Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K092175
    Manufacturer
    Date Cleared
    2010-02-17

    (211 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    THE CROSSER S6 CATHETER, THE CROSSER SYSTEM ELECTRONICS, MODEL CRUS6, GEN200

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CROSSER S6 System is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions. The device is contraindicated for use in carotid arteries.

    The CROSSER S6 Catheter is only intended for use with the CROSSER Electronics System. Refer to the CROSSER Electronics System Manual of Operations for proper use.

    Device Description

    The CROSSER S6 System consists of a re-usable electronic Generator, Foot Switch, high-frequency Transducer, and single-use CROSSER S6 Catheter.

    Each Catheter is connected to the electronic Generator through the Transducer. The Foot Switch is used to activate the system. The Generator and Transducer convert AC power into high-frequency mechanical vibrations which are propagated through a Nitinol core wire to the stainless steel tip of the CROSSER Catheter. The main body of the catheter is constructed from Pebax and a hydrophilic coating which covers the distal end of the catheter.

    AI/ML Overview

    This document describes the FlowCardia CROSSER S6 System, a medical device designed to assist in navigating guidewires beyond chronic total occlusions in peripheral arteries. The provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than presenting a standalone clinical study with detailed acceptance criteria and performance metrics.

    Based on the provided information, a clinical study proving the device meets specific acceptance criteria in the context of human performance, as would be typical for an AI/ML device, was not conducted or reported in this 510(k) summary. This summary focuses on physical and bench testing to demonstrate substantial equivalence to previously cleared devices.

    Here's an breakdown of the information that can be extracted and what is NOT available:

    1. Table of acceptance criteria and reported device performance:

    TestAcceptance CriteriaReported Device Performance
    Design analysisMet relevant guidance documents to demonstrate substantial equivalence to predicate device.Performed similarly to predicate device.
    Bench testingMet relevant guidance documents to demonstrate substantial equivalence to predicate device.Performed similarly to predicate device.
    Biocompatibility testingMet relevant guidance documents to demonstrate substantial equivalence to predicate device.Met acceptance criteria.
    Dimensional VerificationNot explicitly stated, but implied to meet design specifications.Not explicitly stated, but implied to meet design specifications.
    Tensile StrengthNot explicitly stated, but implied to meet safety and performance standards.Tested
    Torque StrengthNot explicitly stated, but implied to meet safety and performance standards.Tested
    TorqueabilityNot explicitly stated, but implied to meet safety and performance standards.Tested
    Tip FlexibilityNot explicitly stated, but implied to meet safety and performance standards.Tested
    Coating Adherence/IntegrityNot explicitly stated, but implied to meet safety and performance standards.Tested
    Bench top Simulated EfficiencyNot explicitly stated, but implied to meet safety and performance standards.Tested
    Catheter Fatigue TestingNot explicitly stated, but implied to meet safety and performance standards.Tested
    Shelf Life TestingNot explicitly stated, but implied to meet safety and performance standards.Tested
    Animal studies (using predicate device)Established initial safety and performance.Performed (with CROSSER LP predicate device).

    Note: The acceptance criteria are broadly described as meeting "relevant guidance documents" and performing "similarly to the predicate device." Specific numerical or qualitative targets for each test are not provided in this summary.


    Regarding a "study that proves the device meets the acceptance criteria" in the context of an AI/ML device (which this is not) or human clinical performance, the following points are largely absent or not applicable:

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not applicable for a clinical test set in this 510(k) summary. The testing documented here is primarily physical and bench testing. Animal studies were conducted with a predicate device, but no details on sample size or provenance are provided for those.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable for this 510(k) summary. This type of information would be relevant for clinical studies involving expert interpretation, which is not the focus here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable for this 510(k) summary. Adjudication methods are typically used in clinical trials to establish ground truth for ambiguous cases, which is not described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done or reported. This device is a physical catheter system, not an AI/ML diagnostic tool meant to assist human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a medical device (catheter system), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the physical tests, the "ground truth" would be established by engineering specifications, material science standards, and performance characteristics of the predicate device.
    • For animal studies (with the predicate device), the ground truth for safety and performance would have been established through direct observation, physiological measurements, and potentially necropsy, but no details are provided.

