Search Results
Found 1 results
510(k) Data Aggregation
(86 days)
Tango Neo, Tango Reflex Neo, Ultra Q Reflex Neo
In the YAG mode (Tango Neo, Tango Reflex Neo, Ultra Q Reflex Neo): -Iridotomy and iridectomy. -Posterior capsulotomy. -Posterior membranectomy.
In the SLT mode (Tango Neo, Tango Reflex Neo) : -Selective Laser Trabeculoplasty (SLT)
These devices are slit lamp based surgical laser instruments designed for use by ophthalmologists in clinics or an outpatient facility in a hospital or surgery. These systems are used to perform both Photodisruption procedures such as posterior capsulotomies, iridotomies, posterior membranectomies in the photodisruptor (YAG mode) and Selective Laser Trabeculoplasty (SLT mode) for the treatment of chronic open angle glaucoma.
All these devices consist of the below main parts: Delivery Head, Microscope and Slit lamp Illumination, Console and Table Top, Joystick fire button, Total solution table, Display unit.
This FDA 510(k) summary does not contain the information required to populate a table of acceptance criteria and reported device performance related to clinical effectiveness. This submission is for an ophthalmic laser system (Tango Neo, Tango Reflex Neo, Ultra Q Reflex Neo) and the FDA has determined that clinical data is not required because the new devices use the same technology and principles as a previously cleared predicate device (Ellex YAG/SLT, K212630).
Therefore, many of the requested fields cannot be filled. However, based on the non-clinical testing performed, we can infer some "acceptance criteria" related to safety and performance specifications, though not in the typical format of clinical outcomes.
Here's a summary of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
As clinical data was not required, there are no specific clinical acceptance criteria or reported clinical device performance metrics in this summary. The "acceptance criteria" are implied by compliance with various safety and performance standards for medical devices and lasers.
Category | Acceptance Criteria (Implied by Standards Compliance) | Reported Device Performance (Compliance Status) |
---|---|---|
Electrical Safety | Conformance to ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012 | Conforms |
EMC (Electromagnetic Compatibility) | Conformance to IEC 60601-1-2:2014 | Conforms |
Usability | Conformance to IEC 60601-1-6:2010+A1:2013 | Conforms |
Alarm Systems | Conformance to IEC 60601-1-8: 2006 (2nd Ed) + A1:2012 | Conforms |
Surgical Laser Safety | Conformance to IEC 60601-2-22: 2007 (Third Edition) + A1:2012 | Conforms |
Laser Product Safety | Conformance to IEC 60825-1 2nd Edition Part I | Conforms |
Biocompatibility | (For chin rest/head rest): Classified as "surface contacting devices in contact with intact skin" with cumulative use up to 24 hours. Existing predicate clearance of slit lamps applies. | Not explicitly stated as "conforming" for the new device, but deemed acceptable by leveraging predicate slit lamp clearances. |
Software V&V | Fulfillment of software requirement specifications for a device deemed "Major" level of concern. | Confirmed |
Risk Analysis | Device is safe for intended use and does not pose unacceptable risks (ISO-14971 compliance). | Conclusion: Safe for intended use |
Technological Equivalence (Key Parameters) | Same wavelength, laser class, pulse duration, energy, pulse setting, repetition rate, spot size for treatment beam as predicate. Same aiming beam safety class, wavelength, power. | Confirmed (stated as "same") |
Technological Equivalence (Components) | Same slit lamp microscope, magnification settings, working length, controls, illumination mirror, chinrest, joystick, laser firing mechanism as predicate. | Confirmed (stated as "same") |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided. As per Section IX, "clinical data is not required" for the substantial equivalence determination. The "tests" mentioned are non-clinical safety and performance tests against standards, not clinical studies with patients.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable, as no clinical ground truth was established for a test set in this submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as no clinical test set requiring adjudication was used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a laser device, not an AI-assisted diagnostic tool, and no MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical laser device, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable, as no clinical ground truth was established for a clinical study in this submission. For the non-clinical tests, the "ground truth" is adherence to international and federal safety and performance standards.
8. The sample size for the training set
Not applicable. No AI/ML component requiring a training set is mentioned.
9. How the ground truth for the training set was established
Not applicable. No AI/ML component requiring a training set is mentioned.
Ask a specific question about this device
Page 1 of 1