510(k) Data Aggregation

    K Number
    K142466
    Date Cleared
    2014-10-03

    (31 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    UGEO PT60A Diagnostic Ultrasound System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UGEO PT60A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

    Device Description

    The UGEO PT60A is a general purpose, hand-held, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode. M mode, Color Doppler imaging, Power Doppler imaging, PW Spectral Doppler mode, Harmonic imaging or as a combination of these modes. The UGEO PT60A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The UGEO PT60A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the UGEO PT60A Diagnostic Ultrasound System. It aims to demonstrate substantial equivalence to previously cleared devices rather than proving the device meets acceptance criteria through clinical studies.

    Therefore, many of the requested sections regarding acceptance criteria and study details are not explicitly present in the provided text. However, based on the information provided, here's what can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are implicitly tied to demonstrating substantial equivalence to its predicate devices by meeting the same safety, effectiveness, and functionality standards. The device performance is generally stated as being comparable to these predicates across various features and clinical applications.

    Acceptance Criteria (Implied)Reported Device Performance
    Safety (e.g., meeting medical device safety standards)Conforms to UL 60601-1, CSA C22.2 No. 601.1, IEC60601-2-37, EN/IEC60601-1, EN/IEC60601-1-2 (EMC), NEMA UD-2 & UD-3, ISO10993-1 (Biocompatibility), ISO14971 (Risk Management).
    Effectiveness (e.g., comparable diagnostic imaging capabilities)Acquires ultrasound data and displays it as B mode, M mode, Color Doppler, Power Doppler, PW Spectral Doppler, Harmonic imaging, or combinations. Provides measurement and analysis packages.
    Functionality (e.g., presence of key features and clinical applications)Supports clinical applications including Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal, Urology, Cardiac Adult/Pediatric, Peripheral vessel (See detailed tables in original text for specific transducer capabilities). Features like Quick Scan, Spatial Compound Imaging (MultiVision), and ClearVision are present.

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. The submission states: "Summary of Clinical Tests: Not applicable. The subject of this submission, UGEO PT60A, did not require clinical studies to support substantial equivalence." This is a regulatory clearance based on substantial equivalence to existing devices, not a study demonstrating performance on a test set.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable, as no clinical studies were required or conducted to establish new ground truth for this submission.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical studies were required or conducted.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This device is an ultrasound system, not an AI-assisted diagnostic tool. No AI-related MRMC studies were conducted or mentioned.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable, as this is a medical imaging device, not an algorithm, and the submission explicitly states clinical studies were not required.

    7. The Type of Ground Truth Used

    Not applicable. The basis for clearance is substantial equivalence to predicate devices, implying that the established performance and safety of those predicates serve as the "ground truth" for comparison.

    8. The Sample Size for the Training Set

    Not applicable. This is a conventional medical imaging device, not a machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132228
    Date Cleared
    2013-08-20

    (34 days)

    Product Code
    Regulation Number
    892.1560
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    UGEO PT60A DIAGNOSTIC ULTRASOUND SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UGEO PT60A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

    The clinical applications include:Fetal, Abdominal, Small Organs, Adult Cephalic, Muscular-Skeletal (Conventional, Superficial), Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

    Device Description

    The UGEO PT60A is a general purpose, hand-held, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode, M mode, Color Doppler imaging, Power Doppler imaging, PWSpectral Doppler mode, Harmonic imagingor as a combination of these modes. The UGEO PT60A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The UGEO PT60A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    The provided 510(k) Premarket Notification for the UGEO PT60A Diagnostic Ultrasound System does not contain specific acceptance criteria or details of a study proving the device meets acceptance criteria in the way that an AI CADe/CADx device submission would. This document is for a general-purpose diagnostic ultrasound system, and the regulatory pathway relies on demonstrating substantial equivalence to legally marketed predicate devices, rather than establishing specific performance metrics against a defined ground truth through a clinical study.

    Therefore, many of the requested details about acceptance criteria, study design, sample sizes, expert qualifications, and ground truth establishment are not typically found in submissions for general ultrasound devices.

    Here's a breakdown of the information available based on the provided text, and where gaps exist:

    1. Table of Acceptance Criteria and Reported Device Performance

    Not applicable in the context of a general ultrasound system's 510(k) submission.
    The submission focuses on demonstrating substantial equivalence in terms of technological characteristics and safety standards to predicate devices. There are no specific performance metrics (e.g., sensitivity, specificity, AUC) for diagnostic tasks defined as "acceptance criteria" for this type of device in the provided document.

    The comparison table in section {2} and {3} shows the subject device's features and its predicates, highlighting congruency rather than meeting specific quantifiable performance targets.

    • Feature/Characteristic | UGEO PT60A (Subject Device) | Predicate Device (UGEO H60 / MySono U6 / SonoSite Maxx Series) | Equivalence Claim
    • Indications for Use | Matches predicate devices for most applications (Fetal, Abdominal, Small Organ, Adult Cephalic, Muscular-Skeletal, Cardiac Adult/Pediatric, Peripheral vessel). | Substantially equivalent in scope of indications.
    • Scanhead Types | Linear Array, Phased Array | Matches predicate devices.
    • Modes of Operation | B-mode, M-mode, PW Doppler, CW Doppler, Color Doppler, Power Amplitude Doppler, Tissue Harmonic Imaging, 3D/4D imaging, Combined modes | Matches predicate devices.
    • Safety & EMC Compliance | IEC60601-1, IEC 60601-2-37, IEC 60601-1-2, NEMA UD-2, NEMA UD-3, ISO10993-1, ISO14971 | Conforms to the same safety and EMC standards as predicate devices.
    • Functionality | Quick Scan, Spatial Compound Imaging, SMDR (Dynamic MR Plus), Auto IMT, Needle Mate | Substantially equivalent; some features are present in the subject device and some predicates.

    2. Sample size used for the test set and the data provenance

    Not applicable. The submission states in section {3}: "Summary of Clinical Tests: Not applicable. The subject of this submission, UGEO PT60A, did not require clinical studies to support substantial equivalence." This means there was no "test set" in the context of a clinical performance evaluation. The device was evaluated against existing standards and predicate devices (as described under 'Technological Characteristics').

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No clinical studies requiring ground truth establishment were conducted for this submission.

    4. Adjudication method for the test set

    Not applicable. No clinical studies or test sets with ground truth requiring adjudication were conducted.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The UGEO PT60A is a diagnostic ultrasound system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The UGEO PT60A is a diagnostic ultrasound system, not an algorithm being evaluated for standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable. No clinical studies requiring ground truth were conducted for this device. The regulatory pathway relied on demonstrating that the device's technical specifications, safety, and intended use were substantially equivalent to existing, legally marketed ultrasound systems.

    8. The sample size for the training set

    Not applicable. This device is a general-purpose ultrasound system, not an AI or machine learning algorithm that requires a training set in this context.

    9. How the ground truth for the training set was established

    Not applicable. As above, there was no training set for an AI algorithm.