Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K150343
    Manufacturer
    Date Cleared
    2015-04-03

    (51 days)

    Product Code
    Regulation Number
    870.1255
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    V8 Transluminal BAV Catheter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V8 Transluminal BAV Catheter is indicated for Balloon Aortic Valvuloplasty.

    Device Description

    The V8 Transluminal BAV Catheter System features an hour glass shaped dilatation balloon on the distal end of a catheter. The catheter is inserted through a percutaneous entry site into the common femoral artery via an introducer sheath and advanced retrograde to the aortic valve. The catheter is always delivered over a guidewire. The balloon is then inflated to dilate the stenotic aortic valve leaflets in an effort to increase valve opening dimensions and systemic blood flow by improving leaflet mobility. The hour glass shaped balloon, with the undersized waist segment, is intended to minimize over-dilatation of the valve annulus while allowing the full dilation of the valve leaflet. The bulbous proximal balloon segment is appropriately sized for the patient's aortic root dimensions to maximize valve leaflet opening.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the "V8 Transluminal BAV Catheter" (K150343). The purpose of this submission is not to introduce a new device, but rather to update the labeling of an existing device (V8 Transluminal BAV Catheter, K133607) and establish its substantial equivalence to another predicate (Z-Med/Z-Med II, K122012).

    The key finding is that no new performance data or studies were conducted for this specific submission (K150343) because the device design, materials, and function remain identical to the predicate V8 device (K133607). Therefore, the acceptance criteria and the studies proving the device meets them refer to previous submissions (K133607 and K123111).

    Given this information, I will synthesize the available information regarding the acceptance criteria and prior studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this submission explicitly states "no design changes have been made to the predicate K133607 V8 device" and "performance testing performed during design verification and shelf life testing were not repeated because the design and materials of the device remain the same as the V8 predicate," the acceptance criteria and performance are referred to the previous submissions. The document lists the types of tests performed in the predicate design verification. Specific quantitative acceptance criteria are not explicitly stated in this document but are implied by the successful completion of these tests in the predicate device.

    Acceptance Criteria (Implied by Predicate Verification)Reported Device Performance (from Predicate)
    Meet specified balloon rated burst pressurePerformance met for predicate V8 (K133607)
    Meet specified balloon compliancePerformance met for predicate V8 (K133607)
    Conform to critical dimensionsPerformance met for predicate V8 (K133607)
    Compatible with guidewire and introducerPerformance met for predicate V8 (K133607)
    Achieve specified deflation timesPerformance met for predicate V8 (K133607)
    Maintain integrity over repeat inflationsPerformance met for predicate V8 (K133607)
    Exhibit no leaksPerformance met for predicate V8 (K133607)
    Withstand specified tensile forcesPerformance met for predicate V8 (K133607)
    Resist kinkingPerformance met for predicate V8 (K133607)
    Exhibit adequate torque performancePerformance met for predicate V8 (K133607)
    Biocompatible for patient contactAchieved (test data provided in K123111)
    Sterile for patient-contacting componentsAchieved (sterilization validation in K123111)
    Maintain shelf-lifeAchieved (shelf life testing data in K133607)
    Demonstrate safety and effectiveness for device delivery, inflation, deflation, and removalAchieved (design verification in K133607, additional bench testing summarized in Section 11.5 of this submission)
    Demonstrate performance, including ability to treat aortic stenosis (in vivo)Achieved (based on predicate device in K132728, and identical shape/principle of operation)

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not provide details on sample sizes for test sets for the initial predicate device studies. It only states that the performance testing was not repeated because the design and materials are the same. Therefore, the sample sizes and data provenance would refer to the original K133607 and K123111 submissions. The nature of these tests (design verification, shelf life, biocompatibility) implies bench and potentially in-vitro/animal studies, rather than human clinical trials for this type of device. There is no mention of country of origin or retrospective/prospective for the previous studies within this document.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This request is primarily applicable to studies involving interpretation of data (e.g., medical images). The listed tests are primarily engineering and material performance tests (e.g., burst pressure, tensile strength, biocompatibility, sterility). For such tests, "ground truth" is typically established by objective physical measurements and adherence to recognized standards, rather than expert consensus on interpretive data. Therefore, this question is not directly applicable in the context of the provided document.

    4. Adjudication Method for the Test Set

    As explained above, the tests are primarily objective physical and material performance evaluations. Adjudication methods like 2+1 or 3+1 are used for subjective assessments (e.g., radiographic interpretations). This is not relevant to the types of tests described in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence of a medical device (a catheter) through bench testing, material characterization, and reference to prior regulatory submissions. MRMC studies are typically performed for diagnostic devices, especially those involving human interpretation of complex data (like imaging algorithms), to assess the impact of a device on reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the loop performance) Was Done

    No, this question refers to the performance of an algorithm without human intervention, which is relevant for AI/ML devices. The V8 Transluminal BAV Catheter is a medical instrument (hardware), not an algorithm or AI system. Therefore, a "standalone algorithm performance" study is not applicable.

