Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K231490
    Manufacturer
    Date Cleared
    2023-08-07

    (76 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Celerity 20 HP Biological Indicator; VERIFY V24 Self-Contained Biological Indicator

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VERIFY® V24 Self-Contained Biological Indicator is intended for routine monitoring of the following sterilizer cycles:

    · Lumen, Non Lumen, Flexible, Fast, Fast Non Lumen and Specialty Cycles of the V-PRO® 1, 1 Plus, maX, maX 2, 60 and s2 Low Temperature Sterilization Systems

    · STERRAD® 100S Sterilizer (Default Cycle)

    · STERRAD® 200 Sterilizer (Default Cycle)

    · Standard and Advanced Cycles of the STERRAD® NX Sterilizer to include sterilizers with ALLClear® Technology

    · Express, Flex Scope and Standard Cycles of the STERRAD® 100NX Sterilizer to include sterilizers with ALLClear® Technology.

    The Celerity 20 HP Biological indicator is intended for the routine monitoring of the following sterilizer cycles:

    · Lumen, Non Lumen, Fast, Flexible and Specialty Cycles of the V-PRO 1, 1 Plus, maX, maX, maX, maX 2, 60 and s2 Low Temperature Sterilization Systems

    · STERRAD® 100S Sterilizer (Default Cycle)

    • Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without ALLClear® Technology
      · Standard, Flex Scope, Express, and DUO Cycles of the STERRAD® 100NX Sterilizer with or without ALLClear® Technology

    When used in conjunction with the Celerity HP Incubator or the Celerity Incubator, the Celerity HP BI provides a fluorescent result within 20 minutes.

    Device Description

    The product is intended to monitor the vapor phased hydrogen peroxide sterilization cycles described in the indication for use. It produces an optical change in signal that is detected by the STERIS proprietary reader, Celerity 20 HP Incubator, within 20 minutes to confirm the viability of the biological indicator at the end of a sterilization process. The product consists of Geobacillus stearothermophilus spores and a defined nutrient media in a plastic vial. A reporter enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.

    The VERIFY V24 Self-Contained Biological Indicator (SCBI) is used by healthcare facilities to monitor the V-PRO® Low Temperature Sterilization Systems. It is designed to accompany medical devices placed in the sterilizer.

    The user places the VERIFY V24 SCBI into the V-PRO® Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.

    The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activator automatically seals the SCBI vial and releases the growth media.

    The activated SCBI is incubated at 55-60℃ for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

    AI/ML Overview

    This document is a 510(k) premarket notification for two biological indicators: the Celerity 20 HP Biological Indicator and the VERIFY V24 Self-Contained Biological Indicator, both manufactured by STERIS Corporation. The purpose of the submission is to obtain clearance for these devices to routinely monitor additional sterilizer cycles, specifically the "Specialty Cycle" of the V-PRO® maX 2 Low Temperature Sterilization System, as well as several other specified STERRAD® cycles.

    It's important to note that this document is for a sterilization process indicator, not a medical device in the typical sense (e.g., a diagnostic AI tool). Therefore, many of the typical acceptance criteria and study designs associated with AI/ML-driven medical devices (like MRMC studies, expert consensus for imaging, etc.) do not apply here. This is a technical performance study for a biological measurement device, focusing on its ability to accurately reflect sterilization efficacy.

    Based on the provided text, here's a breakdown of the acceptance criteria and the study that proves the device meets them for each of the two devices mentioned:


    For the Celerity 20 HP Biological Indicator:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Performance (Conclusion)
    Verification of Celerity HP BI Performance after V-PRO Specialty Cycle Extended AerationCelerity BIs exposed to the Specialty Cycle demonstrate essentially equivalent or not significantly different resistance as compared to SCBIs exposed to the 136L V-PRO Sterilizer Fast Non Lumen Cycle.Pass
    Final Process Qualification of the VPRO maX2 Sterilizer Specialty CycleThe V-PRO maX 2 Sterilizer Specialty Cycle final process qualification was successful.Pass

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated in the provided text. The term "Celerity BIs" and "SCBIs" are used, implying multiple biological indicators were tested, but the exact number is not given.
    • Data Provenance: The studies were non-clinical tests conducted by STERIS Corporation. The location (Mentor, OH) of the manufacturing and sponsor facility would suggest the studies were conducted in the USA. The data appears to be prospective as it's specifically for a new cycle qualification.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not applicable in the context of this device. Ground truth for a biological indicator is typically established by the physical inactivation of bacterial spores under controlled, validated sterilization conditions, and the subsequent growth/no-growth determination. This is a direct measurement of sterility, not an interpretation of an image or signal that would require human expert consensus in the medical imaging sense.

