Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140085
    Manufacturer
    Date Cleared
    2014-11-18

    (309 days)

    Product Code
    Regulation Number
    876.5090
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    UNIVERSA PERCULTANEOUS DRAINAGE CATHETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Universa® Loop Drainage Catheter Set is intended to provide percutaneous urine drainage from the genitourinary system.

    The Universa® Malecot Drainage Catheter Set is intended to provide percutaneous urine drainage from the genitourinary system.

    Device Description

    The Universa® Percutaneous Drainage Catheter Sets include introductory and exchange sets available with either a loop (6-14 Fr) or a Malecot catheter (8-24 Fr). Suprapubic sets are available with a Malecot catheter (8-16 Fr). Each set includes a drainage catheter and connecting tube, and may also include straightening stylets, trocar needle with obturator, hollow needle, wire guide, dilators, silicone retention disc with pull tie, or one-way stopcock. The devices will be supplied sterile and intended for one-time use.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for two medical devices: the Universa® Loop Drainage Catheter Set and the Universa® Malecot Drainage Catheter Set. It does not describe an AI/ML device or a study involving such a device. Therefore, many of the requested categories (such as sample sizes for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, or standalone algorithm performance) are not applicable to this document.

    However, I can extract the information related to acceptance criteria and the studies performed for these physical medical devices.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test PerformedPredetermined Acceptance CriteriaReported Device Performance
    Mechanical IntegrityTensile Strength TestingNot explicitly stated but implied to ensure catheter materials/bonds do not fracture during proper clinical use.The predetermined acceptance criteria were met.
    Functional RetentionLoop and Malecot Retention TestingNot explicitly stated but implied to ensure retention features resist dislodgement/removal.The predetermined acceptance criterion was met.
    Fluid DynamicsGravity Flow Rate, Lumen Blockage, and Leakage TestingNot explicitly stated but implied to ensure consistent water flow without blockage or leakage.The predetermined acceptance criteria were met.
    BiocompatibilityBiocompatibility TestingConformance with ISO 10993-1.The predetermined acceptance criteria were met.
    Usability/PerformanceSimulated Use TestingNot explicitly stated but implied to ensure compatibility and performance according to instructions for use.The results of the study are acceptable for clinical practice.

    2. Sample size used for the test set and the data provenance

    • Sample size: Not specified in the provided document.
    • Data provenance: Not applicable. These are physical device tests, not data-driven studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of experts: Not applicable. Ground truth for these physical device tests is established through engineering and biological standards, not expert consensus on data interpretation.
    • Qualifications of experts: Not applicable.

    4. Adjudication method for the test set

    • Adjudication method: Not applicable. Performance is measured against physical test standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC study: No. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone study: No. This is not an AI/ML device.

    7. The type of ground truth used

    • The "ground truth" for these physical device tests is based on:
      • Engineering specifications and material properties (for tensile strength, flow rates).
      • Functional design requirements (for retention).
      • International standards (ISO 10993-1 for biocompatibility).
      • Instructions for use for simulated clinical performance.

    8. The sample size for the training set

    • Sample size: Not applicable. This is not an AI/ML device, so there is no training set in that context.

    9. How the ground truth for the training set was established

    • Ground truth establishment: Not applicable. No training set for an AI/ML context.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1