Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K223578
    Manufacturer
    Date Cleared
    2023-07-07

    (219 days)

    Product Code
    Regulation Number
    870.2200
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    T3 Platform software

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T3 Platform™ software features the T3 Data Aggregation & Visualization software module version 5.0 and the T3 Risk Analytics Engine software module version 9.0.

    The T3 Data Aggregation & Visualization software module is intended for the recording and display of multiple physiological parameters of the adult, pediatric, and neonatal patients from supported bedside devices. The software module is not intended for alarm notification or waveform display, nor is it intended to control any of the independent bedside devices to which it is connected. The software module is intended to be used by healthcare professionals for the following purposes:

    • To remotely consult regarding a patient's status, and
    • To remotely review other standard or critical near real-time patient data in order to aid in clinical decisions and deliver patient care in a timely manner.

    The T3 Data Aggregation & Visualization software module can display numeric physiologic data captured by other medical devices:

    • · Airway flow, volume, and pressure
    • · Arterial blood pressure (invasive and non-invasive, systolic, diastolic, and mean)
    • Bispectral index (BIS, signal quality index, suppression ratio) .
    • Cardiac Index
    • Cardiac output .
    • Central venous pressure .
    • . Cerebral perfusion pressure
    • End-tidal CO2 .
    • · Heart rate
    • Heart rate variability .
    • Intracranial pressure .
    • . Left atrium pressure
    • Oxygen saturation (intravascular, regional, SpO2) .
    • Premature ventricular counted beats .
    • Pulmonary artery pressure (systolic, diastolic, and mean) .
    • Pulse pressure variation
    • · Pulse Rate
    • · Respiratory rate
    • Right atrium pressure .
    • Temperature (rectal, esophageal, tympanic, blood, core, nasopharyngeal, skin)
    • · Umbilical arterial pressure (systolic, diastolic, and mean)

    The T3 Data Aggregation & Visualization software module can display laboratory measurements including arterial and venous blood gases, complete blood count, and lactic acid. T3 Data Aggregation & Visualization software module can display information captured by the T3 Risk Analytics Engine software module.

    The T3 Risk Analytics Engine software module calculates four indices: the IDO2 Index™ for inadequate delivery of oxygen, the IVCO2 Index™ for inadequate ventilation of carbon dioxide, the ACD Index™ for acidemia, and the HLA Index™ for hyperlactatemia.

    The IDO2 Index™ is indicated for use by health care professionals with post-surgical patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The IDO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module. When the IDO2 Index™ is increasing, it means that there is an increasing risk of inadequate oxygen delivery and attention should be brought to the patient. The IDO2 Index™ presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements.

    The IVCO2 Index™ is indicated for use by health care professionals with invasively ventilated patients 0 to 12 years of age under intensive care. The IVCO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module. When the IVCO2 Index™ is increasing, it means that there is an increasing risk of inadequate carbon dioxide ventilation and attention should be brought to the patient. The IVCO2 Index™ presents partial quantitative information about the patient's respiratory condition, and no therapy or drugs can be administered based solely on the interpretation statements.

    The ACD Index™ is indicated for use by health care professionals with invasively ventilated patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The ACD Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module. When the ACD Index™ is increasing, it means that there is an increasing risk of acidemia and attention should be brought to the patient. The ACD Index™ presents partial quantitative information about the patient's respiratory condition, and no therapy or drugs can be administered based solely on the interpretation statements.

    The HLA Index™ is indicated for use by health care professionals with post-surgical patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The HLA Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module. When the HLA Index™ is increasing, it means that there is an increasing risk of hyperlactatemia and attention should be brought to the patient. The HLA Index™ presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements.

    Device Description

    The Tracking, Trajectory, Trigger (73) intensive care unit software solution allows clinicians and quality improvement teams in the ICU to aggregate data from multiple sources, store it in a database for analysis, and view the streaming data. System features include:

    • Adjunctive status indicators ●
    • Customizable display of physiologic parameters over the entire patient stay
    • . Configurable annotation
    • Web-based visualization that may be used on any standard browser
    • Minimal IT footprint ●
    • Software-only solution no new bedside hardware required ●
    • Highly reliable and robust operation
    • . Auditable data storage
    AI/ML Overview

    The T3 Platform™ software calculates four indices: the IDO2 Index™ for inadequate delivery of oxygen, the IVCO2 Index™ for inadequate ventilation of carbon dioxide, the ACD Index™ for acidemia, and the HLA Index™ for hyperlactatemia. The document specifies that studies were performed to validate the IVCO2 Index.

    Here's an analysis of the acceptance criteria and study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document explicitly states that the acceptance criteria for the IVCO2 Index™ are: discriminatory power, range utilization, resolution/limitation, and robustness. It is stated that "All results met the same acceptance criteria as the predicate device".

