510(k) Data Aggregation
(24 days)
TEXAS T - PLATE
The Texas T – Plate Distal Radius Plating System is indicated for use in open reduction and internal fixation of fresh fractures, osteotomies, Arthrodesis, and revision procedures of the distal radius.
The Texas T – Plate is an Anatomical Plating System for fractures of the distal radius. All components will be manufactured of 316 LVM Stainless Steel in accordance with ASTM F – 138 standards.
The provided text describes the Biomet Texas T-Plate, a device used for internal fixation of distal radius fractures. The primary evidence presented for its acceptance is a "Non-Clinical cantilever bending test."
Here's an analysis of the acceptance criteria and the study as per your request, based only on the provided text:
Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Sustain a load during cantilever bending test | Able to sustain a higher load than either the IQL Small Epiphysis Plate or the IQL T-Plate (predicate devices) |
Study Details:
-
Sample size used for the test set and the data provenance:
- Sample Size: Not specified in the provided text. The text only mentions "the Texas T -- Plate device" (singular) and the predicate devices (also singular descriptions: "either the IQL Small Epiphysis Plate or the IQL T - Plate"). It's likely a small number of physical samples were tested, but the exact count is not given.
- Data Provenance: The study was a "Non-Clinical" test. No information is provided regarding the country of origin of the data. It's a laboratory test, not patient data, so "retrospective or prospective" is not applicable in the usual clinical sense.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. This was a non-clinical, mechanical bending test. "Ground truth" in this context would be the measured load sustained before failure, which is determined by instrumentation, not expert consensus.
-
Adjudication method for the test set:
- Not applicable. As a mechanical test, the results are objectively measured.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a non-clinical mechanical test of a bone plate, not an AI-assisted diagnostic or treatment device involving human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical medical device, not an algorithm.
-
The type of ground truth used:
- The "ground truth" for this non-clinical test would be the direct measurement of load sustained during a cantilever bending test, likely measured by specialized mechanical testing equipment.
-
The sample size for the training set:
- Not applicable. This is a non-clinical mechanical test, not a machine learning study that requires a training set.
-
How the ground truth for the training set was established:
- Not applicable.
Summary of the Study:
The "study" described is a Non-Clinical cantilever bending test. Its purpose was to compare the mechanical strength (specifically, the ability to sustain a load) of the Texas T-Plate against two legally marketed predicate devices: the IQL Small Epiphysis Plate and the IQL T-Plate. The test found that the Texas T-Plate was able to sustain a higher load than both predicate devices, indicating superior mechanical performance in this specific test. This type of non-clinical data is commonly used for Class II surgical implants to demonstrate substantial equivalence in performance characteristics.