated patient simulation models" but does not detail a "test set" in the context of patient data with a specific sample size or provenance. It explicitly states: "Clinical data for the VPAP ST-A is not required as the predicate devices have been subjected to clinical trial requirements or validated patient simulation models have been used during the bench testing phases."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Cannot be provided. As there is no mention of a clinical test set requiring expert ground truth establishment for patient outcomes, this information is not present.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Cannot be provided. Similar to point 3, without a clinical test set subject to human interpretation, adjudication methods are not relevant here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC study was not done. The document explicitly states: "Clinical data for the VPAP ST-A is not required". The device described is a medical device (ventilator) and not an AI or imaging diagnostic tool that would typically involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Yes, in essence, standalone testing was done. The "bench tests" and testing "against the predicate devices using common protocols" along with the use of "validated patient simulation models" can be considered a form of standalone performance assessment against engineering and functional specifications, rather than human-in-the-loop performance. However, this is for the device's core functions, not an 'algorithm' in the typical AI sense.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Objective engineering/functional specifications and predicate device performance. For the bench tests, the "ground truth" would be the pre-defined engineering design specifications for parameters like pressure, flow, trigger, cycling, and apnea/hypopnea detection, as well as the performance characteristics of the predicate devices. These are objective, measurable parameters rather than clinical outcomes or pathology.

    8. The sample size for the training set

    • Not applicable/Cannot be provided. The device is a ventilator, not an AI model that undergoes "training" in the machine learning sense from a dataset.

    9. How the ground truth for the training set was established

    • Not applicable/Cannot be provided. As there is no training set for an AI model, this question is not relevant.

    Summary of what the document does state regarding testing:

    • Type of Testing: Non-clinical bench tests and simulation models.
    • Confirmation: "All bench tests confirmed the product met the predetermined acceptance criteria."
    • Criteria Areas: Pressure, Flow, Pressure Support, Trigger and Cycling, Hypopnea and Apnea tests.
    • Comparison Basis: Against predicate devices.
    • Protocols: Common protocols for both devices (new and predicate).
    • Clinical Data: "Clinical data for the VPAP ST-A is not required."
    • Standards: Tested to appropriate FDA consensus standards and other applicable requirements, including IEC 60601-1-2:2007, IEC 60601-1:2005, and IEC 60601-1-8:2006.
    Ask a Question

    Ask a specific question about this device

    K Number
    K102513
    Manufacturer
    Date Cleared
    2011-01-05

    (126 days)

    Product Code
    Regulation Number
    868.5905
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    S9 VPAP ST WITH H5I

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The S9 VPAP ST is indicated for the treatment of Obstructive Sleep Apnea.(OSA).in.patients.weighing. . more than 66 lb (30 kg). The S9 VPAP ST is intended for use in the hospital and home.

    Device Description

    S9 VPAP ST System (S9 VPAP ST with H5i) is similar to the predicate device(s), using a blower based positive pressure system with an integrated heated humidifier and heater controller. The device platform, both hardware and firmware come from the S8 Aspen with H4i Plus (K091947) and contains a micro-processor controlled blower system that generates controlled positive airway pressure between 3-25 cmH2O as required to maintain an "air splint" for effective treatment of OSA. The system comprises the flow generator, patient tubing, mask (patient interface) and humidifier. The S9 VPAP ST contains treatment modes that come from earlier versions of ResMed Bilevel devices, these modes are known as Spontaneous (S Mode), Timed (T Mode), Spontaneous/Timed (S/T Mode) and VAuto Mode. Therapy modes come from the VPAP Tx system (K092186), VPAP ST (K080131) for S, S/T, T and VAuto and CPAP Mode comes from the S8 Aspen (K091947).

    AI/ML Overview

    This document is a 510(k) summary for the ResMed S9 VPAP ST, a non-continuous ventilator intended for the treatment of Obstructive Sleep Apnea (OSA). It outlines the device's substantial equivalence to predicate devices and the non-clinical testing performed.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly state quantitative acceptance criteria in a table format with specific performance metrics and their corresponding values for each test. Instead, it broadly states that "All bench tests confirmed the predetermined acceptance criteria." The performance is generally reported as meeting these criteria through various tests against predicate devices.

    Acceptance Criteria CategoryReported Device Performance
    Bench Tests (General)Confirmed predetermined acceptance criteria.
    Specific Bench TestsPressure, Flattening, Snore, Hypopnea, and Apnea tests passed against predicate devices using common protocols.
    Compliance with StandardsPassed all test protocols for:
    IEC 60601-1-2:2007 (EMC)Met requirements.
    IEC 60601-1:1988 Ed 2 (Basic Safety)Met requirements.
    ISO 17510-1:2007 (Sleep Apnea Therapy)Met requirements.
    ISO 8185:2007 (Respiratory Humidifiers)Met requirements.
    Safety and Effectiveness"Analysis of comparison of design, function and features of the predicate devices... together with the results of testing demonstrates that the S9 VPAP ST to be substantially equivalent to the predicate devices in terms of meeting performance criteria and function as intended."

    2. Sample Size Used for the Test Set and Data Provenance

    The document indicates that clinical data for the S9 VPAP ST is not required. Therefore, there is no human patient test set as the device's safety and effectiveness were established through comparison to predicate devices and bench testing.

    • Sample Size: Not applicable for a human test set. Bench tests were performed on the device itself.
    • Data Provenance: Not applicable for patient data. The tests were non-clinical (bench testing) and likely performed internally by ResMed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    As no clinical data was required and testing was done non-clinically, there was no expert adjudication process to establish ground truth for a patient test set. The "ground truth" for the non-clinical tests would have been the specifications and performance characteristics of the predicate devices and relevant engineering standards.

