Search Results
Found 2 results
510(k) Data Aggregation
(81 days)
LapBox Power Tissue Containment System
The LapBox Power Tissue Containment System is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue considered benign, resected during single-port or multi-site laparoscopic surgery during power morcellation and removal. The LapBox is compatible with electromechanical laparoscopic power morcellators that are between 15 mm and 18 mm in shaft outer diameter and 135 mm and 180 mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation. When used in women with fibroids, the LapBox Power Tissue Containment System is for women who are pre-menopausal and under age 50.
The LapBox Power Tissue Containment System is a single use sterile device. It is comprised of a double wall inflatable polyurethane chamber which is mounted on an insertion shaft and is provided with two port sizes. Once the shaft is inserted to the abdominal cavity, the chamber is deployed and the organ to be morcellated is placed within the chamber. The chamber is then inflated using an external handpump and the sleeve of the chamber is exteriorized. The selected port is then placed over the sleeve in the incision site and the organ can be morcellated using a power morcellator. Once morcellation is complete, the port is removed, and the chamber is deflated and removed from the patient.
The provided FDA 510(k) clearance letter describes a medical device, the LapBox Power Tissue Containment System, which falls under the regulation of medical devices, not AI/ML-based software. Therefore, the information typically found in a study proving an AI device meets acceptance criteria (such as sample sizes for training/test sets, expert adjudication, MRMC studies, or specific ground truth methodologies for AI models) is not applicable here.
However, I can extract the acceptance criteria and performance data for the LapBox Power Tissue Containment System based on the non-clinical performance testing described in the document.
Overview of Device and Study Context
The LapBox Power Tissue Containment System is a medical device intended for use during minimally invasive gynecologic laparoscopic surgery to contain and isolate tissue during power morcellation. It is a physical device, not an AI/ML software. The provided document details the non-clinical testing performed to demonstrate its safety and effectiveness for 510(k) clearance, asserting substantial equivalence to a predicate device (PneumoLiner).
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a physical medical device and not an AI/ML model, the "acceptance criteria" and "device performance" are focused on engineering, biological, and usability metrics rather than AI-specific performance metrics like sensitivity, specificity, or AUC. The document outlines various tests conducted and confirms whether the device met the prespecified acceptance criteria for each test. The actual quantitative "performance" data (e.g., specific burst pressures, detailed force measurements) are generally not fully presented in a 510(k) summary but rather summarized as meeting criteria.
Acceptance Criterion (Category / Test) | Reported Device Performance (Summary from Document) |
---|---|
Sterilization & Shelf-Life | |
Sterility Assurance Level (SAL) | Validated in accordance with ISO 11135 to a SAL of $10^{-6}$. |
Shelf-Life (18 months) | Validation performed following accelerated aging and simulated transit conditioning. Testing evaluated packaging integrity and device functionality, which met acceptance criteria. |
Biocompatibility | |
Cytotoxicity | Device was non cytotoxic. |
Sensitization | Device was non-sensitizer. |
Intracutaneous Irritation | Device was non-irritant. |
Acute Systemic Toxicity | Device was not systemically toxic. |
Pyrogenicity | Device was non-pyrogenic. |
Non-Clinical Performance Testing (Bench) | |
Bacterial Immersion Test (Impermeability) | Test samples did not show any evidence of growth of Brevundimonas Diminuta after simulated use, disinfection, sterilization, and immersion. (Positive and negative controls performed as intended). |
Dimensional Verification (Initial) | A total of 30 non-aged units tested for proximal handle outer diameter, inner tube inner diameter, outer tube inner diameter, and inflation chamber inner diameter. All samples met the prespecified acceptance criteria. (Supplemental to K221365 tests). |
Pressure Relief Valve Testing | A total of 30 aged and 30 non-aged valves tested to ensure relief above set target. All samples met the prespecified acceptance criteria. |
Design and Performance Validation Test | 30 aged and 27 non-aged units tested by 2 trained surgeons in a simulator with bovine tongue. All samples passed a dye leak test. (Design changes validated in supplemental tests). |