    8. The sample size for the training set:

    • Not applicable. This is a physical device, not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, no training set.

    In summary: The provided document is a 510(k) summary for a physical medical device. It details a series of physical, bench, and biocompatibility tests designed to demonstrate substantial equivalence to predicate devices, rather than a clinical study evaluating human-machine interaction or AI performance against specific clinical acceptance criteria. The acceptance criteria are generally framed as meeting relevant guidance documents and performing similarly to the predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091119
    Manufacturer
    Date Cleared
    2009-05-15

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    MODIFICATION TO THE CROSSER SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CROSSER® System is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions. The device is contraindicated for use in carotid arteries.

    The CROSSER® Catheter is only intended for use with the CROSSER Electronics System. Refer to the CROSSER Electronics System Manual of Operations for proper use.

    Device Description

    The CROSSER® System consists of a re-usable electronic Generator, Foot Switch, highfrequency Transducer, and single-use CROSSER Catheter. The CROSSER Over the Wire Catheter is available in two variations:

    • CROSSER 14P OTW (014 guidewire compatible flexible), .
    • CROSSER 14S OTW (014 guidewire compatible supportive). .

    Each Catheter is connected to the electronic Generator through the Transducer. The Foot Switch is used to activate the system. The Generator and Transducer convert AC power into high frequency mechanical vibrations which are propagated through a Nitinol core wire to the Titanium tip of the CROSSER Catheter. The main body of the catheter is constructed from Pebax and a hydrophilic coating which covers the distal end of the catheter.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the FLOWCARDIA, INC. CROSSER System, based on the provided text:

    Acceptance Criteria and Device Performance

    The documents do not provide a table with specific numerical or quantified acceptance criteria for each test (e.g., "Tensile Strength must be > X Newtons"). Instead, they state that the tests were conducted "to demonstrate that the FlowCardia CROSSER Over the Wire Catheters met the acceptance criteria and performed similarly to the predicate device."

    Therefore, the table below reflects the types of tests performed and the general statement of meeting criteria, rather than specific performance metrics.

    Acceptance Criterion (Test Type)Reported Device Performance
    Dimensional VerificationMet acceptance criteria (implicitly, as it passed testing)
    Tensile StrengthMet acceptance criteria
    Torque StrengthMet acceptance criteria
    TorqueabilityMet acceptance criteria
    Tip FlexibilityMet acceptance criteria
    Coating Adherence/IntegrityMet acceptance criteria
    Benchtop Simulated EfficiencyMet acceptance criteria
    Catheter Fatigue TestingMet acceptance criteria
    BiocompatibilityMet acceptance criteria
    Shelf Life TestingMet acceptance criteria
    Similarity to Predicate DevicePerformed similarly to the predicate device

    Study Details

    1. Sample size used for the test set and the data provenance:

      • The document mentions "Design analysis, bench, and biocompatibility testing," but does not specify the sample sizes for these tests.
      • The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. It appears to be bench testing performed by the manufacturer.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided. The testing appears to be primarily engineering and physical testing, not dependent on expert medical interpretation for ground truth.
    3. Adjudication method for the test set:

      • This information is not provided, as the tests are physical/bench tests rather than clinical evaluations requiring adjudication of expert opinions.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This device is a mechanical catheter system, not an AI-powered diagnostic or assistive technology for human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This question is not applicable. The device is a mechanical catheter and does not involve an algorithm for standalone performance.
    6. The type of ground truth used:

      • For the physical tests, the "ground truth" would be established by engineering specifications, standard test methods, and potentially the performance of the predicate device. For biocompatibility, it would be based on established international standards (e.g., ISO 10993).
    7. The sample size for the training set:

      • This question is not applicable. This device is a mechanical catheter and does not involve a "training set" in the context of machine learning or AI.
    8. How the ground truth for the training set was established:

      • This question is not applicable for the reasons mentioned above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K072776
    Manufacturer
    Date Cleared
    2007-12-07

    (70 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    THE CROSSER SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CROSSER System is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions. The device is contraindicated for use in carotid arteries.