    4. Adjudication Method for the Test Set

    Not applicable, as there was no patient test set requiring expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical data for the S9 VPAP ST is not required as the predicate devices have been subjected to clinical trial requirements or validated patient simulation models have been used during the bench testing phases."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The S9 VPAP ST is a physical medical device (a ventilator), not a software algorithm providing diagnostic or treatment recommendations. Therefore, the concept of "standalone performance" in the context of an algorithm's without-human-in-the-loop performance is not directly applicable. The device itself performs its intended function (delivering positive airway pressure). Its performance was evaluated through non-clinical bench testing.

    7. The Type of Ground Truth Used

    For the non-clinical testing, the "ground truth" was based on:

    • The performance specifications and established safety/effectiveness of the predicate devices.
    • Engineering parameters and requirements defined by international standards (IEC, ISO) and internal design requirements.
    • Validated patient simulation models used during bench testing.

    8. The Sample Size for the Training Set

    The device is a hardware-based medical device with integrated software, not a machine learning algorithm that undergoes a "training phase" with a distinct "training set" of data in the typical AI/ML sense. Therefore, no specific sample size for a training set is applicable or mentioned. Its design and functionality are based on established engineering principles and prior device iterations, not statistical learning from a dataset.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" in the context of machine learning, this question is not applicable. The device's functionality and "knowledge" are engineered into its design, operating principles, and firmware, validated against industry standards and predicate device performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K080131
    Device Name
    VPAP ST
    Manufacturer
    Date Cleared
    2008-04-01

    (74 days)

    Product Code
    Regulation Number
    868.5905
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    VPAP ST

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VPAP ST is indicated for the treatment of Obstructive Sleep Apnea (OSA) in patients weighing more than 66 lb (30 kg). The optional integrated HumidAire 4i is indicated for humidification of the air delivered from a ResMed compatible CPAP therapy device. The VPAP ST and HumidAire 4i are intended for use in the hospital and home.

    Device Description

    The VPAP ST is similar to the predicate devices, VPAP Malibu (K062291), S8 Pioneer (K041209), Sullivan VPAP II ST (K961783), and VPAP Auto (K071171). The VPAP ST provides CPAP, Bilevel and VAuto therapy modes to adult patients with OSA. This is achieved through the use of a micro-processor controlled blower system that generates airway pressures as required to maintain an "air splint" for effective treatment of OSA.

    The system comprises the Flow Generator, HumidAire 4i, patient tubing, mask (patient interface), Smart Card and USB Adapter.

    The performance and functional characteristics of the VPAP ST includes all the clinician and user friendly features of the predicate devices, VPAP Malibu (K062291), S8 Pioneer (K041209), Sullivan VPAP II ST (K961783), and VPAP Auto (K071171).

    AI/ML Overview

    It appears there has been a misunderstanding. The provided document is a 510(k) Premarket Notification for the ResMed VPAP ST, a medical device used for treating Obstructive Sleep Apnea (OSA).

    This type of document is a regulatory submission to the FDA to demonstrate substantial equivalence to a predicate device, not a study report detailing acceptance criteria for performance and experimental results.

    Therefore, the requested information regarding a study proving acceptance criteria for device performance metrics (like sensitivity, specificity, accuracy, etc., which are common for AI/diagnostic devices) is not present in this document. The document confirms that design and verification activities were performed and that the VPAP ST met pre-determined acceptance criteria, but it does not provide the specifics of those criteria or the detailed results of a performance study in the format typically associated with AI or diagnostic device evaluation.

    The key statement is: "Design and Verification activities were performed on the VPAP ST as a result of the risk analysis and product requirements. All tests confirmed the VPAP ST met the predetermined acceptance criteria." This refers to engineering and functional testing, not a clinical performance study with human readers, AI, FPs, FNs, etc.

    Below is a breakdown of why I cannot fulfill your request based on the provided text, and what information is available:

    • 1. A table of acceptance criteria and the reported device performance: Not provided. The document states that the device met "predetermined acceptance criteria" resulting from risk analysis and product requirements, but it doesn't list them or specific performance metrics from a clinical study.
    • 2. Sample size used for the test set and the data provenance: Not provided, as this is not a study report.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided.
    • 4. Adjudication method for the test set: Not provided.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This device is a ventilator, not an AI diagnostic tool. Therefore, an MRMC study comparing human readers with and without AI assistance is irrelevant and not present.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: This device is a piece of hardware, a ventilator. The concept of "standalone algorithm performance" does not apply.
    • 7. The type of ground truth used: Not applicable in the context of this device's regulatory submission.
    • 8. The sample size for the training set: Not applicable. This is not an AI/machine learning device.
    • 9. How the ground truth for the training set was established: Not applicable.

    Information that is available from the document:

    • Device Name: VPAP ST
    • Intended Use: For the treatment of Obstructive Sleep Apnea (OSA) in patients weighing more than 66 lb (30 kg). The optional integrated HumidAire 4i is indicated for humidification of the air. Intended for use in the hospital and home.
    • Regulatory Basis: Substantial equivalence to predicate devices (VPAP Malibu, S8 Pioneer, Sullivan VPAP II ST, VPAP Auto).
    • Key Design Aspects: Provides CPAP, Bilevel, and VAuto therapy modes through a micro-processor controlled blower system.
    • Regulatory Clearance: K080131, cleared on April 1, 2008.

    To reiterate, this document is focused on regulatory clearance for a physical medical device (a ventilator) based on substantial equivalence, not on the performance evaluation of a diagnostic or AI-driven system against specific acceptance criteria like sensitivity or specificity.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1