    However, specific quantitative values for these criteria (e.g., a specific threshold for discriminatory power, or a defined range utilization percentage) and the exact performance metrics obtained are not provided in the given text. The document confirms that the device met these criteria without detailing the quantitative results of the performance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set 1 (Neonatal ICU patients):
      • Sample Size: 180 patients, contributing 1108 PaCO2 measurements from arterial blood gases.
      • Data Provenance: Retrospective, obtained from two Level IV regional NICUs in the US.
    • Test Set 2 (Non-NICU patients):
      • Sample Size: 2090 patients, contributing 29,841 PaCO2 measurements from arterial blood gases.
      • Data Provenance: Retrospective, from clinical sites in the US.
      • Demographics: 42% neonates, 32% infants, and 26% children. 46% female and 54% male.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts or their qualifications used to establish the ground truth for the test set. It mentions "PaCO2 measurements from arterial blood gases," which implies that the ground truth was based on laboratory measurements rather than expert review.

    4. Adjudication Method for the Test Set

    The document does not mention an adjudication method for the test set. Given that the ground truth appears to be objective laboratory measurements (PaCO2 from arterial blood gases) and not subjective expert interpretations, a traditional adjudication process (like 2+1 or 3+1) would not be applicable or necessary.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned. The study described focuses on the device's performance against objective physiological measurements, not on its impact on human reader performance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, a standalone performance study was done. The IVCO2 Index™ was "retrospectively computed on all de-identified patients" and "evaluated against the same acceptance criteria as the predicate device." This indicates an algorithm-only evaluation, without human-in-the-loop assistance.

    7. Type of Ground Truth Used

    The ground truth used was objective physiological data/laboratory measurements, specifically PaCO2 measurements from arterial blood gases.

    8. Sample Size for the Training Set

    The document mentions "Development test sets are used to evaluate the impact of the development changes during the development process." However, it does not specify the sample size for the training set(s) used to develop the IVCO2 Index™ model. It only provides details for the validation sets.

    9. How the Ground Truth for the Training Set Was Established

    The document states that the "adjunctive status indicators are produced by a model-based approach. The model-based approach is designed based on principles of physiology with parameters chosen to reflect those specified in the medical literature." It does not explicitly detail how ground truth was established for the training data beyond implying it was based on physiological principles and medical literature, and presumably also included patient data similar to the validation sets.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213423
    Manufacturer
    Date Cleared
    2023-01-06

    (443 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    T3 Platform software

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T3 Platform™ software features the T3 Data Aggregation & Visualization software module version 5.0 and the T3 Adult Risk Analytics Engine software module version 1.0.

    The T3 Data Aggregation & Visualization software module is intended for the recording and display of multiple physiological parameters of the adult, pediatric, and neonatal patients from supported bedside devices. The software module is not intended for alarm notification or waveform display, nor is it intended to control any of the independent bedside devices to which it is connected. The software module is intended to be used by healthcare professionals for the following purposes:

    • . To remotely consult regarding a patient's status, and
    • To remotely review other standard or critical near real-time patient data in order to aid in . clinical decisions and deliver patient care in a timely manner.

    The T3 Data Aggregation & Visualization software module can display numeric physiologic data captured by other medical devices:

    • · Airway flow, volume, and pressure
    • · Arterial blood pressure (invasive and non-invasive, systolic, diastolic, and mean)
    • · Bispectral index (BIS, signal quality index, suppression ratio)
    • · Cardiac Index
    • . Cardiac output
    • Central venous pressure .
    • . Cerebral perfusion pressure
    • · End-tidal CO2
    • · Heart rate
    • · Heart rate variability
    • Intracranial pressure .
    • . Left atrium pressure
    • Oxygen saturation (intravascular, regional, SpO2) .
    • · Premature ventricular counted beats
    • · Pulmonary artery pressure (systolic, diastolic, and mean)
    • · Pulse pressure variation
    • · Pulse Rate
    • · Respiratory rate
    • · Right atrium pressure
    • · Temperature (rectal, esophageal, tympanic, blood, core, nasopharyngeal, skin)
    • · Umbilical arterial pressure (systolic, diastolic, and mean)

    The T3 Data Aggregation & Visualization software module can display laboratory measurements including arterial and venous blood gases, complete blood count, and lactic acid.

    The T3 Data Aggregation & Visualization software module can display information captured by the T3 Adult Risk Analytics Engine software module.

    The T3 Adult Risk Analytics Engine software module calculates the Adult IDO2 Index for inadequate delivery of oxygen. The Adult IDO2 Index is indicated for use by health care professionals with post-surgical patients 18 years of age or older under intensive care and not on Mechanical Circulatory Support. The Adult IDO2 Index is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module. When the Adult IDO2 Index is increasing, it means that there is an increasing risk of madequate oxygen delivery and attention should be brought to the patient. The Adult IDO2 Index presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements.