    7. The Type of Ground Truth Used

    For the performance data referred to from the predicate submissions (K133607 and K123111), the ground truth for tests like burst pressure, compliance, critical dimensions, tensile strength, kink resistance, and torque would be established through:

    • Objective physical measurements: Using calibrated equipment and standardized test methods.
    • Chemical/Biological analysis: For biocompatibility and sterility, adhering to ISO standards and laboratory protocols.
    • Engineering specifications: The device is designed to meet specific engineering tolerances and performance parameters.

    Ground truths like expert consensus, pathology, or outcomes data would typically be used for diagnostic or treatment effectiveness studies involving patient data, which are not detailed here as the primary evidence for this 510(k) submission.

    8. The Sample Size for the Training Set

    This question is only relevant for artificial intelligence or machine learning models. The V8 Transluminal BAV Catheter is a physical medical device. Therefore, the concept of a "training set" is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    As explained above, there is no training set for this type of physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133607
    Manufacturer
    Date Cleared
    2014-02-11

    (78 days)

    Product Code
    Regulation Number
    870.1255
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    V8 TRANSLUMINAL BAV CATHETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V8 Transluminal BAV Catheter is indicated for Balloon Aortic Valvuloplasty.

    Device Description

    The V8 Transluminal BAV Catheter System features a figure-8 shaped dilatation balloon on the distal end of a catheter. The catheter is inserted through a percutaneous entry site into the common femoral artery via an introducer sheath and advanced retrograde to the aortic valve. The catheter is always delivered over a guidewire. The balloon is then inflated to dilate the stenotic aortic valve leaflets in an effort to increase valve opening dimensions and systemic blood flow by improving leaflet mobility. The figure-8 shaped balloon with the undersized waist segment is intended to minimize over-dilatation of the valve annulus while allowing the full dilation of the valve leaflet. The bulbous proximal balloon segment is appropriately sized for the patient's aortic root dimensions to maximize valve leaflet opening.

    AI/ML Overview

    This K133607 submission is for a device modification (V8 Transluminal BAV Catheter System), not for a new AI/ML-driven device. Therefore, the traditional acceptance criteria and study designs typically associated with AI/ML device evaluations (e.g., performance metrics like sensitivity/specificity, reader studies, etc.) are not directly applicable in this context.

    The submission focuses on demonstrating substantial equivalence to predicate devices based on design, materials, technological characteristics, function, and identical intended use. The "acceptance criteria" here are the Special Control Requirements for Balloon Aortic Valvuloplasty (BAV) devices, and the "study" is the collection of design verification and shelf-life testing.

    Here's the information broken down based on the provided document, adapted to the context of a device modification and substantial equivalence:


    1. Table of Acceptance Criteria and Reported Device Performance

    Special Control Requirement (Acceptance Criteria)Reported Device Performance (Evidence of Conformity)
    The device should be demonstrated to be biocompatible.The proposed change does not affect the biocompatibility of the device, as no new materials were added. Biocompatibility test data from predicate device K123111 were provided and deemed applicable.
    Sterility and shelf life testing should demonstrate the sterility of patient-contacting components and the shelf-life of these components.The sterile packaging and shelf carton are identical to predicate V8 devices. The worst-case device is identical, so sterility data from K123111 is applicable. Shelf life testing was repeated and provided in this submission, and aging data for sterile barrier seal from the predicate device submission was also provided.
    Non-clinical performance evaluation of the device should demonstrate substantial equivalence in terms of safety and effectiveness for device delivery, inflation, deflation, and removal.Design verification testing was repeated and provided in the submission, including data for delivery, inflation, deflation, and removal. Specific tests performed before and after accelerated aging included: Balloon rated burst pressure, Balloon compliance, Critical dimension verifications, Guidewire and introducer compatibility, Deflation times, Repeat inflation, Leak, Tensile, Kink, and Torque. The testing showed the device meets its specifications both before and after aging.
    In vivo evaluation of the device should demonstrate device performance, including the ability of the device to treat aortic stenosis.The general shape and principles of operation have not changed. Device dimensions are identical to the predicate device in K132728. The submission states that the slight changes in balloon geometry and compliance are within the bounds of original V8 balloon dimensions, and nominal dimensions are identical to V8 from K132728, possessing the same technical characteristics of cleared predicate V8 balloons. This implies that previous in vivo evaluations for the predicate devices are considered sufficient and applicable, given the demonstrated equivalence in key performance characteristics.
    Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device.The Instructions For Use (IFU) contains a "Potential Complications, Outcomes, Adverse Events" section.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not applicable in the context of an AI/ML test set. The "testing" refers to bench testing (design verification and shelf life testing) of the physical device and its components. The document does not specify exact sample sizes for each type of bench test, but indicates tests were repeated and performed before and after accelerated aging. This implies multiple units were tested.
    • Data Provenance: Not applicable for a traditional test set as described for AI/ML. All data generated came from in-house device testing (bench testing and compatibility assessments). No patient data or external datasets were used to "test" the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The acceptance criteria and performance are based on engineering specifications and physical measurements, not on expert interpretations of data like in medical imaging.