    The CROSSER Catheter is only intended for use with the CROSSER Electronics System. Refer to the CROSSER Electronics System Manual of Operations for proper use.

    Device Description

    The CROSSER System consists of a re-usable electronic Generator, Foot Switch, highfrequency Transducer, and single-use CROSSER Catheter. The CROSSER Catheter is available in three sizes:

    • CROSSER 14P (014 guidewire compatible), .
    • CROSSER 14S (014 guidewire compatible supportive), .
    • CROSSER 18 (018 guidewire compatible). .

    Each Catheter is connected to the electronic Generator through the Transducer. The Foot Switch is used to activate the system. The Generator and Transducer convert AC power into high-frequency mechanical vibrations which are propagated through a Nitinol core wire to the Titanium tip of the CROSSER Catheter. The main body of the catheter is constructed from Pebax and a hydrophilic coating which covers the distal end of the catheter.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the CROSSER System, as much as can be extracted from the given document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Physical Test DataTensile StrengthMet relevant guidance documentsPerformed similarly to predicate device
    Torque StrengthMet relevant guidance documentsPerformed similarly to predicate device
    TorqueabilityMet relevant guidance documentsPerformed similarly to predicate device
    Tip FlexibilityMet relevant guidance documentsPerformed similarly to predicate device
    Coating Adherence/IntegrityMet relevant guidance documentsPerformed similarly to predicate device
    BiocompatibilityMet relevant guidance documentsPerformed similarly to predicate device
    Benchtop Simulated EfficiencyMet relevant guidance documentsPerformed similarly to predicate device
    Catheter Fatigue TestingMet relevant guidance documentsPerformed similarly to predicate device
    Shelf Life and Package Integrity TestingMet relevant guidance documentsPerformed similarly to predicate device
    Dimensional VerificationMet relevant guidance documents(Implied to be met for substantial equivalence)
    Clinical Test DataIntended UseFacilitate intra-luminal placement of guidewires beyond peripheral artery CTOs."The clinical performance of the CROSSER System was verified through a human clinical study designed to demonstrate the intended use of the device." (Specific performance metrics are not detailed in this summary.)

    Important Note: The document heavily relies on "performed similarly to the predicate device" and "met the acceptance criteria" without explicitly stating numerical or qualitative thresholds for each specific test. This is common in 510(k) summaries where detailed raw data and specific thresholds are typically in the full submission, not the public summary.

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size for Test Set: The document mentions "a human clinical study" but does not specify the sample size for this study.
    • Data Provenance: The document does not specify the country of origin of the data. It indicates the study was a "human clinical study," implying prospective human data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not provide any information about experts used to establish ground truth in the clinical study, nor their number or qualifications. Given the nature of the device (facilitating guidewire placement), the primary "ground truth" would likely be successful guidewire crossing and subsequent procedural outcomes, observed and documented by the interventionalists performing the procedures.

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is not an imaging AI diagnostic system that would typically involve a multi-reader study. The clinical study was designed to verify the performance of the device itself in facilitating guidewire placement.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This question is not applicable to the CROSSER System. The device is a physical medical device (catheter system), not a standalone algorithm. Its function is to be used with a human operator (interventionalist).

    7. The Type of Ground Truth Used

    The ground truth for the clinical study was likely procedural success and outcomes data. The "intended use" of the device is to facilitate guidewire placement beyond peripheral artery chronic total occlusions. Therefore, a successful outcome would be the ability to achieve this objective. This would be clinical observation by the treating physicians, rather than pathology or expert consensus on image interpretation.

    8. The Sample Size for the Training Set

    The document does not mention or imply a "training set" as this is not an AI/machine learning device. The design, bench, and biocompatibility testing inform the device's development (analogous to "training" in a broad sense), but there isn't a specific data set called a "training set" in the context of this traditional device submission.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as there is no "training set" in the context of an AI/machine learning algorithm for which ground truth would be established. The "ground truth" for the device's design and manufacturing revolves around meeting established engineering, material science, and safety standards during its development.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1