Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    Chemical Indicator; Celerity™ HP Multivariable Chemical Indicator; VERIFY HPU Chemical Indicator; VERIFY
    VH2O2 Indicator Tape

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Celerity™ HP Chemical Indicator is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units, when placed within packs to be sterilized, through a visible change from red to orange/yellow or lighter, when the device has been exposed to the:

    • · Lumen, Non Lumen, Flexible, Fast Non Lumen, Fast or Specialty sterilization cycle of a V-PRO® Low Temperature Sterilization System, or
    • · Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP STERRAD® System, including those systems with ALLClear Technology.

    The Celerity™ HP Multivariable Chemical Indicator is a vaporized hydrogen peroxide multivariable chemical indicator. It is designed for routine monitoring of the following cycles:

    • Lumen, Non Lumen, Flexible, Fast Non Lumen, Fast or Specialty sterilization ● cycle of a V-PRO Low Temperature Sterilization System, or
    • Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP STERRAD ● System, including those systems with ALLClear Technology.

    The VERIFY® HPU Chemical Indicator is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units when placed within packs to be sterilized to indicate, through a visible change from magenta to yellow, when the device has been exposed to the Lumen, Non Lumen, Flexible, Fast Non Lumen, or Specialty sterilization cycle of a V-PRO® Low Temperature Sterilization System.

    The Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink) to peach, yellow or lighter. The Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches.
    The tape may be used in the following sterilization cycles:

    • . Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible, and Specialty Cycles of the V-PRO: 1, 1 Plus, maX, maX 2, 60, and s2 Low Temperature Sterilization Systems.
    • STERRAD® 100S Sterilizer (Default Cycle)
    • . Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without ALLClear
    • Standard, Flex Scope, Express and DUO Cycles of the STERRAD® 100NX . Sterilizer without ALLClear.
    Device Description

    The Celerity "M HP Chemical Indicator is an ISO 11140-1:2014 Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units through a visible change from magenta to orange/vellow or lighter. when the device has been exposed. No changes have been made to the device other than additional testing and updating the labeling for the Specialty Cycle.

    The Celerity™ HP Multivariable Chemical Indicator is a chemical indicator strip consisting of indicator ink containing the reactive chemicals printed on one end of a polypropylene strip. If the critical variables are achieved, the color of the indicator ink changes from magenta to orange/yellow or lighter yellow when exposed to the V-PRO® Low Temperature Sterilization System cycles or ASP STERRAD® System cycles. The indicator is validated to function as a multiple variable indicator with increased resistance characteristics similar to ISO 11140-1:2014 end points for a Type 4 Vaporized Hydrogen Peroxide (VHP) Chemical Indicator (CI). No changes have been made to the device other than additional testing and updating the labeling for Specialty cycle.

    The VERIFY HPU Chemical Indicator (Chemical Indicator) is a Type 1 process indicator in accordance with ANSI/AAMI/ISO 11140-1:2014. The indicators are used in the processing cycles to indicate exposure to the following sterilization cycles in the V-PRO Low Temperature Sterilization Systems. The indicator ink spot on the proposed Chemical Indicator undergoes a color change from magenta to vellow when exposed to the defined processing conditions.

    The VERIFY VH2O2 Indicator Tape (Indicator Tape) is a 3/" wide crepe paper tape printed with diagonal stripes of a vaporized hydrogen peroxide (VHP) reactive ink. The ink is sealed with a varnish, the function of which is to inhibit removal of the ink via transference or by the adhesive (as the tape is dispensed). The reactive ink meets the performance specifications for a Type 1 process indicator for vaporized hydrogen peroxide as defined in ANSI/AAMI/ISO 11140-1:2014.

    AI/ML Overview

    The provided text describes premarket notification (510(k)) submissions for four chemical indicators (CIs) used in sterilization processes:

    1. Celerity™ HP Chemical Indicator (CI) (K231488) on pages 3-5.
    2. Celerity™ HP Multivariable Chemical Indicator (CI) (K231488) on pages 6-9.
    3. VERIFY HPU Chemical Indicator (CI) (K231488) on pages 10-12.
    4. VERIFY VH2O2 Indicator Tape (K231488) on pages 13-16.

    For each of these devices, the core purpose of the 510(k) submission is to demonstrate substantial equivalence to a legally marketed predicate device, primarily by showing that the indicators function as intended even with an expanded "Indications for Use" to include the "Specialty" sterilization cycle of the V-PRO® Low Temperature Sterilization System.