    Device Description

    The Tracking, Trajectory, Trigger (73) intensive care unit software solution allows clinicians and quality improvement teams in the ICU to aggregate data from multiple sources, store it in a database for analysis, and view the streaming data. System features include:

    • · Customizable display of physiologic parameters over entire patient stay
    • Configurable annotation
    • Web-based visualization that may be used on any standard browser
    • Minimal IT footprint
    • Software-only solution no new bedside hardware required
    • Highly reliable and robust operation
    • Auditable data storage
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the T3 Platform™ software:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Discriminatory PowerMet acceptance criteria
    Range UtilizationMet acceptance criteria
    Resolution/LimitationMet acceptance criteria
    RobustnessMet acceptance criteria

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: 4251 mixed venous oxygen saturation measurements from 634 patients.
    • Data Provenance: Clinical data from different clinical sites in the US (retrospective). The data were obtained by the T3 Platform software and were de-identified.

    3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)

    This information is not explicitly provided in the document. The document states that the Adult IDO2 Index was "retrospectively computed on all de-identified patients" and "evaluated against the same acceptance criteria as the supportive predicate device," implying an objective ground truth related to mixed venous oxygen saturation, but not explicitly stating an expert consensus process for the test set.

    4. Adjudication Method for Test Set

    This information is not explicitly provided in the document. The evaluation was against objective acceptance criteria and not described as involving human adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned for this device. The study focused on the standalone performance of the Adult IDO2 Index against established acceptance criteria.

    6. Standalone Performance Study

    Yes, a standalone study was done. The Adult IDO2 Index algorithm's performance was validated using clinical datasets.

    7. Type of Ground Truth Used

    The ground truth used was based on mixed venous oxygen saturation measurements, a physiological parameter that the Adult IDO2 Index aims to reflect (likelihood of inadequate delivery of oxygen, specifically defined as mixed venous oxygen saturation below a threshold of 50%).

    8. Sample Size for Training Set

    The document does not explicitly state the sample size used for the training set. It mentions that the Adult Risk Analytics Engine version 1.0 employs the "same model of human physiology as the one utilized in Risk Analytics Engine version 8.0 cleared under K213230 for the computation of the IDO2 index in pediatric patients (0 to 12 years of age), however, the physiology model has been modified to extend the age-based parameterization." This suggests that a previous model was adapted, but details of the original or updated training data size are not provided.

    9. How Ground Truth for Training Set Was Established

    The document does not explicitly describe how the ground truth for the training set was established. It states that the Adult Risk Analytics Engine version 1.0 "employs the same model of human physiology" as its predecessor, which was "modified to extend the age-based parameterization." This implies the model was developed based on physiological principles and likely validated against clinical data that included mixed venous oxygen saturation measurements. However, the specific process for establishing ground truth during the training phase is not detailed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213230
    Manufacturer
    Date Cleared
    2022-06-22

    (266 days)

    Product Code
    Regulation Number
    870.2200
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    T3 Platform Software

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T3 Platform™ software features the T3 Data Aggregation & Visualization software module version 5.0 and the T3 Risk Analytics Engine software module version 8.0.

    The T3 Data Aggregation & Visualization software module is intended for the recording and display of multiple physiological parameters of the adult, pediatric, and neonatal patients from supported bedside devices. The software module is not intended for alarm notification or waveform display, nor is it intended to control any of the independent bedside devices to which it is connected. The software module is intended to be used by healthcare professionals for the following purposes:

    • To remotely consult regarding a patient's status, and
    • To remotely review other standard or critical near real-time patient data in order to aid in clinical decisions and deliver patient care in a timely manner.

    The T3 Data Aggregation & Visualization software module can display numeric physiologic data captured by other medical devices:

    • Airway flow, volume, and pressure
    • Arterial blood pressure (invasive and non-invasive, systolic, diastolic, and mean)
    • Bispectral index (BIS, signal quality index, suppression ratio) .
    • Cardiac Index
    • . Cardiac output
    • Central venous pressure .
    • . Cerebral perfusion pressure
    • End-tidal CO2 .
    • · Heart rate
    • Heart rate variability .
    • Intracranial pressure .
    • . Left atrium pressure
    • Oxygen saturation (intravascular, regional, SpO2) .
    • Premature ventricular counted beats .
    • · Pulmonary artery pressure (systolic, diastolic, and mean)
    • Pulse pressure variation
    • · Pulse Rate
    • · Respiratory rate
    • Right atrium pressure .
    • Temperature (rectal, esophageal, tympanic, blood, core, nasopharyngeal, skin)
    • · Umbilical arterial pressure (systolic, diastolic, and mean)

    The T3 Data Aggregation & Visualization software module can display laboratory measurements including arterial and venous blood gases, complete blood count, and lactic acid. T3 Data Aggregation & Visualization software module can display information captured by the T3 Risk Analytics Engine software module,

    The T3 Risk Analytics Engine software module calculates four indices: the IDO2 Index™ for inadequate delivery of oxygen, the IVCO2 Index™ for inadequate ventilation of carbon dioxide, the ACD Index™ for acidemia, and the HLA Index™ for hyperlactatemia.

