Search Results
Found 5 results
510(k) Data Aggregation
(90 days)
t:slim Insulin Delivery System, t:flex Insulin Delivery System, Tandem Device Updater
The t:slim® Insulin Delivery System is indicated for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 6 years of age and greater. The device is indicated for use with NovoLog® or Humalog® U-100 insulin.
The t:flex™ Insulin Delivery System is indicated for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 12 years of age and greater. The device is indicated for use with NovoLog® or Humalog® U-100 insulin.
The Tandem® Device Updater™ System consists of software which allows for communication between a computer and a t:slim® or t:flex™ Insulin Delivery System. It allows for remote software installation and update of a Tandem Insulin Delivery System.
The t:slim® and t:flex™ Insulin Delivery Systems facilitate the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.
The accessory Tandem Device Updater™ System is comprised of a personal computer application. a web server, an embedded firmware application, and a Tandem pump (t:slim or t:flex). The goal of the Tandem Device Updater System is to provide a secure process for software installation and update.
This document is regarding a 510(k) premarket notification for an insulin delivery system and its updater. It primarily focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a study with specific acceptance criteria and detailed performance data in the typical sense of a diagnostic or screening device.
The "acceptance criteria" and "device performance" in this context are related to the regulatory process of demonstrating safety and effectiveness for substantial equivalence, particularly for software verification/validation and hardware changes.
Here's an attempt to extract and interpret the information based on the provided text, acknowledging that it's a regulatory submission for device changes and not a clinical study report for AI performance.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with corresponding device performance metrics in a quantitative format typical for a diagnostic study. Instead, "acceptance criteria" are implied by successful completion of various verification and validation activities. The "reported device performance" is a statement that these tests were performed and supported substantial equivalence.
Acceptance Criteria (Implied by Regulatory Requirements) | Reported Device Performance (Summary from Text) |
---|---|
Software Verification and Validation (V&V) per FDA guidance. | Software V&V testing was performed. |
Hardware changes supported by verification testing. | Hardware verification testing was performed per previously established acceptance criteria. |
Tactile feedback verification. | Testing includes tactile feedback. |
Fluid ingress verification. | Testing includes fluid ingress. |
Performance testing of hardware changes. | Testing includes performance testing. |
Simulated use testing for hardware changes. | Testing includes simulated use. |
Pump software, hardware, EMC, and electrical safety testing. | Previously completed testing supported substantial equivalence determination. |
Validation of Tandem Device Updater System's user interface. | Validation was completed through human factor formative and summative studies. |
Pediatric Human Factors testing (Ages 6-12) for t:slim Insulin Delivery System. | Pediatric Human Factors testing was completed to support use in individuals 6 years and greater. |
Assurance case addressing operational, environmental, electrical, hardware, software, mechanical, biological/chemical, and use hazards. | An updated safety assurance case was provided, defining the system and covering various attributes and hazards, including new elements like LCD screen reports, Tandem Device Updater, wake button, fuel gauge register, and updated t:slim indications. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not explicitly stated for any specific test. The document refers to "testing" and "studies" but does not quantify the number of devices or participants used for each.
- Data Provenance: Not explicitly stated as country of origin. The studies were conducted by Tandem Diabetes Care, Inc. for submission to the FDA (USA). The nature of most tests described (software V&V, hardware verification, human factors) suggests a mix of in-house testing and possibly external human factors studies. The document does not specify if the human factors studies were prospective or retrospective, but typically, these are prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
There is no mention of "experts" in the context of establishing a ground truth for a test set, as this is not a diagnostic device relying on expert interpretation for its output. The "ground truth" for the device's function is its ability to correctly deliver insulin as programmed and function safely according to its specifications.
4. Adjudication Method for the Test Set
Not applicable. This is not a study requiring adjudication of expert interpretations or clinical outcomes.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an insulin delivery system and its updater, not an AI-assisted diagnostic or screening tool. Therefore, MRMC studies and "human reader improvement with AI" are not relevant to this submission.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
The "Tandem® Device Updater™ System" consists of software, and "Software verification and validation testing" was performed. This implies standalone testing of the software component, although the details of such testing are not provided beyond the statement that it was done per FDA guidance. The insulin delivery systems themselves are hardware-software integrated devices, so "standalone" in the AI sense isn't directly applicable.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The ground truth for this device's performance is based on:
- Engineering Specifications: Successful operation within pre-defined hardware and software specifications.