    The information provided does not pertain to a medical device that uses AI or machine learning, nor does it involve human readers/experts interpreting medical images. Instead, it describes performance testing of chemical indicators, which are physical objects designed to change color under specific sterilization conditions. Therefore, the requested information elements such as "Number of experts used to establish ground truth," "Adjudication method," "MRMC comparative effectiveness study," "Effect size of human readers improvement with AI," "Stand-alone algorithm performance," "Training set sample size," and "How ground truth for training set was established" are not applicable to this type of device and study.

    The acceptance criteria and reported performance for each device are derived from "Simulated Use Testing" and are consistently focused on the physical color change and the indicator's integrity.

    Here's a summary of the relevant information provided for each product:


    1. Celerity™ HP Chemical Indicator (CI) (K231488)

    Acceptance Criteria and Reported Device Performance:

    TestingAcceptance CriteriaStudy Result
    Simulated Use TestingComplete color change No reversionComplete color change No reversion

    Study that proves the device meets the acceptance criteria:

    • Sample Size: Not explicitly stated, but the context implies a sufficient number of indicators were tested to demonstrate consistent performance.
    • Data Provenance: Not explicitly stated, but the manufacturing facility is in the United Kingdom, and the testing simulates "typical in-use applications in a V-PRO maX 2 Sterilizer Specialty cycle." This suggests laboratory-based, prospective testing designed to validate performance under specified conditions.
    • Type of Ground Truth: The ground truth is the physical observation of the color change of the chemical indicator after exposure to the V-PRO maX 2 Sterilizer Specialty cycle, indicating that the sterilization conditions were met. This is a direct physical outcome.

    2. Celerity™ HP Multivariable Chemical Indicator (CI) (K231488)

    Acceptance Criteria and Reported Device Performance:

    TestingAcceptance CriteriaStudy Result
    Simulated Use TestingComplete color change No reversionComplete color change No reversion

    Study that proves the device meets the acceptance criteria:

    • Sample Size: Not explicitly stated.
    • Data Provenance: Not explicitly stated, but similar to the previous Celerity CI, implies laboratory-based, prospective testing.
    • Type of Ground Truth: The ground truth is the physical observation of the color change of the chemical indicator after exposure to the V-PRO maX 2 Sterilizer Specialty cycle, confirming proper function as a multivariable indicator.

    3. VERIFY HPU Chemical Indicator (CI) (K231488)

    Acceptance Criteria and Reported Device Performance:

    TestingAcceptance CriteriaStudy Result
    Simulated Use TestingComplete color change No reversionComplete color change No reversion

    Study that proves the device meets the acceptance criteria:

    • Sample Size: Not explicitly stated.
    • Data Provenance: Not explicitly stated, but implies laboratory-based, prospective testing.
    • Type of Ground Truth: The ground truth is the physical observation of the color change of the chemical indicator (magenta to yellow) after exposure to the V-PRO maX 2 Sterilizer Specialty cycle.

    4. VERIFY VH2O2 Indicator Tape (K231488)

    Acceptance Criteria and Reported Device Performance:

    TestingAcceptance CriteriaStudy Result
    Simulated Use TestingComplete color change Remained adhered after cycleComplete color change Remained adhered after cycle

    Study that proves the device meets the acceptance criteria:

    • Sample Size: Not explicitly stated.
    • Data Provenance: Not explicitly stated, but implies laboratory-based, prospective testing. The manufacturing facility is in Buena Park, CA (USA).
    • Type of Ground Truth: The ground truth is the physical observation of the color change of the indicator tape (pink to peach, yellow or lighter) and its physical integrity (remaining adhered) after exposure to the V-PRO maX 2 Sterilizer Specialty cycle.

    Summary of non-applicable AI/ML specific criteria:

    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth is based on physical/chemical reaction.
    • Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. No human adjudication of complex data; rather, direct observation of state change.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted diagnostic device.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical chemical indicator, not an algorithm.
    • The sample size for the training set: Not applicable. These are physical devices, not AI models requiring training data.
    • How the ground truth for the training set was established: Not applicable for the same reason.

    In all cases, the primary purpose of the testing described is to confirm that the chemical indicators reliably undergo the expected color change (and adhere, for the tape) when exposed to the specific V-PRO maX 2 Sterilizer Specialty cycle, thereby demonstrating their continued safety and effectiveness for an expanded indication.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172753
    Manufacturer
    Date Cleared
    2018-02-09

    (150 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    VERIFY VH2O2 Indicator Tape

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VERIFY™ VH2O2 Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink) to peach, yellow or lighter. The VERIFY™ VH2O2 Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches.