    The IDO2 Index™ is indicated for use by health care professionals with post-surgical patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The IDO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module. When the IDO2 Index™ is increasing, it means that there is an increasing risk of inadequate oxygen delivery and attention should be brought to the patient. The IDO2 Index™ presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements.

    The IVCO2 Index™ is indicated for use by health care professionals with invasively ventilated patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The IVCO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module. When the IVCO2 Index™ is increasing, it means that there is an increasing risk of inadequate carbon dioxide ventilation and attention should be brought to the IVCO2 Index™ presents partial quantitative information about the patient's respiratory condition, and no therapy or drugs can be administered based solely on the internets.

    The ACD Index™ is indicated for use by health care professionals with invasively ventilated patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The ACD Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module. When the ACD Index™ is increasing, it means that there is an increasing risk of acidemia and attention should be brought to the patient. The ACD Index™ presents partial quantitative information about the patient's respiratory condition, and no therapy or drugs can be administered based solely on the interpretation statements.

    The HLA Index™ is indicated for use by health care professionals with post-surgical patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The HLA Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module. When the HLA Index™ is increasing, it means that there is an increasing risk of hyperlactaternia and attention should be brought to the patient. The HLA Index™ presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements.

    Device Description

    The Tracking, Trajectory, Trigger (73) intensive care unit software solution allows clinicians and quality improvement teams in the ICU to aggregate data from multiple sources, store it in a database for analysis, and view the streaming data. System features include:

    • Adjunctive status indicators
    • Customizable display of physiologic parameters over entire patient stay
    • Configurable annotation
    • Web-based visualization that may be used on any standard browser
    • Minimal IT footprint
    • Software-only solution no new bedside hardware required
    • Highly reliable and robust operation
    • Auditable data storage
    AI/ML Overview

    Here's an analysis of the acceptance criteria and study proving the device meets those criteria, based on the provided text:

    Device: T3 Platform™ software (specifically the T3 Risk Analytics Engine software module version 8.0, which calculates the IDO2 Index™, IVCO2 Index™, ACD Index™, and HLA Index™).

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are broadly defined as "discriminatory power, range utilization, resolution/limitation, and robustness." While specific numerical thresholds for these criteria are not provided in the document, the conclusion states that "All results met acceptance criteria for discriminatory power, range utilization, resolution/limitation, and robustness."

    Acceptance CriteriaReported Device Performance
    Discriminatory PowerMet acceptance criteria (all indices)
    Range UtilizationMet acceptance criteria (all indices)
    Resolution/LimitationMet acceptance criteria (all indices)
    RobustnessMet acceptance criteria (all indices)

    2. Sample Size Used for the Test Set and Data Provenance

    The document provides sample sizes for the validation of the new indices (ACD and HLA Indexes):

    • HLA Index Validation: 58,168 whole blood lactate measurements from 3,496 patients.
    • ACD Index Validation: 24,431 arterial blood pH measurements from 1,858 patients.

    Data Provenance:

    • Country of Origin: United States (data from eleven different clinical sites in the US).
    • Retrospective or Prospective: Retrospectively computed on all de-identified patients.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The document does not specify the number of experts or their qualifications used to establish the ground truth for the test set. It states that the indices are "derived by mathematical manipulations of the physiologic data and laboratory measurements" and that the "underlying physiology model" for the indices has been updated. The ground truth for the ACD Index is defined as "arterial pH less than 7.25" and for the HLA Index as "whole blood lactate level concentration above 4 mmol / L." These appear to be objective thresholds based on laboratory measurements rather than expert consensus on interpretations of images or complex clinical states.

    4. Adjudication Method for the Test Set

    No adjudication method (e.g., 2+1, 3+1) is mentioned or appears to be relevant given the nature of the ground truth (objective laboratory measurements).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study was done, as the device is a risk analytics engine providing calculated indices, not an imaging AI designed to assist human readers in interpreting medical images. The primary use is for "remotely consult regarding a patient's status" and "remotely review other standard or critical near real-time patient data in order to aid in clinical decisions and deliver patient care." The indices provide "partial quantitative information" and "no therapy or drugs can be administered based solely on the interpretation statements." Furthermore, the warnings state: "Do not rely on the T3 Platform software as the sole source of patient status information." This indicates an adjunctive role rather than one where human interpretation is directly assisted in a comparative effectiveness study.