Puncturing Force Comparative Test | Results obtained with LapBox supported non-inferiority when compared to the predicate (72 predicate material samples; 72 aged and 72 non-aged LapBox material samples tested). |
Air Tube Bond Connection & Handpump Assembly (Bond Strength) | 30 aged units tested. All samples met the prespecified acceptance criteria. (Supplemental to K221365 tests). |
Burst Pressure Evaluation (Modified Air Tube) | 10 non-aged units with modified new air tube tested. All samples met the prespecified acceptance criteria. (Supplemental to K221365 tests). |
Inflation Pressure Range (Shape Maintenance) | Customized jig developed. One LapBox unit used. Demonstrated that the chamber will maintain its shape at a pressure below that relieved by the pressure relief valve. |
Comparison Force Test (Outward Force) | Customized jig developed. One predicate and one LapBox utilized. Results demonstrate that a uniform distribution of force by LapBox does not pose a greater risk of compression on adjacent organs when compared to the predicate. |
Corrosion Resistance (K221365) | 30 units of 2 different metal components, double sterilized and 18 months accelerated aged. No corrosion was noted on any samples. |
Burst Pressure Evaluation (K221365) | 30 aged and 30 non-aged units tested for burst pressure of the chamber. All tested units met the prespecified acceptance criteria. |
Bond Strength Tests (K221365) | 30 non-aged units and 30 aged units tested for tensile strength of various bonds. All samples met the prespecified acceptance criteria. |
Dimensional Verification (K221365) | 36 non-aged units tested for shaft effective length, inner sleeve height, chamber double wall height, chamber outer diameter, and chamber opening. All samples met the prespecified acceptance criteria. |
Viral Penetration (ASTM F 1671) (K221365) | 30 material samples with 12 months equivalent accelerated aging. All samples met the prespecified acceptance criteria. |
Maximum Organ Size Testing (K221365) | 2 aged devices utilized. (Specific findings not detailed, but considered in the clearance). |
Training Validation and Usability Studies | |
Initial Usability Study (31 physicians) | 31 physicians (varying levels of experience) performed 40 procedures in a simulator with bovine tongue. No device samples had observed leakage of dye. Minor device issues noted and addressed via manufacturing/labeling changes. |
Supplementary Usability Study (6 physicians) | 6 physicians performed 33 procedures in a simulator. (Units used in bacterial immersion testing). No evidence of any device leak. |
Training Validation (Porcine Model) | 16 physicians (varying experience) completed 32 morcellation procedures. Device samples subjected to dye leak test. No samples leaked. Minor issues with IFU adherence noted, leading to revision of training and IFU. |
Follow-up Usability Study (15 physicians) | 15 physicians performed 30 power morcellation procedures in a simulator. Demonstrated that revisions to training and IFU were effective, ensuring safe and effective use. |
Additional Animal Studies | |
Performance Characterization (In-Vivo) | Part 1: One animal, 3 devices. Part 2: One animal, 5 devices. (Purpose: define performance characteristics; specific quantitative results not detailed in summary, but support clearance). |
2. Sample Sizes Used for the Test Set and the Data Provenance
As this is a physical device, the "test set" refers to the samples of the device itself and the human users/animals involved in performance testing. There isn't "data provenance" in the sense of patient data for AI.
- Bacterial Immersion Test: 32 device samples. Data provenance: Controlled laboratory test.
- Dimensional Verification Tests (Supplemental): 30 non-aged units. Data provenance: Controlled laboratory test.
- Pressure Relief Valve Testing: 30 aged and 30 non-aged units (total 60). Data provenance: Controlled laboratory test.
- Design and Performance Validation Test: 30 aged and 27 non-aged units (total 57). Data provenance: Simulated environment using bovine tongue.
- Puncturing Force Comparative Test: 72 predicate material samples, 72 aged LapBox material samples, and 72 non-aged LapBox material samples (total 216 material samples). Data provenance: Controlled laboratory test.
- Bond Strength Tests (Supplemental): 30 aged units. Data provenance: Controlled laboratory test.
- Burst Pressure Evaluation (Modified Air Tube): 10 non-aged units. Data provenance: Controlled laboratory test.
- Inflation Pressure Range: One LapBox unit. Data provenance: Controlled laboratory test.
- Comparison Force Test: One predicate device and one LapBox device. Data provenance: Controlled laboratory test.