- Safety Standards: Compliance with relevant safety standards (e.g., electrical safety, EMC, hazard analysis).
- User Performance: Demonstrated ability of users (including pediatric users) to safely and effectively interact with the device as assessed through human factors studies.
- Regulatory Guidance: Adherence to FDA guidance documents for infusion pumps and software V&V.
8. The Sample Size for the Training Set
Not applicable. This document describes a medical device undergoing 510(k) clearance, not a machine learning or AI model that typically requires a "training set" in the computational sense. The software is part of an embedded system, thoroughly tested for compliance with specifications, rather than "trained" on data.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" in the context of an AI/ML model for this device's clearance.
Ask a specific question about this device
(195 days)
t:slim Insulin Delivery System
The t:slim® Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 6 years of age and greater. The device is indicated for use with NovoLog or Humalog U-100 insulin.
The t:slim® Insulin Delivery System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The scope of this Traditional 510(k) is limited to expansion of the indications for use to include individuals 6 years of age and greater. The device is indicated for use with NovoLog or Humalog U-100 insulin. No changes have been made to the device components to accommodate the modified indications for use.
The t:slim System, like the predicate device, consists of the following components and accessories:
-
- a software-controlled, programmable insulin infusion pump ("t:slim Pump" or "Pump");
-
- a dedicated, disposable 3.0 mL (300 unit) insulin cartridge ("cartridge");
-
- an UnoMedical Comfort™ Infusion Set (K051264), or an equivalently FDA cleared infusion set ("infusion set"); and
-
- accessories, including a 3 mL sterile syringe and 26 gauge sterile needle, as well as an AC power supply and DC car adapter power supply with USB for charging the Pump's internal battery, cartridge Instructions for Use, and User's Guide,
The t:slim Pump is a battery operated infusion pump capable of both basal and bolus delivery of insulin. The Pump utilizes a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient's subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). The graphical user interface ("GUI") of the t:slim Pump is a touch screen that displays information used to control the t:slim System. The delivery of insulin is accomplished through a micro-syringe within the head of the cartridge.
Identical to the predicate device, the t:slim cartridge is a single-use device, individually packaged and sealed, and is provided sterile. The cartridge attaches to the t:slim Pump and is designed to hold up to 3.0 mL (300 units) of insulin. Both in the predicate and in the device that is the subject of this 510(k) notice, the delivery of insulin is accomplished through a micro-syringe within the head of the cartridge. The Pump has a pressure measurement sensor to estimate the volume of insulin in the insulin reservoir by measuring air pressure in the cartridge. This information is used to provide feedback to the software to determine the insulin volume, detect occlusions and cartridge removal.
The provided text describes a 510(k) submission for the t:slim® Insulin Delivery System, focusing on expanding its indications for use to include individuals 6 years of age and greater. This is a regulatory document rather than a study report detailing specific performance criteria and a study to prove they are met in a quantitative sense as might be expected for an AI/ML medical device.
Therefore, many of the requested elements for acceptance criteria and study details are not directly applicable or available in the provided text, as this document is about establishing substantial equivalence based on existing data and human factors testing for a wider age range, rather than presenting a novel clinical performance study with specific device metrics.
However, I can extract information related to the safety assurance case and human factors testing, as these touch upon aspects of "acceptance criteria" in a broader sense for this type of regulatory submission.
Here's an attempt to answer based on the provided text, noting where information is not available:
1. A table of acceptance criteria and the reported device performance
The document does not provide a table with specific quantitative acceptance criteria or reported device performance metrics in the way one would see for a diagnostic device (e.g., sensitivity, specificity thresholds). Instead, the "acceptance criteria" are framed in terms of general safety, reliability, and substantial equivalence, particularly concerning the expanded age range.
