Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K250236
    Manufacturer
    Date Cleared
    2025-05-30

    (123 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Swoop**®** Portable MR Imaging**®** System (V2)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for use vary depending on the specific product and its intended application. These products are designed for use in medical or laboratory settings by trained professionals. Depending on the device, intended uses may include:

    • Diagnostic purposes: Analyzing biological samples (e.g., blood, urine, tissue) to identify diseases, conditions, or other health markers. This can include detecting infections, monitoring chronic illnesses, or screening for genetic predispositions.
    • Therapeutic procedures: Assisting in or performing medical interventions, such as administering medications, delivering fluids, or providing respiratory support.
    • Research and development: Used in laboratory experiments and studies to investigate biological processes, test new drugs, or develop new medical technologies.
    • Monitoring physiological parameters: Measuring heart rate, blood pressure, oxygen saturation, or other vital signs.
    • Sample collection and preparation: Gathering, processing, and storing biological samples for further analysis.

    Specific indications for use are provided in the product's labeling, instructions for use (IFU), or accompanying documentation. Users should always refer to the manufacturer's provided information for the most accurate and complete indications.

    Device Description

    {{overview}}

    AI/ML Overview

    The FDA Clearance Letter for the Swoop® Portable MR Imaging® System (V2) provides details on the acceptance criteria and the studies conducted to demonstrate the device meets these criteria, particularly focusing on the "Advanced Reconstruction" feature which likely incorporates deep learning for image quality optimization.

    Here's a breakdown of the requested information:

    1. Acceptance Criteria and Reported Device Performance

    The core performance of the device's "Advanced Reconstruction" was evaluated through three studies: Performance Analysis, Contrast-to-Noise Ratio (CNR) Validation, and Advanced Reconstruction Image Validation.

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Performance Analysis (Robustness, Stability, Generalizability)Quantitative: Reduced Normalized Mean Squared Error (NMSE) and improved Structural Similarity Index (SSIM) for Advanced Reconstruction compared to Linear Reconstruction.
    Qualitative: Preservation of motion and zipper artifacts, and no unexpected output.Quantitative: "For all models and all test datasets NMSE was reduced and SSIM was improved for Advanced Reconstruction test images compared to Linear Reconstruction test images."
    Qualitative: "Advanced Reconstruction preserved the presentation of motion and zipper artifacts, and no unexpected output was observed."
    Contrast-to-Noise Ratio (CNR) ValidationMean CNR of Advanced Reconstruction required to be greater than the mean CNR of baseline Linear Reconstruction at a statistical significance level of 0.05 for each sequence type."In all cases, CNR of Advanced Reconstruction was greater than or equal to Linear Reconstruction for both hyper- and hypo-intense pathologies. The study result demonstrates that Advanced Reconstruction does not unexpectedly modify, remove, or reduce the contrast of pathology features."
    Advanced Reconstruction Image Validation (Human Reader Study)Advanced Reconstruction required to perform at least as well as Linear Reconstruction in all categories (median score ≥0 on Likert scale) and perform better (≥1 on Likert scale) in at least one of the quality-based categories (noise, sharpness, contrast, geometric fidelity, artifact, overall image quality)."Advanced Reconstruction achieved a median score of 2 (the most positive rating scale value) in all categories. This scoring indicates reviewers found Advanced Reconstruction improved image quality while maintaining diagnostic consistency relative to Linear Reconstruction."

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes three distinct test sets for different validation studies.

    • Performance Analysis (Robustness, Stability, Generalizability):
      • Sample Size:
        • T1, T2, FLAIR group: 40 patients, 111 images.
        • DWI group: 29 patients, 94 images.
      • Data Provenance: Not explicitly stated regarding country of origin. The test set was "entirely independent from the dataset used for model training." The "Equipment Type" is listed as "Swoop v2" (with
    Ask a Question

    Ask a specific question about this device

    K Number
    K251276
    Manufacturer
    Date Cleared
    2025-05-21

    (27 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Swoop**®** Portable MR Imaging**®** System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Swoop Portable MR Imaging System is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

    Device Description

    The Swoop system is portable, ultra-low field MRI device that enables visualization of the internal structures of the head using standard magnetic resonance imaging contrasts. The main interface is a commercial off-the-shelf device that is used for operating the system, providing access to patient data, exam setup, exam execution, viewing MRI image data for quality control purposes, and cloud storage interactions. The system can generate MRI data sets with a broad range of contrasts. The Swoop system user interface includes touch screen menus, controls, indicators, and navigation icons that allow the operator to control the system and to view imagery. The Swoop System image reconstruction algorithm utilizes deep learning to provide improved image quality for T1W, T2W, FLAIR, and DWI sequences.