    4. Adjudication method for the test set

    Not applicable. There was no need for expert adjudication, as the performance was evaluated against predefined engineering specifications for the physical device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is a physical medical device (catheter), not an AI/ML-driven diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's performance is established by engineering specifications and ISO/ASTM standards for medical device components and performance (e.g., burst pressure, compliance, tensile strength, sterility, etc.). Data is compared against these predetermined, objective, quantitative metrics.

    8. The sample size for the training set

    Not applicable, as this is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as this is a physical medical device, not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132728
    Manufacturer
    Date Cleared
    2013-11-22

    (80 days)

    Product Code
    Regulation Number
    870.1255
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    V8 TRANSLUMINAL BAV CATHETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V8 Transluminal BAV Catheter is indicated for Balloon Aortic Valvuloplasty.

    Device Description

    The V8 Transluminal BAV Catheter System features an hour-glass shaped dilatation balloon on the distal end of a catheter. The catheter is inserted through a percutaneous entry site into the common femoral artery via an introducer sheath and advanced retrograde to the aortic valve. The catheter is always delivered over a guidewire. The balloon is then inflated to dilate the stenotic aortic valve leaflets in an effort to increase valve opening dimensions and systemic blood flow by improving leaflet mobility. The hour-glass shaped balloon with the undersized waist segment is intended to minimize over-dilatation of the valve annulus while allowing the full dilation of the valve leaflet. The bulbous proximal balloon segment is appropriately sized for the patient's aortic root dimensions to maximize valve leaflet opening.

    AI/ML Overview

    The provided text describes a 510(k) submission for the InterValve V8 Transluminal BAV Catheter. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting new clinical trials to establish efficacy and safety from scratch. Therefore, the information typically found in acceptance criteria tables and detailed study descriptions for novel devices is not fully present.

    Based on the provided text, here's what can be extracted and inferred regarding acceptance criteria and the supporting study:

    1. A table of acceptance criteria and the reported device performance

    The submission demonstrates substantial equivalence based on technological characteristics and performance data from predicate devices, with relevant design verification testing repeated for minor modifications.

    Acceptance CriterionReported Device Performance
    Material Composition (V8 balloon)Made of clear non-compliant polymeric material. (Identical to predicate K123111)
    Balloon Sizes (Available diameters)22mm, 24mm, 28mm. (New dimensions for 24mm and 26mm are within bounds of original K123111 and do not change technological characteristics).
    Balloon Waist DesignHour-glass shaped with undersized waist segment, smaller than bulb diameter up to rated burst pressure. (Identical to predicate K123111)
    Catheter Working LengthsStandard working lengths (107cm - 113cm). (Identical to predicate K123111)
    Introducer Sheath CompatibilityCompatible with a 12F or 14F introducer sheath. (Identical to predicate K123111)
    Guidewire CompatibilityTracked over a 0.035" wire. (Identical to predicate K123111)
    Radiopaque MarkersPlatinum iridium marker bands on inner shaft beneath the balloon (center of waist, outside edges of proximal and distal balloon shoulders). (Identical to predicate K123111)
    Packaging & SterilizationPackaged in heat-sealed Tyvek pouch, provided sterilized. (Identical to predicate K123111)
    Intended UseBalloon Aortic Valvuloplasty. (No change from predicate K123111 and K082776)
    Safety & Effectiveness ProfileThe clinical experience used to support the change to the device labeling indicates that the change does not affect the safety and effectiveness profile of the device. The minor differences do not raise new questions of safety or efficacy.
    Physician Discretion for Rapid Ventricular PacingRevised wording on labeling to allow physician discretion (consistent with how devices are used, similar to predicate devices, and accepted industry practice).
    Bonds and Balloon Characteristics (following redesign)"Relevant design verification testing was repeated. These tests included all bonds and balloon characteristics of the redesigned balloon." "The results demonstrated that the device functions as intended."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The submission does not detail a specific "test set" in the context of a new clinical study. Instead, it relies on:

