Search Results
Found 5 results
510(k) Data Aggregation
(169 days)
CMO8 Adult PMP Oxygenator with Tubing Pack (Model: CME40009)
The CMO8 Adult PMP Oxygenator is a hollow fiber membrane oxygenator intended for physiologic gas exchange in adults and small adults undergoing cardiopulmonary bypass surgery. The integrated heat exchanger makes it possible to regulate the blood temperature. The CMO8 Adult PMP Oxygenator is intended for up to 6 hours. The Tubing Pack is intended for use in extracorporeal crculation during cardiopulmonary bypass procedures, lasting 6 hours or less.
The CMO8 Adult PMP oxygenator (CMO8) is a hollow fiber membrane oxygenator with an integrated heat exchanger. The CMO8 facilitates the gas exchange into and out of the blood, and the regulation of blood temperature during cardiopulmonary bypass. The CMO8 is supplied with a Rheopak surface coating that reduces platelet adhesion to coated surfaces. The qas exchanger part of the CMO8 is formed of plasma tight hollow fiber membranes. The gas flow takes place through the inner lumen of the fibers. Blood is in contact with the outer side of the membranes, so that oxygen can diffuse into the venous blood, while carbon dioxide diffuses out of the blood. The heat exchanging part of the CMO8 is made of non-porous hollow fiber membranes. Water flows through the inner lumen of the fibers, so that blood temperature flowing outside is regulated. The Tubing Pack, which contains blood lines, gas lines, a venous reservoir and the CMO8 Oxygenator, is a component of a cardiopulmonary bypass procedure that facilitates extracorporeal circulatory support for a term of up to 6 hours.
The provided text is a 510(k) summary for a medical device (CMO8 Adult PMP Oxygenator with Tubing Pack) and not a report or study describing the acceptance criteria or performance of an AI/ML powered device.
Therefore, I cannot extract the requested information regarding acceptance criteria and a study proving an AI/ML device meets them from this document. The document primarily discusses the substantial equivalence of a physical medical device (an oxygenator) to predicate devices through in vitro and non-clinical testing.
The questions you asked are highly relevant to AI/ML device submissions, particularly regarding:
- Acceptance Criteria and Performance Table: This is typically a key component of AI/ML device validation.
- Sample Size and Data Provenance: Crucial for understanding the representativeness and generalizability of AI/ML model testing.
- Expert Ground Truth (Number of Experts, Qualifications, Adjudication): Essential for establishing reliable labels for AI/ML training and testing data.
- MRMC Studies: Often used to evaluate the impact of AI on human reader performance.
- Standalone Performance: Evaluating the AI algorithm's capabilities independently.
- Type of Ground Truth: The gold standard used to assess AI accuracy.
- Training Set Details: Information about the data used to develop the AI model.
To answer your request, I would need a document detailing the clinical validation or performance study for an AI/ML-powered medical device.
Ask a specific question about this device
(93 days)
Paragon Adult Maxi PMP Oxygenator with Tubing Pack
The Paragon Adult Maxi PMP Oxygenator is a hollow fiber membrane oxygenator intended for physiologic gas exchange in adults and small adults undergoing cardiopulmonary bypass surgery. The integrated heat exchanger makes it possible to regulate the blood temperature. The Paragon Adult Maxi PMP Oxygenator is intended for up to 6 hours. The Tubing Pack is intended for use in extracorporeal circulation during cardiopulmonary bypass procedures, lasting 6 hours or less.
The Paragon Adult Maxi PMP oxygenator (Paragon Maxi) with Tubing Pack is a hollow fiber membrane oxygenator with an integrated heat exchanger pre-connected to blood line and gas line tubing. The Paragon Maxi facilitates the gas exchange into and out of the blood, and the requlation of blood temperature during cardiopulmonary bypass. The Paragon Maxi is supplied with a Rheopak surface coating that reduces platelet adhesion to coated surfaces.
The gas exchanger part of the Paragon Maxi is formed of plasma tight hollow fiber membranes. The gas flow takes place through the inner lumen of the fibers. Blood is in contact with the outer side of the membranes, so that oxygen can diffuse into the venous blood, while carbon dioxide diffuses out of the blood. The heat exchanging part of the Paragon Maxi is made of non-porous hollow fiber membranes. Water flows through the inner lumen of the fibers, so that blood temperature flowing outside is regulated.
The Tubing Pack is a component of a cardiopulmonary bypass procedure that facilitates extracorporeal circulatory support for a term of up to 6 hours.
