Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K212488
    Date Cleared
    2021-11-26

    (109 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The subject device is powder free vinyl examination gloves, adding 1%-3% nitrile to improve the tensile strength and ultimate elongation. The subject device is blue color. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for Synthetic Nitrile Examination Gloves. It details the device's characteristics, its intended use, and comparative testing against a predicate device to establish substantial equivalence.

    Based on the document, I can provide the acceptance criteria and study details for the device as follows:

    Acceptance Criteria and Device Performance

    The acceptance criteria for the Synthetic Nitrile Examination Gloves are primarily based on performance standards and biocompatibility. The reported device performance is compared against these criteria.

    Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device PerformanceResult
    ASTM D5250Physical Dimensions Test - Length$\ge$ 230 mm> 240 mmPass
    ASTM D5250Physical Dimensions Test - Width (S)85 ± 5 mm87-88 mmPass
    ASTM D5250Physical Dimensions Test - Width (M)95 ± 5 mm96-98 mmPass
    ASTM D5250Physical Dimensions Test - Width (L)105 ± 5 mm105-107 mmPass
    ASTM D5250Physical Dimensions Test - Width (XL)115 ± 5 mm116 mmPass
    ASTM D5250Physical Dimensions Test - Thickness (Finger)$\ge$ 0.08 mm0.12-0.13 mmPass
    ASTM D5250Physical Dimensions Test - Thickness (Palm)$\ge$ 0.08 mm0.08-0.09 mmPass
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125 (implies 0 defects in 125 samples, meeting AQL 2.5 criteria)Pass
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 ($
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1