Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K980014
    Manufacturer
    Date Cleared
    1998-05-15

    (133 days)

    Product Code
    Regulation Number
    880.6880
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    THE SYSTEM 2S STEAM STERILIZER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The System 2S Steam Sterilizer is designed for efficient, point-of-use sterilization of nonporous and porous, heat and moisture-stabile materials used in outpatient surgery, ambulatory care and urgent care centers. The System 2S Steam Sterilizer is available in one configuration: Prevacuum 16"diameter x 26"length chamber size.

    The System 2S Steam Sterilizer is equipped with the following factory-programmed set sterilization cycles and cycle values:

    | CYCLES | RECOMMENDED
    LOAD | STERILIZE
    TEMP. | STERILIZE
    TIME | DRY TIME |
    |----------|-------------------------------------------------------------|--------------------|-------------------|------------|
    | FLASH | Unwrapped
    Instrument Tray with
    single instrument | 270°F (132°C) | 3 minutes | 1 minute |
    | GRAVITY* | Fabric Pack · | 250°F (121°C) | 30 minutes | 15 minutes |
    | PREVAC** | Double Wrapped
    Instrument Tray | 270°F (132°C) | 4 minutes | 30 minutes |
    | EXPRESS | Single Wrapped
    Instrument Tray with
    single instrument | 270°F (132°C) | 4 minutes | 3 minutes |

    *250°F Cycle can be adjusted from factory-set value of 15 minutes dry time to 30 minutes dry time for instrument trays.

    ** 270°F Cycle can be adjusted from factory-set value of 30 minutes dry time to 5 minutes dry time for single fabric pack.

    Device Description

    The System 2S Steam Sterilizer is a Class II medical device as defined by 21 CFR Part 880.6880. The System 2S Steam Sterilizer is a sliding door sterilizer intended for the terminal sterilization of surgical instruments, various hospital utensils and other surgical materials and supplies to be used for human diagnostic or therapeutic procedures.

    The System 2S Steam Sterilizer cycle descriptions are as follows:

    • FLASH Cycle is provided for quick turnaround processing of nonporous goods (e.g. dropped instruments) at 270°F (132°C) for 3 minutes sterilize and 1 minute dry;
    • GRAVITY Cycle is provided for the sterilization of heat and moisture stabile goods at 250°F (121°C) for 30 minutes sterilize and 15 minutes dry;
    • PREVAC Cycle is provided for the sterilization of heat and moisture stabile goods at 270°F (132°C) for 4 minutes sterilize and 30 minutes dry;
    • EXPRESS Cycle is offered for rapid processing of single-wrapped instrument trays at 270°F (132°C) for 4 minutes sterilize and 3 minutes dry;
    • DART (Bowie-Dick Test) Cycle is provided for verification of effective removal of residual air in the chamber and load during testing. This cycle is pre-programmed and fixed; and
    • LEAK Test Cycle is provided for verification of door seal and piping system integrity and is pre-programmed and fixed.
    AI/ML Overview

    Here's an analysis of the provided text regarding the STERIS System 2S Steam Sterilizer, focusing on acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Effectiveness: Complete kill of Bacillus stearothermophilus spores and a sterility assurance level (SAL) of
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1