510(k) Data Aggregation
(265 days)
The Heart Seat is a replacement for a standard toilet seat that is indicated for use in a home environment. The Heart Seat is intended to be used for measuring, reviewing and storing non-invasive functional oxygen saturation of arterial hemoglobin (SpO2) and heart rate (HR) in adults of at least 22 years of age with weight ranging from 90 to 350 pounds. Data from the Heart Seat are collected whenever the seat is used and are automatically uploaded to the Casana Cloud where they can be viewed by the healthcare provider. The Heart Seat is not intended for continuous monitoring.
The Heart Seat™ is a prescription use remote monitoring system built into a toilet seat. It is intended for physiological monitoring in the home setting. The device and platform are tools intended to support clinicians by providing them with data to help them better manage patients. Monitored users (or Patients) sit on the seat for their data to be captured and sent to the cloud. Clinical users interact with the clinical cloud-based application that provides the patient measurements.
Here's a summary of the acceptance criteria and the studies that prove the device meets those criteria, based on the provided text:
Acceptance Criteria and Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
SpO2 Measurement Accuracy (ARMS) | Acceptance Criteria: Not explicitly stated as a numerical value in the "Performance Data" section, but the predicate device had: |
±2% (90-100%) | |
±4% (70-89%). However, the comparison table states that the proposed device is ±3.5% (70-100%) and that "clinical data demonstrates equivalent performance" to the predicate. This implies the device aims to meet or be comparable to the predicate's accuracy. | Reported Device Performance: ARMS of 2.7% over the range of 70-100%. (This meets or exceeds the implicit expectation of equivalence to the predicate's ±3.5% (70-100%) as listed for the proposed device in the comparison table.) |
Heart Rate Measurement Accuracy | Acceptance Criteria: Absolute Accuracy |