Search Results
Found 1 results
510(k) Data Aggregation
(59 days)
VALIDATE GC1, GC2, GC3 AND GC4 CALIBRATION VERIFICATION/LINEARITY TEST SET
Validate® GC Calibration Verification / Linearity Test Sets are intended for in vitro diagnostic use in the quantitative determination of linearity, calibration verification and verification of reportable range in automated, semi automated, and manual chemistry systems. See attached Package Insert labeling for all analytes claimed.
VALIDATE® GC Calibration Verification / Linearity Test Sets are human and aqueous based calibration verification materials containing multiple levels used to establish the relationship between theoretical operation and actual performance of the included analytes. There exists a linear relationship among each set of solutions.
The provided text is a 510(k) summary for a medical device called "VALIDATE® GC Calibration Verification / Linearity Test Sets." This device is a quality control material intended for in vitro diagnostic use to determine linearity, calibration verification, and reportable range in chemistry systems.
It is crucial to understand that this submission is for a quality control material, not an AI device or an imaging device. Therefore, the concepts of human readers, AI assistance, complex imaging, and expert adjudication (as typically applied in imaging or AI-driven diagnostic tools) are not applicable here. The "device" in this context is a set of chemical solutions with known properties, and its performance is evaluated based on its chemical stability, consistency, and how accurately it can verify the calibration and linearity of laboratory instruments.
Given this context, I will address the questions as they pertain to a quality control material and the information provided in the 510(k) summary. Many of the questions are specifically designed for AI/imaging devices and cannot be directly answered from the provided text for a chemical control.
Here's the analysis:
Acceptance Criteria and Study for VALIDATE® GC Calibration Verification / Linearity Test Sets
The core of the study for this type of device lies in demonstrating its equivalence to existing predicate devices and its ability to accurately assess the linearity and calibration of various chemistry analyzers.
1. A table of acceptance criteria and the reported device performance
The provided 510(k) summary does not include a detailed table of specific acceptance criteria or reported device performance data. It states:
- "The VALIDATE® GC Calibration Verification / Linearity Test Sets behave in a manner suitable for the evaluation of calibration, verification of reportable range, and the linear response of the listed analytes over the ranges tested when compared to the predicate devices."
- "VALIDATE® GC Calibration Verification / Linearity Test Sets are as safe, as effective, and perform as well as or better than the predicate device."
Such specific data would typically be found in the full 510(k) submission, including the package insert and detailed study reports, which are not part of this summary excerpt. For a calibration verification material, acceptance criteria would typically include:
* Linearity: The observed values for each level across the concentration range should demonstrate a linear relationship. This is often assessed by linear regression analysis, with criteria for correlation coefficient (e.g., R > 0.99) and acceptable deviation from linearity (e.g., % bias at specific points).
* Accuracy/Bias: The expected values of the control material should be accurately recovered by the chemistry systems, often compared to target values with a specified allowable bias.
* Stability: The material must maintain its stated characteristics over its shelf life and during use.
* Matrix Effects: Minimal interference from the control material matrix on various analytical methods.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The summary does not specify the sample size (e.g., number of runs, number of instruments, number of analyte levels tested on each instrument) or the provenance (country, retrospective/prospective) of the data. For this type of device, a "test set" would refer to the various levels of the calibration verification material across different analytes, tested on multiple instrument platforms.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This question is not applicable to a chemical quality control material. "Ground truth" for these materials is established through rigorous manufacturing processes, formulation, and analytical testing by qualified chemists and laboratory personnel using highly accurate reference methods and/or validated assays. It's not typically established by clinical "experts" in the sense of physicians or radiologists.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This question is not applicable to a chemical quality control material. Adjudication methods like 2+1 are used for human interpretation of clinical data (e.g., image reading) where disagreement among experts might occur. The "truth" for a chemical control is determined by its inherent chemical properties and the accuracy of its manufacture and assay.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is completely inapplicable. MRMC studies and "human readers improving with AI" are concepts specific to AI-assisted diagnostic imaging or similar interpretation tasks. This device is a chemical control, not an AI or imaging diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is completely inapplicable. There is no "algorithm" in the sense of AI or software for this chemical control material. The device itself is a physical substance.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
As mentioned in #3, "ground truth" for this device would be established by:
- Reference Methods: Highly accurate and precise laboratory methods used to assign target values to the various levels of the control material.
- Gravimetric and Volumetric Standards: Precise measurements during formulation.
- Interlaboratory Consensus: In some cases, target values might be confirmed through round-robin studies among multiple qualified laboratories.
- Traceability: Ensuring the values are traceable to internationally recognized reference materials or methods.
It is not based on expert consensus in a clinical diagnostic sense, pathology, or outcomes data.
8. The sample size for the training set
This concept is not applicable here. A "training set" refers to data used to train a machine learning model. This device is a chemical control material, not a machine learning algorithm. Its development involves chemical formulation and analytical testing, not statistical training data sets in the AI sense.
9. How the ground truth for the training set was established
Not applicable, as there is no "training set" in the context of this chemical control material.
Ask a specific question about this device
Page 1 of 1