510(k) Data Aggregation

    K Number
    K133849
    Manufacturer
    Date Cleared
    2014-10-22

    (307 days)

    Product Code
    Regulation Number
    862.2570
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    VANTERA CLINICAL ANALYZER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vantera® Clinical Analyzer is an automated laboratory test analyzer which measures the 400 MHz proton nuclear magnetic resonance (NMR) spectrum of clinical samples to produce signal amplitudes, converting these signal amplitudes to analyte concentration. The device includes a 400 MHz NMR spectrometer and software to analyze digitized spectral data. This instrumentation is intended to be used with NMR based assays to detect multiple analytes from clinical samples.

    The NMR LipoProfile® test, when used with the Vantera® Clinical Analyzer, an automated NMR spectrometer, measures lipoprotein particles to quantify LDL particle number (LDL-P), HDL cholesterol (HDL-C), and triglycerides in human serum and plasma using nuclear magnetic resonance (NMR) spectroscopy. LDL-P and these NMR-derived concentrations of HDL-C and triglycerides are used in conjunction with other lipid measurements and clinical evaluation to aid in the management of lipoprotein disorders associated with cardiovascular disease.

    Device Description

    The Vantera Clinical Analyzer is a clinical laboratory analyzer that employs nuclear magnetic resonance spectroscopic detection to quantify multiple analytes in biological fluid specimens, specifically blood plasma and serum.

    The Vantera Clinical Analyzer system design is divided into 3 major subassemblies: a sample handling assembly, an NMR subassembly, and an enclosure. The Vantera Clinical Analyzer control system is distributed across three separate computers:

    • The Host (1 U) controls user interface, data handling, results calculation, system startup and shutdown.
    • The Process Control (4U) schedules and manages all activities required to process a sample, controls all hardware in the sample handling subsystem.
    • The NMR Control Computer controls all magnet operations. Two of these computers are contained within the Sample Handling Subassembly (1 U and 4U) and one in the NMR Subassembly (NMR Console).

    The NMR LipoProfile test involves measurement of the 400 MHz proton NMR spectrum of a plasma/serum sample, deconvolution of the composite signal at approximately 0.8 ppm to produce signal amplitudes of the lipoprotein subclasses that contribute to the composite plasma/serum signal, and conversion of these subclass signal amplitudes to Lipoprotein subclass concentrations. The -0.8 ppm plasma NMR signal arises from the methyl group protons of the lipids carried in the LDL, HDL and VLDL subclasses of varying diameters. The NMR signals from the various lipoprotein subclasses have unique and distinctive frequencies and lineshapes, each of which is accounted for in the deconvolution analysis model. Each subclass signal amplitude is proportional to the number of subclass particles emitting the signal, which enables subclass particle concentrations to be calculated from the subclass signal amplitudes derived from the spectral deconvolution analysis. LDL subclass particle concentrations, in units of nanomoles of particles per liter (nmol/L), are summed to give the reported total LDL particle concentration (LDL-P). By employing conversion factors assuming that the various lipoprotein subclass particles have cholesterol and triglyceride contents characteristic of normolipidemic individuals, HDL cholesterol and triglyceride concentrations are also derived.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Vantera Clinical Analyzer and NMR LipoProfile® test, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (K113830) rather than explicitly stating pre-defined "acceptance criteria" for novel performance features. However, the comparisons in the tables serve as the de-facto acceptance criteria: the candidate device's performance must be comparable to or better than the predicate device across various analytical metrics.

