510(k) Data Aggregation

    K Number
    K230217
    Date Cleared
    2023-05-13

    (107 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Sterile Polyisoprene Powder Free Surgical Glove, Tested for Use with Chemotherapy Drugs

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This glove is intended to be worn by operating room personnel to protect a surgical wound from contamination. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05(Reapproved 2019).

    Device Description

    Sterile Polyisoprene Powder Free Surgical Glove, tested for Use with Chemotherapy Drugs is a disposable singleuse, sterile, Cream-colored and powder-free surgical glove made from synthetic polyisoprene latex.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for a medical device – "Sterile Polyisoprene Powder Free Surgical Glove, Tested for Use with Chemotherapy Drugs." This device is a Class I, reserved product, meaning it's subject to general controls but does not require extensive clinical data for approval if it demonstrates substantial equivalence to a legally marketed predicate device.

    The provided text focuses on the device's technical specifications and non-clinical testing to demonstrate substantial equivalence to a predicate device, not on proving an AI algorithm's performance in a diagnostic context. Therefore, many of the requested elements for describing an AI acceptance criteria study (e.g., sample size for test set, data provenance, expert ground truth, MRMC study, effect size, standalone performance, training set details) are not applicable to this submission.

    The "acceptance criteria" here refer to the product's physical, chemical, and performance standards as determined by established industry and regulatory test methods, not diagnostic performance metrics of an AI.

    Here's an interpretation of the "acceptance criteria" and "study" based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The device's performance is demonstrated through adherence to various ASTM and ISO standards. The acceptance criteria are the minimum or maximum values specified by these standards, and the reported device performance indicates whether the tested product met these thresholds.

    CharacteristicAcceptance Criteria (Standard/Guidance)Reported Device PerformanceConclusion
    DimensionsASTM D3577-19Meets ASTM D3577 requirements for length, width, and thicknessPass
    LengthMinimum 265mm280-305mmPass
    Palm Width (size)Various (70±6mm to 114±6mm)Average values within range (e.g., 73mm for size 5.5, 114mm for size 9.0)Pass
    Thickness (Finger, Palm, Cuff)Minimum 0.10 mmFinger: 0.22-0.24 mm, Palm: 0.22 mm, Cuff: 0.17-0.18 mmPass
    Physical PropertiesASTM D3577-19Meets ASTM D3577-19Pass
    Tensile Strength, Before Aging≥ 17 MPaAverage 20-21 MPaPass
    Ultimate Elongation, Before Aging≥ 650 %Average 780-823 %Pass
    Stress at 500% Elongation7.0 MPa, maxAverage 2.5-3.1 MPaPass
    Tensile Strength, After Accelerated Aging≥ 12 MPaAverage 18-20 MPaPass
    Ultimate Elongation, After Accelerated Aging≥ 490 %Average 747-809 %Pass
    Freedom from holesASTM D3577-19, ASTM D 5151-19 (AQL 1.5)Meets AQL 1.5Pass
    Powder-FreeASTM D3577-19, ASTM D 6124-06(2022) (≤ 2 mg per glove)0.10-0.27 mg per glovePass
    Aqueous Extractable Protein ContentASTM D3577-19, ASTM D5712-15 (≤ 200 ug/dm2)≤ 50 ug/dm2Pass
    Sterility10-6 SAL10-6 SALPass
    BiocompatibilityISO 10993 standards (10, 23, 5, 11)Results indicate compliance or acceptable levelsPass
    Skin Sensitization TestNo significant evidence of causing skin sensitizationNo significant evidence of causing skin sensitizationPass
    Intracutaneous Reactivity TestScore
    Ask a Question