    4. Adjudication Method for the Test Set:

    • Not applicable as this is a direct biological and physical process validation, not a diagnostic interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    • No, an MRMC study was not done. This type of study is relevant for diagnostic devices where human readers interpret output (e.g., images) with and without AI assistance. This device is a sterility indicator; its performance is measured by its biological response to sterilization, not by how humans interpret its output in a diagnostic scenario.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was Done:

    • Yes, in principle. The performance of the biological indicator itself is "standalone." Its function is to respond directly to the sterilization process. The "reading" of the Celerity 20 HP BI is performed by the Celerity HP Incubator which detects a fluorescent signal within 20 minutes to confirm viability. While an instrument is involved, it's not an "algorithm" in the AI sense, but rather a direct measurement device. The studies described are assessing the BI's intrinsic performance against the sterilization cycle.

    7. The Type of Ground Truth Used:

    • Biological Inactivation/Sterility: The ground truth is the actual biological state of the spores on the biological indicator (e.g., killed or survived) after exposure to varying sterilization conditions. This is determined by observing the growth (or lack thereof) of the Geobacillus stearothermophilus spores under specific culture conditions, which then produces a fluorescent moiety if growth occurs. The "successfully qualified" aspect of the sterilizer cycle itself is the ultimate ground truth for the process the BI is monitoring.

    8. The Sample Size for the Training Set:

    • Not applicable. Biological indicators are not "trained" in the machine learning sense. Their function relies on the pre-defined biological characteristics of the Geobacillus stearothermophilus spores and the biochemical reaction designed into the device.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. (See #8)

    For the VERIFY V24 Self-Contained Biological Indicator:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Performance (Conclusion)
    Verification of V24 SCBI Performance after V-PRO Specialty Cycle Extended AerationSCBIs exposed to the Specialty Cycle demonstrate essentially equivalent or not significantly different resistance as compared to SCBIs exposed to the 136L V-PRO Sterilizer Fast Non Lumen Cycle.Pass
    Final Process Qualification of the VPRO maX2 Sterilizer Specialty CycleThe V-PRO maX 2 Sterilizer Specialty Cycle final process qualification was successful.Pass

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated. Similar to the Celerity BI, "SCBIs" (Self-Contained Biological Indicators) are mentioned, implying multiple units were tested, but the exact count is not provided.
    • Data Provenance: The studies were non-clinical tests conducted by STERIS Corporation (likely in the USA). The data appears to be prospective for the new cycle qualification.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not applicable. Ground truth for a biological indicator is typically established by the direct biological response of the bacterial spores to sterilization. While there's a visual detection of growth (color change/turbidity), this is a direct observation of a biological event, not a subjective interpretation requiring expert consensus.

    4. Adjudication Method for the Test Set:

    • Not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    • No, an MRMC study was not done.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was Done:

    • Yes, in principle. The performance of the biological indicator itself is "standalone." Its function is to respond directly to the sterilization process. The "reading" of the VERIFY V24 SCBI involves visual detection of growth (media color change/turbidity) after incubation. While this involves human observation, it's observing a biological fact (growth/no growth), not interpreting an image or signal in a diagnostic sense. The studies described are assessing the BI's intrinsic performance against the sterilization cycle.

    7. The Type of Ground Truth Used:

    • Biological Inactivation/Sterility: The ground truth is the actual biological state of the spores on the biological indicator (e.g., killed or survived) after exposure to varying sterilization conditions. This is determined by observing the growth (color change from orange to yellow and/or turbidity) or no-growth (media remains orange and non-turbid) of the Geobacillus stearothermophilus spores after incubation. The "successfully qualified" aspect of the sterilizer cycle itself is the ultimate ground truth for the process the BI is monitoring.

    8. The Sample Size for the Training Set:

    • Not applicable. Biological indicators are not "trained."

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Celerity HP Chemical Indicator, Celerity Vaporized VH2O2 Process Indicator Adhesive Label, VERIFY V24
    Self-Contained Biological Indicator Vial Label

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Celerity™ HP Chemical Indicator is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units, when placed within packs to be sterilized, through a visible change from red to orange/yellow, when the device has been exposed to the:

    · Lumen. Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or

    · Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP STERRAD® System, including those systems with ALLClear Technology.

    The Celerity™ Vaporized VH2O2 Process Indicator Adhesive Label is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed units, when affixed to packs to be sterilized, through a visible change from magenta to orange/yellow or lighter, when the pack has been exposed to the:

    · Lumen, Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or

    · Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP STERRAD® System, including STERRAD NX and 100NX with ALLClear Technology.