    Metric (Analyte)Predicate Device (K113830) PerformanceCandidate Device Performance
    LDL-P (nmol/L)
    Measuring Range300 - 3500300-3500
    LoB00
    LoD40.750.1
    LoQ132154.7
    Linearity Regressiony= 1.02x + 7.82y= 0.99x + 106.6
    Linearity R²0.9950.997
    Linear Range225 - 4322290 - 3524
    Within-Run Precision (CV%)Level 1: 5.8, Level 2: 3.0, Level 3: 2.7Level 1: 7.7, Level 2: 5.5, Level 3: 2.6
    Within-Lab Precision (CV%)Level 1: 5.3, Level 2: 4.0, Level 3: 3.9Level 1: 7.0, Level 2: 6.8, Level 3: 2.7
    Method ComparisonLinear regression: y=1.03x - 36.60, r=0.978Deming fit: y= 43.44 + 0.98x r = 0.988
    Medical Decision Limits1000, 1300 and 1600same
    Sample TypeSerum and Plasmasame
    CarryoverNo significant trending of results and no persistent biassame
    Interference StudySalicylic acid at ≥ 1.3mmol/L, Clopidogrel hydrogensulfate at ≥ 95.7 µmol/LClopidogrel (Plavix) at 95.7 µmol/L, Salicylic acid at 1.3 mmol/L, Fenofibrate at 31 µmol/L, Menhaden oil at 0.6 mg/mL. (Note: Candidate device lists more interferences than predicate for LDL-P)
    TG (mg/dL)
    Measuring Range5- 110010 - 1100
    LoB1.11.2
    LoD2.32.3
    LoQ44.8
    Linearity Regression$y= 1.01x - 0.40$$y= 1.01x - 1.7$
    Linearity R²1.01.0
    Linear Range4 - 13464 - 1355
    Within-Run Precision (CV%)Level 1: 2.3, Level 2: 2.1, Level 3: 1.2Level 1: 2.7, Level 2: 0.9, Level 3: 0.6
    Within-Lab Precision (CV%)Level 1: 2.3, Level 2: 2.4, Level 3: 2.7Level 1: 3.3, Level 2: 1.5, Level 3: 2.5
    Method ComparisonLinear regression: $y=1.00x + 0.92, r=0.998$Deming fit: $Y= 1.01x +0.30 r=1.00$
    Sample TypeSerum and Plasmasame
    CarryoverNo significant trending of results and no persistent biassame
    Interference Study7 Endogenous and 23 Exogenous substances tested, no interference foundsame (no interference found with 7 Endogenous and 23 Exogenous substances)
    HDL-C (mg/dL)
    Measuring Range7 - 1407 - 140
    LoB2.73.2
    LoD3.54.4
    LoQ44.4
    Linearity Regression$y= 1.04x - 0.35$$y= 1.02x - 0.63$
    Linearity R²1.01.0
    Linear Range6 - 1485 - 168
    Within-Run Precision (CV%)Level 1: 4.0, Level 2: 2.8, Level 3: 2.6Level 1: 1.5, Level 2: 0.7, Level 3: 1.8
    Within-Lab Precision (CV%)Level 1: 2.8, Level 2: 2.0, Level 3: 1.8Level 1: 2.7, Level 2: 1.9, Level 3: 2.9
    Method ComparisonLinear regression: $y=1.04x-1.20, r=0.989$Deming fit: $y= -1.36 + 1.01x - r=0.998$
    Sample TypeSerum and Plasmasame
    CarryoverNo significant trending of results and no persistent biassame
    Interference Study7 Endogenous and 23 Exogenous substances tested, no interference foundsame (no interference found with 7 Endogenous and 23 Exogenous substances)

    2. Sample Size and Data Provenance for the Test Set

    The document does not explicitly state the specific sample sizes for each analytical validation study for the "test set" in terms of number of patient samples. It mentions "Level 1, Level 2, Level 3" for precision studies, implying control samples at different concentrations. For linearity, there are ranges of values. For the method comparison, it gives regression parameters, which typically require a reasonable number of samples, but the exact count isn't specified.

    The data provenance (country of origin, retrospective/prospective) is not provided in the document.

    3. Number of Experts and Qualifications for Ground Truth of Test Set

    This information is not applicable and not provided. The device (Vantera Clinical Analyzer with NMR LipoProfile® test) is an IVD for quantitative measurement of lipoprotein particles, HDL cholesterol, and triglycerides. The ground truth for these measurements in analytical validation studies would typically be established by:

    • Reference methods (e.g., ultracentrifugation for lipoproteins, enzymatic assays for cholesterol/triglycerides).
    • Certified reference materials.
    • Highly qualified laboratory personnel following established protocols for the reference methods.

    Experts in the sense of clinical reviewers (e.g., radiologists for imaging devices) are not typically involved in establishing ground truth for this type of analytical device.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a human interpretation-based diagnostic device requiring adjudication of expert opinions.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is an automated analytical laboratory device, not an imaging or interpretation aid for human readers. Therefore, an MRMC study and effects of AI assistance on human readers are not relevant.

    6. Standalone Performance Study

    Yes, the studies presented are standalone (algorithm only) performance studies. The results in the tables (Measuring Range, LoB, LoD, LoQ, Linearity, Precision, Method Comparison, Carryover, Interference) directly reflect the performance of the Vantera Clinical Analyzer and NMR LipoProfile® test itself, without any human-in-the-loop interaction for interpretation, beyond the standard operation of an automated laboratory instrument.

