510(k) Data Aggregation

    K Number
    K032490
    Date Cleared
    2003-08-21

    (9 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    TOSHIBA EXCELART VANTAGE XGV/AGV MRI DIAGNOSTIC SYSTEM, MODELS MRT-1503/P3 AND MRT-1503/P2

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Imaging of:

    • The Whole Body (including head, abdomen, pelvis, limbs and extremities, spine, . neck, TMJ, heart, blood vessels). [Application terms include MRCP (MR Cholanqiopancreatography), MR Cisternography, MR Urography, MR Myelography, MR Fluoroscopy, SAS (Surface Anatomy Scan), Dynamic Scan,Cine Imaging and Cardiac tagging.]
    • Fluid Visualization .
    • 2D / 3D Imaging ●
    • MR Angiography / MR Vascular Imaging .
    • . Blood Oxygenation Level Dependent (BOLD) imaging
    • Perfusion / Diffusion Imaging .
    • Proton Spectroscopy .
    Device Description

    The EXCELART Vantage™ XGV/AGV system has the following features compared to the current EXCELART™ with Pianissimo XG/AG SPIN Edition system.

    • New gantry design as Ultra-short bore. .
    • . New patient couch design as wide couch top.
    • It is possible to combine detachable couch top and gurney of exclusive use. .
    • . The CPU of computer system was changed from RISC type to Xeon.
    • . dB/dt enabled use by 1st operating mode specified in IEC 60601-2-33(2002) by some sequences.
    • . SAR enabled use by normal operating mode specified in IEC 60601-2-33(2002).
    • The marginal performance (Min.TR/Min.TE / Min.Slice thickness / Imaging area) of a . sequence has been improved.
    • The gating waveform can be displayed by the side of the gantry. .
    AI/ML Overview

    This document describes a 510(k) submission for the EXCELART Vantage™ XGV/AGV MRI system. The submission focuses on demonstrating substantial equivalence to a predicate device, the EXCELART™ with Pianissimo XG/AG SPIN Edition.

    Here's an analysis of the provided information concerning acceptance criteria and device performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission primarily focuses on comparing the safety and imaging performance parameters of the new device to its predicate. The acceptance criteria essentially align with maintaining or improving upon the predicate device's specifications, particularly within regulated safety limits.

    ParameterAcceptance Criteria (Predicate Device)Reported Device Performance (New Device)
    Safety Parameters
    Static field strength1.5 TSame (1.5 T)
    Peak & A-weighted acoustic noise95.4 dB (A-weighted)110 dB (A-weighted)
    Operational modes1st operating mode for dB/dt, SAR, dB/dtSame
    Safety parameter displaySameSame
    Operating mode access requirementsNot applicable (used only in normal operating mode)Same
    Maximum SAR