    The VERIFY V24 Self-Contained Biological Indicator Vial Label is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed SCBIs, whilst affixed to SCBI vials to be sterilized, through a visible change from red to orange/yellow, when the SCBI has been exposed to the:

    · Lumen. Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or

    • Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP STERRAD® System, including STERRAD NX and 100NX with ALLClear Technology.

    Device Description

    The Celerity HP Chemical Indicator is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units through a visible change from magenta to orange/yellow or lighter, when the device has been exposed to the:

    • . Lumen, Non-Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or
    • . Standard. Advanced, Express, Flex or Duo cycles of an ASP STERRAD® System, including those systems with ALLClear Technology.

    The Celerity "10 HP Chemical indicator is provided in three formats:

    • Version 1: Celerity™ Chemical Indicator (CI) .
    • . Version 2: Celerity™ Vaporized |VH2O2 Process Indicator Adhesive Label (PI)
    • Version 3: VERIFY V24 Self-Contained Biological Indicator Vial Label (PI)

    The three formats differ in their intended location inside the load to be sterilized. Version 1 is to be placed inside a pack to be sterilized. Version 2 is to be affixed to the outside of a pack by means of the adhesive back. Version 3 is to be affixed to the outside of a Self-Contained Biological Indicator by means of the adhesive back.

    The proposed devices are manufactured by application of the indicator ink by printing onto a substrate in a variety of shapes in three constructions:

    • Construction C Polypropylene ●
    • Construction B - Spun-bonded polyolefin with an adhesive and a glassine backing paper
    • Construction D - polypropylene with an adhesive and a glassine backing paper.

    The proposed devices will consist of one of the above substrates with an indicator ink printed thereon.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Testing CategoryMethodologyAcceptance CriteriaStudy Result
    Type 1 Performance TestingANSI/AAMI/ISO 11140-1:2014- When exposed to VHP Resistometer cycles at the Fail parameters (Test Points 2 and 3): 100% Celerity™ HP Chemical Indicator stripes must show a FAIL result.
    • When exposed to VHP Resistometer cycles at the Pass condition (Test Point 1): Minimum 90% Celerity™ HP Chemical Indicator stripes must show a PASS result. | Pass |
      | Simulated Use Testing | Internal Method Used | - When exposed to V-PRO maX Non Lumen 1/84 cycles with ≥50 lb load (Fail condition): 100% Celerity™ HP Chemical Indicator stripes must show a FAIL result.
    • When exposed to V-PRO and STERRAD cycles with the worst case loads: Minimum 90% Celerity™ HP Chemical Indicator stripes must show a PASS result. | Pass |
      | Fluorescent Light Stability Testing | Internal Method & ANSI/AAMI/ISO 11140-1:2014 | - Indicator start color to remain unchanged after exposure to fluorescent lighting for ≥ 30 days before processing.
    • When exposed to VHP Resistometer cycles at the Fail parameters following exposure to fluorescent lighting for ≥ 30 days: 100% of Indicators tested must show a FAIL results (non-visible color change or color change to another color except orange/yellow).
    • When exposed to VHP Resistometer cycles at the PASS parameters following exposure to fluorescent lighting for ≥ 30 days: Minimum 90% Indicators tested must show a PASS result (orange/yellow or lighter). | Pass |
      | Temperature Extremes Exposure (Freeze/Thaw) Testing | Internal Method & ANSI/AAMI/ISO 11140-1:2014 | - Indicator start color to remain unchanged after exposure to extreme temperatures before processing.
    • When exposed to VHP Resistometer cycles at the Fail parameters following exposure to extreme temperatures: 100% of Indicators tested must show a FAIL results (non-visible color change or color change to another color except orange/yellow).
    • When exposed to VHP Resistometer cycles at the PASS parameters following exposure to extreme temperatures: Minimum 90% Indicators tested must show a PASS result (orange/yellow or lighter). | Pass |
      | Transference Testing | ANSI/AAMI/ISO 11140-1:2014 | - Indicators placed between a piece of the same substrate and held together with tape, then processed in a double V-PRO maX 2 Lumen cycle (no load) and evaluated for off-set, bleed and migration. | Pass |
      | | | - Samples with reference and text ink processed in a double V-PRO maX 2 Lumen cycle (no load) and evaluated for smearing, discoloration and fading. | Pass |
      | Adhesion Stability (Vials) Testing | Internal Method Used | - When exposed to two consecutive V-PRO maX 2 Lumen cycles:
      • 0% samples show bleeding of the adhesive
      • 100% must be firmly affixed
      • 0% must show signs of peeling
      • Minimum 90% must show a PASS result | Pass |
      | Shelf Life Testing | ANSI/AAMI/ISO 11140-1:2014 | - After the process indicator has been aged: Minimum of 90% of indicators must PASS Type 1 PASS conditions.
    • After the process indicator has been aged: 100% of indicators must FAIL. | Pass |
      | Post-Processing Stability Testing | Internal Method Used | - Indicator processed ink color remains unchanged for a minimum of 6 months following exposure to a PASS or FAIL cycle. | Pass |

