510(k) Data Aggregation

    K Number
    K242860
    Date Cleared
    2025-01-17

    (119 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Type 5 Integrating Indicator for Steam (CSPN-15)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The True Indicating Type 5 Integrating Indicator for Steam CSPN-15 is designed to chemically react over time with the critical parameters of steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:

    Gravity Displacement: 121° C for 30 minutes, 132° C for 15 minutes, 132° C for 25 minutes, 132° C for 3 minutes. 135° C for 10 minutes;

    Dynamic Air Removal (Vacuum Assist): 132° C for 4 minutes, 135° C for 3 minutes.

    Minimum Stated Values as determined in a resistometer:

    SV121°C/16.5 min., SV132°C/2.0min. SV134°C/1.4 min. SV135°C/1.2 min.

    Device Description

    The True Indicating Chemical Indicator for Steam CSPN-15 consists of a white polyester substrate with acrylic adhesive married to a paper substrate. The paper substrate (with acrylic adhesive) has True Indicating water-based steam indicating ink, ISPN-01 (pink to dark brown/black), and general artwork printed. After printing, the top layer of the Type 5 Integrating Indicator strip is covered with a clear colorless polyester laminate with acrylic adhesive. The Integrating Indicators have been designed for Steam sterilization processes, ensuring the effectiveness of sterilization by monitoring all critical process variables (time, temperature and steam). The pink indicating ink was developed to turn to dark and conceal the word "NOT" when a theoretical spore population reaches its kill time, indicating integration condition has been reached. This condition is calibrated with the kill time of a 105 Geobacillus stearothermophilus spore population, calculated in BIER (Biological Indicator Evaluator Resistometer).

    AI/ML Overview

    The provided document describes the acceptance criteria and study results for the True Indicating Type 5 Integrating Indicator for Steam (CSPN-15). This device is a chemical indicator designed to monitor steam sterilization cycles.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Name of TestAcceptance CriteriaSubject Device Result
    Steam Resistometer TestingPass result (end signal color achieved for all product codes) at the value for each temperature claimed:
    121°C for 16.5 minutes
    132°C for 2.0 minutes
    134°C for 1.4 minutes
    135°C for 1.2 minutes
    Fail result (end signal color not achieved) at the value for each temperature claimed:
    120°C for 14.025 minutes
    131°C for 1.7 minutes
    133°C for 1.19 minutes
    134°C for 1.02 minutes
    Integrator temperature coefficient: 10 - 27 °C
    Correlation coefficient: ≥ 0.9PASS
    Hospital Steam Sterilizer TestingPass result (signal color achieved for all product codes) at the value for each temperature claimed in combination of a sterilization load:
    Gravity Displacement: 121° C for 30 minutes, 132° C for 15 minutes, 132° C for 25 minutes, 132° C for 3 minutes, 135° C for 10 minutes;
    Dynamic Air Removal (Vacuum Assist): 132° C for 4 minutes, 135° C for 3 minutes.PASS
    Dry Heat TestingFail result when exposed to dry heat alone for 30 minutes (±1 minute) at 140°C (±2°C)PASS
    End Point Stability6 MonthsPASS
    Side-by-Side Testing with a Biological IndicatorChemical Integrator does not reach endpoint before BIs are inactivated.PASS
    Offset/TransferenceThe chemical indicators shall not offset or bleed, penetrate the substrate to which it is applied, or materials in which it is in contact before, during or after the sterilization cycles for which it is designed.PASS

    2. Sample size used for the test set and the data provenance

    The document states that "Testing was performed for steam sterilization processes using multiple lots of True Indicating Chemical Indicator for Steam CSPN-15 over the range of the shelf-life". Specific numerical sample sizes for each test are not explicitly provided in this summary. The data provenance is derived from nonclinical tests conducted by or for True Indicating, LLC, following FDA and ISO guidelines. The context suggests prospective testing specific to the device's development and validation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This device is not an AI/imaging device requiring expert interpretation for ground truth. The ground truth for chemical indicators is established by objective physical and chemical measurements (e.g., specific temperatures, times, and presence/absence of steam for sterilization cycles, and color change observation). There is no mention of human experts being used to establish ground truth for this type of device.

    4. Adjudication method for the test set

    Not applicable. This device is a chemical indicator, and its performance is assessed by objective chemical/physical reactions and visual color changes, rather than subjective interpretation requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, in a sense. The testing performed on the chemical indicator (Type 5 Integrating Indicator for Steam CSPN-15) is a standalone evaluation of its performance without human-in-the-loop interaction for interpretation, except for visual confirmation of the color change. The indicator itself is designed to provide a direct visual output.

    7. The type of ground truth used

    The ground truth used for this chemical indicator is based on:

    • Physical and Chemical Parameters: Precisely controlled steam sterilization conditions (temperature, time, and steam presence) established in resistometers and hospital sterilizers.
    • Biological Indicator Inactivation: For the "Side-by-Side Testing with a Biological Indicator," the ground truth for successful sterilization is the inactivation of biological indicators (BIs), which contain a known population of resistant bacterial spores. The chemical indicator's performance is compared against this biological standard.
    • Color Change: The specific, objectively defined color change from pink to dark brown/black that indicates successful exposure to sterilization parameters.

    8. The sample size for the training set

    Not applicable. This is a chemical indicator, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As noted above, this device does not utilize a training set in the way an AI model would.