The document provided is a 510(k) summary for a medical device (Paragon Adult Maxi PMP Oxygenator with Tubing Pack) and does not contain information about an AI/ML powered device or study that determines acceptance criteria based on AI/ML performance.
The provided text details the regulatory review of a conventional medical device (oxygenator and tubing pack) and focuses on its substantial equivalence to predicate devices, in vitro and non-clinical test results. There is no mention of acceptance criteria related to a study proving an AI/ML device meets them. Therefore, I cannot extract the requested information from the provided document.
Ask a specific question about this device
(119 days)
Tubing Pack
This product is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
Medtronic Tubing Packs are used in cardiopulmonary bypass procedures for connecting the primary devices of the bypass circuit. Medtronic Tubing Packs consist of coated and uncoated tubing, connectors and various medical devices that are pre-assembled into user-specified extracorporeal cardiopulmonary bypass perfusion circuits. Tubing Packs are intended for use within the extracorporeal blood pathway during cardiopulmonary bypass surgical procedures. Medtronic Tubing Packs are currently packaged into qualified clear tray and sterile packaging systems. This submission is to add additional tubing and connectors for use in Tubing Pack and is changing as part of continual process improvement efforts. The component additions include tubing and connectors that may be coated or uncoated. The coating types include: Cortiva, Trillium and Balance and these three types of coatings are identical to the cleared existing components for Tubing Packs (K891687, K012538 and K122811). The components being added do not have any changes in materials, technology or labeling than previously cleared components for Tubing Packs. These additional components are representative of currently cleared components and do not add any additional risk to patients therefore, Medtronic believes these components may be added to the Tubing Packs. All packaging components remain unchanged (peel pouches, ties, bands, trays, lids and shipper). There is no change to the sterile barrier, sterilization and test methods used for these components.
This looks like information for a regulatory submission for a medical device. Based on the provided text, here's a breakdown of the acceptance criteria and the study conducted to prove the device meets those criteria:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Performance Requirements) | Reported Device Performance |
---|---|
Pressure Integrity | Met (Verification testing demonstrated) |
Pressure Decay | Met (Verification testing demonstrated) |
Spallation (Tubing) | Met (Verification testing demonstrated) |
Kink Resistance (Tubing) | Met (Verification testing demonstrated) |
Pull Force | Met (Verification testing demonstrated) |
Tubing Life | Met (Verification testing demonstrated) |
Biocompatibility | Met (Verification testing demonstrated) |
Mechanical Requirements | Met (Rationalization documentation completed) |
Coating Requirements | Met (Rationalization documentation completed) |
Design Verification for Non-stacked conditioning and Pressure Integrity | Met (Rationalization documentation completed) |
2. Sample Size and Data Provenance
The document does not explicitly state the sample size used for the test set in quantitative terms (e.g., number of units tested). Instead, it mentions that "worse case testing for tubing and connectors" was performed and "verification results were leveraged based on most challenging components." This suggests that a subset of components representing the most challenging scenarios was selected for testing.
The data provenance is from Medtronic, Inc.'s internal testing. It is prospective in the sense that the testing was conducted specifically for this submission to verify the performance of the added components. The country of origin of the data is implicitly the United States, as Medtronic is based there and submitting to the FDA.
3. Number of Experts and Qualifications for Ground Truth
This device (Medtronic Tubing Pack) is a Class II medical device, specifically "Cardiopulmonary Bypass Vascular Catheter, Cannula, or Tubing." The "ground truth" in this context refers to established engineering and biocompatibility standards and performance requirements for such medical components, rather than clinical interpretation by medical experts (like radiologists for imaging devices).
The document does not mention the use of experts to establish a "ground truth" for a test set in the way one would for an AI-powered diagnostic device. The "ground truth" for this device's performance is derived from established industry standards, internal product specifications, and regulatory requirements that define acceptable limits for pressure integrity, biocompatibility, etc. The document refers to "product specification" and "rationalization documentation" as the basis for evaluation.
4. Adjudication Method
An adjudication method (like 2+1, 3+1) is typically used in clinical studies where there's a need to resolve discrepancies in expert opinions on imaging or clinical assessments. For this type of device (tubing and connectors), the "adjudication" is based on objective measurement against predefined specifications and standards. There is no mention of human adjudication in the traditional sense for reconciling expert interpretations for a test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. MRMC studies are relevant for devices that involve human interpretation, particularly in diagnostic imaging, to assess improvements in reader performance with AI assistance. This device is a passive component for cardiopulmonary bypass and does not involve human interpretation in a diagnostic capacity or AI assistance.