    The subject Swoop System described in this submission includes software modifications related to the pulse sequences and image processing.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the studies that prove the device meets them, based on the provided FDA 510(k) clearance letter for the Swoop® Portable MR Imaging® System:


    1. Table of Acceptance Criteria and Reported Device Performance

    Study ComponentAcceptance CriteriaReported Device Performance
    Performance AnalysisNMSE (Normalized Mean Squared Error) should be reduced and SSIM (Structural Similarity Index) should be improved for Advanced Reconstruction test images compared to Linear Reconstruction test images. Advanced Reconstruction must preserve the presentation of motion and zipper artifacts and no unexpected output should be observed.For all models and all test datasets, NMSE was reduced and SSIM was improved for Advanced Reconstruction test images compared to Linear Reconstruction test images. Advanced Reconstruction preserved the presentation of motion and zipper artifacts, and no unexpected output was observed.
    Contrast-to-Noise Ratio (CNR) ValidationThe mean CNR of Advanced Reconstruction was required to be greater than the mean CNR of the baseline Linear Reconstruction at a statistical significance level of 0.05 for each sequence type. The study result must demonstrate that Advanced Reconstruction does not unexpectedly modify, remove, or reduce the contrast of pathology features.In all cases, CNR of Advanced Reconstruction was greater than or equal to Linear Reconstruction for both hyper- and hypo-intense pathologies. This demonstrates that Advanced Reconstruction does not unexpectedly modify, remove, or reduce the contrast of pathology features.
    Advanced Reconstruction Image ValidationAdvanced Reconstruction was required to perform at least as well as Linear Reconstruction in all categories (median score ≥0 on a Likert scale) and perform better (median score ≥1 on a Likert scale) in at least one of the quality-based categories (noise, sharpness, contrast, geometric fidelity, artifact, and overall image quality) when reviewed by external ABR-certified radiologists.Advanced Reconstruction achieved a median score of 2 (the most positive rating scale value) in all categories. This indicates reviewers found Advanced Reconstruction improved image quality while maintaining diagnostic consistency relative to Linear Reconstruction.

    2. Sample Sizes and Data Provenance

    The provided document does not explicitly state the country of origin for the data or whether it was retrospective or prospective for the training or test sets.

    Study ComponentSample Size (Test Set)Data Provenance (Country, Retrospective/Prospective)
    Performance AnalysisTotal Subjects: 118
    Total Unique Images: 378
    Per Model/Sequence Group:
    • T1, T2, FLAIR: 44 patients, 92 images
    • DWI: 34 patients, 65 images | Not specified in the provided document. |
      | Contrast-to-Noise Ratio (CNR) Validation | Patients: 43
      Images: 95
      ROIs (Regions of Interest): 316 | Not specified in the provided document. |
      | Advanced Reconstruction Image Validation | Patients: 46
      Images: 177
      Per Sequence: At least 16 cases per sequence (with at least 4 cases per sequence-available image orientation) | Not specified in the provided document. |

    3. Number of Experts and Qualifications for Ground Truth

    Study ComponentNumber of ExpertsQualifications of Experts
    Performance AnalysisNot applicable for direct expert review; ground truth was generated synthetically or from high-field/synthetic contrast images.N/A (reference-based metrics comparing reconstructed images to ground truth images).
    Contrast-to-Noise Ratio (CNR) Validation2American Board of Radiology (ABR) certified radiologists.
    Advanced Reconstruction Image Validation5External, American Board of Radiology (ABR) certified radiologists representing clinical users.

    4. Adjudication Method

    Study ComponentAdjudication Method
    Performance AnalysisNot applicable; objective metrics (NMSE, SSIM) compared reconstructed images to synthetic/derived ground truth. Qualitative assessment for motion and zipper artifacts.
    Contrast-to-Noise Ratio (CNR) ValidationROI annotations were reviewed by two ABR-certified radiologists, and inaccurate annotations were excluded. This implies a form of consensus or expert reconciliation for the ROIs.
    Advanced Reconstruction Image ValidationReviewers rated images using a five-point Likert scale. Individual ratings were used to derive a median score for each category. No explicit adjudication method (e.g., 2+1) for discrepant reader opinions is described beyond deriving a median.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • A form of MRMC study was conducted for the "Advanced Reconstruction Image Validation" where five ABR-certified radiologists reviewed images.
    • Effect Size with AI vs. without AI assistance: The study compared Advanced Reconstruction (which utilizes deep learning) to Linear Reconstruction (without advanced AI assistance). Advanced Reconstruction achieved a median score of 2 (the most positive rating scale value) in all categories (noise, sharpness, contrast, geometric fidelity, artifact, and overall image quality), indicating "improved image quality" relative to Linear Reconstruction. The Likert scale used was not detailed, but a score of 2 on a 5-point scale (where 0 might be "no difference" and higher values indicate improvement) suggests a significant positive effect.

