Search Results
Found 1 results
510(k) Data Aggregation
(82 days)
VIDAS Troponin I Assay is an automated quantitative test for use on the VIDAS instruments for the determination of human cardiac troponin 1 in serum or plasma (lithium heparinate) using the ELFA (Enzyme-Linked Fluorescent Assay) technique
The VIDAS Troponin I (TNI) Assay is an enzyme-linked fluorescent immunoassay (ELFA) performed in an automated VIDAS® instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR), serves as the solid phase as well as a pipettor for the assay. Reagents for the assay are in the sealed TNI Reagent Strips.
The sample is transferred into the wells containing anti-cardiac troponin I antibodies labeled with alkaline phosphatase (conjugate). The sample/conjugate mixture is cycled in and out of the SPR for a specified length of time. Troponin I present in the specimen will bind to the anti-cardiac troponin I immunoglobulin coating the interior of the SPR. Unbound sample components are washed away.
A fluorescent substrate, 4-methylumbelliferyl phosphate, is introduced into the SPR. Enzyme remaining on the SPR wall will catalyze the conversion of the substrate to the fluorescent product 4-methylumbelliferone. The optical scanner in the instrument measures the intensity of fluorescence. When the VIDAS TNI assay is completed, the results are analyzed automatically by the computer, a test value is generated, and a report is printed for each sample.
Here's a breakdown of the acceptance criteria and study details for the VIDAS Troponin I (TNI) Assay, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" as a separate, pre-defined column. Instead, it presents performance data for both the VIDAS TNI assay and its predicate device, the Stratus CS Cardiac Troponin I Testpak, implying that the VIDAS TNI assay needed to demonstrate comparable performance to the predicate. The "Reported Device Performance" column reflects the performance of the VIDAS TNI assay.
Performance Characteristic | Predicate Device (SCS cTnl) Performance | VIDAS TNI Assay Performance |
---|---|---|
Nonclinical Testing | ||
Within-run precision (% CV) | 2.7 - 4.3 | 1.3 - 2.5 |
Dilution % Recovery | 98.1 - 102.0 | 97 - 111 |
Analytical Sensitivity | 0.03 µg/l |
Ask a specific question about this device
Page 1 of 1