    The tape may be used in the following sterilization cycles:

    • Non Lumen, Flexible, Lumen and Fast Non Lumen Cycles of the V-PRO 1, 1 Plus, maX, 60 and maX 2 Sterlizers
    • · Standard Cycle of the STERRAD® 100S Sterilizer
    • · Standard and Advanced Cycles of the STERRAD® NX Sterilizer
    • · Express, Flex Scope, Duo and Standard Cycles of the STERRAD® 100 NX Sterilizer
    Device Description

    The VERIFY VH2O2 Indicator Tape consists of a special coated crepe paper with diagonal lines of a chemical indicator specific for hydrogen peroxide sterilization. It meets the performance specifications for a Type 1 process indicator as defined in ANSI/AAMI/ISO 11140-1:2014.

    The VERIFY | VH2O2 Indicator Tape and its packaging are manufactured in the absence of lead, heavy metals, natural rubber latex and dry natural rubber.

    AI/ML Overview

    The provided document is a 510(k) summary for the VERIFY™ VH2O2 Indicator Tape, a Class II medical device (Sterilization Process Indicator). It details the device's characteristics and compares it to a predicate device (VERIFY™ VH2O2 Indicator Tape, K160485) to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of formal acceptance criteria with specific numerical targets. However, it implies acceptance based on compliance with standards and successful performance in specific tests. The key performance indicators evaluated are:

    Acceptance Criteria (Implied)Reported Device Performance
    Conforms to ISO 11140-1:2014 Type 1 requirements"Proprietary - conforms with ISO 11140-1:2014 Type 1 requirements"
    Distinguishes between processed and unprocessed units"color change from the start color (pink) to peach, yellow or lighter"
    Maintains performance in "Fast Non Lumen Cycle" of V-PRO maX 2"Simulated Use Testing in Fast Non Lumen Cycle" resulted in "Pass"
    Maintains performance over an 18-month shelf-life"Stability Testing for 18 month Shelf Life" resulted in "Pass"
    Consistent adhesive performance"Testing included in this submission demonstrating" (adhesive performance)
    Endpoint color stability"endpoint color stability" tested during shelf-life
    BiocompatibilityReferenced from predicate device (K160485)
    Light stabilityReferenced from predicate device (K160485)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide specific sample sizes (e.g., number of indicator tapes tested) for the simulated use testing or stability testing. It also does not specify the country of origin of the data or whether the studies were retrospective or prospective.

    The testing for the current submission focused on:

    • "performance was tested in the V-PRO maX 2's Fast Non Lumen Cycle with loads under simulated use conditions under pass and fail conditions."
    • "Shelf-life testing evaluated CI performance per ANSI/AAMI/ISO 11140-1:2014, endpoint color stability and tape adhesion after various times of storage within the labeled storage conditions."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable to the VERIFY™ VH2O2 Indicator Tape. This device is a chemical indicator that undergoes a physical color change. Its performance is evaluated objectively against chemical parameters and color charts, not through expert interpretation of complex medical images or diagnoses. Therefore, there's no mention of experts establishing a "ground truth" in the way it would be for a diagnostic AI device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, the device's performance is based on a direct chemical reaction resulting in a color change, which is evaluated against established physical criteria (e.g., an industry standard color chart or specific chemical exposure levels), rather than subjective interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The VERIFY™ VH2O2 Indicator Tape is a standalone chemical indicator, not an AI-assisted diagnostic tool. Therefore, no MRMC study or evaluation of human reader improvement with AI assistance was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is inherently a "standalone" indicator. Its performance is entirely determined by its chemical reaction to hydrogen peroxide and the resulting color change, without any human-in-the-loop interaction beyond observation. The reported "Simulated Use Testing" and "Stability Testing" are essentially standalone performance evaluations of the indicator tape itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance relies on objective chemical and physical parameters as defined by:

    • ANSI/AAMI/ISO 11140-1:2014 Type 1 requirements: This standard defines the performance characteristics for process indicators, including the specific conditions (e.g., concentration of sterilant, temperature) under which the indicator must change color, and the expected color change.
    • Sterilization cycle parameters: The device is tested against the actual sterilization cycles of specific sterlizers (e.g., V-PRO maX 2, STERRAD 100S, STERRAD NX, STERRAD 100 NX) under "pass and fail conditions." The "ground truth" here is whether the sterilization process itself was successful or not, based on the validated parameters of the sterilizer, and whether the indicator accurately reflected that state.
    • Visual color change: The ground truth for distinguishing processed from unprocessed units is the observed color change from pink to peach, yellow, or lighter. This is a direct physical observable.

    8. The sample size for the training set

    Not applicable. This device is a chemical indicator, not a machine learning or AI algorithm that requires a "training set." Its design and performance are based on chemical engineering and adherence to international standards, not data-driven learning.

    9. How the ground truth for the training set was established

    Not applicable, for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160485
    Manufacturer
    Date Cleared
    2016-06-01

    (100 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    VERIFY VH2O2 Indicator Tape

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VERIFY™ VH2O2 Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink) to peach, yellow or lighter.