    7. Type of Ground Truth Used

    The ground truth for the analytical validation studies would be established using reference methods or reference materials. For example:

    • Method Comparison: Comparison against a recognized reference method for lipoprotein quantification, HDL-C, and triglycerides. While not explicitly stated, standard practice would involve a comparison to a well-characterized, clinically accepted method. The "Deming fit" and "Linear regression" indicate comparison to another quantitative measurement.
    • Linearity, LoB, LoD, LoQ: These are typically established using characterized control materials (known concentrations) or serial dilutions of patient samples.
    • Precision: Established using control materials at different concentrations.

    8. Sample Size for the Training Set

    The document does not specify a separate "training set" or its size. This device is an analytical instrument based on Nuclear Magnetic Resonance (NMR) spectroscopy and a deconvolution analysis model. While the deconvolution model itself would have been developed and "trained" or optimized using a dataset of known NMR spectra and corresponding reference measurements, the document focuses on the validation of the integrated device. The details of the dataset used for the initial development/training of the deconvolution algorithm are not provided here.

    9. How the Ground Truth for the Training Set Was Established

    As above, the document does not provide details on the training set for the deconvolution model. However, for a device based on a physical principle like NMR spectroscopy, the ground truth for establishing the deconvolution model and conversion factors would likely involve:

    • Carefully characterized samples with known concentrations of lipoproteins, HDL-C, and triglycerides, determined by highly accurate reference methods (e.g., ultracentrifugation for lipoprotein subfractions, established enzymatic methods for cholesterol and triglycerides).
    • Spectral data from these characterized samples would be used to build and optimize the deconvolution algorithm (i.e., establish unique frequencies and lineshapes) and the conversion factors from signal amplitudes to concentrations.
    Ask a Question

    Ask a specific question about this device

    K Number
    K113830
    Manufacturer
    Date Cleared
    2012-08-30

    (247 days)

    Product Code
    Regulation Number
    862.1705
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    VANTERA CLINICAL ANALYZER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vantera® Clinical Analyzer is an automated laboratory test analyzer which measures the 400 MHz proton nuclear magnetic resonance (NMR) spectrum of clinical samples to produce signal amplitudes, converting these signal amplitudes to analyte concentration. The device includes a 400 MHz NMR spectrometer and software to analyze digitized spectral data. This instrumentation is intended to be used with NMR based assays to detect multiple analytes from clinical samples.

    The NMR LipoProfile® test, when used with the Vantera® Clinical Analyzer, an automated NMR spectrometer, measures lipoprotein particles to quantify LDL particle number (LDL-P), HDL cholesterol (HDL-C), and triglycerides in human serum and plasma using nuclear magnetic resonance (NMR) spectroscopy. LDL-P and these NMR-derived concentrations of HDL-C and triglycerides are used in conjunction with other lipid measurements and clinical evaluation to aid in the management of lipoprotein disorders associated with cardiovascular disease.

    Device Description

    The Vantera Clinical Analyzer is a clinical laboratory analyzer that employs nuclear magnetic resonance spectroscopic detection to quantify multiple analytes in biological fluid specimens, specifically blood plasma and serum. The Vantera Clinical Analyzer system design is divided into 3 major subassemblies: a sample handling assembly, an NMR subassembly, and an enclosure. The Vantera Clinical Analyzer control system is distributed across three separate computers: The Host (1U) controls user interface, data handling, results calculation, system startup and shutdown. The Process Control (4U) schedules and manages all activities required to process a sample, controls all hardware in the sample handling subsystem, and manages remote access to the system. The NMR Control Computer controls all magnet operations. Two of these computers are contained within the Sample Handling Subassembly (1U and 4U) and one in the NMR Subassembly (NMR Console).

    The NMR LipoProfile test involves measurement of the 400 MHz proton NMR spectrum of a plasma/serum sample, deconvolution of the composite signal at approximately 0.8 ppm to produce signal amplitudes of the lipoprotein subclasses that contribute to the composite plasma/serum signal, and conversion of these subclass signal amplitudes to lipoprotein subclass concentrations.

    AI/ML Overview

    The provided 510(k) summary focuses on the analytical performance of the Vantera® Clinical Analyzer and the NMR LipoProfile® test compared to predicate devices, establishing substantial equivalence rather than providing explicit acceptance criteria as would be typical for a novel device. The study described primarily demonstrates that the proposed device performs comparably to its predicate devices in terms of analytical accuracy and precision.

    Here's an analysis of the acceptance criteria and study information provided:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a pass/fail format. Instead, it demonstrates performance by comparing the analytical results of the Vantera® Clinical Analyzer with the NMR LipoProfile® test to its predicate device (NMR Profiler for the assay) across various metrics. The unstated acceptance criteria for each analytical performance metric would be that the proposed device's performance must be comparable to or better than the predicate device's performance.