Acceptance Criteria (Implied from Safety Assurance Case & Substantial Equivalence Claim) | Reported Device Performance (Summary) |
---|---|
Safety risk reduced to acceptable levels (ALARP) | A comprehensive safety assurance case was provided, addressing operational, environmental, electrical, hardware, software, mechanical, biological/chemical, and use hazards. Mitigating technologies are used, and risks are deemed substantially equivalent to the predicate device. |
Adequate device reliability | Documented as part of the assurance case. |
Adequate design verification and validation of device specifications | Documented as part of the assurance case. |
Maintain substantially equivalent use for individuals 6 years and greater | Human Factors testing supports substantially equivalent use of the device in individuals from 6 years of age and greater. Updated labeling, warnings, and precautions provide further information regarding the risks for pediatric users. The device has identical technological characteristics to the predicate device for this expanded indication. |
Supported by peer-reviewed literature for CSII in ages 6-11 | A summary of peer-reviewed literature supports the use of Continuous Subcutaneous Insulin Infusion (CSII) in ages 6 through 11. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not specified for the human factors testing or the peer-reviewed literature summary.
- Data Provenance: Not specified for human factors testing. The peer-reviewed literature would be drawn from various studies, likely international, but this is not detailed. The human factors testing is implied to be relevant to the U.S. market.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable/Not specified. This document describes a 510(k) submission for an insulin pump, not for a diagnostic device that requires expert-established ground truth for a test set in the same manner as an AI/ML imaging system. The "ground truth" here is the established safety and efficacy of CSII, and the device's equivalent performance to its predicate and suitability for a broader age range, assessed through engineering principles, human factors, and literature review.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable/None specified. No adjudication method is mentioned as this is not a study involving human reader interpretations that require adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. An MRMC study was not done. This device is an insulin pump, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable/Not directly addressed in this framework. The device itself is a standalone insulin pump. The "performance" being evaluated is its safety and functionality for an expanded age range, not an algorithm's diagnostic accuracy in isolation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" used for this submission is multifaceted:
- Peer-reviewed medical literature: Supporting the general efficacy and safety of CSII in pediatric populations (ages 6-11).
- Device specifications and design verification/validation: Demonstrating the device meets its intended functional requirements.
- Human Factors testing: Confirming that the device can be used safely and effectively by the intended user population, including the expanded pediatric group, and that updated labeling is sufficient.
- Predicate device's established safety and performance: The t:slim System is shown to have identical technological characteristics and similar intended use to the predicate device, implying its "ground truth" is aligned with the predicate's proven track record.
8. The sample size for the training set
- Not applicable/Not specified. This document does not describe a machine learning model or a training set in the conventional sense. The "training" for the device's expanded indication relies on existing engineering documentation, human factors insights, and clinical literature.
9. How the ground truth for the training set was established
- Not applicable. As there is no "training set" for a machine learning model described, this question is not relevant to the provided text.
Ask a specific question about this device
(142 days)
t:slim Insulin Delivery System, t:flex Insulin Delivery System, Tandem Device Updater
The t:slim® Insulin Delivery System is indicated for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 12 years of age and greater. The device is indicated for use with NovoLog or Humalog U-100 insulin.
The t:flex™ Insulin Delivery System is indicated for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 12 years of age and greater. The device is indicated for use with NovoLog or Humalog U-100 insulin.
The Tandem® Device Updater™ System consists of software which allows for communication between a computer and a t:slim® or t:flex™ Insulin Delivery System. It allows for remote software installation and update of a Tandem Insulin Delivery System.
The t:slim® and t:flex™ Insulin Delivery Systems facilitate the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.
The accessory Tandem Device Updater™ System consists of software which allows for communications between a computer and a Tandem Diabetes Care. Inc. Insulin Delivery System. The Tandem Device Updater™ allows for software installation and update.
The Tandem Device Updater System is comprised of a personal computer application, a web server (HAL), an embedded firmware application (Homer), and a Tandem pump (t.slim or t.flex). The goal of the Tandem Device Updater System is to provide a process for the user to perform updates to the firmware components of a Tandem pump. The Tandem Device Updater System provides communication security consisting of message authentication and encryption, two commonly used methods to secure communication protocols (i.e. USB). Communication between the pump and the Tandem Device Updater System can only occur while the pump is connected to a computer via USB cable.
