Search Results
Found 3 results
510(k) Data Aggregation
(29 days)
VERIFY SPORE TEST STRIP for S40 STERILANT CONCENTRATE
The VERIFY Spore Test Strip for S40 Sterilant Concentrate is intended to provide users with a means to assess spore kill by S40 Sterilant use dilution in the STERIS automated Liquid Chemical Sterilant Processing Systems (SYSTEM 1E and SYSTEM 1® endo Liquid Chemical Sterilant Processing Systems and enspire 3000 Series Cleaning and Liquid Chemical Sterilant Processing System). A "no growth" result from the VERIFY Spore Test Strip for S40 Sterilant Concentrate after 24 hours of incubation indicates that the liquid chemical sterilization process achieved the conditions necessary to kill at least 1 x 105 viable spores (5 logs) on the spore test strip (STS). It does not confirm the expected full performance of the liquid chemical sterilization cycle.
The Spore Test Strip for S40 Sterilant Concentrate (STS) consists of a 1 % in. x 14 in. filter paper-based strip inoculated with Geobacillus stearothermophilus spores and is enclosed in a glassine envelope. The STS are provided with media vials containing a modified tryptic soy broth with phenol red pH indicator, and a transfer clip. The transfer clip is used to remove the STS from the glassine envelope and serves to hold the strip in a fixed location in either the enspire 3000 Series Cleaning and Liquid Chemical Sterilant Processing System (enspire CLCSPS) or the SYSTEM 1E or SYSTEM 1 endo Liquid Chemical Sterilant Processing system. After the cycle concludes, the transfer clip is used to aseptically transfer the STS from the processor into the growth media for incubation at 55-60°C for a minimum of 24 hours but may be incubated for up to 7 days. If the media remains red and non-turbid, the user interprets the results as a pass. If the media color turns yellow or is turbid, the user interprets the result as a fail. The shelf life is 12 months when stored at 2-24°C and 30-80% relative humidity (RH) away from sterilizing agents and excessive heat.
There are no changes to the strip itself for this submission; only an expansion to the exposure time to the chemical sterilant.
The provided document, an FDA 510(k) Premarket Notification summary for the "VERIFY SPORE TEST STRIP for S40 STERILANT CONCENTRATE," details the acceptance criteria and the study proving the device meets these criteria.
Here's the breakdown:
1. Table of Acceptance Criteria and the Reported Device Performance:
The primary acceptance criteria for this device revolve around demonstrating its continued performance as a spore test strip, specifically that the expansion of contact time with the liquid chemical sterilant does not negatively impact its ability to accurately assess spore kill.
Test | Acceptance Criteria | Result |
---|---|---|
Population Wash Off Testing | The modification does not affect the performance of the device. | Pass |
Bacteriostasis Testing | The modification does not affect the performance of the device. | Pass |
Implied Performance Criteria (from Intended Use & Predicate Comparison): | ||
Ability to indicate ≥ 5 logs kill of Geobacillus stearothermophilus spores | A "no growth" result from the VERIFY Spore Test Strip for S40 Sterilant Concentrate after 24 hours of incubation indicates that the liquid chemical sterilization process achieved the conditions necessary to kill at least 1 x 105 viable spores (5 logs) on the spore test strip (STS). | Demonstrated as substantially equivalent to predicate, which has this capability. |
Shelf Life | 12 months | Identical to predicate. |
Resistance Characteristics (at 1635 ppm PAA) | D-value 12 – 26 seconds; Survival Time ≥ 38 seconds; Kill Time ≥ 239 seconds | Identical to predicate. |
2. Sample Size Used for the Test Set and the Data Provenance:
The document does not explicitly state the numerical sample size for the "Population Wash Off Testing" and "Bacteriostasis Testing." It only indicates that these tests were performed and passed.
The data provenance is implied to be prospective testing conducted by the manufacturer, STERIS Corporation, in support of this 510(k) submission. The country of origin of the data is not specified, but given that STERIS Corporation is based in Mentor, Ohio, USA, it is highly probable the testing was conducted in the USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
This information is not provided in the document. The nature of these tests (Population Wash Off and Bacteriostasis Testing for a biological indicator) suggests laboratory-based assessments rather than expert interpretation of complex medical images, therefore, detailed information on expert consensus is not typically required or provided for this type of device.