    2. Sample Size and Data Provenance for Test Set:

    The document does not explicitly state the specific sample sizes (number of indicators) used for each individual test in the performance studies. However, for "Type 1 Performance Testing", "Simulated Use Testing", "Fluorescent Light Stability Testing", "Temperature Extremes Exposure Testing", "Shelf Life Testing", and "Adhesion Stability (Vials) Testing", the acceptance criteria mention percentages (e.g., "100% Celerity™ HP Chemical Indicator stripes", "Minimum 90% Celerity™ HP Chemical Indicator stripes"). This implies that multiple indicators were tested for each condition.

    The data provenance is not explicitly stated in terms of country of origin. The manufacturing facility is in the United Kingdom, but the testing may have occurred elsewhere. The studies are described in the context of a 510(k) submission, which typically involves prospective testing designed to meet regulatory requirements.

    3. Number of Experts and Qualifications for Ground Truth of Test Set:

    This information is not provided. The performance testing for these chemical indicators relies on objective color changes and adherence to established standards (ANSI/AAMI/ISO 11140-1:2014), rather than expert interpretation of the results. The 'ground truth' is the objective physical change of the chemical indicator based on exposure to sterilization conditions.

    4. Adjudication Method for the Test Set:

    Not applicable. The performance testing for chemical indicators is based on objective observable color changes and compliance with quantitative standards, not subjective interpretations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic devices where human readers interpret images or other data, and the comparison is how AI assistance affects their diagnostic accuracy. Chemical indicators, like the Celerity™ HP, are designed for direct visual interpretation of a color change, not complex interpretation by multiple human readers.

    6. Standalone (Algorithm Only) Performance Study:

    Yes, the studies described are standalone performance studies of the chemical indicator itself. The device is a "Type 1 vaporized hydrogen peroxide sterilization process indicator" that provides a direct visual cue (color change) to indicate exposure to sterilization conditions. Its performance is evaluated based on its intrinsic ability to change color accurately under specific conditions, without human interpretation beyond basic color recognition.

    7. Type of Ground Truth Used:

    The ground truth used is primarily based on physical exposure to defined sterilization conditions (VHP Resistometer cycles at specific "Fail parameters" and "Pass condition", and various V-PRO and STERRAD cycles with worst-case loads). The observable color change (or lack thereof) in the chemical indicator is then compared against these known exposure conditions to determine if the indicator provides an accurate visual signal according to the standard.

    8. Sample Size for Training Set:

    Not applicable. Chemical indicators are passive devices that rely on a chemical reaction for their function, not on machine learning algorithms that require a "training set" of data.

    9. How Ground Truth for Training Set was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183300
    Date Cleared
    2019-01-03

    (37 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    VERIFY V24 Self-Contained Biological Indicator, VERIFY V24 Biological Indicator Challenge Pack

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VERIFY® V24 Self-Contained Biological Indicator is intended for routine monitoring of the following sterilizer cycles:

    · Lumen, Non Lumen, Flexible, Fast and Fast Non Lumen Cycles of the V-PRO® 1, 1 Plus, maX, maX 2, 60 and s 2 Sterilizers

    · STERRAD® 100S Sterilizer (default cycle)

    • STERRAD® 200 Sterilizer (default cycle)

    · Standard and Advanced Cycles of the STERRAD® NX Sterilizers with All Clear Technology

    · Express, Flex Scope and Standard Cycles of the STERRAD® 100NX Sterilizers with All Clear technology

    The VERIFY V24 BI Challenge Pack is intended for qualification testing of the V PRO 1 Plus, VPRO maX, V-PRO maX 2, V-PRO 60 and V PRO s 2 Low Temperature Sterilization Systems following installation, relocation, malfunctions or major repairs.

    The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.

    The challenge pack is not intended for routine of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.