The HAL Web Server (HAL) is a web server that is responsible for storing and delivering the firmware components of a pump.
The Homer Firmware Application (Homer) is a software application which is part of Tandem Device Updater System that runs on the main MCU incorporated in the pump. Homer is used to update the Bootloader Software in each MCU, essentially rewriting the contents in the flash memory of the MCU. In addition to updating the Bootloader Software, Homer will provide an interface for the secure exchange of security keys during the enabling of communication security. Homer authenticates critical messages received from the PC application.
The Tandem pumps (i.e., t.flex and t:slim described above) incorporate three separate microcontroller units (MCUs), and each MCU runs two separate pieces of firmware, known as the Bootloader Software, and Application Software. The Application Software running on the main MCU, utilizes a graphics library to generate the screens for the pump display. The Tandem Device Updater downloads files from Tandem and installs them on the pump. All binary images for use on the pump transmitted by the Tandem Device Updater are encrypted.
The software components on the pump that may be updated by the Tandem Device Updater System include: the Bootloader Software stored in each MCU, the Application Software stored in each MCU, and the graphics library files stored in the nonvolatile flash memory module.
The software version of the t:slim and t:flex Insulin Delivery Systems for this submission is version 4.3.5.1. The software version of the Tandem Device Updater System for this submission is version 1.0.
This document is a 510(k) summary for the Tandem® Insulin Delivery Systems and the Tandem® Device Updater™ System. It discusses the substantial equivalence to predicate devices and provides performance data primarily in the form of assurance cases and verification/validation testing. However, it does not provide specific numerical acceptance criteria or reported device performance in the format of a table as requested. It also lacks details about specific studies with sample sizes, ground truth establishment, or expert involvement as typically found in diagnostic device evaluations.
Here's an attempt to answer your questions based only on the provided text, but please note the significant limitations due to the nature of this type of submission (a 510(k) for an infusion pump system, not a diagnostic device with specific performance metrics like sensitivity/specificity).
1. A table of acceptance criteria and the reported device performance
The document does not provide a table of numerical acceptance criteria or detailed reported device performance in the context of diagnostic metrics (e.g., sensitivity, specificity, accuracy). The performance data cited are primarily related to safety risks and compliance with guidance documents for infusion pumps.
The "acceptance criteria" are implied by the stated goals of the assurance cases:
- "The safety risk resulting from use of t:slim/t:flex pump is reduced to acceptable levels and is As Low As Reasonably Practicable (ALARP) through use of mitigating technologies and the risks are substantially equivalent to the predicate device."
- "The safety risk resulting from use of Tandem Device Updater is reduced to acceptable levels and is As Low As Reasonably Practicable (ALARP) through use of mitigating technologies and the risks are substantially equivalent to the predicate device."
Reported device performance focuses on successful verification and validation testing, risk mitigation, and substantial equivalence to predicate devices. Specific quantitative results (e.g., "device achieved X% accuracy") are not present.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify a "test set" sample size in the context of diagnostic performance. It refers to "Software verification and validation testing" and "Hardware changes were supported by verification testing." These are likely internal engineering tests, and details on sample sizes or data provenance (country, retrospective/prospective) are not provided. "Human factor formative and summative studies" were conducted for the Tandem Device Updater System's user interface, but details on participant numbers, data type, or provenance are not given.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable in the context of this document. The device is an insulin pump system and a software updater, not a diagnostic tool where expert ground truth establishment for a test set is typically required. The "ground truth" for these types of devices is typically compliance with technical specifications, safety standards, and functional requirements, established through engineering and quality assurance processes, not expert review of medical images or patient data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable. There is no mention of a "test set" requiring adjudication in the diagnostic sense within this document. The testing described is verification and validation, which would follow established engineering protocols rather than adjudication by medical experts.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No such study was done or is mentioned. This is an insulin pump system, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
While the device's software components (including the Tandem Device Updater) operate "standalone" in the sense of executing code, the performance is evaluated in the context of its function within the medical device system. Standalone "algorithm only" performance in the diagnostic sense (e.g., classifying images) is not relevant or described here. Instead, software verification and validation testing was performed on the device's software.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The concept of "ground truth" in the diagnostic sense (e.g., pathology, outcomes data) is not applicable to this submission. For an insulin pump and its updater, the "truth" against which performance is measured is its adherence to design specifications, safety requirements, and proper functionality as determined by engineering tests and regulatory standards. For example, for software, "truth" would be its adherence to specified algorithms and absence of critical bugs.