    6. Standalone (Algorithm Only) Performance Study

    Yes, the validation of the ACD Index and HLA Index appears to be a standalone (algorithm only) performance study against objective ground truth derived from laboratory measurements. The indices were "retrospectively computed on all de-identified patients" and evaluated against the defined acceptance criteria.

    7. Type of Ground Truth Used

    The ground truth used is primarily outcomes data / objective laboratory measurements.

    • For the ACD Index: Arterial blood pH measurements, with acidemia defined as arterial pH less than 7.25.
    • For the HLA Index: Whole blood lactate measurements, with hyperlactatemia defined as whole blood lactate level concentration above 4 mmol/L.

    8. Sample Size for the Training Set

    The document does not explicitly state the sample size for a dedicated "training set." It mentions that the "approach is designed based on principles of physiology, and parameters are chosen to reflect those specified in the medical literature and employed development testing data sets and validation sets." It distinguishes between "Development testing sets" (used to evaluate impact of changes during development) and "Validation sets" (used after development is complete). The sample sizes provided (3,496 patients for HLA, 1,858 patients for ACD) are for the validation sets. The size of the "development testing sets" is not specified.

    9. How Ground Truth for the Training Set Was Established

    The document implies that the model's development and parameter choices were guided by "principles of physiology" and "medical literature," which would have formed the basis for the ground truth during the development ("training") phase. For the actual performance testing, the ground truth was established by objective laboratory measurements (arterial pH and whole blood lactate levels) with pre-defined thresholds for acidemia and hyperlactatemia. The details of how ground truth was precisely established for the "development testing data sets" are not explicitly detailed, but it would logically follow the same principles of using objective physiological and laboratory data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202306
    Manufacturer
    Date Cleared
    2020-11-25

    (103 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    T3 Platform software

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T3 Data Aggregation & Visualization software module is intended for the recording and display of multiple physiological parameters of the adult, pediatric, and neonatal patients from supported bedside devices. The software module is not intended for alarm notification or waveform display, nor is it intended to control any of the independent bedside devices to which it is connected. The software module is intended to be used by healthcare professionals for the following purposes: To remotely consult regarding a patient's status, and To remotely review other standard or critical near real-time patient data in order to aid in clinical decisions and deliver patient care in a timely manner.

    The T3 Data Aggregation & Visualization software module can display numeric physiologic data captured by other medical devices: Airway flow, volume, and pressure Arterial blood pressure (invasive and non-invasive, systolic, diastolic, and mean) Bispectral index (BIS, signal quality index, suppression ratio) Cardiac Index Cardiac output Central venous pressure Cerebral perfusion pressure End-tidal CO2 Heart rate Heart rate variability Intracranial pressure Left atrium pressure Oxygen saturation (intravascular, regional, SpO2) Premature ventricular counted beats Pulmonary artery pressure (systolic, diastolic, and mean) Pulse pressure variation Pulse Rate Respiratory rate Right atrium pressure Temperature (rectal, esophageal, tympanic, blood, core, nasopharyngeal, skin) Umbilical arterial pressure (systolic, diastolic, and mean)

    The T3 Data Aggregation & Visualization software module can display laboratory measurements including arterial and venous blood gases, complete blood count, and lactic acid.

    T3 Data Aggregation & Visualization software module can display information captured by the T3 Risk Analytics Engine software module.

    The T3 Risk Analytics Engine software module calculates two indices: the IDO2 Index™ for inadequate delivery of oxygen and the IVCO2 Index™ for inadequate ventilation of carbon dioxide.

    IDO2 Index™ is indicated for use by health care professionals with post-surgical patients aged zero days to twelve years weighing 2 kg or more under intensive care. IDO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module, version 5.0 or higher. When the IDO2 Index™ is increasing, it means that there is an increasing risk of inadequate oxygen delivery and attention should be brought to the IDO2 Index™ presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements.

    IVCO2 Index™ is indicated for use by health care professionals with invasively ventilated patients aged 29 days to 12 years weighing 2 kg or more under intensive care. IVCO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module, version 5.0 or higher. When the IVCO2 Index™ is increasing, it means that there is an increasing risk of inadequate carbon dioxide ventilation and attention should be brought to the IVCO2 Index™ presents partial quantitative information about the patient's respiratory condition, and no therapy or drugs can be administered based solely on the interpretation statements.

