Search Results
Found 3 results
510(k) Data Aggregation
(24 days)
The UNIPLATE Anterior Cervical Plate System is intended for anterior cervical intervertebral body fixation. This system is indicated for patients in which stability is desired following anterior cervical fusion for the indications listed below. The intended levels for treatment range from C2-T1. Indications include symptomatic cervical spondylosis, trauma, fracture, post-traumatic kyphosis or lordosis, tumor, degenerative disc disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, re-operation for failed fusion, or instability following surgery for the above indications.
The UNIPLATE Anterior Cervical Plate System consists of an assortment of plates and screws. The UNIPLATE Anterior Cervical Plate System also contains Class I manual surgical instruments and cases that are considered exempt from premarket notification. Manufactured from ASTM F 136 implant grade titanium alloy.
The provided text is a 510(k) summary for the UNIPLATE Anterior Cervical Plate System. It describes the device, its intended use, and indicates that performance data were submitted to characterize the device. However, it does not contain detailed information about specific acceptance criteria or the study data proving the device meets those criteria in the way typically expected for an AI/ML device.
This document is for a medical implant (spinal fixation system), not a diagnostic AI/ML device. Therefore, the questions regarding AI/ML performance metrics like sample sizes for test/training sets, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies are not applicable to this submission.
Here's a breakdown of what can be extracted from the provided text, and where it falls short for an AI/ML device context:
1. A table of acceptance criteria and the reported device performance
The document states: "Performance data were submitted to characterize the subject UNIPLATE Anterior Cervical Plate System components addressed in this notification."
However, it does not provide a table of acceptance criteria or reported device performance in the typical sense of accuracy, sensitivity, specificity, etc., that would be applicable to an AI/ML diagnostic or measurement device. For an implantable device like this, "performance data" likely refers to biomechanical testing (e.g., fatigue, static strength, subsidence testing) to ensure it can withstand the forces in the spine and maintain stability. These specific results and their acceptance criteria are not detailed in this summary.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable and not present as this is a medical implant, not a diagnostic AI/ML device. There is no "test set" in the context of image analysis data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable and not present as this is a medical implant, not a diagnostic AI/ML device. Ground truth as typically defined for AI/ML (e.g., expert consensus on images) is not relevant here.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable and not present as this is a medical implant, not a diagnostic AI/ML device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable and not present as this is a medical implant, not a diagnostic AI/ML device. MRMC studies are used for evaluating diagnostic performance of human readers with and without AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable and not present as this is a medical implant, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
This information is not applicable and not present directly in the provided text. For an implant, "ground truth" might pertain to established biomechanical standards or clinical outcomes from prior similar devices (often addressed through substantial equivalence).
8. The sample size for the training set
This information is not applicable and not present as this is a medical implant, not a machine learning algorithm.
9. How the ground truth for the training set was established
This information is not applicable and not present as this is a medical implant, not a machine learning algorithm.
Ask a specific question about this device
(30 days)
The UNIPLATE Anterior Cervical Plate System is intended for anterior cervical intervertebral body fixation. This system is indicated for patients in which stability is desired following anterior cervical fusion for the indications listed below. The intended levels for treatment range from C2-T1.
Indications include symptomatic cervical spondylosis, trauma, fracture, posttraumatic kyphosis or lordosis, turnor, degenerative disc disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, re-operation for failed fusion, or instability following surgery for the above indications.
The UNIPLATE Anterior Cervical Plate System consists of an assortment of plates and screws. The UNIPLATE Anterior Cervical Plate System also contains Class I manual surgical instruments and cases that are considered exempt from premarket notification. Manufactured from ASTM F 136 implant grade titanium Materials: alloy.
The provided text describes a 510(k) submission for a medical device and does not contain information about a study proving the device meets acceptance criteria or details about AI/algorithm performance. It is a regulatory submission for a spinal implant, which typically relies on mechanical testing and comparison to predicate devices, rather than AI performance studies or human reader studies.
Therefore, I cannot fulfill the request as it asks for information not present in the provided context. The document focuses on regulatory approval based on substantial equivalence and mechanical performance data, not a study evaluating algorithm performance or human reader improvement with AI.
To directly answer the prompt based on the absence of information:
- Table of acceptance criteria and reported device performance: Not applicable. The document refers to "Performance data per ASTM F 1717" being submitted to characterize the device but does not provide specific acceptance criteria or detailed results of those tests.
- Sample size used for the test set and the data provenance: Not applicable. This document is for a spinal implant, not an AI/algorithm-based device. "Test set" in this context would refer to mechanical testing of the implant, not a data set for an algorithm. Data provenance is not mentioned.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically defined for AI/algorithm studies (e.g., expert consensus on medical images) is not relevant to the regulatory submission for this physical medical device.
- Adjudication method: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, an MRMC study was not done, as this is a physical medical device (spinal plate) and not an AI or imaging diagnosis device.
- If a standalone (i.e. algorithm only without human-in-the loop performance) was done: No, a standalone algorithm performance study was not done, as this is a physical medical device.
- The type of ground truth used: Not applicable in the context of AI/algorithm ground truth. The "ground truth" for this device would be its mechanical properties, established through standardized ASTM testing.
- The sample size for the training set: Not applicable, as this is not an AI/algorithm device.
- How the ground truth for the training set was established: Not applicable, as this is not an AI/algorithm device.
Ask a specific question about this device
(64 days)
The Uniplate Anterior Cervical Plate System is intended for anterior cervical intervertebral body fixation. This system is indicated for patients in which stability is desired following anterior cervical fusion for the indications listed below. The intended levels for treatment range from C2 to T1.
Indication includes symptomatic cervical spondylosis, trauma, fracture, post-traumatic kyphosis or lordosis, tumor, degenerative disc disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, re-operation for failed fusion, or instability following surgery for the above indications.
Uniplate Anterior Cervical Plate System consists of an assortment of plates and screws.
The Uniplate Anterior Cervical Plate System also contains Class 1 manual surgical instruments and cases that are considered exempt from premarket notification.
This 510(k) summary describes a spinal implant device, not a device that measures or reports performance metrics like accuracy, sensitivity, or specificity. Therefore, the information requested in the prompt (acceptance criteria, study details, human reader improvement, etc.) is not applicable to this submission.
The document focuses on demonstrating substantial equivalence to a predicate device based on design, materials, and intended use, rather than on performance metrics obtained from human subject studies or algorithmic evaluations.
Here's why the specific questions cannot be answered from the provided text:
- No reported performance data: The "PERFORMANCE DATA:" section states "Performance data were submitted to characterize the Uniplate Anterior Cervical Plate System" but does not provide any specific metrics or results. This is common for mechanical devices where testing often involves benchtop mechanical evaluations rather than clinical performance metrics.
- No AI/Algorithm component: The device is a physical spinal implant (plates and screws). It does not involve any artificial intelligence, algorithms, or image interpretation components that would require a test set, ground truth, expert consensus, or MRMC studies.
- Nature of the device: This is a Class II medical device, specifically a "Spinal Intervertebral Body Fixation Orthosis." The regulatory pathway for such devices often involves demonstrating mechanical performance and biocompatibility, along with substantial equivalence to a legally marketed predicate device, rather than diagnostic accuracy.
Therefore, a table of acceptance criteria and reported device performance with the requested details cannot be generated from the given text.
Ask a specific question about this device
Page 1 of 1