8. The sample size for the training set
The document does not mention a "training set." This type of submission does not involve machine learning algorithms that typically require training sets. The software development likely follows traditional software engineering paradigms, including design, coding, verification, and validation.
9. How the ground truth for the training set was established
As there is no mention of a "training set," this question is not applicable.
Ask a specific question about this device
(123 days)
T:SLIM INSULIN DELIVERY SYSTEM
The t:slim® Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 12 years of age and greater.
The t:slim® Insulin Delivery System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The scope of this 510(k) is limited to a new software version and minor changes to the pump component of the t:slim System.
The modified t:slim System, like the predicate device, consists of the following components and accessories:
- a software-controlled, programmable insulin infusion pump ("t:slim Pump" or "Pump");
- a dedicated, disposable 3mL (300 unit) insulin cartridge ("cartridge");
- an UnoMedical Comfort™ Infusion Set (K051264), or an equivalently FDA cleared infusion set ("infusion set"); and
- accessories, including a Becton Dickenson 3mL sterile syringe (K941562) and 26 gauge sterile needle, or equivalently cleared syringe and needle, as well as an AC power supply and DC car adapter power supply with USB for charging the Pump's internal battery, cartridge Instructions for Use, and User's Guide.
The t:slim Pump is a battery operated infusion pump capable of both basal and bolus delivery of insulin. It utilizes a motor-driver mechanism to deliver insulin from within a disposable cartridge. through an infusion set, into a patient's subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). The graphical user interface ("GUI") of the t:slim Pump is a capacitive touch screen that displays information used to control the t:slim System. The delivery of insulin is accomplished through a micro-syringe within the head of the cartridge.
The t:slim cartridge is a single-use device. individually packaged and provided sterile. The cartridge attaches to the t:slim Pump and is designed to hold up to 3 mL, or 300 units, of insulin. The delivery of insulin is accomplished through a micro-syringe within the head of the cartridge. The Pump has a pressure measurement sensor to estimate the volume of insulin in the insulin reservoir by measuring air pressure in the cartridge. This information is used to provide feedback to the software to determine the insulin volume, detect occlusions and cartridge removal.
This document is a 510(k) premarket notification for the t:slim Insulin Delivery System, specifically for a new software version and minor hardware changes. As such, it focuses on demonstrating substantial equivalence to a predicate device rather than presenting a de novo study with explicit acceptance criteria and device performance tables in the format typically used for novel AI/ML devices.
Therefore, much of the requested information regarding acceptance criteria, study details, and ground truth establishment, particularly in the context of an AI/ML device, is not directly available or applicable within this specific 510(k) summary.
However, I can extract information related to the performance data as described for this modified insulin pump.
Here's a breakdown of the available information based on your request:
1. A table of acceptance criteria and the reported device performance:
This document does not provide a table of acceptance criteria and reported device performance in a quantitative manner as you might expect for an AI/ML imaging device. Instead, it states that:
- "Hardware changes were supported by verification testing per previously established acceptance criteria."
- "The new software version and minor changes to the pump component of the t:slim System were verified and/or validated per Tandem's design control process and shown to meet design requirements."
The document focuses on the types of testing performed rather than specific quantitative performance metrics against acceptance criteria.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
The document mentions "human factor formative and summative studies" for verifying the user interface. However, it does not specify:
- The sample size used for these studies.