    Device Description

    The VERIFY V24 Self-Contained Biological Indicator (SCBI) is used by healthcare facilites to monitor the V-PRO® Low Temperature Sterilization Systems. It is designed to accompany medical devices placed in the sterilizer. The user places the VERIFY V24 SCBI into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation. The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

    The VERIFY V24 Biological Indicator Challenge Pack is used by healthcare facilities for qualification testing of the V-PRO Low Temperature Sterilization Systems following installation, relocations, malfunctions or major repairs. The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital defined challenge load is not included. The user places the VERIFY V24 Biological Indicator Challenge Pack into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the VERIFY HPU Chemical Indicator (CI) and the VERIFY V24 Self-Contained Biological Indicator (SCBI) contained in the challenge pack are retrieved. The CI is assessed for a passing color change immediately and the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation. The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the VERIFY V24 Self-Contained Biological Indicator and VERIFY V24 Biological Indicator Challenge Pack, based on the provided document.

    1. Table of Acceptance Criteria and Reported Device Performance

    It's important to note that the document presents separate acceptance criteria and testing for the "Self-Contained Biological Indicator (SCBI)" and the "Challenge Pack."

    For the VERIFY V24 Self-Contained Biological Indicator (SCBI):

    TestAcceptance CriteriaReported Device Performance
    ½ & ¾ Cycle PerformanceAll test SCBIs exposed in either the ½ or ¾ cycle display passing resultsPASS
    Simulated UseAll processed SCBIs exhibit negative growth results
    All processed SCBI Label PIs exhibit a “pass” result
    All processed CIs exhibit a “pass” resultPASS
    Verification of Viable Spore Population after exposure to ALLClearThe mean population of V24 SCBI exposed to the ALLClear pre-conditioning was -50% to +300% of the unexposed V24 SCBI populationPASS
    Reduced Incubation Time (RIT) TestingMeets FDA's requirement of > 97% alignment of the 20-minute results with the conventional incubation time of 7 daysPASS
    Viable spore population2.0 - 3.4 x 10^6 spore/SCBIPASS (stated to contain "greater than 10^6 spores/BI" and "Same" as predicate in Table 5-1)
    D-value4-8 secPASS (stated "Same" as predicate in Table 5-1)
    Survival Time4-30 secPASS (stated "Same" as predicate in Table 5-1)
    Kill Time≤ 16 minPASS (stated "Same" as predicate in Table 5-1)
    Hold TimePerformance not affected if incubated within 72 hours of exposure to VHP sterilizationPASS

    For the VERIFY V24 Biological Indicator Challenge Pack:

    TestAcceptance CriteriaReported Device Performance
    Comparative Dose Response to Biological ModelThe challenge pack shall demonstrate equal or greater resistance as compared to the worst case biological modelPASS
    Simulated UseAll processed SCBIs exhibit negative growth results
    All processed SCBI Label PIs exhibit a “pass” result
    All processed CIs exhibit a “pass” resultPASS

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size used for the test set in any of the studies. It refers to "All test SCBIs" and "All processed SCBIs" without providing a count.

    The data provenance is not specified beyond being "nonclinical tests" and "performance testing to demonstrate substantial equivalence to the predicate." There is no mention of country of origin or whether the studies were retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The document describes tests related to the biological and chemical indicators themselves and their performance in sterilization cycles, not a diagnostic or interpretive task that would typically involve human expert adjudication for ground truth.

    4. Adjudication Method for the Test Set

    This information is not provided as the tests described are objective measures of biological indicator performance (e.g., growth/no growth, color change, spore population) rather than subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This is not applicable. The device described (Biological Indicator and Challenge Pack) is a sterilization process indicator, not an AI or imaging device that would typically be involved in human-in-the-loop diagnostic tasks or benefit from a reader study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable as the device is a physical biological indicator, not an algorithm. Its performance is evaluated through laboratory and simulated use testing.

    7. The Type of Ground Truth Used

    The ground truth for the performance tests appears to be based on:

    • Biological viability: Growth or lack thereof of Geobacillus stearothermophilus spores after exposure to sterilization cycles, as determined by a color change in the growth medium.
    • Established physical and chemical properties: D-value, survival time, kill time, and viable spore population are laboratory-measured physical characteristics against established standards.
    • Chemical indicator color change: A visible, objective change in color on the indicator.
    • Comparative resistance: The challenge pack's resistance compared to a "worst case biological model."

    Essentially, the "ground truth" is defined by the expected biological or chemical response to a sterilization process.

