Search Results
Found 1 results
510(k) Data Aggregation
(309 days)
UNIVERSA PERCULTANEOUS DRAINAGE CATHETER
The Universa® Loop Drainage Catheter Set is intended to provide percutaneous urine drainage from the genitourinary system.
The Universa® Malecot Drainage Catheter Set is intended to provide percutaneous urine drainage from the genitourinary system.
The Universa® Percutaneous Drainage Catheter Sets include introductory and exchange sets available with either a loop (6-14 Fr) or a Malecot catheter (8-24 Fr). Suprapubic sets are available with a Malecot catheter (8-16 Fr). Each set includes a drainage catheter and connecting tube, and may also include straightening stylets, trocar needle with obturator, hollow needle, wire guide, dilators, silicone retention disc with pull tie, or one-way stopcock. The devices will be supplied sterile and intended for one-time use.
The provided text describes a 510(k) premarket notification for two medical devices: the Universa® Loop Drainage Catheter Set and the Universa® Malecot Drainage Catheter Set. It does not describe an AI/ML device or a study involving such a device. Therefore, many of the requested categories (such as sample sizes for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, or standalone algorithm performance) are not applicable to this document.
However, I can extract the information related to acceptance criteria and the studies performed for these physical medical devices.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Test Performed | Predetermined Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Mechanical Integrity | Tensile Strength Testing | Not explicitly stated but implied to ensure catheter materials/bonds do not fracture during proper clinical use. | The predetermined acceptance criteria were met. |
Functional Retention | Loop and Malecot Retention Testing | Not explicitly stated but implied to ensure retention features resist dislodgement/removal. | The predetermined acceptance criterion was met. |
Fluid Dynamics | Gravity Flow Rate, Lumen Blockage, and Leakage Testing | Not explicitly stated but implied to ensure consistent water flow without blockage or leakage. | The predetermined acceptance criteria were met. |
Biocompatibility | Biocompatibility Testing | Conformance with ISO 10993-1. | The predetermined acceptance criteria were met. |
Usability/Performance | Simulated Use Testing | Not explicitly stated but implied to ensure compatibility and performance according to instructions for use. | The results of the study are acceptable for clinical practice. |
2. Sample size used for the test set and the data provenance
- Sample size: Not specified in the provided document.
- Data provenance: Not applicable. These are physical device tests, not data-driven studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of experts: Not applicable. Ground truth for these physical device tests is established through engineering and biological standards, not expert consensus on data interpretation.
- Qualifications of experts: Not applicable.
4. Adjudication method for the test set
- Adjudication method: Not applicable. Performance is measured against physical test standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC study: No. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone study: No. This is not an AI/ML device.
7. The type of ground truth used
- The "ground truth" for these physical device tests is based on:
- Engineering specifications and material properties (for tensile strength, flow rates).
- Functional design requirements (for retention).
- International standards (ISO 10993-1 for biocompatibility).
- Instructions for use for simulated clinical performance.
8. The sample size for the training set
- Sample size: Not applicable. This is not an AI/ML device, so there is no training set in that context.
9. How the ground truth for the training set was established
- Ground truth establishment: Not applicable. No training set for an AI/ML context.
Ask a specific question about this device
Page 1 of 1