- The data provenance (e.g., country of origin, retrospective or prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not provided in the document. For an insulin pump, "ground truth" would likely refer to clinical outcomes or accurate insulin delivery, but the document doesn't detail how this was established or by whom for the purpose of testing the modifications.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This type of study is relevant for AI/ML diagnostic tools. The t:slim Insulin Delivery System is an infusion pump, not a diagnostic imaging AI/ML device. Therefore, an MRMC comparative effectiveness study is not applicable and was not performed in this context.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
The device is an insulin pump with human interaction. The "software-controlled, programmable insulin infusion pump" operates based on user input. While the software has standalone functions (e.g., controlling insulin delivery rates), the overall system performance is inherently with human-in-the-loop. The document confirms "Tandem completed verification of the t:slim System's user interface through a series of human factor formative and summative studies," indicating that human interaction was part of the testing.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The document does not explicitly state the "type of ground truth" used for performance assessment in terms of a single, definitive reference standard. For an insulin pump, performance is assessed through various engineering and functional tests:
- Verification and validation testing: Ensures the device meets design requirements.
- Hardware changes supported by verification testing: implies comparison against expected device behavior and specifications.
- Software verification and validation testing: per FDA guidance document for infusion pumps.
- Other specific tests: mechanical, environmental conditioning, water ingress, simulated aging testing, electrical safety, and electromagnetic compatibility testing.
The "ground truth" for these tests would be the established specifications, regulatory standards, and expected physical/electrical behavior of the device.
8. The sample size for the training set:
The document describes a modification to an existing device, not the development of an AI/ML model that typically involves a distinct "training set." Therefore, a "training set sample size" in that context is not applicable and not provided.
9. How the ground truth for the training set was established:
As above, this concept of "training set" and its ground truth establishment is not applicable to this type of device modification submission.
Ask a specific question about this device
(122 days)
T:SLIM INSULIN DELIVERY SYSTEM
The t:slim™ Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 12 years of age and greater.
The t:slim™ Insulin Delivery System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The scope of this Special 510(k) is limited to changes to the cartridge configuration of the t:slim System.
The t:slim System with the modified cartridge, like the predicate device, consists of the following components and accessories:
-
- a software-controlled, programmable insulin infusion pump ("t:slim Pump" or "Pump"),
-
- a dedicated, disposable 3mL (300 unit) insulin cartridge ("cartridge");
-
- an UnoMedical Comfort™ Infusion Set (K051264), or an equivalently FDA cleared infusion set ("infusion set"); and
-
- accessories, including a Becton Dickenson 3mL sterile syringe and 26 gauge sterile needle, or equivalently cleared syringe and needle, as well as an AC power supply with USB for charging the Pump's internal battery, cartridge Instructions for Use, and User's Guide.
The t:slim Pump is a battery operated infusion pump capable of both basal and bolus delivery of insulin. It utilizes a motor-driven mechanism to deliver insulin from within a disposable cartridge. through an infusion set, into a patient's subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). The graphical user interface (GUI) of the t:slim Pump is a capacitive touch screen that displays information used to control the t:slim System. The delivery of insulin is accomplished through a micro-syringe within the head of the cartridge.
The t:slim cartridge is a single-use device, individually packaged and sealed, and provided sterile. The cartridge attaches to the t:slim Pump and is designed to hold up to 3 mL, or 300 units, of insulin. The delivery of insulin is accomplished through a micro-syringe within the head of the cartridge. The Pump has a pressure measurement sensor to estimate the volume of insulin in the insulin reservoir by measuring air pressure in the cartridge. This information is used to provide feedback to the software to determine the insulin volume, detect occlusions and cartridge removal.
- accessories, including a Becton Dickenson 3mL sterile syringe and 26 gauge sterile needle, or equivalently cleared syringe and needle, as well as an AC power supply with USB for charging the Pump's internal battery, cartridge Instructions for Use, and User's Guide.