    8. The Sample Size for the Training Set

    This is not applicable. There is no concept of a "training set" for a biological indicator. The tests described are performance and validation tests, not machine learning model training.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172748
    Date Cleared
    2018-02-09

    (150 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    VERIFY V24 Self-Contained Biological Indicator, VERIFY V24 Biological Indicator Challenge Pack

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VERIFY® V24 Self-Contained Biological Indicator is intended for routine monitoring of the following sterilizers and cycles:

    • · Non Lumen, Flexible, Lumen and Fast Non Lumen Cycles of the V-PRO 1, 1 Plus, maX, 60 and maX 2 Sterilizers
    • · Standard Cycle of the STERRAD® 100S Sterilizer
    • · Standard and Advanced Cycles of the STERRAD® NX Sterilizer
    • · Express, Flex Scope and Standard Cycles of the STERRAD® 100 NX Sterilizer

    The VERIFY V24 Challenge Pack is intended for qualification testing of the V-PRO 1 Plus, V-PRO maX, V-PRO maX 2 and V-PRO 60 Low Temperature Sterilization Systems following installation, relocation, malfunctions or major repairs.

    The VERIFY V24 Challenge Pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.

    The VERIFY V24 Challenge Pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing the Sterilizers.

    Device Description

    The VERIFY V24 Self-Contained Biological Indicator (SCBI) is used by healthcare facilites to monitor the V-PRO® Low Temperature Sterilization Systems. It is designed to accompany medical devices placed in the sterilizer.

    The user places the VERIFY V24 Self-Contained Biological Indicator into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the VERIFY V24 Self-Contained Biological Indicator can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.

    The VERIFY V24 Self-Contained Biological Indicator is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activator automatically seals the VERIFY V24 Self-Contained Biological Indicator vial and releases the growth media.

    The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The VERIFY V24 Self-Contained Biological Indicator indicates a pass if the media remains orange and non-turbid. The VERIFY V24 Self-Contained Biological Indicator indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

    The VERIFY V24 Challenge Pack is used by healthcare facilities for qualification testing of the V-PRO Low Temperature Sterilization Systems following installation, relocations, malfunctions or major repairs. The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital defined challenge load is not included.

    The user places the VERIFY V24 Challenge Pack into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the VERIFY HPU Chemical Indicator (CI) and the VERIFY V24 Self-Contained Biological Indicator (SCBI) contained in the challenge pack are retrieved. The CI is assessed for a passing color change immediately and the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.

    The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activator automatically seals the SCBI vial and releases the growth media.

    The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

    AI/ML Overview

    This document describes the acceptance criteria and study proving the VERIFY® V24 Self-Contained Biological Indicator (SCBI) and the VERIFY® V24 Challenge Pack meet these criteria.

    VERIFY® V24 Self-Contained Biological Indicator

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance/Conclusion
    1/4, 1/2 & 3/4 Cycle PerformancePartial positive results obtained in 1/4 and 1/2 cycles; all negative results obtained in 3/4 cycle evaluations.PASS
    Growth InhibitionUninoculated VERIFY V24 SCBIs processed in a Fast Non Lumen Cycle and inoculated with low numbers (
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    VERIFY V24 SELF-CONTAINED BIOLOGICAL INDICATOR FOR VAPORIZED VH202 STERILIZATION PROCESSES

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VERIFY® V24 Self-Contained Biological Indicator is intended for routine monitoring of the Lumen, Non Lumen and Flexible cycles of V-PRO® Low Temperature Sterilization Systems.

    The VERIFY V24 Biological Indicator Challenge Pack is intended for qualification testing of the Lumen. Non Lumen and Flexible Cycles in V-PRO Low Temperature Sterilization Systems following installation, relocation, malfunctions or major repairs.

    The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.

    The challenge pack is not intended for routine monitoring of V-PRO Sterliizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.

    Device Description

    The VERIFY V24 Self-Contained Biological Indicator (SCBI) is used by healthcare providers to monitor the V-PRO® Low Temperature Sterilization Systems. It is designed to accompany medical devices placed in the sterilizer.

    The user places the VERIFY V24 Self-Contained Biological Indicator into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.

    The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activator automatically seals the SCBI vial and releases the growth media.

    The activated SCBI is incubated at 55 - 60 °C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

    The VERIFY V24 Biological Indicator Challenge Pack is used by healthcare providers for qualification testing of the V-PRO Low Temperature Sterilization Systems following installation, relocation, malfunctions or major repairs. The challenge pack is placed in an otherwise empty sterilizer chamber; a hospitaldefined challenge load is not included.

    The user places the VERIFY V24 Biological Indicator Challenge Pack into the V-PRO Sterilizer and performs a sterilization cycle. After cvcle completion, the VERIFY HPU Chemical Indicator (CI) and the VERIFY V24 Self-Contained Biological Indicator (SCBI) contained in the challenge pack are retrieved. The CI is assessed for a passing color change immediately and the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.

    The SCBI is activated by sealing the vial and rupturing the media ampoule using the STERIS VERIFY SCBI HP activator. The activator automatically seals the SCBI vial and releases the growth media.