6. Standalone Performance Study
The studies described are essentially standalone performance studies for the device components. The verification and validation testing assesses the "algorithm only" (if one considers the manufacturing process and design as the "algorithm" for a physical device) against predetermined performance specifications. The tests for pressure integrity, spallation, kink resistance, pull force, tubing life, and biocompatibility are all evaluations of the device's inherent performance characteristics without human interaction being part of the primary performance metric.
7. Type of Ground Truth Used
The ground truth used is primarily engineering specifications, material science standards, and biocompatibility standards, as outlined in the "product specification" and various "rationalization documentation" (e.g., Mechanical Requirements, Coating Requirements, Biocompatibility per product specifications). Essentially, the "ground truth" is established industry and regulatory benchmarks for the safety and efficacy of cardiopulmonary bypass components.
8. Sample Size for the Training Set
This document does not mention a "training set" as this is not an AI/machine learning device. The "training" for a physical medical device typically refers to the design, manufacturing process, and quality control procedures, which are developed and refined over time based on engineering principles and previous device iterations. There isn't a "training data" set in the context of an algorithm.
9. How Ground Truth for the Training Set Was Established
As there is no "training set" in the AI/machine learning sense, there is no ground truth established for it. The development and refinement of the device's design and manufacturing processes (analogous to "training" for a physical product) would be based on:
- Engineering design principles
- Material science knowledge
- Previous iterations of similar devices (predicates)
- User feedback
- Regulatory requirements and standards for medical devices.
Ask a specific question about this device
(30 days)
Tubing Pack
This product is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
Medtronic Tubing Packs are used in cardiopulmonary bypass for connecting the primary devices of the bypass circuit. The functionality and intended use of these devices are the same as those for the coated (Cortiva - Coated Tubing Packs(K891687), Trillium BioPassive Surface Coated Tubing Packs (K012538), and Balance BioSurface Coated Tubing Packs(K122811) and uncoated tubing, connectors, and accessories that are FDA cleared.
Medtronic Tubing Packs consist of coated and uncoated tubing, connectors and various medical devices that are pre-assembled into user-specified extracorporeal cardiopulmonary bypass perfusion circuits. Tubing Packs are intended for use within the extracorporeal blood pathway during cardiopulmonary bypass surgical procedures.
The provided text describes a 510(k) premarket notification for a medical device called "Tubing Pack" by Medtronic, Inc. This submission focuses on a packaging change and does not involve the development or evaluation of an AI/ML powered device. Therefore, many of the requested elements for describing the acceptance criteria and study proving an AI device meets them are not applicable to this document.
However, I can extract and present the information that is relevant to the device and its testing as described in the document, even though it's not specific to AI.
Here's a breakdown based on the information available:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state quantitative "acceptance criteria" in a table format with corresponding "reported device performance" in the context of an AI/ML model's output metrics (e.g., sensitivity, specificity, AUC). Instead, it discusses demonstrating equivalence to a predicate device through various tests, implying that the acceptance criterion is successful completion of these tests without adverse outcomes.
Test Category | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Process Characterization | Adequacy of new tray manufacturing process | Successfully characterized process |
Shelf Life | Maintenance of product integrity and sterility over time | Successfully demonstrated shelf life |
Process Validation | New tray manufacturing process consistency and reliability | Successfully validated process |
Packaging Performance | Integrity of packaging against relevant standards | Successfully met packaging performance requirements |
Sterilization | Achieved required sterility assurance level (SAL) | Successfully sterilized (Ethylene oxide sterilization methods unchanged) |
Biocompatibility | No adverse biological reactions caused by the new material | Successfully passed biocompatibility testing |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not provide details on sample sizes for these tests, nor does it specify data provenance in terms of country of origin or retrospective/prospective nature. These kinds of details are typically provided for clinical studies involving patient data, not for engineering and material testing of packaging components.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. The tests performed are engineering and material science evaluations, not clinical assessments requiring expert human interpretation of data for ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are typically use for resolving discrepancies in expert interpretations in clinical studies.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device, and no MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML device.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
The "ground truth" for these tests would be established through scientific and engineering standards and validated test methods. For example:
- Shelf Life: Stability of the product over time, measured through established degradation testing.
- Sterilization: Achievement of a specified Sterility Assurance Level (SAL) confirmed by validated biological indicators and methods.
- Biocompatibility: Absence of adverse biological reactions as per ISO 10993 standards and associated testing.
- Packaging Performance: Adherence to ISO or ASTM packaging integrity test standards.
8. The sample size for the training set
Not applicable. There is no AI/ML model involved, and thus no training set.
9. How the ground truth for the training set was established
Not applicable. There is no AI/ML model involved.