This 510(k) submission (K133593) is for a Special 510(k) notice regarding changes to the cartridge of the t:slim™ Insulin Delivery System. The primary focus of the performance data and acceptance criteria is to ensure that the modifications to the cartridge materials do not negatively impact the device's safety and effectiveness compared to the predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criterion | Reported Device Performance / Evaluation Method |
---|---|---|
Material Safety & Compatibility | Biocompatibility with human tissue | Confirmation of biocompatibility testing per ISO 10993-1, repeated cytotoxicity testing. |
Insulin compatibility (no degradation or interaction with insulin) | Insulin Compatibility testing. | |
Extractables and Leachables within acceptable limits | Extractable and Leachables testing. | |
Chemical composition of new materials matches specifications (or is acceptable) | Chemical composition comparison (FTIR). | |
Cartridge Structural Integrity | Patient line bond strength sufficient | Patient line bond strength testing. |
Cartridge integrity (no leaks under pressure) | Cartridge Pressure/leak testing. | |
Package seal integrity maintained | Package seal integrity testing. | |
Functional Performance (Delivery) | Infusion flow accuracy meets specifications | Infusion flow accuracy testing. |
Fill volume accuracy meets specifications | Fill volume accuracy testing. | |
Manufacturing Process & Quality | Production processes validated (Installation, Operational, Performance) | Production IQ/OQ/PQ (Installation Qualification, Operational Qualification, Performance Qualification). |
Product inspection and release testing standards met | Product inspection and release testing. | |
Risk Mitigation | All identified risks mitigated, no new risks introduced | Evaluation of risks for the type of change involved, confirmation that all changes made to the cartridge were assessed to make sure all the risks were mitigated (e.g., infusion flow accuracy, fill volume accuracy, leakage, bond strength and tubing integrity, shelf life, biocompatibility, insulin compatibility) and that no new risks were introduced as a result of these modifications. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state specific sample sizes for each test mentioned (e.g., for bond strength, leak testing, flow accuracy, etc.). It refers generally to "testing" and "evaluation". Given the nature of a Special 510(k) for material and supplier changes, the testing would typically involve a statistically relevant number of units to demonstrate equivalence to the previously cleared predicate device.
The data provenance is not explicitly stated as country of origin but is implicitly domestic, as the submitter, Tandem Diabetes Care, Inc., is based in San Diego, CA, USA, and the submission is to the U.S. FDA. The data is retrospective in the sense that it's based on internal verification and qualification testing performed by the manufacturer to support the safety and effectiveness of the modified device before market introduction. It is not clinical trial data involving patients.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of submission (Special 510(k) for material changes) does not typically involve a "test set" with "ground truth" established by external experts in the same way clinical studies for diagnostic devices might. The "ground truth" here is adherence to pre-defined engineering specifications, regulatory standards (like ISO 10993-1), and the performance characteristics of the predicate device. The experts involved would be the manufacturer's internal R&D, engineering, quality, and regulatory personnel responsible for designing, testing, and verifying the changes, along with potentially external lab specialists for specific tests like biocompatibility or extractables/leachables. Their qualifications would be relevant scientific and engineering degrees and experience in medical device development and testing. The document does not specify the number or detailed qualifications of these individuals.
4. Adjudication Method for the Test Set
Not applicable in the context of this Special 510(k). Adjudication methods like "2+1" or "3+1" are typically used in clinical studies or image-based diagnostic device evaluations where multiple human readers interpret data to establish a consensus ground truth. Here, the "ground truth" is defined by established engineering specifications and compliance with validated test methods.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This submission is for modifications to an insulin pump cartridge, not an AI-powered diagnostic or assistive technology. Therefore, no MRMC study or AI-related effectiveness assessment was performed or is relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is an insulin pump, not an algorithm. Performance is tested for the physical device components and their functionality.
7. The Type of Ground Truth Used
The ground truth used in this context is primarily:
- Engineering Specifications and Design Requirements: The performance of the modified cartridge must meet or exceed the original design specifications for the predicate device.
- Regulatory Standards: Compliance with relevant international standards, e.g., ISO 10993-1 for biocompatibility.
- Predicate Device Performance: The modified device must demonstrate substantial equivalence to the predicate device, meaning its performance, safety, and effectiveness are comparable.
- Validated Test Methods: The results are derived from validated test methods and protocols.
8. The Sample Size for the Training Set
Not applicable. This device does not have an "algorithm" in the sense of a machine learning model that requires a training set. The "training" here refers to the development and manufacturing processes which are subject to qualification (IQ/OQ/PQ) and ongoing quality control.
9. How the Ground Truth for the Training Set was Established
Not applicable for the reasons stated in point 8.
Ask a specific question about this device
Page 1 of 1