    The activated SCBI is incubated at 55-60 ℃ for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure if the media changes from orange to vellow and/or if the media is turbid.

    AI/ML Overview

    This document describes two devices, the VERIFY® V24 Self-Contained Biological Indicator (SCBI) and the VERIFY® V24 Biological Indicator Challenge Pack. Both submissions are focused on qualifying these devices for use with the new V-PRO 60 Low Temperature Sterilization System, demonstrating substantial equivalence to their respective predicate devices.

    VERIFY® V24 Self-Contained Biological Indicator (SCBI)

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance Criteria (Implied)Reported Device Performance
    SCBI Half Cycle Performance Evaluation in the V-PRO 60 SterilizerSCBIs must be sterile after half-cycle testing.Pass (SCBIs were sterile)
    SCBI Growth Inhibition Following Exposure to V-PRO 60 CycleNo growth inhibition or effect on media detected.Pass (No growth inhibition)
    SCBI Simulated Use in the V-PRO 60 SterilizerSuccessful demonstration of simulated use performance.Pass
    SCBI Worst Case Location in the V-PRO 60 SterilizerThe worst-case location for SCBI placement must be determined.Pass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document summarizes nonclinical tests without providing specific sample sizes for each test. The studies were conducted by STERIS Corporation. The data provenance is described as being for "the new V-PRO 60 unit," implying prospective testing for this specific sterilizer. The K numbers (K073244, K090514) indicate prior clearances and associated data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not provided within the document. The "ground truth" for biological indicators is inherently based on the viability of Geobacillus stearothermophilus spores, which is determined by culturing and observing growth or no-growth.

    4. Adjudication Method for the Test Set:

    This information is not provided within the document. The outcome (Pass/Fail, or sterile/non-sterile) for biological indicators is typically a direct observation of microbial growth or lack thereof.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    This type of study is not applicable as the device is a biological indicator, not an imaging or diagnostic AI tool that involves human readers.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance):

    This is a standalone performance study. The device, a biological indicator, functions independently without a human in the loop for its direct performance, though human observation is required for interpreting its results (color change, turbidity).

    7. Type of Ground Truth Used:

    The ground truth used is the viability of Geobacillus stearothermophilus ATCC 7953 spores. A successful sterilization cycle should kill all spores, resulting in no growth (sterile). If spores survive, they will grow, indicating a sterilization failure. This is effectively a biological assay for sterility.

    8. Sample Size for the Training Set:

    The document does not specify a separate "training set" in the context of machine learning, as this is a biological indicator. All listed tests are performance evaluations for the device itself rather than training data for an algorithm.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set in the machine learning sense. The biological principles of spore inactivation and growth underpin the ground truth for biological indicators.


    VERIFY® V24 Biological Indicator Challenge Pack

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance Criteria (Implied)Reported Device Performance
    Resistance CharacterizationChallenge Pack resistance must be equivalent or greater than the biological models used to validate the V-PRO 60 Low Temperature Sterilization System cycles.Pass
    Simulated Use EvaluationVERIFY HPU Chemical Indicator and VERIFY V24 SCBI must yield passing results under worst-case simulated use conditions.Pass
    Worst Case LocationThe worst-case location within the V-PRO 60 Sterilizer Chamber must be identified.Pass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document summarizes nonclinical tests without providing specific sample sizes for each test. The studies were conducted by STERIS Corporation. The data provenance is described as being for "the new V-PRO 60 Sterilizer," implying prospective testing for this specific sterilizer. K103331 is cited as the predicate.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not provided within the document. Similar to the SCBI, the ground truth for the challenge pack relies on the biological viability of Geobacillus stearothermophilus spores and the visual change of the chemical indicator.

    4. Adjudication Method for the Test Set:

    This information is not provided within the document. The outcome (Pass/Fail) is based on the biological indicator's growth/no growth and the chemical indicator's color change.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    This type of study is not applicable as the device is a biological indicator challenge pack.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance):

    This is a standalone performance study. The challenge pack, containing a biological indicator and a chemical indicator, functions independently in the sterilizer without a human in the loop for its direct performance. Human observation is required for interpreting the results of both indicators.

    7. Type of Ground Truth Used:

    The ground truth used is the viability of Geobacillus stearothermophilus ATCC 7953 spores for the biological indicator component and the color change of the chemical indicator. Both are indicators of effective sterilization.