- Corrosion Resistance (from K221365): 30 units of 2 different metal components (total 60). Data provenance: Controlled laboratory test.
- Burst Pressure Evaluation (from K221365): 30 aged and 30 non-aged units (total 60). Data provenance: Controlled laboratory test.
- Bond Strength Tests (from K221365): 30 non-aged units and 30 aged units (total 60). Data provenance: Controlled laboratory test.
- Dimensional Verification (from K221365): 36 non-aged units. Data provenance: Controlled laboratory test.
- Viral Penetration ASTM Method F 1671 (from K221365): 30 material samples. Data provenance: Controlled laboratory test.
- Maximum Organ Size Testing (from K221365): 2 aged devices. Data provenance: Controlled laboratory test.
- Initial Usability Study: 31 physicians, performing a total of 40 procedures. Data provenance: Simulator with bovine tongue.
- Supplementary Usability Study: 6 physicians, performing a total of 33 procedures. Data provenance: Simulator.
- Training Validation: 16 physicians, completing 32 procedures. Data provenance: Porcine model with bovine tongue.
- Follow-up Usability Study: 15 physicians, performing 30 procedures. Data provenance: Simulator.
- Performance Characterization Animal In-Vivo Study: Part 1: 1 animal, 3 devices. Part 2: 1 animal, 5 devices. Data provenance: Porcine model (in-vivo).
Data provenance details (e.g., country of origin, retrospective/prospective) are not directly applicable as these are laboratory, simulator, or animal studies of a physical device, not patient data for an AI model.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
For a physical medical device, "ground truth" often relates to objective measurements (e.g., integrity, dimensions) or expert assessment of functionality and usability.
- Design and Performance Validation Test: 2 trained surgeons. Qualifications not explicitly detailed beyond "trained surgeons."
- Usability Studies (Initial, Supplementary, Training Validation, Follow-up): Physicians of "varying levels of experience" were involved. The exact number of "experts" (or the most senior/experienced) who formally established performance metrics or approved the "ground truth" (i.e., whether a procedure was successful or a leak occurred) is not specified beyond their participation as test users. The methods (e.g., dye leak test) were objective.
4. Adjudication Method for the Test Set
Adjudication methods (like 2+1, 3+1) are typically used for subjectively labeling data in AI ground truth establishment. For this physical device, the "adjudication" is inherent in the test protocols and measurements.
- Dye leak tests: This is an objective visual assessment.
- Dimensional, pressure, strength tests: These rely on objective measurement against prespecified criteria, not human consensus on subjective labels.
- Usability Studies: While surgeons performed the procedures, the outcomes like "observed leakage of dye" or "device leak" are objective observations rather than subjective interpretations requiring adjudication. "Minor device issues noted" would have been observed directly by the testers or study coordinators.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. MRMC studies are specific to evaluating human reader performance (e.g., radiologists interpreting images) with and without AI assistance on complex medical cases. This device is a physical surgical tool, not an AI software tool for interpretation.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
This question is not applicable. The device is a physical product, not an algorithm. Its performance is always in the context of human use (e.g., surgeons performing procedures with it). The "non-clinical performance testing (bench)" and "animal studies" could be considered "standalone" in that they test the device's inherent properties and function, but not in the AI sense of algorithm-only performance.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is established through a combination of:
- Objective Measurement against Engineering Specifications: For tests like dimensional verification, burst pressure, puncturing force, bond strength, pressure relief valve performance, corrosion resistance.
- Controlled Biological/Physical Outcomes: For bacterial immersion (absence of growth) and viral penetration (meeting criteria).
- Expert Observational Assessment of Functionality: In simulated and in-vivo environments, such as dye leak tests after morcellation, maintenance of pneumoperitoneum, adequate space/visualization, and observation of proper deployment and removal without spillage. Usability studies assessed successful completion of tasks and adherence to IFU.
There is no "pathology" or "outcomes data" in the sense of clinical patient outcomes or histopathology for establishing ground truth for individual performance evaluations outlined.
8. The Sample Size for the Training Set
This question is not applicable for a physical device. There is no "training set" in the context of machine learning. The studies were validation tests.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable, as there is no training set for an AI model.