    6. Standalone Performance Study

    • Yes, a standalone (algorithm only) performance study was conducted.
    • The "Performance Analysis" section describes evaluating Advanced Reconstruction's ability to reproduce ground truth images using objective metrics (NMSE, SSIM) without human reader involvement for the primary comparison. The "Contrast-to-Noise Ratio Validation" also measured objective image characteristics (CNR) of the algorithm's output.

    7. Type of Ground Truth Used

    Study ComponentType of Ground Truth
    Performance AnalysisReference-based metrics: A set of images including Swoop data, high field images, and synthetic contrast images was used as ground truth target images. Test input data (synthetic k-space generated from the target images) was reconstructed and compared to this ground truth. This is a form of derived/computed ground truth based on ideal or high-quality reference scans and synthetic generation.
    Contrast-to-Noise Ratio (CNR) ValidationPathologies in images were annotated and reviewed by two ABR-certified radiologists. The CNR was measured between these annotated pathologies and healthy white matter. This implicitly uses expert consensus/annotation for identifying and defining the regions of interest for ground truth comparison. However, the "ground truth" for the improvement in CNR is the Linear Reconstruction itself.
    Advanced Reconstruction Image ValidationThe reference standard for comparison was Linear Reconstruction. The "ground truth" here is human expert assessment (radiologists' ratings) of relative image quality and diagnostic consistency between the Advanced and Linear reconstructions, treating Linear Reconstruction as the baseline for comparison.

    8. Sample Size for the Training Set

    • The document states that the test dataset was "entirely independent from the dataset used for model training."
    • However, the specific sample size or characteristics of the training set are not provided in this document.

    9. How the Ground Truth for the Training Set was Established

    • The document states: "In all cases, models are trained and validated with MRI data and images as the only inputs and outputs" and "Advanced Reconstruction was performed using a test dataset entirely independent from the dataset used for model training."
    • Similar to the training set sample size, the establishment of ground truth for the training set is not detailed in the provided document. It can be inferred that if high-field or synthetically derived images were used for validation, similar methods might have been used for training, but this is not explicitly stated.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240944
    Manufacturer
    Date Cleared
    2024-07-16

    (102 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Swoop**®** Portable MR Imaging**®** System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Swoop Portable MR Imaging System is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

    Device Description

    The Swoop system is portable, ultra-low field MRI device that enables visualization of the internal structures of the head using standard magnetic resonance imaging contrasts. The main interface is a commercial off-the-shelf device that is used for operating the system, providing access to patient data, exam setup, exam execution, viewing MRI image data for quality control purposes, and cloud storage interactions. The system can generate MRI data sets with a broad range of contrasts. The Swoop system user interface includes touch screen menus, controls, indicators, and navigation icons that allow the operator to control the system and to view imagery. The Swoop System image reconstruction algorithm utilizes deep learning to provide improved image quality for T1W, T2W, FLAIR, and DWI sequences.

    AI/ML Overview

    The provided document is a 510(k) Summary for the Hyperfine Swoop Portable MR Imaging System (K240944). It describes the device, its intended use, and compares it to a predicate device (K232760) to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and the study information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list "acceptance criteria" for the device, but rather describes the tests performed and the standards met for demonstrating substantial equivalence. The provided information focuses on engineering and software validation rather than clinical performance metrics such as sensitivity, specificity, or accuracy for a specific diagnostic task.