    Device Description

    The Tracking, Trajectory, Trigger (T3) intensive care unit software solution allows clinicians and quality improvement teams in the ICU to aggregate data from multiple sources, store it in a database for analysis, and view the streaming data. System features include: Customizable display of physiologic parameters over entire patient stay, Configurable annotation, Web-based visualization that may be used on any standard browser, Minimal IT footprint, Software-only solution - no new bedside hardware required, Highly reliable and robust operation, Auditable data storage.

    AI/ML Overview

    The provided text is a 510(k) summary for the Etiometry T3 Platform™ software. It describes the device, its intended use, and a comparison to a predicate device, focusing on changes made since the prior clearance. However, it does not contain the detailed acceptance criteria or the specifics of a study proving the device meets those criteria, especially in the context of an AI/ML device.

    The document states: "The changes to the algorithm were validated using clinical data sets. The software verification and validation testing were conducted for the subject device, and documentation was provided in accordance with FDA's 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices', May 11, 2005." This indicates that some validation was performed, but the specific metrics, methodologies, and results that would constitute "acceptance criteria" and "detailed performance study" for an AI/ML algorithm are not included in this publicly available summary.

    The device includes an "T3 Risk Analytics Engine software module" which calculates two indices: the IDO2 Index™ and the IVCO2 Index™. These are described as being derived by "mathematical manipulations of the physiologic data and laboratory measurements" and they provide "partial quantitative information about the patient's cardiovascular condition" or "respiratory condition." While these are algorithmic calculations, the document does not present them as a typical AI/ML system that would require a ground truth study with human expert readers for "acceptance criteria" based on metrics like sensitivity, specificity, or AUC, as is common for diagnostic AI.

    Therefore,Based on the provided text, I cannot extract the specific information requested in your prompt regarding acceptance criteria and a study that proves the device meets them, especially in the context of AI/ML performance. The document focuses on showing substantial equivalence to a predicate device via comparative technological characteristics and a general statement of software verification and validation. It does not contain the detailed breakdown of the
    AI/ML specific performance study as requested.

    Here's a breakdown of why each point cannot be robustly addressed from the text:

    1. A table of acceptance criteria and the reported device performance: Not provided. The document states "The performance of the IDO2 Index™ was evaluated with and without regional oximetry to confirm the utility of the measurement as an input," but no specific performance metrics or acceptance criteria are listed.
    2. Sample sizes used for the test set and the data provenance: Not mentioned. It only states "validated using clinical data sets."
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned. The ground truth method is not explicitly detailed beyond the indices being "derived by mathematical manipulations of the physiologic data and laboratory measurements."
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not mentioned.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not mentioned. The IDO2 and IVCO2 indices are described as providing "partial quantitative information" and not something on which therapy or drugs can be administered solely. There is no indication of a human-in-the-loop study to demonstrate improvement.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The indices are calculated by the software ("algorithm only"), but the performance metrics for these calculations (e.g., accuracy against a gold standard for the derived "risk") are not presented.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The indices are mathematically derived from physiological data and lab measurements. The "ground truth" for these indices would likely be the accurate calculation based on these inputs, or potentially a correlation with actual patient outcomes (though no such correlation study is detailed here). The text mentions the indices represent "increasing risk of inadequate oxygen delivery" or "inadequate carbon dioxide ventilation," but how this "risk" itself is ground-truthed is not explained.
    8. The sample size for the training set: Not mentioned.
    9. How the ground truth for the training set was established: Not mentioned.

    In summary, the provided FDA 510(k) summary focuses on demonstrating substantial equivalence based on the device's intended use and technological characteristics compared to a previously cleared predicate and generalized statements of V&V. It does not provide the specific performance study details, particularly those related to acceptance criteria, ground truth establishment, and reader studies that are typically associated with AI/ML devices making diagnostic assertions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190273
    Manufacturer
    Date Cleared
    2019-12-08

    (303 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    T3 Platform software

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T3 Data Aggregation & Visualization software module is intended for the recording and display of multiple physiological parameters of the adult, pediatric, and neonatal patients from supported bedside devices. The software module is not intended for alarm notification or waveform display, nor is it intended to control any of the independent bedside devices to which it is connected. The software module is intended to be used by healthcare professionals for the following purposes:
    · To remotely consult regarding a patient's status, and
    • To remotely review other standard or critical near real-time patient data in order to aid in clinical decisions and deliver patient care in a timely manner.