    MetricAcceptance Criteria (Implied)Reported Proposed Device Performance (Vantera® Clinical Analyzer)Predicate Device Performance (NMR Profiler)
    LDL-P
    LoBComparable to predicate0 nmol/L0 nmol/L
    LoDComparable to predicate40.7 nmol/L41 nmol/L
    LoQComparable to predicate132 nmol/L157 nmol/L
    Measuring RangeComparable to predicate300-3500 nmol/L300-3500 nmol/L
    Linearity Regression (Y=mX+b)R² comparable to predicatey=1.02x+7.82, R²=0.9949y=0.99x-22.37, R²=0.9979
    Within-Run Precision (CV%)Comparable to predicateLevel 1: 5.8%, Level 2: 3.0%, Level 3: 2.7%Level 1: 5.0%, Level 2: 4.3%, Level 3: 3.7%
    Within-Lab Precision (CV%)Comparable to predicateLevel 1: 5.3%, Level 2: 4.0%, Level 3: 3.9%Level 1: 7.6%, Level 2: 4.5%, Level 3: 4.3%
    Method Comparison (Correlation R)Comparable to predicateR=0.978R=0.973
    Interference StudyNo significant interference for tested substancesSalicylic acid (≥ 1.3mmol/L) and Clopidogrel hydrogensulfate (≥ 95.7 µmol/L) determined to interfere.No interference found for 5 endogenous & 22 exogenous substances.
    Specimen Stability (Refrigerated)Comparable to predicate6 days5 days
    Triglycerides
    LoBComparable to predicate1.1 mg/dL1.4 mg/dL
    LoDComparable to predicate2.4 mg/dL2.6 mg/dL
    LoQComparable to predicate4 mg/dL2.6 mg/dL
    Measuring RangeComparable to predicate5-1100 mg/dL5-1100 mg/dL
    Linearity Regression (Y=mX+b)R² comparable to predicatey=1.008x-0.3979, R²=0.9999y=0.95x-12.21, R²=0.999
    Within-Run Precision (CV%)Comparable to predicateLevel 1: 2.3%, Level 2: 2.1%, Level 3: 1.2%Level 1: 2.6%, Level 2: 1.8%, Level 3: 1.3%
    Within-Lab Precision (CV%)Comparable to predicateLevel 1: 2.3%, Level 2: 2.4%, Level 3: 2.7%Level 1: 3.6%, Level 2: 2.6%, Level 3: 2.5%
    Method Comparison (Correlation R)Comparable to predicateR=0.998R=1.00
    Interference StudyNo significant interference for tested substancesNo interference found for 7 endogenous & 23 exogenous substances.Ibuprofen may interfere with TG measurement at and above 210µg/mL for 5 endogenous & 22 exogenous substances.
    Specimen Stability (Refrigerated)Comparable to predicate6 days10 days
    HDL-C
    LoBComparable to predicate2.7 mg/dL4.3 mg/dL
    LoDComparable to predicate3.5 mg/dL5.2 mg/dL
    LoQComparable to predicate4 mg/dL5.2 mg/dL
    Measuring RangeComparable to predicate7-140 mg/dL7-140 mg/dL
    Linearity Regression (Y=mX+b)R² comparable to predicatey=1.049x-0.3459, R²=0.9961y=1.004x-0.5956, R²=0.9998
    Within-Run Precision (CV%)Comparable to predicateLevel 1: 4.0%, Level 2: 2.8%, Level 3: 2.6%Level 1: 2.0%, Level 2: 1.9%, Level 3: 0.9%
    Within-Lab Precision (CV%)Comparable to predicateLevel 1: 2.8%, Level 2: 2.0%, Level 3: 1.8%Level 1: 3.3%, Level 2: 2.0%, Level 3: 1.8%
    Method Comparison (Correlation R)Comparable to predicateR=0.989R=0.999
    Interference StudyNo significant interference for tested substancesNo interference found for 7 endogenous & 23 exogenous substances.No interference found for 5 endogenous & 22 exogenous substances.
    Specimen Stability (Refrigerated)Comparable to predicate6 days10 days

    Study Proving Acceptance Criteria:

    The study conducted was an analytical validation comparing the performance of the Vantera® Clinical Analyzer with the NMR LipoProfile® test to its predicate device (NMR Profiler for the assay) across various analytical parameters. The overall conclusion is that the new device is "as safe and effective as its predicate device."