4. Adjudication Method for the Test Set:
This information is not provided in the document. For the type of testing described (objective laboratory assays), an adjudication method like 2+1 or 3+1 for human readers is not applicable. The results are likely determined by predefined laboratory protocols and measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic tools where human readers interpret medical images. The device in question is a biological indicator for sterilization monitoring, not a diagnostic imaging device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
Yes, in essence. The "Population Wash Off Testing" and "Bacteriostasis Testing" are standalone tests of the device's physical and biological integrity and performance characteristics, independent of human interpretation beyond reading the final culture results (growth/no growth). The device itself functions as a standalone indicator of sterilization efficacy.
7. The Type of Ground Truth Used:
The ground truth for the performance of this device is established by:
- Laboratory Testing: Directly measuring the
Population Wash Off
(the number of viable spores remaining on the strip after specified processing) andBacteriostasis
(the absence of inhibitory substances affecting subsequent bacterial growth). - Biological Efficacy: The ultimate ground truth for a spore test strip is its ability to accurately reflect the killing of the specified 5 logs of Geobacillus stearothermophilus spores, as confirmed by standard microbiological culture methods (absence of growth after incubation).
8. The Sample Size for the Training Set:
This information is not applicable and therefore not provided. This device is a biological indicator, not a machine learning or AI model that requires training data.
9. How the Ground Truth for the Training Set Was Established:
This information is not applicable. As stated above, this device does not involve a training set as it is not an AI/ML model.
Ask a specific question about this device
(88 days)
VERIFY Spore Test Strip for S40 Sterilant Concentrate
The VERIFY Spore Test Strip for S40 Sterilant Concentrate is intended to provide users with a means to assess spore kill by S40 Sterilant use dilution in the STERIS automated Liquid Chemical Sterilant Processing Systems (SYSTEM 1E and SYSTEM 1® endo Liquid Chemical Sterilant Processing Systems and enspire 3000 Series Cleaning and Liquid Chemical Sterilant Processing System). A "no growth" result from the VERIFY Spore Test Strip for S40 Sterilant Concentrate after 24 hours of incubation indicates that the liquid chemical sterilization process achieved the conditions necessary to kill at least 1 x 105 viable spores (5 logs) on the spore test strip (STS). It does not confirm the expected full performance of the liquid chemical sterilization cycle.
The Spore Test Strip for S40 Sterilant Concentrate (STS) consists of a 1 % in. x 14 in. filter paper-based strip inoculated with Geobacillus stearothermophilus spores and is enclosed in a glassine envelope. The STS are provided with media vials containing a modified tryptic soy broth with phenol red pH indicator, and a transfer clip. The transfer clip is used to remove the STS from the glassine envelope and serves to hold the strip in a fixed location in either the enspire 3000 Series Cleaning and Liquid Chemical Sterilant Processing System (enspire CLCSPS) or the SYSTEM 1E or SYSTEM 1 endo Liquid Chemical Sterilant Processing system. After the cycle concludes, the transfer clip is used to aseptically transfer the STS from the processor into the growth media for incubation at 55-60℃ for a minimum of 24 hours but may be incubated for up to 7 days. If the media remains red and non-turbid, the user interprets the results as a pass. If the media color turns yellow or is turbid, the user interprets the result as a fail. The shelf life is 12 months when stored at 2-24°C and 30-80% relative humidity (RH) away from sterilizing agents and excessive heat.
The provided text describes a 510(k) premarket notification for a medical device called "VERIFY Spore Test Strip for S40 Sterilant Concentrate." This submission is for a device that already has a predicate (K180553) and the main change is an expansion of its indications for use to include a new processing system, the "enspire 3000 Series Cleaning and Liquid Chemical Sterilant Processing System (CLCSPS)."
The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the performance testing conducted to demonstrate that the expanded use of the spore test strip in the new system is safe and effective, performing as well as the previous approved uses.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the successful outcomes ("PASS") of the performance tests, demonstrating that the device functions appropriately under the new use conditions. The "performance" is the conclusion that the device passed these tests.
Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|
Population Wash Off Stability: Evaluate the population of the spores remaining on the STS after processing without sterilant. | PASS (Demonstrated that the population of spores remaining on the STS after processing without sterilant was acceptable.) |
Simulated Use: Demonstrate the STS functions appropriately in the enspire CLCSPS. | PASS (Demonstrated that the STS functions appropriately in the enspire CLCSPS.) |
Bacteriostasis: Ensure no inhibitory effects on the outgrowth of a low number of spores. | PASS (Demonstrated no inhibitory effects on the outgrowth of a low number of spores.) |
2. Sample size used for the test set and the data provenance
The document does not explicitly state the sample sizes (e.g., number of strips, number of runs) used for each of the performance tests (Population Wash Off Stability, Simulated Use, Bacteriostasis).