    • Design Verification Testing: For the minor modifications to the 24mm and 26mm balloons, all bonds and balloon characteristics were tested. The sample size for this specific testing is not provided. This would typically involve engineering tests on a batch of manufactured devices, not patient data.
    • Clinical Experience of Predicate Device: The submission refers to "clinical experience used to support the change to the device labeling" and "Predicate device clinical experience showed that physicians may or may not employ the use of concomitant rapid ventricular pacing." The sample size, country of origin, or retrospective/prospective nature of this predicate clinical experience are not detailed in the provided text, as this information would have been associated with the original clearance of the predicate device (K123111).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided as a formal clinical study with expert adjudication for a ground truth establishing accuracy was not conducted for this 510(k) submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/none as a formal clinical study with expert adjudication was not conducted. The assessment relies on engineering testing and the established safety/efficacy profile of the predicate devices.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This submission is for a medical catheter, not an AI-assisted diagnostic or therapeutic device that would involve human "readers" or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the design verification testing of the modified balloon characteristics, the "ground truth" would be engineering specifications and performance standards for medical balloons (e.g., burst pressure, diameter at specified pressure, material integrity). For the labeling change regarding physician discretion, the ground truth appears to be current clinical practice and accepted industry standard ("The revised wording on the device labeling was changed to be consistent with how the devices are used by physicians, giving the physician the discretion... This is similar to the predicate devices and is accepted practice in the industry.").

    8. The sample size for the training set

    This is not applicable. There is no "training set" in the context of this 510(k) submission, as it does not involve machine learning or AI.

    9. How the ground truth for the training set was established

    This is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123111
    Manufacturer
    Date Cleared
    2013-01-31

    (120 days)

    Product Code
    Regulation Number
    870.1255
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    V8 TRANSLUMINAL BAV CATHETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V8 Transluminal BAV Catheter is indicated for Balloon Aortic Valvuloplasty.

    Device Description

    The V8 Transluminal BAV Catheter System features an hour-glass shaped dilatation balloon on the distal end of a catheter. The catheter is inserted through a percutaneous entry site into the common femoral artery via an introducer sheath and advanced retrograde to the aortic valve. The catheter is always delivered over a guidewire. The balloon is then inflated to dilate the stenotic aortic valve leaflets in an effort to increase valve opening dimensions and systemic blood flow by improving leaflet mobility. The hour-glass shaped balloon with the undersized waist segment is intended to minimize over-dilatation of the valve annulus while allowing the full dilation of the valve leaflet. The bulbous proximal balloon segment is appropriately sized for the patient's aortic root dimensions to maximize valve leaflet opening.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "V8 Transluminal BAV Catheter" and references a predicate device, the "NuMed NuCLEUS-X BAV Catheter (K082776)". The study provided focuses on establishing substantial equivalence to the predicate device rather than presenting a performance study with detailed acceptance criteria against clinical efficacy.

    Therefore, the response will focus on the engineering and biocompatibility aspects, as these are the primary details provided regarding performance.

    Acceptance Criteria for the V8 Transluminal BAV Catheter and Reported Device Performance

    Acceptance CriteriaReported Device Performance/Comparison to Predicate
    Physical and Performance SpecificationsThe V8 Transluminal BAV Catheter System is tested and meets all its physical and performance specifications, including:
    Balloon rated burst pressureMet specifications.
    Balloon complianceMet specifications. The V8 balloon maintains its hour-glass shape at low and rated burst pressures, unlike the predicate device which loses its shape at working pressure. This difference is not deemed to raise new safety/effectiveness questions as the effectiveness is determined by the bulbous section (similar to predicate) and the waist reduces probability of annular distention, hence risk of dissection.
    Critical dimension verificationsMet specifications. The V8 balloon is available in four diameter sizes (22mm - 28mm). The waist is smaller than the bulb diameter up to rated burst pressure. Catheter available in standard working lengths (107cm - 113cm).
    Guidewire and introducer compatibilityMet specifications. Compatible with a 12F or 14F introducer sheath and tracked over a 0.035" wire. (Predicate also compatible with 0.035" wire).
    Fluoroscopic visualizationMet specifications. Radiopaque platinum iridium marker bands are present on the inner shaft beneath the balloon (one at the center of the waist, and one each at the outside edges of the proximal and distal balloon shoulders).
    Deflation timesMet specifications.
    Repeat inflationMet specifications.
    LeakMet specifications.
    TensileMet specifications.
    KinkMet specifications.
    TorqueMet specifications.
    Luer lock compatibilityMet specifications.
    DistributionMet specifications.
    BiocompatibilityTested per ISO 10993-1 for short duration contact with blood (
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1