    The T3 Data Aggregation & Visualization software module can display numeric physiologic data captured by other medical devices:

    • · Airway flow, volume, and pressure
    • · Arterial blood pressure (invasive and non-invasive, systolic, diastolic, and mean)
    • · Bispectral index (BIS, signal quality index, suppression ratio)
    • Cardiac Index
    • Cardiac output
    • Central venous pressure
    • · Cerebral perfusion pressure
    • · End-tidal CO2
    • · Heart rate
    • · Heart rate variability
    • · Intracranial pressure
    • Left atrium pressure
    • · Oxygen saturation (intravascular, regional, SpO2)
    • · Premature ventricular counted beats
    • · Pulmonary artery pressure (systolic, diastolic, and mean)
    • · Pulse pressure variation
    • · Pulse Rate
    • · Respiratory rate
    • · Right atrium pressure
    • · Temperature (rectal, esophageal, tympanic, blood, core, nasopharyngeal, skin)
    • · Umbilical arterial pressure (systolic, diastolic, and mean)

    The T3 Data Aggregation & Visualization software module can display laboratory measurements including arterial and venous blood gases, complete blood count, and lactic acid.

    The T3 Data Aggregation & Visualization software module can display information captured by the T3 Risk Analytics Engine software module.

    The T3 Risk Analytics Engine software module calculates two indices: the IDO2 Index™ for inadequate delivery of oxygen and the IVCO2 Index™ for inadequate ventilation of carbon dioxide.

    IDO2 Index™ is indicated for use by health care professionals with post-surgical patients aged zero days to twelve years weighing 2 kg or more under intensive care. IDO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module, version 3.3 or higher. When the IDO2 Index™ is elevated. it means that there is an increased risk of inadequate oxygen delivery and attention should be brought to the IDO2 Index™ presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements.

    IVCO2 Index™ is indicated for use by health care professionals with invasively ventilated patients aged 29 days to 12 years weighing 2 kg or more under intensive care. IVCO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module, version 3.3 or higher. When the IVCO2 Index™ is increasing, it means that there is an increasing risk of inadequate carbon dioxide ventilation and attention should be brought to the patient.

    Device Description

    The Tracking, Trajectory, Trigger (73) intensive care unit software solution allows clinicians and quality improvement teams in the ICU to aggregate data from multiple sources, store it in a database for analysis, and view the streaming data in real-time. System features include:

    • Customizable display of physiologic parameters over entire patient stay
    • Configurable annotation
    • Web-based visualization that may be used on any standard browser ●
    • . Minimal IT footprint
    • Software-only solution no new bedside hardware required .
    • . Highly reliable and robust operation
    • Auditable data storage

    This release of the T3 Platform Software contains the following software module changes:

    T3 Risk Analytics Engine software module version 5.0

    1. IVCO2 Index™. The subject device is a modification of the T3 Platform software clearance under K163065 that expands the Indications for Use for inadequate ventilation of carbon dioxide index (IVCO2 Index™). The IVCO2 Index™ calculates a patient status index. This is a measure of the likelihood that an arterial blood gas measurement of carbon dioxide would indicate a partial pressure greater than 50 mmHg.
    2. Minor changes to maintain the performance of the IDO2 Index™ when external factors exist. Minor changes to the IDO2 Index™ have been made to maintain the specified performance of the IDO2 Index™ when variations in external factors exist. Specifically, these included changes in the initialization routine to ensure that the algorithm converges to stable behavior within the required time window, and changes in the artifact rejection logic of noninvasive blood pressure measurements to ensure that these measurements are not ignored when only minimum data is available thus providing required minimum data performance.