- Data Provenance: The study was conducted by STERIS Corporation, an American company (Mentor, Ohio). Given the context of a 510(k) submission to the FDA, the data would have been generated in a controlled environment to meet regulatory standards. The document does not specify whether the data was retrospective or prospective, but performance testing for a 510(k) generally involves prospective testing of the device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This type of device (spore test strip for sterilant efficacy) does not typically involve human expert interpretation of images or clinical outcomes in the same way an AI diagnostic device would. Ground truth is established through laboratory methods (e.g., microbial growth/no growth, quantitative spore counts). Therefore, the concept of "experts" for ground truth establishment in the human-in-the-loop sense, or their qualifications, is not applicable directly to this device. Competent laboratory personnel would conduct and interpret the tests according to established protocols.
4. Adjudication method for the test set
Adjudication methods (like 2+1, 3+1) are common in clinical studies involving human interpretation or subjective assessments. For a biological indicator like a spore test strip, the results (growth/no growth, quantitative counts) are objective, and adjudication is not typically applied in the same manner. The outcome is determined by the presence or absence of growth and quantitative measurements.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study is not applicable to this device. This is not an AI diagnostic device that assists human readers. It is a biological indicator for sterilization processes.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The device itself is a standalone biological indicator. It provides a result (growth/no growth) based on a biological process (spore killing by sterilant and subsequent incubation). There is no "algorithm" in the sense of artificial intelligence. The device's performance is its ability to accurately reflect the 5-log kill of spores under specified conditions. The performance tests ("Population Wash Off Stability," "Simulated Use," "Bacteriostasis") directly evaluate this standalone performance.
7. The type of ground truth used
The ground truth for this device is based on biological and chemical efficacy.
- Biological Efficacy: The primary ground truth is the presence or absence of Geobacillus stearothermophilus growth after exposure to the sterilant and incubation, indicating whether a 5-log reduction of spores has occurred. This is a direct measure of microbial inactivation.
- Chemical/Physical Parameters: The ground truth also relates to the specific conditions of the sterilization process (e.g., PAA concentration, temperature, time) as defined for the sterilant and the processing systems.
8. The sample size for the training set
This device does not involve machine learning or AI, and therefore, there is no "training set" in the typical sense for algorithms. The "training" of the product's design would be based on historical data, prior knowledge of sterilant efficacy, and established microbiological methods for biological indicators, rather than a data set used to train a predictive model.
9. How the ground truth for the training set was established
As there is no training set for an AI algorithm, this question is not applicable. The underlying scientific principles for spore test strips and sterilization efficacy are well-established through decades of microbiological and chemical research. The "ground truth" for the development of such a device is rooted in these scientific principles and standard test methods (e.g., those from AAMI, ISO).
Ask a specific question about this device
(89 days)
VERIFY Spore Test Strip for S40 Sterilant
The VERIFY® Spore Test Strip for S40 Sterilant is intended to provide users with a means to assess spore kill by S40 sterilant use dilution in the STERIS automated Liquid Chemical Sterilant Processing Systems (SYSTEM 1E Liquid Chemical Sterilant Processing System and SYSTEM 1 endo Liquid Chemical Sterilant Processing System). A "no growth" result from the VERIFY Spore Test Strip for S40 Sterilant after 24 hours of incubation indicates that the liquid chemical sterilization process achieved the conditions necessary to kill at least 1 x 10E5 viable spores (5 logs) on the test strip. It does not confirm the expected full performance of the liquid chemical sterilization cycle.
The VERIFY® Spore Test Strip for S40 Sterilant consists of a 1 36 in. x 14 in. filter paper-based strip inoculated with Geobacillus stearothermophilus spores and is enclosed in a glassine envelope. The Spore Test Strips are provided with media vials containing a modified tryptic soy broth with phenol red pH indicator, and a transfer clip. Using the clip, the Spore Test Strip is secured on the available post located in the tray of the liquid chemical sterilant processor. The transfer clip thus holds the Spore Test Strip in a fixed location during a processing cycle in the STERIS Liquid Chemical Sterilant Processing System and enables aseptic transfer of the strip from the processor into the growth media vial after the cycle ends.
For use, the Spore Test Strip is removed from the glassine envelope and secured using the transfer clip in the SYSTEM 1E Liquid Chemical Sterilant Processing System or SYSTEM 1 endo Liquid Chemical Sterilant Processing System along with the items to be liquid chemically sterilized. The liquid chemical sterilization cycle is initiated. At the end of the full cycle, the Spore Test Strip is removed and placed into the vial of growth medium for incubation under the specified conditions that will promote spore growth for at least 24 hours. If the media remains red and non-turbid, the user interprets the results as a pass. If the media color turns yellow or is turbid, the user interprets the result as a fail.
