Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K141985
    Manufacturer
    Date Cleared
    2014-09-18

    (58 days)

    Product Code
    Regulation Number
    870.1255
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    TRUE BA V BALLOON VALVULOPLASTY CATHETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The True™ BAV Balloon Valvuloplasty Catheter is indicated for balloon aortic valvuloplasty.

    Device Description

    The True™ BAV Balloon Valvuloplasty Catheter is a high performance balloon catheter consisting of an over-the-wire catheter with a balloon fixed at the distal tip. The proprietary, non-compliant, low profile balloon is designed to provide consistent balloon diameters and lengths. Two radiopaque markers delineate the working length of the balloon and aid in balloon placement. The coaxial catheter includes an atraumatic tip to facilitate advancement of the catheter to and through the valve. The over-the-wire catheter is compatible with .035" quidewire and is available in 100 cm working length. The proximal portion of the catheter includes a female luer-lock hub connected to the inflation lumen, and a female luer-lock hub connected to the guidewire lumen. Packaged with every product is a profile reducing sheath that is positioned over the balloon for protection before use. A re-wrapping tool is also provided on the catheter shaft. This product is not manufactured with any latex.

    AI/ML Overview

    The provided text describes a medical device, the True™ BAV Balloon Valvuloplasty Catheter, and its substantial equivalence to predicate devices, rather than a study evaluating an AI/ML device. Therefore, much of the requested information regarding acceptance criteria and studies for AI/ML performance (e.g., sample sizes for test and training sets, expert qualifications, MRMC studies, standalone performance) is not applicable here.

    However, I can extract the information related to device performance and acceptance criteria as presented in the document for this specific medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the "True™ BAV Balloon Valvuloplasty Catheter met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs."

    However, specific numerical acceptance criteria and their corresponding reported device performance values are not provided in the text. The document lists the types of in-vitro tests performed, implying that for each of these, acceptance criteria existed and were met.

    Here's a table based on the provided text, outlining the tests performed, which inherently means they had acceptance criteria that were met, but the specific numerical criteria and results are not detailed.

    Acceptance Criteria Category (Test Performed)Reported Device Performance (Implied)
    Visual InspectionMet predetermined acceptance criteria
    Balloon DistensibilityMet predetermined acceptance criteria
    Tip LengthMet predetermined acceptance criteria
    Balloon Outer Diameter (Diameter and Profile Test)Met predetermined acceptance criteria
    Balloon Working LengthMet predetermined acceptance criteria
    Catheter Shaft Outer DiameterMet predetermined acceptance criteria
    Catheter Shaft Inner DiameterMet predetermined acceptance criteria
    Tip VisibilityMet predetermined acceptance criteria
    Catheter Shaft VisibilityMet predetermined acceptance criteria
    Marker Band VisibilityMet predetermined acceptance criteria
    Tip MorphologyMet predetermined acceptance criteria
    Tip Tensile (Tip Pull, Torque, and Bond Strength Tests)Met predetermined acceptance criteria
    Balloon to Shaft TensileMet predetermined acceptance criteria
    Hub to Shaft TensileMet predetermined acceptance criteria
    Catheter Shaft ElongationMet predetermined acceptance criteria
    Balloon Nominal (Operating) PressureMet predetermined acceptance criteria
    Rated Burst Pressure (Balloon Minimum Burst Strength)Met predetermined acceptance criteria
    Balloon Burst ModeMet predetermined acceptance criteria
    Fatigue (Repeated Balloon Inflation)Met predetermined acceptance criteria
    Catheter Shaft Leaks (Catheter Body Maximum Pressure Test)Met predetermined acceptance criteria
    Catheter Shaft Burst (Catheter Body Maximum Pressure Test)Met predetermined acceptance criteria
    Media Interaction (Balloon Preparation Test)Met predetermined acceptance criteria
    Catheter Shaft LengthMet predetermined acceptance criteria
    TrackabilityMet predetermined acceptance criteria
    Inflation/Deflation Time (Balloon Inflatability Test)Met predetermined acceptance criteria
    Marker Band AlignmentMet predetermined acceptance criteria
    Sheath CompatibilityMet predetermined acceptance criteria
    Equipment InterfaceMet predetermined acceptance criteria
    Visual Inspection of PackagingMet predetermined acceptance criteria
    Dye PenetrationMet predetermined acceptance criteria
    Pouch Tensile StrengthMet predetermined acceptance criteria

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the document. The text describes in-vitro (bench) testing, not clinical studies with patients or data sets in the typical sense for AI/ML validation. There's no mention of sample sizes for individual tests or the provenance of any "data."

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable to this type of device submission. There is no mention of experts or ground truth establishment as it typically refers to for AI/ML diagnostic devices. The acceptance criteria for this device are based on engineering specifications and performance standards through in-vitro testing.

    4. Adjudication Method

    This information is not applicable. Adjudication methods (like 2+1 or 3+1) are relevant for reconciling differing expert opinions, which isn't part of this device's testing and validation process.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not applicable to a physical medical device like a balloon catheter.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm-only performance study was not done. This device is a physical catheter, not an AI algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's validation is based on engineering specifications, established performance standards, and internal risk assessment procedures as demonstrated through the in-vitro tests listed. It's not based on expert consensus, pathology, or outcomes data in the way an AI/ML diagnostic device's ground truth would be.

    8. The Sample Size for the Training Set

    This information is not applicable. The concept of a "training set" applies to machine learning algorithms, which is not what this device is.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1