    Here's a table summarizing the tests described and the reported outcome:

    CategoryTest DescriptionApplicable Standard(s)Reported Performance/Outcome
    Non-Clinical Performance
    Software VerificationSoftware verification testing in accordance with the design requirements to ensure that the software requirements were met.• IEC 62304:2015
    • FDA Guidance, "Content of Premarket Submissions for Device Software Functions"The subject device passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence.
    Image PerformanceTesting to verify the subject device meets all image quality criteria.NEMA MS 1-2008 (R2020), NEMA MS 3-2008 (R2020), NEMA MS 9-2008 (R2020), NEMA MS 12-2016, American College of Radiology (ACR) Phantom Test Guidance for Use of the Large MRI Phantom for the ACR MRI Accreditation Program, American College of Radiology standards for named sequencesThe subject device passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence.
    CybersecurityTesting to verify cybersecurity controls and management.FDA Guidance, "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions"The subject device passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence.
    Software ValidationValidation to ensure the subject device meets user needs and performs as intended.FDA Guidance, "Content of Premarket Submissions for Device Software Functions"The subject device passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence.
    Leveraged from Predicate
    BiocompatibilityBiocompatibility testing of patient-contacting materials.• ISO 10993-1:2018
    • ISO 10993-5:2009
    • ISO 10993-10:2010Test results from the predicate were used to support the subject device because the conditions were identical or the subject device modifications did not introduce a new worst-case configuration or scenario for testing.
    Cleaning/DisinfectionCleaning and disinfection validation of patient-contacting materials.• FDA Guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling"
    • ISO 17664:2017
    • ASTM F3208-17Test results from the predicate were used to support the subject device because the conditions were identical or the subject device modifications did not introduce a new worst-case configuration or scenario for testing.
    SafetyElectrical Safety, EMC, and Essential Performance testing.• ANSI/AAMI ES 60601-1:2005/(R)2012
    • IEC 60601-1-2:2014
    • IEC 60601-1-6:2013Test results from the predicate were used to support the subject device because the conditions were identical or the subject device modifications did not introduce a new worst-case configuration or scenario for testing.
    PerformanceCharacterization of the Specific Absorption Rate for Magnetic Resonance Imaging Systems.• NEMA MS 8-2016Test results from the predicate were used to support the subject device because the conditions were identical or the subject device modifications did not introduce a new worst-case configuration or scenario for testing.

    The document states that the "subject device passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence." This implies that the acceptance criteria were met, which were defined by the adherence to these standards and the internal requirements.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document details testing for software verification, image performance (phantom-based), cybersecurity, and software validation. It does not describe a clinical test set involving patient data for the subject device to evaluate diagnostic performance. The image performance testing appears to be based on physical phantoms (NEMA, ACR). Therefore, information on sample size for a "test set" in the context of clinical images and data provenance (country of origin, retrospective/prospective) is not provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Since no clinical test set evaluating diagnostic performance with patient images is described for the subject device in this document, there is no mention of experts establishing ground truth for such a set. The image performance testing refers to ACR Phantom Test Guidance and standards, which don't typically involve expert reading of collected patient images.

    4. Adjudication Method for the Test Set

    As no clinical test set for diagnostic performance is described, there is no information on an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance?

    No MRMC comparative effectiveness study is mentioned in this document. The submission focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance and leveraging prior test results from the predicate, not on comparative clinical efficacy or improvement with AI assistance for human readers. The device does utilize deep learning for image reconstruction, but its impact on human reader performance is not evaluated in this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document does not describe a standalone performance study in the context of diagnostic accuracy for the AI component (deep learning for image reconstruction). The deep learning is part of an image reconstruction algorithm, and the "image performance" testing is done against established phantom standards, not against a ground truth for diagnostic accuracy.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests described, the "ground truth" refers to:

    • Software Verification/Validation: Adherence to design requirements and user needs.
    • Image Performance: Adherence to image quality criteria as defined by NEMA and ACR phantom standards. The "ground truth" for these tests would be the known properties of the phantoms and the expected imaging parameters.
    • Cybersecurity, Biocompatibility, Cleaning/Disinfection, Safety, Performance (SAR): Adherence to relevant regulatory standards (IEC, ISO, FDA Guidance, ANSI/AAMI, ASTM).

    There is no mention of expert consensus, pathology, or outcomes data as ground truth because no clinical diagnostic accuracy study is presented.

    8. The Sample Size for the Training Set

    The document states that "The Swoop System image reconstruction algorithm utilizes deep learning to provide improved image quality for T1W, T2W, FLAIR, and DWI sequences." However, it does not provide any information regarding the sample size of the training set used for this deep learning algorithm.

    9. How the Ground Truth for the Training Set was Established

    Since the document does not provide details on the training set for the deep learning algorithm, it also does not specify how the ground truth for that training set was established. Given it's an image reconstruction algorithm, the "ground truth" for training would typically involve pairs of raw MRI data and high-quality reconstructed images (often from different acquisition parameters or iterative reconstruction methods) rather than diagnostic labels from experts.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1