The Spore Test Strip is incubated at 55-60°C for a minimum of 24 hours, but may be incubated for up to 7 days. The shelf life is 12 months when stored at 2-24℃ and 30-80% relative humidity (RH) away from sterilizing agents and excessive heat.
The current submission makes no changes to the strip itself; only the indications for use are changing to encompass the VERIFY Spore Test Strip's broader use in more than one STERIS automated liquid chemical sterilant processing system (SYSTEM 1E Liquid Chemical Sterilant Processing System and SYSTEM 1 endo Liquid Chemical Sterilant Processing System) that uses S40 Sterilant Concentrate under identical exposure conditions.
The provided text describes the 510(k) premarket notification for the VERIFY® Spore Test Strip for S40® Sterilant. This device is a biological indicator designed to assess the effectiveness of liquid chemical sterilization processes. It is important to note that this is not an AI/ML device, and therefore, several requested categories are not applicable.
Here's the breakdown based on the provided document:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria | Reported Device Performance (Table 1 & 2) |
---|---|
Population | |
• At manufacture | $\geq$ 1.5 x 105 cfu/strip |
• After wash-off (end of cycle) | $\geq$ 1.0 x 105 cfu/strip |
Resistance (at 1635 ppm PAA) | |
• D-value | 12-26 sec |
• Survival Time | $\geq$ 38 seconds |
• Kill Time | $\leq$ 239 seconds |
Reduced Incubation Time | 24 hours |
Simulated Use | PASS |
Stability of Media Color | PASS |
Shelf Life | PASS (for 12 months) |
2. Sample size used for the test set and the data provenance
The document indicates that non-clinical testing was performed for "Simulated Use", "Stability of Media Color", "Reduced Incubation Time", and "Shelf Life for 12 months." However, the specific sample sizes used for these tests are not provided in the document.
Regarding data provenance:
- Country of Origin: Not explicitly stated, but the submission is to the U.S. FDA by STERIS Corporation located in Mentor, Ohio, USA. It can be inferred that the testing likely occurred in the USA.
- Retrospective or Prospective: The testing described (simulated use, stability, reduced incubation time, shelf life) appears to be prospective laboratory-based testing designed to validate the device's performance characteristics.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This device is a biological indicator, not an AI/ML diagnostic device that requires expert interpretation of images or other complex data. The "ground truth" for its performance is established through microbiological methods and chemical analysis of the sterilant's effect on spores, rather than expert consensus on complex medical data.
Therefore, this question is not applicable in the context of this device.
4. Adjudication method for the test set
As the "ground truth" for this device is based on objective microbiological and chemical measurements (e.g., spore kill, population counts, D-values, color change indicating growth), an adjudication method involving multiple human experts is not applicable or described. The results (PASS/FAIL, numerical values) are determined by laboratory assays.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question pertains to AI/ML devices involving human readers. The VERIFY® Spore Test Strip is a biological indicator, not an AI/ML device.
Therefore, this question is not applicable.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This question pertains to AI/ML devices. The VERIFY® Spore Test Strip is a biological indicator, not an AI/ML device. While its performance is "standalone" in the sense of being an automated indicator, it does not involve an algorithm generating a diagnostic output without human input in the way a medical AI would. Users visually interpret the media color.
Therefore, this question is not applicable in the context of an AI algorithm.
7. The type of ground truth used
The ground truth used for this device is based on microbiological viability testing and chemical resistance characteristics.
- Spore Kill: Demonstrated by the absence of growth (no color change) in the culture medium, indicating that the targeted 5-log reduction of Geobacillus stearothermophilus spores was achieved.
- Population Counts: Directly measured in colony-forming units (CFU/strip).
- Resistance Characteristics (D-value, Survival Time, Kill Time): These are experimentally determined values that quantify how resistant the spores on the strip are to the sterilant under specific conditions.
- Stability of Media Color: Objective observation of the indicator color over time.
8. The sample size for the training set
This device is not an AI/ML device that requires a training set. Its performance characteristics are inherent to its biological and chemical design, validated through laboratory experiments.
Therefore, this question is not applicable.
9. How the ground truth for the training set was established
This device is not an AI/ML device that requires a training set.
Therefore, this question is not applicable.
Ask a specific question about this device
Page 1 of 1