    The tape may be used in the following sterilization cycles:

    • · Non Lumen, Flexible and Lumen Cycles of the V-PRO 1, 1 Plus, maX and 60 Low Temperature Sterilizers
    • · Standard Cycle of the STERRAD® 100S Sterilizer
    • · Standard and Advanced Cycles of the STERRAD® NX Sterilizer
    • · Express, Flex Scope, Duo and Standard Cycles of the STERRAD® 100 NX Sterilizer
    Device Description

    The VERIFY VH2O2 Indicator Tape consists of a special coated crepe paper with diagonal lines of a chemical indicator specific for hydrogen peroxide sterilization. It meets the performance specifications for a Type 1 process indicator as defined in ANSI/AAMI/ISO 11140-1:2014.

    The VERIFY | VH2O2 Indicator Tape and its packaging are manufactured in the absence of lead, heavy metals, natural rubber latex and dry natural rubber.

    AI/ML Overview

    The medical device is the VERIFY™ VH2O2 Indicator Tape, which is a chemical sterilization process indicator.

    Here's an analysis of the provided information, addressing your specific questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Test)Reported Device Performance (Result)
    ANSI/AAMI/ISO 11140-1:2014 Vaporized Hydrogen Peroxide Process Indicator Performance TestingPass
    Simulated Use: Pass Performance Testing (in V-Pro and STERRAD sterilizers with loads)Pass
    Simulated Use: Fail Performance Testing (in V-Pro and STERRAD sterilizers with loads)Pass
    Endpoint Stability (color of processed tape at various time points)Pass
    Light Stability after Removal from Packaging (before exposure to sterilization)Pass
    Adhesion Evaluation (ASTM adhesion standard and simulated use)Pass
    Shelf Life Evaluation (CI performance and tape adhesion)Pass
    Hydrogen Peroxide Residual TestingPass
    Off-set (transference) Performance TestingPass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes for each test in the test set. It mentions tests being conducted "in an ANSI/AAMI/ISO 18472:2006-conforming resistometer" and "in hydrogen peroxide sterilizers (V-Pro and STERRAD) with loads under simulated use conditions." The provenance of the data is not explicitly stated as country of origin, but it is implied to be from the manufacturer's (STERIS Corporation) testing, which is based in Mentor, Ohio, USA. The testing appears to be prospective as it was conducted to demonstrate the performance of the new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. For chemical indicators like this, ground truth is typically established by physical/chemical properties and standardized tests, not by human expert interpretation in the same way an imaging AI might be evaluated. The "Pass" results are against established standards (ANSI/AAMI/ISO).

    4. Adjudication Method for the Test Set

    This information is not applicable in the traditional sense for a chemical indicator. The "Pass" or "Fail" of the tests is determined by adherence to predefined measurable criteria set forth in the ISO standards and internal performance specifications, rather than by a panel of human adjudicators reviewing outputs.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically conducted for diagnostic devices where human interpretation of results is involved, and the AI's impact on human performance is being assessed. The VERIFY™ VH2O2 Indicator Tape is a chemical indicator that provides a direct visual color change, not an interpretive diagnostic.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, the testing performed can be considered "standalone" in the sense that the device's performance (color change, adhesion, stability) was evaluated directly against scientific and industry standards without a human-in-the-loop component for the device's core function. The device itself is designed to provide a clear, unambiguous signal (color change) that indicates processing. The evaluation focuses on whether this signal occurs accurately and reliably under specified conditions.

    7. The Type of Ground Truth Used

    The ground truth used is based on established industry standards and physical/chemical properties. This includes:

    • ANSI/AAMI/ISO 11140-1:2014: This standard defines the requirements for chemical indicators for sterilization.
    • ANSI/AAMI/ISO 18472:2006: This standard relates to resistometers, which are used to evaluate sterilization processes and indicators.
    • ASTM adhesion standard: Used for evaluating tape adhesion.
    • Measurable parameters for color change (from pink to peach/yellow or lighter) under defined sterilization conditions (pass conditions) and no change under fail conditions.
    • Shelf-life stability of the indicator.
    • Residual chemical levels (hydrogen peroxide).

    8. The Sample Size for the Training Set

    This document describes a physical device, a chemical indicator tape, not an algorithm that requires a "training set" in the context of machine learning or AI. Therefore, the concept of a training set sample size does not apply.

    9. How the Ground Truth for the Training Set Was Established

    As explained above, this device does not utilize a training set in the AI sense. Ground truth relates to the performance standards and physical/chemical characteristics of the indicator itself, established by international and industry standards, and verified through laboratory and simulated-use testing.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1