    T3 Data Aggregation & Visualization software module version 3.3

    1. Index Displayed on Census View. IDO2 Index™ and the IVCO2 Index™ are not displayed on the Census View and the requirement was removed.
    2. Blackout. To mitigate potential misuse of active patient monitoring with Persistent Displays in the ICU, the IDO2 Index™ and the IVCO2 Index™ are blacked out in the near-real-time, a warning label is placed above the Index, and a user is prompted to acknowledge warnings when selecting an Index to be graphed.
    3. Measure synonyms. Certain measures, while not identical to each other, are nevertheless closely related and can be thought of as "synonyms" for each other. For example, "Temp" (Generic Temperature), "Tairwy" (Airway Temperature), "Tesoph" (Esophogeal Temperature), and "Ttymp" (Tympanic Temperature) are all measures of the patient's temperature. In this example, the measure synonym feature would allow T3 Data Aggregation & Visualization software module to continuously report a value for "Temperature" based on whatever values were being received for Temp, Tairwy, Tesoph and/or Ttymp. The raw (source) variables continue to be available to be seen or graphed. In addition, this feature includes an "explain" function that allows the clinician to examine measure synonyms to see the source of the displayed data value at each moment in time.
    4. Respiratory View. T3 Data Aggregation & Visualization software module now includes a screen that displays measures and laboratory results related to the patient's respiratory system. This predefined arrangement, accessed by clicking on the "Respiratory" tab beside the previously released "Hemodynamic" view, saves the clinician the time and effort of graphing these laboratory data and measures in order to understand the status and trend of the patient's respiratory system. The respiratory view includes Smart Visualization Groups that show measures and labs related to respiratory rate, ventilation pressure, respiratory volume, oxygenation, and ventilation.
    5. Placement of IDO2 at the top of the Patient View. In the T3 Data Aggregation & Visualization software module, the display canvas for IDO2 Index™ has been moved to the top of the Patient View.
    6. Current measure value is shown on the graph instead of the legend. When the user positions the mouse cursor in the central graph area, T3 Data Aggregation & Visualization software module displays the value of all graphed measures for the time corresponding to the cursor location. Previous releases of T3 Data Aggregation & Visualization software module displayed this information on the legend (the label for each measure or laboratory result that is shown to the left of the graph area). This release of T3 Data Aggregation & Visualization software module instead shows the value beside a light vertical line that is drawn at the mouse cursor position. This moves the information closer to the user's focus of attention (the mouse cursor).
    7. Ability to receive and merge multiple data interfaces of the same type. T3 Data Aggregation & Visualization software module currently receives three types of data interfaces: patient registration (ADT), laboratory results, and physiometric data. This release allows T3 Data Aggregation & Visualization software module to receive and merge data interfaces of the same type coming from two or more sources. For example, a physiometric data interface for beds from a particular floor of the hospital can now be merged with a separate physiometric data interface with beds from a different floor or unit. This feature allows clinicians to view collections of beds that correspond to patient care trajectories such as the operating room and recovery unit, or intensive care and step-down unit.
    8. Ability to receive data through SQL relational database query. Previous releases of T3 Data Aggregation & Visualization software module received data exclusively through HL7-based data interfaces. However, depending on a hospital's IT infrastructure, data from peripheral devices like ventilators may be sent to the hospital electronic medical record but not broadcast through an HL7 data interface. This release of T3 Data Aggregation & Visualization software module retains its HL7 functionality, but also supplements it with the ability to simultaneously receive other data through periodic SQL relational database queries. (SQL, structured query language, is the ANSI standard language for relational database management systems). For example, T3 Data Aggregation & Visualization software module can query a hospital database periodically to read validated ventilator values that are not otherwise transmitted through HL7 data interfaces, while simultaneously receiving device and lab data through HL7.
    9. Support for an expanded number of beds. This release of T3 Platform software of T3 web server to support up to 128 intensive-care beds. Previously an instance of T3 web server could only support 30 intensive-care beds. The server-side hardware requirements have been modified to request additional hard disk space for T3 web server to support more beds.
    AI/ML Overview

    The T3 Platform software (T3 Data Aggregation & Visualization software module version 3.3 and T3 Risk Analytics Engine software module version 5.0) was evaluated for its IVCO2 Index™ for inadequate ventilation of carbon dioxide.

    1. Table of Acceptance Criteria and Reported Device Performance (Focus on IVCO2 Index™):

    The document does not explicitly state quantitative acceptance criteria for the IVCO2 Index™ performance. However, it states the index's ability to "correctly discriminate the physiologic state of inadequate ventilation of carbon dioxide though its whole range" and its robustness "against different data availability" was validated. The reported performance is that the validation study demonstrated this ability.

    Acceptance Criteria (Inferred from study description)Reported Device Performance
    Ability to correctly discriminate the physiologic state of inadequate ventilation of carbon dioxide through its whole range.The IVCO2 Index™ was validated retrospectively on clinical data and demonstrated its ability to correctly discriminate the physiologic state of inadequate ventilation of carbon dioxide through its whole range.
    Robustness against different data availability.The IVCO2 Index™ was validated retrospectively on clinical data and demonstrated its robustness against different data availability.

    2. Sample Size and Data Provenance for Test Set:

    • Sample Size: 951 patients.
    • Data Provenance: The data was clinical data gathered in the intended patient population, implying it's from a healthcare setting. The document does not specify a country of origin. The study was retrospective.

    3. Number of Experts and Qualifications for Ground Truth:

    The document does not specify the number or qualifications of experts used to establish the ground truth.

    4. Adjudication Method for Test Set:

    The document does not describe an adjudication method for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported in this document. The focus of the validation was on the algorithm's standalone performance against a clinical gold standard.

    6. Standalone Performance Study:

    Yes, a standalone study of the algorithm's performance (without human-in-the-loop) was done. The IVCO2 Index™ was retrospectively computed on patient data and validated against a gold standard.

    7. Type of Ground Truth Used:

    The ground truth used was arterial blood gases.

    8. Sample Size for Training Set:

    The document does not specify a separate sample size for a training set. The descriptions focus on the validation study using clinical data.

    9. How Ground Truth for Training Set Was Established:

    The document does not provide information on how ground truth was established for a training set, as a specific training set size is not mentioned. For the validation set, the ground truth was established using arterial blood gases.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1