510(k) Data Aggregation

    K Number
    K232837
    Manufacturer
    Date Cleared
    2024-06-07

    (267 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Uro-G HD Flexible Cystoscope

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Uro-G HD Flexible Cystoscope is intended for endoscopic diagnosis and infusion of irrigating fluids within the bladder and urethra.

    Generally recognized indications for cystoscopy include:
    • Symptomatic voiding dysfunction
    • Hematuria
    • Bladder tumor surveillance
    • Recurrent lower urinary tract infections
    • Pelvic pain syndromes

    Device Description

    The Uro-G HD Flexible Cystoscope System is a handheld, battery-operated portable cystoscope consisting of a disposable steerable endoscopic cannula with a 2.2 mm channel for fluid infusion and instrument advancement, and a reusable handle with a video monitor. The fully assembled Uro-G HD Cystoscope System—cannula and handle—has an overall length of 630 mm (approximately 25 inches). The disposable cannula's working length is 380.2 mm (approx. 15 inches) and a total length of 406 mm, from tip to connector. The cannula's maximum width is 5.6 mm and it has a connector for attaching and detaching to the handle.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the Uro-G HD Flexible Cystoscope, seeking to prove substantial equivalence to a predicate device (Uro-G Flexible Cystoscope, K202921). As such, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" refer to the demonstration that the modified device (Uro-G HD) is as safe and effective as the predicate device, not necessarily a performance study against a specific clinical endpoint. The primary method for proving this is through comparative testing and analysis of technological characteristics.

    Here's an analysis of the provided text in the context of acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as quantitative targets in a table format within the provided text. Instead, the document uses a comparative approach against a predicate device. The "reported device performance" is essentially the Uro-G HD's characteristics, and "meeting acceptance criteria" means demonstrating that these characteristics are equivalent to or improve upon the predicate without introducing new risks.

    Here’s a table summarizing the key characteristics and the comparison to the predicate device, highlighting what could be implicitly considered an "acceptance criterion" (i.e., equivalence or improvement):

    CharacteristicPredicate Device (Uro-G, K202921) PerformanceSubject Device (Uro-G HD) PerformanceAcceptance Criteria (Implicit)Met?
    Cannula
    Working length380.2 mm380.2 mmNot Changed (Equivalent)Yes
    Total length (tip to connector)406 mm406 mmNot Changed (Equivalent)Yes
    Number of ports22Not Changed (Equivalent)Yes
    Working channel minimum width2.2 mm2.2 mmNot Changed (Equivalent)Yes
    Maximum insertion width (OD)5.5 mm5.6 mmEquivalent (Increased by 0.1mm, deemed not to introduce new risk, as human urethral diameter is larger)Yes
    Tip OD4 mm4 mmNot Changed (Equivalent)Yes
    Camera
    CMOS module pixels0.3 million pixels0.7 million pixelsEquivalent (Higher pixels for HD imaging, does not introduce new risk)Yes
    Focal length5 mm to 50 mm5 mm to 50 mmNot Changed (Equivalent)Yes
    Field of view (in air)120 ° ± 5 °100 ° ± 5 °Equivalent (Smaller FOV but sufficient for HD imaging in confined spaces, does not introduce new risk)Yes
    Direction of view from center axis0 ° (Forward Viewing)0 ° (Forward Viewing)Not Changed (Equivalent)Yes
    Image resolution3 lp/mm on 1551 USAF test chart in air3 lp/mm on 1551 USAF test chart in airNot Changed (Equivalent)Yes
    Image distortion120 mL/min> 120 mL/minNot Changed (Equivalent)Yes
    Light sourceLED, 520 lumens at 25 mm focal lengthLED, 520 lumens at 25 mm focal lengthNot Changed (Equivalent)Yes
    Tip temperature when LEDs are on