In summary, the provided document details a regulatory submission for a packaging change to a medical device. It thoroughly outlines the scope of the change (material, supplier, mold, and slight dimension changes to the tray) and the types of tests performed to ensure that these changes do not negatively impact the device's safety and effectiveness, and that it remains substantially equivalent to its predicate. However, it does not involve an AI/ML component and therefore lacks the information requested for AI device validation.
Ask a specific question about this device
(246 days)
BIOTEQUE 3 IN 1 HEMODIALYSIS BLOOD TUBING PACK
The BIOTEQ® 3 in 1 Hemodialysis Blood Tubing Packs ( BT-102DA · BT-102DB · BT-102DC · BT-102DD) are sets of single-use disposable components intended to provide extracorporeal access to blood of patients suffering from end stage renal disease during Hemodialysis therapy. Each set will include an arterial/venous bloodline set, and any combination of two or three component of the following items: An intravenous administration set for (use in administration of intravenous fluids to a dialysis set in conjunction with hemodialysis procedures.) , AV fistula needle set to (apply on the access site of patient vessel to draw the blood flow adequate to pass through the dialyer and reinfusion of dialysed blood back to patient vessel via the A.V. Fistula Needle Set during hemodialysis procedures.) , and/or A transducer protector for (use as protective devices for pressure monitors and to help protect the sterility of the fluid pathway).
The BIOTEQ® 3 in 1 Hemodialysis Blood Tubing Packs (BT-102DA 、BT-102DB、BT-102DC、BT-102DD) are sets of single-use disposable components intended to provide extracorporeal access to blood of patients suffering from end stage renal disease during Hemodialysis therapy. Each set will include an arterial/venous bloodline set, and any combination of two or three component of the following items: An intravenous administration set for use in administration of intravenous fluids to a dialysis set in conjunction with hemodialysis procedures. , AV fistula needle set to apply on the access site of patient vessel to draw the blood flow adequate to pass through the dialyer and reinfusion of dialysed blood back to patient vessel via the A.V. Fistula Needle Set during hemodialysis procedures. , and/or A transducer protector for use as protective devices for pressure monitors and to help protect the sterility of the fluid pathway. BIOTEQ® 3 in 1 Hemodialysis Blood Tubing Packs includes 4 models (BT-102DA 、BT-102DB、BT-102DC、BT-102DD)
This document describes a 510(k) premarket notification for a medical device and, as such, focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and reported device performance in the way requested for AI/diagnostic devices. The information provided heavily emphasizes physical, chemical, biological, and sterilization specifications rather than clinical performance metrics often associated with "device performance" in the context of diagnostic or AI-driven tools.
Therefore, the requested table of acceptance criteria and reported device performance, and several other points (2, 3, 4, 5, 6, 7, 8, 9), cannot be directly extracted or are not applicable from the provided text in the way they would be for a typical AI/diagnostic device study. The text primarily focuses on conformity to standards and substantial equivalence.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Applicable Standards/Specifications) | Reported Device Performance (as stated in the document) |
---|---|
Physical Specification | Conforms to applicable standards |
Chemical Specification | Conforms to applicable standards |
Biological Specification | Conforms to applicable standards (including ISO 10993 series, USP Pyrogenic standards) |
Sterilization Specification | Conforms to applicable standards (including ISO 11607-1, ISO 11135) |
Substantial Equivalence to Predicate Device | Bench testing demonstrates that differences in technological characteristics do not raise new questions of safety or effectiveness, making it substantially equivalent to Nipro Blood Tubing Set for Hemodialysis with Transducer Protectors and Priming set (K972206) & (K001465). |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document mentions "bench testing," but it does not specify the sample size used for these tests. Data provenance (country of origin, retrospective/prospective) is not provided.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a blood tubing pack, and its evaluation relies on engineering, material, and sterility testing against established standards, not expert interpretation of outputs like an AI diagnostic tool.
4. Adjudication method for the test set
Not applicable. The evaluation is based on conformance to engineering and safety standards, not a diagnostic interpretation process requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI or diagnostic device that involves human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable, as this is not an AI algorithm. The performance evaluation focuses on the physical and biological characteristics of the blood tubing pack itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device is based on established international and national standards for medical device safety, material compatibility, and sterilization (e.g., ISO 10993 series for biological evaluation, ISO 11607-1 for packaging, ISO 11135 for sterilization, USP Pyrogenic standards).
8. The sample size for the training set
Not applicable. This is not an AI/machine learning device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This is not an AI/machine learning device.
Ask a specific question about this device
Page 1 of 1