Ask a specific question about this device
(116 days)
TISSUE CONTAINMENT SYSTEM; 10, 14, 17, 25
The Applied Medical Tissue Containment System is indicated to contain and isolate tissue during, or prior to, surgical removal and/or extracorporeal manual morcellation.
The Tissue Containment System is contraindicated for laparoscopic power morcellation during gynecologic procedures.
The Tissue Containment System is contraindicated for use with powered cutting devices (e.g., power morcellators, electrosurgical and laser instruments), and when, in the judgment of the physician, use of such a device would be contrary to the best interest of the patient.
The Applied Medical Tissue Containment System is a flexible tissue bag that includes an attached tether and guard accessory. The subject system is provided sterile. The Tissue Containment System is used to contain and isolate specimens for surgical removal and/or manual morcellation. After the device is fully inserted into the abdominal or pelvic cavity, the mouth of the bag returns to its original, circular shape, facilitating placement of the specimen in the bag. When the specimen is ready for removal and/or manual morcellation, the tether is used to maneuver the ring to the surface of the extraction site. If the specimen requires manual morcellation, the ring may be repeatedly flipped to shorten the bag and consequently bring the specimen closer to the extraction site. The guard is placed in the bag opening prior to manual morcellation, providing a robust barrier between the bag and sharp instruments.
The FDA 510(k) summary for the Applied Medical Tissue Containment System (K142427) provides information about its performance testing. However, it does not include specific acceptance criteria or detailed results in a tabular format as you've requested for direct comparison.
Instead, the document states that "Results of testing demonstrates that the subject Tissue Containment System is substantially equivalent to the predicates... and that the subject device performs comparably to the current marketed device for the same intended use." This indicates that the device met the performance expectations derived from its predicate devices and relevant standards, but the exact quantitative thresholds are not provided in this public summary.
Based on the provided text, here's what can be extracted and inferred regarding your request:
1. A table of acceptance criteria and the reported device performance
The document does not provide a table with specific quantitative acceptance criteria or reported performance values. It generally states that "All materials were found to be biocompatible" and that "bench top tests were designed to focus on the functional performance of the specimen containment and retrieval features, as well as its use during morcellation." and "Results of testing demonstrates that the subject Tissue Containment System is substantially equivalent..."
General interpretation of implicit acceptance criteria based on testing categories:
Acceptance Criteria Category | Reported Device Performance (Summary) |
---|---|
Biocompatibility | All materials were found to be biocompatible. |
Functional Performance | Demonstrated substantial equivalence to predicate devices. |
Bag Seam Integrity | Performed comparably to predicate. |
Tether Break Force | Performed comparably to predicate. |
Puncture Resistance | Performed comparably to predicate. |
Dye Penetration | Successfully demonstrated device functionality. |
Viral Penetration | Successfully demonstrated device functionality. |
Bubble Leak | Successfully demonstrated device functionality. |
Simulated Use | Successfully demonstrated device functionality. |
2. Sample size used for the test set and the data provenance
The document does not specify the sample sizes for any of the tests conducted (biocompatibility or mechanical). It also does not specify the data provenance (e.g., country of origin, retrospective or prospective) for any of the tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable as the device is a medical device for tissue containment, not an AI or diagnostic device that relies on expert interpretation for ground truth establishment. The testing described (biocompatibility and mechanical) uses objective physical measurements and standardized protocols, not human expert consensus for ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable for the type of testing conducted. Adjudication methods are typically used in clinical studies involving interpretation of medical images or patient outcomes, which is not the case here.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study was not done. This device is a physical tissue containment system, not an AI or imaging device, so such a study would not be relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable for a physical medical device. "Standalone" performance refers to algorithm performance, which is not relevant here.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the mechanical and biocompatibility testing is based on:
- Standardized laboratory methods and established criteria for biocompatibility: Referring to ISO 10993-1, this involves objective measures for cytotoxicity, irritation, and sensitization.
- Physical performance metrics and comparison to predicate devices: For mechanical tests, the ground truth is the measurable physical properties of the device (e.g., actual break force, absence of leakage, resistance to puncture) and its ability to perform its intended function in simulated use, as well as its comparability to the performance of predicate devices.
8. The sample size for the training set
This information is not applicable. There is no "training set" as this is not an AI or statistical modeling device.
9. How the ground truth for the training set was established
This information is not applicable as there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1