    8. Sample Size for the Training Set:

    The document does not specify a separate "training set" in the context of machine learning, as this is a biological indicator challenge pack. All listed tests are performance evaluations for the device itself.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set in the machine learning sense. The biological principles of spore inactivation and growth, along with established chemical indicator responses to sterilization processes, underpin the ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140708
    Manufacturer
    Date Cleared
    2014-06-16

    (87 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    VERIFY V24 SELF-CONTAINED BIOLOGICAL INDICATOR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VERIFY® V24 Self-Contained Biological Indicator is intended for routine monitoring of the following sterilizer cycles:

    • Lumen, Non Lumen and Flexible Cycles of the V-PRO® Low Temperature . Sterilization Systems
    • Default Cycle of the STERRAD 100S Sterilizer ●
    • Default Cycle of the STERRAD 200 Sterilizer .
    • Standard and Advanced Cycles of the STERRAD NX Sterilizer .
    • Express, Standard and Flex Scope Cycles of the STERRAD 100NX Sterilizer .
    Device Description

    The VERIFY V24 Self-Contained Biological Indicator (SCBI) is used by healthcare providers to monitor the V-PRO® Low Temperature Sterilization Systems and the STERRAD® 100S, STERRAD 200, STERRAD NX and STERRAD 100NX (Express, Standard and Flex Scope Cycles) Sterilizers. It is designed to accompany medical devices placed in the sterilizer.
    The user places the VERIFY V24 Self-Contained Biological Indicator into the V-PRO Low Temperature Sterilization System or STERRAD Sterilizer and performs a sterilization cycle. After cycle completion, the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.
    The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activator automatically seals the SCBI vial and releases the growth media.
    The activated SCBI is incubated at 55 – 60 °C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

    AI/ML Overview

    The VERIFY® V24 Self-Contained Biological Indicator (SCBI) is intended to monitor specific cycles of V-PRO® Low Temperature Sterilization Systems and various STERRAD® Sterilizers. The acceptance criteria and supporting study are outlined below:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test & Acceptance CriteriaReported Device Performance
    SCBI Half Cycle Performance
    SCBIs must be sterile after half-cycle testing.Pass
    SCBIs were sterile in half-cycle testing.
    SCBI Growth Inhibition Following Exposure to STERRAD Sterilizer Cycles
    No growth inhibition and no effect of the sterilization process on the media.Pass
    There was no growth inhibition and no effect of the sterilization process on the media.
    SCBI Simulated Use in the STERRAD Sterilizer Cycles
    Successful performance in simulated use.Pass
    Simulated use performance has been successfully demonstrated.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample size (number of VERIFY V24 SCBIs) used for each of the performance tests (Half Cycle, Growth Inhibition, Simulated Use).

    The data provenance is for testing the device's compatibility with STERRAD Sterilizers. The context implies that the testing was likely conducted in a controlled laboratory or clinical setting in the United States, given that the sponsor and manufacturing facilities are in Mentor, OH, USA, and the submission is to the FDA. The study appears to be prospective, specifically designed to demonstrate the device's performance with the new sterilizer indications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. For biological indicators, the ground truth is typically established by observing the growth or non-growth of the bacterial spores, which is a direct, objective biological outcome. It doesn't usually rely on expert interpretation in the same way an imaging diagnostic might.

    4. Adjudication method for the test set

    The document does not describe an adjudication method for the test set. Given the nature of a biological indicator, where the outcome is a clear biological response (growth or no growth, indicated by a color change), an adjudication method by multiple experts is generally not applicable. The results are typically objectively measured.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation and often AI assistance. The VERIFY V24 Self-Contained Biological Indicator is a direct biological indicator without a human-in-the-loop component for reading the primary outcome.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device itself is a standalone indicator. Its performance is evaluated directly based on the biological outcome (spore inactivation and subsequent growth or no growth of Geobacillus stearothermophilus). This is essentially a standalone performance evaluation in that there is no "human-in-the-loop" for the primary interpretation of the device's result (color change or turbidity).

    7. The type of ground truth used

    The ground truth used is the biological outcome – specifically, the successful inactivation of Geobacillus stearothermophilus spores and the presence or absence of their growth after exposure to a sterilization process. This is determined by observing the color change of the media (orange to yellow) and/or turbidity, which indicates viable spores. The specification ranges for D-value, survival time, and kill time are also integral to establishing the efficacy (ground truth) of the indicator.

    8. The sample size for the training set

    The document does not detail a separate "training set" or its size, as this device is a biological indicator and not an AI/ML-based system that typically requires a training set. The existing biological indicator has been previously cleared, and the current submission focuses on extending its use to additional sterilizers.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of a "training set" in the context of an AI/ML model for this biological indicator. The existing performance characteristics of the Geobacillus stearothermophilus spores and the indicator's components (media, etc.) were established through previous studies (K073244, K090514) according to FDA guidance for biological indicators.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1