    2. Sample size used for the test set and the data provenance

    • Test Sets (Analytical Studies):
      • Analytical Sensitivity (LoB, LoD, LoQ): Five serum pools for low concentrations tested in replicates of 4 for 3 days. Non-lipoprotein specimens analyzed 60 consecutive times for 3 days for LoB.
      • Assay Precision (Within-run & Within-Laboratory): 20 replicates of three patient serum pools in the same run and in 20 different runs over 20 days. Reproducibility study used 5 levels of serum panels tested for 5 days, 6 runs per day, 2 replicates per run at 3 sites (n=60 per panel per site, total N=177-180 across all sites for each panel).
      • Linearity: Three serum pools prepared from patient specimens mixed and diluted to produce eleven (for LDL-P) or twelve (for TG and HDL-C) different samples, with four replicates of each pool analyzed.
      • Method Comparison:
        • LDL-P: n=1483 serum samples.
        • HDL-C: n=1518 serum samples.
        • Triglycerides: n=1520 serum samples.
      • Interfering Substances: 7 endogenous agents and 23 drugs were screened.
      • Reference Range: Serum samples (n=452) from apparently healthy men (n=158) and women (n=294).
    • Data Provenance: The document does not specify the country of origin. The studies were described as "analytical validations" and included testing using "patient specimens" and "serum pools." There is no explicit mention of the data being either retrospective or prospective, but the nature of the analytical studies suggests controlled laboratory environments rather than a large-scale clinical trial with patient follow-up. For the reference range, it states "serum samples... were analyzed from apparently healthy men and women," which implies a prospective collection for this specific purpose or a well-characterized existing cohort. The MESA (Multi-Ethnic Study of Atherosclerosis) is mentioned for the predicate device's reference range, suggesting a US context for that, but it's not explicitly stated for the proposed device's reference population.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not describe the use of human experts to establish "ground truth" for the test set in the context of interpretation or diagnosis. This device is an automated laboratory analyzer for quantifying analytes. The "ground truth" for its analytical performance studies (e.g., precision, linearity, method comparison) is established by comparing its measurements to a reference method or known concentrations, or by assessing consistency internally.

    4. Adjudication method for the test set

    Not applicable. There was no clinical study involving human interpretation or diagnosis that would require an adjudication method. The testing involved direct analytical measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in-vitro diagnostic (IVD) device for quantifying analytes (LDL-P, HDL-C, Triglycerides), not an AI-assisted diagnostic imaging or interpretation device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the primary performance studies presented are "standalone" in the sense that they assess the device's analytical performance (algorithm + instrument) in quantifying the specified analytes without human-in-the-loop performance for diagnosis or interpretation. The device's output is explicit numerical analyte concentrations.

    7. The type of ground truth used

    The ground truth used for verifying the analytical performance of the device was:

    • Reference Methods/Known Concentrations: For analytical sensitivity (LoB, LoD, LoQ), linearity, and precision, the "ground truth" was established through precisely prepared serum pools and non-lipoprotein specimens with known or target concentrations, or through statistical determination methods (e.g., EP17-A).
    • Comparison to Predicate Device: For method comparison studies, the "ground truth" or reference was the measurements obtained from the legally marketed predicate device (NMR Profiler) using patient samples. The goal was to show high correlation and similar results between the two devices.
    • CLSI Guidelines: Standardized guidelines (e.g., EP5-A2, EP6-A, EP7-A2, EP9-A2, EP17-A) from the Clinical and Laboratory Standards Institute (CLSI) were referenced for establishing protocols for these analytical validations.

    8. The sample size for the training set

    The document does not explicitly mention a "training set" in the context of machine learning or AI models. This device is an automated NMR spectrometer that measures signals and converts them to concentrations based on specified deconvolution analysis models. The development of these deconvolution models would have involved a form of "training" or optimization, but the document does not detail the dataset size or methodology used for this prior model development. The document focuses on the analytical validation of the manufactured device.

    9. How the ground truth for the training set was established

    As there is no explicit mention of a "training set" in the conventional AI/ML sense, this question cannot be directly answered from the provided text. However, the assay description mentions:
    "The NMR signals from the various lipoprotein subclasses have unique and distinctive frequencies and lineshapes, each of which is accounted for in the deconvolution analysis model. Each subclass signal amplitude is proportional to the number of subclass particles emitting the signal, which enables subclass particle concentrations to be calculated from the subclass signal amplitudes derived from the spectral deconvolution analysis."
    This suggests that the deconvolution analysis model was developed using a "ground truth" based on the established biophysical properties of lipoprotein subclasses and their NMR spectral characteristics. This likely involved:

    • Carefully characterized lipoprotein samples with known subclass concentrations.
    • Expert knowledge of NMR spectroscopy and signal processing.
    • Calibration against established reference methods for lipoprotein analysis.

    The document does not provide details on the specific data sets or expert consensus used for the initial development and establishment of this deconvolution model.